K Number
K163407

Validate with FDA (Live)

Date Cleared
2017-09-29

(298 days)

Product Code
Regulation Number
870.2800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

Device Description

The St. Jude Medical Confirm Rx™ system consists of the following key features and components:

  • Confirm RxTM ICM Model DM3500 Implantable Device: The ICM is intended as a minimally invasive, implantable diagnostic monitoring device, with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability, which help physicians monitor, diagnose, and document patients who are susceptible to cardiac arrhythmias. It is predicated on SJM Confirm™ DM2102 with a two-year longevity, MR conditional labeling, and identical sensing and detection algorithms as SJM Confirm™ DM2102, but with downsized hardware and Bluetooth communication.
  • Implant Tools: Model DM3520 incision tool and Model DM3510 insertion tool to implant the device subcutaneously.
  • Model 3111 Magnet (existing SJM donut magnet) facilitates faster startup of Bluetooth connection and provides user authentication (required for programmer sessions).
  • Clinician Programmer (Merlin PCS Programmer Model 3650): The Merlin PCS Programmer 3650 operates using Merlin PCS Model 3330 software and provides the means for the physician to program device parameters and retrieve diagnostic information from the device, including electrograms, heart rate history, episode duration and trend information.
  • myMerlin™ Patient App (Model APP1000 Android): The Patient App provides the means for the patient to activate EGM recording in the Confirm Rx™ device, with data pass-through functionality to enable physician follow-up via the Merlin.net Patient Care Network.
  • Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator): The Merlin.net MN5000 system allows physicians to remotely monitor and diagnose patients' cardiac events.
AI/ML Overview

The provided text describes the Confirm Rx™ Insertable Cardiac Monitor (ICM) System, Model DM3500.

Here's an analysis of the acceptance criteria and the studies mentioned:

1. Table of Acceptance Criteria and Reported Device Performance:

The document mentions that the Confirm Rx™ ICM leverages the existing market-cleared SJM Confirm™ (DM2102) algorithms and functionality. Therefore, the "acceptance criteria" for the detection algorithms are implicitly inherited from the predicate device. While specific numerical targets for metrics like sensitivity or specificity are not explicitly stated as acceptance criteria for the Confirm Rx™ ICM in this summary, the key performance aspects verified through testing are:

Acceptance Criterion (Implicitly inherited/verified)Reported Device Performance
R-wave detectionVerified via bench and animal testing.
Device migrationVerified via bench and animal testing.
Algorithms and FunctionalityIdentical to SJM Confirm™ DM2102, which previously met clinical testing requirements (IDE G080090).
Intended Performance"The results of testing show that the Confirm Rx™ ICM performs as intended and is safe for its intended use."
Safety and Effectiveness (overall)"The Confirm Rx™ ICM is substantially equivalent to the predicate devices in terms of technological characteristics, intended use, and performance."

Important Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices. It typically does not contain the detailed clinical study reports or comprehensive performance metrics that would be found in a full PMA application or a detailed peer-reviewed publication. The "acceptance criteria" here are therefore inferred from the context of substantial equivalence to a previously cleared device.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The document does not explicitly state a specific sample size for a clinical test set used to evaluate the Confirm Rx™ ICM's diagnostic performance for arrhythmia detection. It states: "R-wave detection and device migration were verified via bench and animal testing for the Confirm Rx™ ICM DM3500." and "The Confirm Rx™ ICM leverages the existing market cleared SJM Confirm™ (DM2102) algorithms and functionality for which the clinical testing, performed on the SJM Confirm™ (DM2102) device per St. Jude Medical IDE G080090, is applicable and was not repeated."
  • Data Provenance: The document does not specify the country of origin for the data, but it refers to an IDE (Investigational Device Exemption) study (G080090) conducted for the predicate device, SJM Confirm™ (DM2102). IDE studies are typically prospective clinical trials.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

  • This information is not provided in the 510(k) summary for the Confirm Rx™ ICM. Since the clinical performance data is leveraged from the predicate device, the details about ground truth establishment would reside in the original IDE G080090 submission for the SJM Confirm™ (DM2102).

4. Adjudication Method for the Test Set:

  • This information is not provided in the 510(k) summary. Details on adjudication methods would typically be found in the full clinical study report for the predicate device (IDE G080090).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:

  • No MRMC comparative effectiveness study is mentioned. The Confirm Rx™ ICM is an insertable cardiac monitor that automatically detects and records cardiac events. It is not an AI-assisted diagnostic tool designed to improve human reader performance in interpreting medical images or complex waveforms. Its primary function is automated detection and recording, with physician review of the recorded electrograms (EGMs).

