K Number
K163522
Date Cleared
2017-03-14

(89 days)

Product Code
Regulation Number
866.3920
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FilmArray Warrior Control Panel M290 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bacillus anthracis (Chromosomal element, pXO1, Coxiella burnetii, Francisella tularensis, Yersinia pestis, Ebola virus and Marburg virus on the FilmArray NGDS Warrior Panel performed on FilmArray 2.0 systems. FilmArray Warrior Control Panel M290 is composed of synthetic DNA specifically designed for and intended to be used solely with the FilmArray NGDS Warrior Panel assay. This product is not intended to replace manufacturer internal FilmArray NGDS Warrior Panel controls provided with the device.

Device Description

FilmArray Warrior Control Panel M290, P/N M290, is a quality control panel consisting of 2 controls, FilmArray Warrior Positive Control, P/N M291 and FilmArray Warrior Negative Control, P/N M292. The Positive Control contains non-infectious surrogate control material; a solution of synthetic DNA in buffers, stabilizers and preservatives. The DNA carries segments of all biothreat agents detected by the FilmArray NGDS Warrior Panel assay (Table 1. below) on FilmArray 2.0 systems. The Negative Control is comprised of non-specific DNA in buffers, stabilizers and preservatives. Each liquid control of FilmArray Warrior Control Panel M290 is processed separately according to FilmArray NGDS Warrior Panel assay manufacturer's Instructions for Use for Blood Samples.

AI/ML Overview

The document describes the performance study for the FilmArray Warrior Control Panel M290, an assayed quality control material for clinical microbiology assays.

Here's the breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document implies an acceptance criterion for "Percent Correct" results of 95% or greater for the external clinical study.

Acceptance Criteria (Implied)Reported Device Performance (Table 2: All Sites)
External Site Performance:99.4% Total Correct Results
Percent Correct Positive Control ≥ 95%98.8%
Percent Correct Negative Control ≥ 95%100%

2. Sample Size and Data Provenance

Test Set (External Study):

  • Total Tests: 310 (163 positive controls, 147 negative controls)
  • Data Provenance: Retrospective and prospective. The study was conducted over 6 months at 5 clinical laboratory sites, incorporating 12 unique pouch lots and multiple operators. The document does not specify countries of origin, but "clinical laboratory sites" typically refers to facilities within the country of the product's intended market (USA, based on FDA review).

3. Number of Experts and Qualifications for Ground Truth

Not applicable. This device is a quality control material, not an AI or diagnostic device that requires expert interpretation for establishing ground truth. The "ground truth" for the controls is their known composition (positive for specific pathogens, negative for all pathogens).

4. Adjudication Method

Not applicable, as this is a quality control material with known positive and negative components, not a diagnostic device requiring expert adjudication of results.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This study focuses on the performance of a quality control material in monitoring the FilmArray NGDS Warrior Panel assay, not on the effectiveness of human readers, with or without AI assistance.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire study describes the performance of the FilmArray Warrior Control Panel M290 itself, demonstrating its ability to produce expected positive and negative results when run on the FilmArray NGDS Warrior Panel.

7. Type of Ground Truth Used

Known Composition: The ground truth is the known synthetic DNA composition of the positive and negative control panels.

  • Positive Control: Contains synthetic DNA segments for all biothreat agents detected by the FilmArray NGDS Warrior Panel assay (Bacillus anthracis: Chromosomal element, pXO1, pXO2, Coxiella burnetii, Francisella tularensis, Yersinia pestis, Ebola virus and Marburg virus).
  • Negative Control: Comprised of non-specific DNA.

