(225 days)
The Prizma device is indicated to measure, record and transmit ECG one-lead data, and to measure, record, display and transmit heart rate, peripheral oxygen saturation and skin temperature data. The device utilizes a mobile platform to initiate user actions (test, display and data transfer by email) through the mobile application. The device does not perform diagnostic functions.
Intended population: adult patients.
The Prizma is designed in the form of a Mobile Device jacket. The device consists of:
- A sensor unit (Prizma device) embedded with electronics and physiological sensors, in . polycarbonate housing.
- A silicone cover to mechanically receive the sensor unit and form a "Jacket" that is attached to the ● back of the smartphone. Cover design can be modified from one smartphone to another to fit the different dimensions.
- Prizma mobile App, installed on a smartphone, for operating the Prizma device.
- A medical power adapter
The physiological sensors housed in the Medical Capsule are as follows
- One lead ECG sensor, recording + heart rate measurement, derived out of ECG data; ●
- Photo-plethysmography, Peripheral capillary Oxygen Saturation (SPO2) measurement; .
- Skin temperature thermometer using IR; .
Here's a breakdown of the acceptance criteria and study information for the Prizma device, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance:
The provided document details the performance of the SpO2 function against the ISO 80601-2-61 standard. Performance data for ECG and Temperature are described in relation to standards and predicate devices, but specific acceptance criteria and detailed quantitative results beyond comparative statements are not explicitly provided in a tabular format within this document.
| Parameter | Acceptance Criteria (Standard Reference) | Reported Device Performance |
|---|---|---|
| SpO2 | ISO 80601-2-61: a-rms ≤ 3.5% SpO2 over the range of 70% to 100% SpO2 | a-rms = 2.33% over the range of 70% to 100% SpO2 |
Note: For ECG and Temperature, the document primarily makes comparative statements to predicates and compliance with standards rather than numerical acceptance criteria and specific performance metrics in this section.
2. Sample Size Used for the Test Set and Data Provenance:
-
SpO2 Clinical Validation:
- Sample Size: 12 subjects (including dark skin).
- Data Provenance: Prospective clinical study conducted at the Bickler Ye Hypoxia Research Laboratory, Shenzhen Medical University, China.
-
Other Tests (ECG, Temperature, Usability): The document mentions various other tests (e.g., QRS detection, HR measurement, usability per IEC 62366-1:2015) but does not specify the sample sizes or data provenance for these particular tests.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:
This information is not explicitly stated in the provided document for any of the tests. The SpO2 clinical validation used blood gas analysis as the reference, which implies laboratory personnel expertise in obtaining and interpreting these measurements, but specific details about "experts" determining primary ground truth for the device's output are not given.
4. Adjudication Method for the Test Set:
This information is not explicitly stated in the provided document. The SpO2 clinical validation involved comparing the device's readings to a reference device (OSM-3R multi-wavelength oximeter) and blood gas analysis, but details on how discrepancies or disagreements, if any, were adjudicated are not provided.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned for this device. The device does not perform diagnostic functions and is not an AI-assisted diagnostic tool that would typically involve human reader performance studies. The improvements mentioned for the Prizma (e.g., larger screen, animations, simplified use via mobile app) are related to usability and user interaction rather than direct AI assistance to human diagnostic interpretation.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
Yes, the performance data presented (e.g., SpO2 a-rms) reflects the standalone performance of the device's sensing and processing capabilities without human diagnostic intervention. The device's indications specifically state it "does not perform diagnostic functions," implying its role is to measure, record, and transmit data.
7. Type of Ground Truth Used:
-
SpO2: The ground truth for the SpO2 clinical validation was established using blood gas analysis and a reference device (OSM-3R multi-wavelength oximeter). The blood gas analysis serves as the more direct physiological ground truth.
-
ECG (QRS detection and HR measurement): Ground truth was established by means of a signal simulator (Prizma QRS Detection Validation by means of a Signal Simulator).
-
Temperature: While not explicitly detailed as "ground truth" for a specific test, the comparison to the predicate (Health-ePod) and its reliance on a "look-up table which was developed based on a clinical trial to convert the skin temperature to body temperature" suggests that reference measurements from clinical scenarios were used to validate the predicate, which the Prizma then references.