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, the document implies standalone performance evaluation of the detection algorithms. It states that the Confirm Rx™ ICM utilizes "identical sensing and detection algorithms as SJM Confirm™ DM2102." These algorithms are designed for automated triggering of EGM storage based on detected events like tachycardia, pauses, and atrial fibrillation. The verification of R-wave detection (a fundamental aspect of these algorithms) was performed via bench and animal testing, suggesting an evaluation of the algorithm's performance independent of a human in the loop for the core detection task. Physicians later review the automatically captured EGMs.

7. The Type of Ground Truth Used:

  • For the R-wave detection and device migration verification in the Confirm Rx™ ICM, the ground truth would likely be based on direct physiological measurements from bench testing and animal studies (e.g., direct electrical signals or physical tracking of device position).
  • For the leveraged clinical performance of the detection algorithms (from SJM Confirm™ DM2102 IDE G080090), the ground truth for arrhythmia detection would typically involve expert consensus of ECG/EGM interpretation by cardiologists or electrophysiologists, potentially corroborated with other clinical data or findings.

8. The Sample Size for the Training Set:

  • The document does not specify a training set sample size. Given that the device leverages "identical sensing and detection algorithms" from a predicate device, the algorithms were likely developed and "trained" (or more accurately, designed and optimized) at an earlier stage with data that is not detailed in this 510(k) summary. This summary focuses on verifying the new device's implementation of those existing algorithms, not on new algorithm development.

9. How the Ground Truth for the Training Set Was Established:

  • This information is not provided in the 510(k) summary. As mentioned, the algorithms are inherited from a predicate. Details on their initial development and the establishment of ground truth for any underlying "training" or optimization data would be found in the documentation for the predicate device (SJM Confirm™ DM2102).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right, representing the department's focus on health, human services, and well-being. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 29, 2017

St. Jude Medical, Inc. Jennifer Dunham Sr. Regulatory Affairs Specialist 15900 Valley View Ct. Sylmar, California 91342

Re: K163407

Trade/Device Name: Confirm Rx™ Insertable Cardiac Monitor (ICM) System, Model DM3500 Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MXC Dated: August 28, 2017 Received: August 29, 2017

Dear Jennifer Dunham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

M.A. Hillebrand

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163407

Device Name

Confirm RxTM Insertable Cardiac Monitor (ICM) System, Model DM3500

Indications for Use (Describe)

The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arthythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

The Confirm Rx ICM has not been specifically tested for pediatric use.

Type of Use (Select one or both, as applicable)
< Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

K163407

Date Prepared:December 2, 2016
Submitter:St. Jude Medical, Cardiac Rhythm Management Division
Address:15900 Valley View Ct.Sylmar, CA 91342USA
Phone:(818) 362-6822
Contact Person:Jennifer DunhamSr. Regulatory Affairs Specialist(818) 493-2363jdunham02@sjm.comColleen CananSr. Regulatory Affairs Manager(818) 493 2960ccanan@sjm.com
Trade Name/ProprietaryName:Confirm Rx™ Insertable Cardiac Monitor (ICM) System, Model DM3500
Common Name:Insertable Cardiac Monitor
Model Number:DM3500
Classification Name:Medical magnetic tape recorder (21 CFR 870.2800)
Product Code:MXC
Classification:Class II
Pediatric Use:The Confirm Rx ICM has not been specifically tested in pediatricpatients below the age of 18 years.

LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED

510(k) K133481: SJM Confirm TM (Model DM2102) Implantable Cardiac Monitor
510(k) K132649: Medtronic Reveal LINQ TM (Model LNQ11) Insertable Cardiac Monitor

INDICATIONS FOR USE

The Indications for Use for the Confirm Rx™ ICM are as follows:

The Confirm Rx™ ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

The Confirm RxTM ICM has not been specifically tested for pediatric use.

{4}------------------------------------------------

510(k) SUMMARY

PRODUCT DESCRIPTION

The St. Jude Medical Confirm Rx™ system consists of the following key features and components:

