K Number
K163252

Validate with FDA (Live)

Date Cleared
2017-01-10

(53 days)

Product Code
Regulation Number
892.1715
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Mammomat Fusion system generates digital mammographic images that can be used for screening, diagnosis and stereotactic biopsies of the breast under supervision of medical professionals. The Mammomat Fusion system is intended to be used in the same clinical applications as traditional film-screen mammography systems. The Mammomat Fusion system may also be used for additional diagnostic workup of the breast.

Device Description

Mammomat Fusion is a floor-mounted full field digital mammography system for screening and diagnostic procedures on standing, seated or recumbent patients.
The stereotactic biopsy accessory uses a computer and imaging performed in at least two planes to localize a target lesion (such as a tumor or micro calcifications in the breast) in three-dimensional space and guide the removal of tissue for examination by a pathologist under a microscope.
The biopsy is attached to the Mammomat Fusion examination stand and comprises a control panel for automatic motorized movement of the needle holder to the pre-set position. There are special biopsy compression plates as well as a face shield to protect the patient from movements of the swivel arm.
The display and control of the biopsy unit are performed via the workstation. Biopsy examinations can be executed with vertical and horizontal needle guidance. For stereo exposures the swivel arm of the Mammomat Fusion is tilted +/- 15 degrees to both sides.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Mammomat Fusion with Stereotactic Biopsy, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Bench Test: Targeting Accuracy)Reported Device Performance (Bench Test: Targeting Accuracy)
The needle tip must be no more than +/-1 mm in x, y, z direction from the selected target point.Within 1 mm of target

Note: The document primarily focuses on demonstrating substantial equivalence to predicate devices and adherence to voluntary standards. The "Targeting accuracy" bench test is the only specific performance criteria with an explicit numerical acceptance criterion and reported result.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not specify a "test set" in the context of clinical data or a large dataset of patient images. The performance evaluation discussed relates to a bench test, not a clinical study with a patient-derived test set.
  • Data Provenance: Not applicable as the reported performance comes from bench testing of the device itself, not from retrospective or prospective patient data from specific countries.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • Not applicable. The "ground truth" for the targeting accuracy bench test would be established by the physical measurements during the test, not by expert consensus on clinical data.

4. Adjudication Method for the Test Set

  • Not applicable. The targeting accuracy bench test would have defined measurement protocols, not an adjudication method for conflicting interpretations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Information

  • No mention of an MRMC comparative effectiveness study being performed. The document focuses on demonstrating substantial equivalence of the device's technical specifications and safety to predicate devices.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Performance

  • Not applicable. The device is a full-field digital mammography system with a stereotactic biopsy attachment. Its performance is inherent to the physical system and its software controls, not a standalone algorithm that could be evaluated without human interaction (as a medical professional operates the device).

7. Type of Ground Truth Used

  • For the "Targeting accuracy" bench test, the ground truth is based on physical measurement standards. The accuracy of the needle tip's position relative to a selected target point is determined by direct measurement, not by expert consensus, pathology, or outcomes data.

8. Sample Size for the Training Set

  • Not applicable. This document describes a medical device (hardware and integral software), not a machine learning algorithm that requires a separate "training set" in the conventional sense of AI development. The software capabilities are validated through engineering tests and adherence to standards.

9. How Ground Truth for the Training Set Was Established

  • Not applicable, as there is no mention of a traditional training set for a machine learning algorithm. The "training" for the device's functionality is inherent in its design, manufacturing, and calibration processes, validated through compliance with standards and bench tests.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 10, 2017

Siemens Medical Solutions USA, Inc. % Ms. Denise Adams Regulatory Affairs Specialist 40 Liberty Boulevard, Mail Code 65-1A MALVERN PA 19355

Re: K163252

Trade/Device Name: Mammomat Fusion with Stereotactic Biopsv Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: November 17, 2016 Received: November 18, 2016

Dear Ms. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K163252

Device Name Mammomat Fusion with Stereotactic Biopsy

Indications for Use (Describe)

The Mammomat Fusion system generates digital mammographic images that can be used for screening, diagnosis and stereotactic biopsies of the breast under supervision of medical professionals. The Mammomat Fusion system is intended to be used in the same clinical applications as traditional film-screen mammography systems. The Mammomat Fusion system may also be used for additional diagnostic workup of the breast.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a teal, sans-serif font. The letters are bold and evenly spaced. The background is plain white, which makes the word stand out. The image is likely a logo or a brand name.