8. Sample Size for the Training Set

Not applicable. This is a quality control material, not a machine learning model, so there is no "training set" in the traditional sense. The controls are designed with known compositions to test the performance of a diagnostic assay.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set. The ground truth for the control material itself is based on its manufactured composition.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

March 14, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Maine Molecular Quality Controls, Inc. Joan Gordon 23 Mill Brook Road Saco, ME 04072

Re: K163522

Trade/Device Name: FilmArray Warrior Control Panel M290 Regulation Number: 21 CFR 866.3920 Regulation Name: Assayed quality control material for clinical microbiology assays Regulatory Class: Class II (Special Controls) Product Code: PMN Dated: December 13, 2016 Received: December 15, 2016

Dear Ms. Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

Page 2 - Ms. Gordon

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Steven R. Gitterman -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163522

Device Name FimlArray Warrior Control Panel M290

Indications for Use (Describe)

FilmArray Warrior Control Panel M290 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bacillus anthracis (Chromosomal element, pXO1, Coxiella burnetii, Francisella tularensis, Yersinia pestis, Ebola virus and Marburg virus on the FilmArray NGDS Warrior Panel performed on FilmArray 2.0 systems. FilmArray Warrior Control Panel M290 is composed of synthetic DNA specifically designed for and intended to be used solely with the FilmArray NGDS Warrior Panel assay. This product is not intended to replace manufacturer internal FilmArray NGDS Warrior Panel controls provided with the device.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a serif font, with the "MM" in a bold, golden color and the "QCI" in a dark brown color. Below the letters, the full name of the company, "MAINE MOLECULAR QUALITY CONTROLS, INC.," is written in a smaller, sans-serif font. A golden, abstract design resembling a strand of DNA or a stylized wave is positioned to the left of the letters, adding a visual element that suggests the company's focus on molecular quality control.

510 (k) Summary

510(k) Number: K163522

Purpose for submission: New product

Applicant Information:

Applicant:Maine Molecular Quality Controls, Inc.
Address:23 Mill Brook RoadSaco, Maine 04072
Contact Person:Joan Gordon, President MMQCI
Phone:207-885-1072 extension 201
Fax:207-885-1079
Email Address:jgordon@mmqci.com

Preparation Date: December 12, 2016

Device

Device Trade Name:FilmArray Warrior Control Panel M290, model M290
Device Common Name:Quality Control Material for Microbiology Assays
Device Type:Assayed quality control material for clinical microbiologyassays
Class:Class II (Special controls)
Regulation:21 CFR 866.3920
Panel:Microbiology - 83
Product code:PMN

Predicate Device

K161573; FilmArray RP EZ Control Panel M265, Maine Molecular Quality Controls, Inc.

Device Description

FilmArray Warrior Control Panel M290, P/N M290, is a quality control panel consisting of 2 controls, FilmArray Warrior Positive Control, P/N M291 and FilmArray Warrior Negative Control, P/N M292. The Positive Control contains non-infectious surrogate control material; a solution of synthetic DNA in buffers, stabilizers and preservatives. The DNA carries segments of all biothreat agents detected by the FilmArray NGDS Warrior Panel assay (Table 1. below) on FilmArray 2.0 systems. The Negative Control is comprised of non-specific DNA in buffers, stabilizers and preservatives. Each liquid control of FilmArray Warrior Control Panel M290 is

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. (MMQCI). The logo features the letters "MMQCI" in a stylized font, with the "MM" in a dark red color and the "QCI" in a lighter yellow color. Below the letters, the full name of the company, "MAINE MOLECULAR QUALITY CONTROLS, INC.", is written in a smaller font. A yellow, abstract design resembling a brushstroke or a strand of DNA is positioned to the left of the letters, adding a visual element to the logo.

processed separately according to FilmArray NGDS Warrior Panel assay manufacturer's Instructions for Use for Blood Samples.

Bacillus anthracis: Chromosomal element, pXO1, pXO2
Coxiella burnetii
Francisella tularensis
Yersinia pestis
Ebola Virus
Marburg Virus
Table 1. Biothreat pathogens detected by FilmArray NGDS Warrior Panel
-----------------------------------------------------------------------------------

Device Intended Use

FilmArray Warrior Control Panel M290 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bacillus anthracis (Chromosomal element, pXO1, pXO2), Coxiella burnetii, Francisella tularensis, Yersinia pestis, Ebola virus and Marburg virus on the FilmArray NGDS Warrior Panel performed on FilmArray 2.0 systems. FilmArray Warrior Control Panel M290 is composed of synthetic DNA specifically designed for and intended to be used solely with the FilmArray NGDS Warrior Panel assay. This product is not intended to replace manufacturer internal FilmArray NGDS Warrior Panel controls provided with the device.