  • . Confirm RxTM ICM Model DM3500 Implantable Device: The ICM is intended as a minimally invasive, implantable diagnostic monitoring device, with subcutaneous electrodes, looping memory, and automatic as well as patient-activated EGM storage capability, which help physicians monitor, diagnose, and document patients who are susceptible to cardiac arrhythmias. It is predicated on SJM Confirm™ DM2102 with a two-year longevity, MR conditional labeling, and identical sensing and detection algorithms as SJM Confirm™ DM2102, but with downsized hardware and Bluetooth communication. Specific features include:
    • Patient-initiated triggering of EGM storage using the myMerlin™ Patient App for o mobile devices. This includes capability for the patient to identify symptoms, which are stored with the EGM for physician review.
    • Automated triggering of EGM storage when tachycardia, or pauses are o detected; with physician-programmable values for pause duration, bradycardia rate, tachycardia rate, and number of tachycardia intervals.
    • Automated triggering of EGM storage when atrial fibrillation (AF) is detected, with O physician programmable values for AF duration.
    • The ability to identify EGM anomalies as a consequence of noise or vigorous activity o and inhibit EGM storage as applicable.
    • The addition of remote care capabilities, previously unavailable in the predicate SJM O Confirm™ DM2102 device.
  • Implant Tools: Model DM3520 incision tool and Model DM3510 insertion tool to 트 implant the device subcutaneously. The implantable device is pre-loaded into the insertion tool and packaged together with the incision tool.
    • The DM3520 incision tool is used to make an angled cut, which is the sole incision O required to implant the ICM. The introducer end of the DM3510 insertion tool is inserted, creating the initial pocket (Figure 1).
    • With the insertion tool in place, the plunger is withdrawn to drop the pre-loaded o device into the insertion channel. The plunger is pushed forward to insert the device into the pocket (Figure 1). This completes insertion (implantation) of the ICM, and the incision is closed per standard of care.

{5}------------------------------------------------

K163407 510(k) SUMMARY

Image /page/5/Figure/3 description: The image contains two figures labeled (a) and (b). Both figures show a hand holding a surgical instrument, possibly a scalpel or similar tool, making an incision on what appears to be skin. In figure (b), there is an arrow indicating the direction of movement of the instrument.

Figure 1: Insertion Process

  • Model 3111 Magnet (existing SJM donut magnet) facilitates faster startup of Bluetooth 트 connection and provides user authentication (required for programmer sessions).
  • ' Clinician Programmer (Merlin PCS Programmer Model 3650): The Merlin PCS Programmer 3650 operates using Merlin PCS Model 3330 software and provides the means for the physician to program device parameters and retrieve diagnostic information from the device, including electrograms, heart rate history, episode duration and trend information. The Merlin PCS programmer, using the Model BLU1000 Bluetooth dongle, an off the shelf component, communicates with the Confirm RxTM device with Bluetooth telemetry (also referred to as Bluetooth Low Energy or BLE). Programmer software Model 3330 v23.0.1 and later will contain support for the Confirm RxTM device, adding support for the Model BLU1000 Bluetooth dongle, and new tabs of Implant View and Reason for Monitoring features.
    • Implant View is designed to streamline programming at the time of implant. Upon o initial interrogation at implant, the programmer automatically displays the Implant View in which the user can immediately input device and patient information to be stored onto the device, as well as set the Reason for Monitoring.
    • o Reason for Monitoring allows the user to select from a list of possible conditions for which the patient is receiving the device (such as Syncope. Ventricular Tachycardia, Palpations, etc). The programmer then sets the AF duration parameter and EGM storage priority based on the reason selected. These parameters can be manually adjusted by the user later, if customization is preferred.
  • myMerlin™ Patient App (Model APP1000 Android): The Patient App provides the 트 means for the patient to activate EGM recording in the Confirm Rx™ device, with data pass-through functionality to enable physician follow-up via the Merlin.net Patient Care Network. Patients who do not supply their own mobile device will be provided a Model MTX1000 mobile device. The MTX1000 is an off-the-shelf unit, a Samsung J3, and is not part of the medical device.
  • Remote Care/Clinician Portal (Merlin.net MN5000 Report Generator): The Merlin.net MN5000 system allows physicians to remotely monitor and diagnose patients' cardiac events. The Merlin.net MN5000 v7.5 contains updates that are specific to Confirm RxTM.

{6}------------------------------------------------

K163407 510(k) SUMMARY

Image /page/6/Figure/3 description: This image shows a diagram of an insertable cardiac monitor system. The diagram includes components such as a clinician, programmer, external magnet, ICM insertable cardiac monitor, heart, patient application, clinician portal, patient activator, and patient transmitter. The diagram shows the flow of information between these components, including telemetry communications, sensing, and internet connectivity. The ICM insertable cardiac monitor is connected to the heart for sensing and to the patient application for telemetry communications.

Figure 2: Block Diagram of Confirm Rx™ ICM System

The Confirm RxTM ICM, DM3500, and the myMerlin™ mobile application, APP1000, are the subject of this premarket notification. The Merlin PCS programmer is already FDA approved. The Model 3330 v23.0.1 software for the Model 3650 Merlin PCS programmer and the Merlin.net MN5000 v7.5 software will be submitted under separate RTR PMA/S submissions. The magnet, Model 3111, is Class I exempt MDDS.