8. Sample Size for the Training Set:
The document does not provide information regarding the sample size for a training set. This is consistent with the device's stated function as a measurement and data transmission device rather than a diagnostic AI algorithm that typically undergoes distinct training and test phases with large datasets.
9. How the Ground Truth for the Training Set Was Established:
Since no information on a training set is provided, the method for establishing its ground truth is also not available in this document.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2017
G-Medical Innovations Ltd. % Jonathan Kahan Partner Hogan Lovells US LLP Columbia Square, 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K170181
Trade/Device Name: Prizma Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II Product Code: DXH, DQA, FLL Dated: August 1, 2017 Received: August 1, 2017
Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M& Millelemen
for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known)
Device Name
Prizma
Indications for Use (Describe)
The Prizma device is indicated to measure, record and transmit ECG one-lead data, and to measure, record, display and transmit heart rate, peripheral oxygen saturation and skin temperature data. The device utilizes a mobile platform to initiate user actions (test, display and data transfer by email) through the mobile application. The device does not perform diagnostic functions.
Intended population: adult patients.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
□ Over-The-Counter Use (21 CFR 801
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
K170181
510(k) SUMMARY
G-Medical Innovations Ltd., Prizma Device
| Submitter | ||
|---|---|---|
| Company: | G-Medical Innovations Ltd. | |
| Address:Phone:Facsimile:Contact Person:Date Prepared: | 5 Openheimer St., Park Rabin, Rehovot, 7670105, Israel972-8-9584777972-8-9584783N. EpsteinSeptember 1, 2017 | |
| Name of Device: Prizma | ||
| Common or Usual Name: | Transmitters and Receivers, Electrocardiograph, Telephone | |
| Oximeter | ||
| Classification Name: | 870.2920 Telephone electrocardiograph transmitter and receive | |
| 870.2700 Oximeter | ||
| Requlatory Class: | 2 | |
| Product Code: | DXH, DQA, FLL |
PREDICATE DEVICES
Card Guard Scientific Survival, Health-ePod, K083174
SpO Medical, PulseOx 5500 Finger Device, K040178
Radiometer, OSM-3 Hemoximeter, K853990 (reference device in oximeter trial)
DEVICE DESCRIPTION
The Prizma is designed in the form of a Mobile Device jacket. The device consists of:
- A sensor unit (Prizma device) embedded with electronics and physiological sensors, in . polycarbonate housing.
- A silicone cover to mechanically receive the sensor unit and form a "Jacket" that is attached to the ● back of the smartphone. Cover design can be modified from one smartphone to another to fit the different dimensions.
- Prizma mobile App, installed on a smartphone, for operating the Prizma device.
- A medical power adapter
The physiological sensors housed in the Medical Capsule are as follows
- One lead ECG sensor, recording + heart rate measurement, derived out of ECG data; ●
- Photo-plethysmography, Peripheral capillary Oxygen Saturation (SPO2) measurement; .
{4}------------------------------------------------
- Skin temperature thermometer using IR; .
INDICATIONS FOR USE
The Prizma device is indicated to measure, record and transmit ECG one-lead data, and to measure, record, display and transmit heart rate, peripheral oxygen saturation and skin temperature data. The device utilizes a mobile platform to initiate user actions (test, display and data transfer by email) through the mobile application. The device does not perform diagnostic functions.
Intended population: adult patients.
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The following technological differences exist between the subject and predicate devices:
The main difference is the addition of the mobile application that controls the device.
The user downloads the application, and pairs the Prizma BLE with smartphone. To run tests, the Prizma application is touched, whereby main screen open with large test icons. Upon touching a test icon, the test screen opens with instructions how to make contact with sensor. Upon test end, the user can add comments to tests and send results via email.
In addition, at end of the tests, the data is automatically transferred over secured BLE protocol from the Prizma to the smartphone where they are stored. Browsing test results history (last five tests) can be performed by touching the Browse icon.
The performance of the sensors against their predicate is summarized below.
ECG
From technology perspective and parameters defined by the standards, both devices are essentially identical. There are some small differences between the devices, such as the dynamic range, which is better in the Prizma due to the fact that the Prizma was designed to comply with IEC-60601-2-47 which is a newer standard that requires better performance, compare to the EC-38 (1998) that the HealthePod was designed to comply with.
In addition, for some parameters, the Prizma performed even better than the Health-ePod (e.g. sampling rate, resolution, heart rate range), which is due to technology and chip improvements over time.
