(150 days)
The Radiopaque Tissue Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.
The device is a sterile, single-patient-use, barium sulfate infused polypropylene suture that is radiopaque using standard radiographs (x-ray, mammography). The device is placed into soft tissue sites during open, percutaneous, or endoscopic procedures and standard surgeon's knots are tied to quickly and inexpensively mark the soft tissue so that the integrity and location of the marked soft tissue can be evaluated.
This document is a 510(k) summary for a medical device called a "Radiopaque Tissue Marker." It outlines the device's characteristics, its intended use, and the non-clinical testing performed to establish substantial equivalence to a predicate device.
Here's an analysis based on the provided text, addressing your questions:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a quantitative format with corresponding reported performance metrics for the device. Instead, it describes non-clinical tests performed to demonstrate substantial equivalence to a predicate device. The performance is assessed against the predicate device, not against specific, pre-defined numerical thresholds for accuracy, sensitivity, or specificity.
However, based on the non-clinical testing mentioned, we can infer performance aspects:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Biocompatibility: Device is safe for implantation and does not cause adverse biological reactions. | Biocompatibility: Showed that the subject device is biocompatible and non-pyrogenic, based on ISO 10993 testing (cytotoxicity, intracutaneous irritation, sensitization, intramuscular implantation, pyrogenicity, and acute systemic toxicity) and toxicological risk assessment of extractables. |
Sterility: Device is sterile at the time of use. | Sterility: Ethylene oxide sterilization parameters were determined to achieve a sterilization assurance level of 10^-6, supported by sterilization resistance and bioburden testing. |
Radiographic Visualization: Device is visible under standard radiological imaging (x-ray, mammography). | Radiographic Visualization: CT/x-ray radiographic imaging at low, medium, and high doses and mammography of implanted subject and predicate devices showed "substantially equivalent radiographic visualization" to the Cassi Beacon Tissue Marker. |
Mechanical Properties (Implied for Suture): Sufficient tensile strength for placement and knot tying. | Mechanical Properties: Polypropylene was chosen as the polymer base due to its "ability to be extruded and processed to increase tensile strength, which is required for placement and knot tying." (This is a design choice to meet a functional requirement, rather than a direct test result). |
Shelf-Life and Packaging Integrity: Device maintains performance and sterility over its shelf life and is protected by its packaging. | Shelf-Life and Packaging Integrity: Packaging validation and post shelf-life performance was also tested. (No specific results provided, only that it was tested). |
Chemical Composition/Safety of Extractables: No harmful leachable substances. | Chemical Composition/Safety of Extractables: Extractables were analyzed using GC-MS, LC-MS, and ICP-MS, and a toxicological risk assessment was performed. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document describes non-clinical laboratory testing. It does not mention a "test set" in the context of patient data, nor does it specify any sample sizes for patients or data provenance (country of origin, retrospective/prospective). The testing involved physical devices and materials, not patient data in the way one would analyze for an AI algorithm.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable to the study described. The study involves non-clinical testing of a physical device. There is no "ground truth" establishment in the context of expert consensus on medical images or diagnoses.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for the study described. There is no "test set" of patient data requiring adjudication of ground truth by multiple experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was performed. The device is a physical radiopaque tissue marker, not an AI algorithm. Therefore, an "effect size of human readers improve with AI vs without AI assistance" is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No standalone algorithm performance study was done. The device is a physical radiopaque tissue marker, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This is not applicable for the study described. The "ground truth" for non-clinical device testing would be the physical and chemical properties of the materials and their behavior under controlled laboratory conditions, as measured by standard testing methods (e.g., ISO 10993 for biocompatibility, specific imaging protocols for radiopacity). It is not based on expert consensus, pathology, or outcomes data related to patient diagnosis.
8. The sample size for the training set
This is not applicable. The device is a physical medical device, not an AI algorithm that requires a "training set."
9. How the ground truth for the training set was established
This is not applicable. There is no AI algorithm or training set associated with this device.
§ 878.4300 Implantable clip.
(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.