TECHNOLOGICAL CHARACTERISTICS

The Confirm RxTM ICM (DM3500) is a smaller version of the predicate SJM Confirm™ ICM (DM2102) device (5.63 x 1.85 x 0.80 cm vs 4.95 x 0.33 cm in dimension) and uses Bluetooth® wireless telemetry instead of inductive telemetry to communicate with external devices, including the Merlin PCS programmer and the myMerlin™M mobile application. In correlation with the reduced size of the Confirm Rx™ ICM, the electrode spacing and surface area is decreased as compared to the SJM Confirm™ ICM (DM2102) (refer to Section 12 for details).

The patient activator and remote monitoring equipment for the predicate SJM Confirm™ ICM (DM2102) device is St. Jude Medical-provided custom hardware (patient activator) using analog telephone connectivity, whereas the remote monitoring equipment for the Confirm RxTM ICM is the myMerlin™ mobile application, installed on a patient's or St. Jude Medical-provided mobile device, using built-in cellular and Wi-Fi connectivity.

Both the predicate SJM Confirm™ ICM (DM2102) and the Confirm Rx™ ICM are MR Conditional.

The Confirm Rx™ ICM firmware including device parameters, features, and detection algorithm are based on the predicate SJM Confirm™ ICM (DM2102) firmware.

Like the predicate SJM Confirm™ ICM (DM2102) device, the Confirm Rx™ ICM is encased in parylene-coated titanium that incorporates two subcutaneous electrodes. The header material on the Confirm RxTM ICM is molded thermoplastic polyurethane (TPU), as is the predicate Medtronic Reveal LINQTM (LNQ11), whereas the header material of the predicate SJM

{7}------------------------------------------------

Confirm™ ICM (DM2102) device is epoxy. The battery chemistry of the predicate SJM Confirm™ ICM (DM2102) device is lithium thionyl chloride. while the battery chemistry of the Confirm Rx™ device is lithium carbon monofluoride, as is the battery chemistry of the predicate Medtronic Reveal LINQ™ (LNQ11) device.

The implant tools, an incision and an insertion tool, are similar to the predicate Medtronic Reveal LINQ™ (LNQ11) system. The incision tool is comprised of polycarbonate and ABS material with a white colorant and has a stainless steel blade. The insertion tool consists of a housing and plunger component comprised of the same polycarbonate and ABS material with white colorant as the incision tool. Both the incision tool and insertion tool are external communicating devices with limited (< 24 hour) contact with tissue or bone.

Performance data is provided to support the determination of substantial equivalence.

SUBSTANTIAL EQUIVALENCE

The Confirm Rx™ ICM is substantially equivalent to the SJM Confirm™ (DM2102) ICM (K133481), and the Medtronic Reveal LINQ™ (LNQ11) ICM (K132649). The subject and the predicate devices are cardiovascular monitoring devices used to record and play back physiological signals. The indications for use for the Confirm RxTM ICM are substantially equivalent to the indications for use for the predicate devices. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the Confirm RxTM ICM system is substantially equivalent to the predicate devices.

TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

All necessary testing was conducted on the Confirm Rx™ ICM to support a determination of substantial equivalence to the predicate devices, including:

  • Software/Firmware Verification and System Validation ●
  • . Cybersecurity
  • . Bluetooth / coexistence
  • Performance Testing (Animal GLP Study)
  • Biocompatibility
  • Sterilization
  • Shelf Life
  • Packaging
  • Electromagnetic compatibility (EMC)
  • Electrical safety ●
  • MRI compliance
  • Mechanical performance
  • Usability testing

The Confirm Rx™ ICM leverages the existing market cleared SJM Confirm™ (DM2102) algorithms and functionality for which the clinical testing, performed on the SJM Confirm™ (DM2102) device per St. Jude Medical IDE G080090, is applicable and was not repeated.

{8}------------------------------------------------

510(k) SUMMARY

R-wave detection and device migration were verified via bench and animal testing for the Confirm Rx™ ICM DM3500.

The results of testing show that the Confirm Rx™ ICM performs as intended and is safe for its intended use.

CONCLUSION

The Confirm Rx™ ICM is an implantable cardiovascular monitoring device and shares its design and mechanism of action with the identified predicate devices. The results of testing demonstrate that the Confirm Rx™ ICM functions to its specifications, performs as intended, and exhibits the appropriate characteristics of an implantable cardiovascular monitoring device. The Confirm Rx™ ICM is substantially equivalent to the predicate devices in terms of technological characteristics, intended use, and performance.

SUMMARY

The Confirm Rx™ ICM System is substantially equivalent to the predicate devices.

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).