The largest difference between the Prizma and the Health-ePod is the use of the mobile application which the Prizma which is expected to simplify device use as cell phones have larger screens than that used in the Health-ePod. Another difference is that the Prizma is intended solely for adults while the Health-ePod is intended also for pediatric use.
SpO2
The SpO2 sensor in the Prizma and the PulseOx 5500 finger device are similar. From a technology point of view, both devices use the finger to measure the oxygen saturation level. Both devices use reflective technology with similar ranges of wavelength. Both devices report 40% to 100% saturation.
The difference between the devices are minor and are primarily related to the fact that the PulseOx 5500 can be used continuously while the Prizma can be used for spot measurement. In addition, like in
{5}------------------------------------------------
the case of the ECG, the use of mobile app is expected to simplify the test by using a larger screen, animation that explain how to do the test, and an indication when the test is over.
Temperature
From technology point of view, the Health-ePod and the Prizma are identical. Both use the Melexis IR sensor. Both devices measure the skin temperature from the same position on the body (behind the Earlobe). The Health-ePod uses the MLX-90614 and the Prizma uses the MLX-90615 which is the newer generation of the MLX 90614 sensor. The relevant differences between the sensors are the thermal stability which is better in the MLX-90614. This is mostly due to the housing size of the 90614, which is bigger but can be easily resolved by using a metal cap on the 90615 as we did in our design to improve the stability.
The resolution, although better in the 90614, doesn't have clinical significance when measuring skin / body temperature. The measurement range is bigger in the 90614 due to its larger size, but in case of human skin and body temperature, the clinically relevant range can be measured by both devices.
Both devices can measure skin temperature accurately. The Health-ePod uses a look-up table which was developed based on a clinical trial to convert the skin temperature to body temperature.
Finally, the Prizma, like in the ECG and SpO₂, uses the Mobile App to initiate the test in a convenient manner.
Conclusion:
The differences between the device and its predicates do not raise different types of safety and effectiveness questions. Differences between the devices are supported by testing at least equivalent performance. Therefore, the device is substantially equivalent to its predicate devices.
PERFORMANCE DATA
The Prizma was tested according to the following standards:
-
- IEC60601-2-47 Ed. 2.0 Medical electrical equipment - Part 2-47: Particular requirements for the safety, including essential performance, of ambulatory electrocardiographic systems.
-
- ISO 80601-2-56 Ed. 1.0 Medical electrical equipment -- Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
-
- ISO 80601-2-61 Ed. 1.0 Medical electrical equipment Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
In addition, the following testing was conducted:
-
- QRS detection and HR measurement test (according to ANSI/AAMI EC57:2012)
-
- Prizma QRS Detection Validation by means of a Signal Simulator
-
- Usability testing per IEC 62366-1:2015
CLINICAL DATA
Clinical Validation Prizma Finger Oximeter
The SpO2 function was validated in a clinical study at the Bickler Ye Hypoxia Research Laboratory, Shenzhen Medical University, China (report # BYL20160415).
{6}------------------------------------------------
The study protocol was established based on the pulse oximetry guidelines of ISO 80601-2-61 and consisted of blood gas analysis to determine oxyhemoglobin saturation in 12 subjects (including dark skin) subiected to hypoxia conditions followed by measurements from the finger. The reference device was OSM-3R multi-wavelength oximeter (Hemoximeter, Radiometer, Copenhagen).
Conclusion of the study
ISO 80601-2-61 specifies that the SpO2 accuracy of pulse oximeter shall be the root-mean-square difference of less than or equal to 3.5 % SpO2 over the range of 70 % to 100 % SpO2. Accuracy of the tested device shall be stated in terms of the rms difference between its measured values and the reference values of the reference device.
In the range of the Prizma device SpO2 function, 70 - 100%, the a-rms was 2.33. Based on the clinical performance, the Prizma has a safety and effectiveness profile that is similar to the predicate device.
CONCLUSIONS
The Prizma device is as safe and effective as the Health-ePod and PulseOx 5500 Finger Devices. The Prizma has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. In addition, the minor technological differences between the Prizma and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Prizma is as safe and effective as the Health-ePod and PulseOx 5500 Finger Device. Thus, the Prizma is substantially equivalent.
§ 870.2920 Telephone electrocardiograph transmitter and receiver.
(a)
Identification. A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.(b)
Classification. Class II (performance standards).