Substantial Equivalence

CharacteristicCandidate Device:FilmArray WarriorControl Panel M290Predicate Device:FilmArray RP EZControl Panel M265(K161573)
Intended UseExternal assayed qualitycontrol to monitor invitro lab nucleic acid testSame
Physical formatReady-to-Use LiquidSame
Directions for UseProcess like patientsampleSame
CompositionSynthetic DNASynthetic RNAtranscripts
Assay Steps MonitoredAmplification, detection,identificationSame plus reversetranscription
Number of targetsmonitored in one assayMultipleSame

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. (MMQCI). The logo features the letters "MMQCI" in a stylized font, with the words "Maine Molecular Quality Controls, Inc." written in a smaller font underneath. A yellow, abstract design element is placed to the left of the text, adding a visual accent to the logo.

Summary Performance Data

All Test Results

Three lots of FilmArray Warrior Control Panel M290, FilmArray Warrior Positive Control and FilmArray Warrior Negative Control, were manufactured by MMQCI and tested using the FilmArray NGDS Warrior Panel on the FilmArray instrument 2.0 system to confirm performance. The lots were manufactured and tested incorporating variables including multiple lots of key manufacturing components, different operators, pouch lots and test sites over six months. Internal and external studies were performed.

The internal study was performed by testing the 3 lots at MMQCI (Saco, Maine) over 23 days with 4 pouch lots by 3 operators. An external study was performed to assess performance of FilmArray Warrior Control Panel M290 in a clinical setting. The external performance study tested the 3 lots of FilmArray Warrior Control Panel M290 at 5 clinical laboratory sites, over 6 months, incorporating 12 unique pouch lots and multiple operators.

All FilmArray Warrior Positive Controls were correctly detected in the internal and external studies (Correct Positive Control Result) except for 2 control samples tested in the external study. One FilmArray Warrior Positive Control test reported Detected for all pathogens except for pXO1 which reported as Not Detected. A second FilmArray Warrior Positive Control test result was Detected for all pathogens except for pXO1 and Yersinia pestis, which reported as Not Detected.

All FilmArray Warrior Negative Controls gave Correct Negative Control Results. There were 4 Invalid results caused by internal control pouch failures; 2 for the FilmArray Warrior Positive Control and 2 for the FilmArray Warrior Negative Control.

Table 1. Summary of All Test Results
Numberof sitesTotalTestsInvalidCorrectPositiveControlResultIncorrectPositiveControlResultPercentCorrect*PositiveControlCorrectNegativeControlResultIncorrectNegativecontrolResultPercentCorrect*NegativeControl
64344221299.1%2070100%

*The 4 Invalid samples were re-tested according to BioFire instructions and were not included in the Percent Correct analysis.

External Site Testing

Three lots of FilmArray Warrior Control Panel M290, FilmArray Warrior Positive Control (Pos) and FilmArray Warrior Negative Control (Neg), were manufactured by MMOCI. They were tested by the FilmArray NGDS Warrior Panel on the FilmArray instrument 2.0 system as follows: Lot 1, Pos C18MAR16/ Neg A02NOV15, was tested at 5 clinical sites over 6 months using 9 pouch lots, incorporating multiple operators. Lots 2 and 3, Pos C01JUL16/ Neg

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a bold, serif font, with "MM" in a larger size than "QCI". Below the letters, there is the text "MAINE MOLECULAR QUALITY CONTROLS, INC." in a smaller, sans-serif font. A stylized, golden swirl design is placed to the left of the text, adding a decorative element to the logo.

B01JUL16, and Pos B15JUL16/ Neg A15JUL16, were tested at 2 of the 5 clinical sites on 10 different days, using 3 pouch lots, incorporating multiple operators.

A total of 310 external controls were tested (163 positive and 147 negative). There were no Invalid results. All controls gave correct results except for 2 FilmArray Warrior Positive Controls at 2 different sites. One sample of Lot 2 Pos C01JUL16 gave a Not Detected for pXO1 and Yersinia pestis with all other biothreat pathogens detected at Site B, and another sample of Lot 2 Pos C01JUL16 gave a Not Detected for pXO1with all other biothreat pathogens reported as Detected at Site E for the same pouch lot. The expected percentage of correct results for the clinical study was 95% or greater.

Table 2. Summary of External Test Results for 3 Control Lots
ExternalSiteTotalTestsPercentTotalCorrectResultsCorrectPositiveControlResultIncorrectPositiveControlResultPercentCorrectPositiveControlCorrectNegativeControlResultIncorrectNegativecontrolResultPercentCorrectNegativeControl
A41100%200100%210100%
B10499.1%58198.3%450100%
C51100%260100%250100%
D60100%310100%290100%
E5498.1%26196.3%270100%
All Sites31099.4%161298.8%1470100%

Reproducibility

NOTE: The FilmArray 2.0 is a random access instrument. Therefore, 'runs' refer to replicates tested on the same day.

Within-run Testing: Within-run reproducibility was demonstrated by 1 operator testing 1 lot each of FilmArray Warrior Positive Control and FilmArray Warrior Negative Control with 1 lot of NGDS Warrior Panel pouches on the FilmArray 2.0, each within one day at MMQCl. All results were correct and reproducible.

Table 3. Summary of Within-run Reproducibility Results
ControlControlLot#No. ofTestsDate ofTestingPouchLotCorrectResults

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a stylized font, with the "MM" in a larger, darker font and the "QCI" in a smaller, lighter font. Below the letters, the full name of the company, "MAINE MOLECULAR QUALITY CONTROLS, INC.", is written in a smaller font. To the left of the letters, there is an abstract design in yellow, resembling a flowing or liquid shape.

FilmArray Warrior Positive ControlC01JUL1668-8-161512166/6
FilmArray Warrior Negative ControlB01JUL1668-19-161512166/6

Precision Testing: Precision was demonstrated by testing 3 lots of FilmArray Warrior Positive Control and 3 lots of FilmArray Warrior Negative Control at MMQCI over multiple days with 4 reagent lots by 3 operators using one FilmArray 2.0 instrument. All FilmArray Warrior Positive Control biothreat pathogens were correctly detected. Four tests were excluded from final data analysis due to Invalid results caused by internal pouch control failures.

Table 4. Summary of Precision Testing for 3 Control Lots at MMQCI
ControlControl Lot #No. of TestsInvalidCorrect ResultsIncorrect ResultsPercent Correct*
FilmArray Warrior Positive ControlC18MAR16200210100%
FilmArray Warrior Negative ControlA02NOV15200200100%
FilmArray Warrior Positive ControlC01JUL16200200100%
FilmArray Warrior Negative ControlB01JUL16211200100%
FilmArray Warrior Positive ControlB15JUL16222200100%
FilmArray Warrior Negative ControlA15JUL16211200100%
TOTAL12441200100 %

*The 4 Invalid samples were re-tested according to BioFire instructions and were not included in the Percent Correct analysis.

Lot-to-Lot Testing: A subset of the above testing in Table 4 was used to demonstrate Lot-to-lot reproducibility by testing 3 lots of FilmArray Warrior Control Panel M290 using the same pouch lot on the same instrument at MMQCI. All results were correct and reproducible.

Table 5. Summary of Lot-to-Lot Testing
Control Lot #Number of TestsPouch LotCorrect Results
FilmArray Warrior Positive ControlC18MAR1671603147/7
FilmArray Warrior Negative ControlA02NOV1571603147/7
FilmArray Warrior Positive ControlC01JUL1671603147/7
FilmArray Warrior Negative ControlB01JUL168 (1 Invalid)1603147/7
FilmArray Warrior Positive ControlB15JUL1671603147/7
FilmArray Warrior Negative ControlA15JUL1671603147/7

§ 866.3920 Assayed quality control material for clinical microbiology assays.

(a)
Identification. An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”