Healthcare

510(k) Summary: Mammomat Fusion with Stereotactic Biopsy

Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355

Date Prepared: January 10, 2017

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

1. General Information:

Importer / Distributor: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355

Establishment Registration Number: 2240869

Manufacturing Site: Siemens Healthcare GmbH Siemensstr. 1

91301 Forchheim Germany Establishment Registration Number:

3004977335

2. Contact:

Siemens Medical Solutions USA, Inc. Denise Adams, RAC Regulatory Affairs Specialist 40 Liberty Boulevard Mail Code: 65-1A Malvern, PA 19355-9998, USA Phone: +1.610.448.6139 Mobile: +1.610.944.4686 adams.denise@siemens.com

3. Device Name and Classification:

Trade Name:Mammomat Fusion with Stereotactic Biopsy
Device:Full Field Digital Mammographic X-Ray system
Reg. Medical Specialty:Radiology
Review Panel:Radiology
Product Code:MUE

Siemens Medical Solutions USA, Inc.

{4}------------------------------------------------

Submission Type:Traditional 510(k)
Regulation Number:21 CFR 892.1715
Device Class:2

4. Legally Marketed Predicate Devices:

Primary predicate
Trade Name:Mammomat Inspiration
Device:Full Field Digital Mammographic X-Ray system
Reg. Medical Specialty:Radiology
Review Panel:Radiology
Product Code:MUE
Submission Type:510(k) (K122286, 02/22/2013)
Regulation Number:21 CFR 892.1715
Device Class:2
Reference predicate
Trade Name:Mammomat Fusion
Device:Full Field Digital Mammographic X-Ray system
Reg. Medical Specialty:Radiology
Review Panel:Radiology
Product Code:MUE
Submission Type:510(k) (K151645, 09/14/2015)
Regulation Number:21 CFR 892.1715
Device Class:

5. Device Description:

Mammomat Fusion is a floor-mounted full field digital mammography system for screening and diagnostic procedures on standing, seated or recumbent patients.

The stereotactic biopsy accessory uses a computer and imaging performed in at least two planes to localize a target lesion (such as a tumor or micro calcifications in the breast) in three-dimensional space and guide the removal of tissue for examination by a pathologist under a microscope.

The biopsy is attached to the Mammomat Fusion examination stand and comprises a control panel for automatic motorized movement of the needle holder to the pre-set position. There are special biopsy compression plates as well as a face shield to protect the patient from movements of the swivel arm.

The display and control of the biopsy unit are performed via the workstation. Biopsy examinations can be executed with vertical and horizontal needle guidance. For stereo exposures the swivel arm of the Mammomat Fusion is tilted ± 15° to both sides.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a teal or turquoise color. The letters are evenly spaced and the word is horizontally oriented.

6. Indication for Use:

The Mammomat Fusion system generates digital mammographic images that can be used for screening, diagnosis and stereotactic biopsies of the breast under supervision of medical professionals. The Mammomat Fusion system is intended to be used in the same clinical applications as traditional film-screen mammography systems. The Mammomat Fusion system may also be used for additional diagnostic workup of the breast.

7. Substantial Equivalence:

The Siemens Mammomat Fusion with Stereotactic Biopsy is substantially equivalent to the commercially available Siemens Mammomat Inspiration (primary predicate) (K122286) and Mammomat Fusion (reference predicate) (K151645).

The Mammomat Fusion with Stereotactic Biopsy is the same system as the reference predicate the Mammomat Fusion with the exception of the biopsy feature. The stereotactic biopsy accessory is identical to the stereotactic biopsy accessory used with the primary predicate the Mammomat Inspiration.

AttributesSubject deviceMammomat Fusion withStereotactic BiopsyPrimary predicateK122286Mammomat InspirationRemarks
Stereotactic Biopsy Option
Indications forUseThe Mammomat Fusion systemgenerates digital mammographicimages that can be used forscreening and diagnosis, andstereotactic biopsies of the breastunder supervision of medicalprofessionals. The MammomatFusion system is intended to beused in the same clinicalapplications as traditional film-screen mammography systems.The Mammomat Fusion systemmay also be used for additionaldiagnostic workup of the breast.The Mammomat Inspiration systemis intended for mammographyexams, screening, diagnosis, andstereotactic biopsies under thesupervision of medicalprofessionals.Mammographic images can beinterpreted by either hard copy filmor soft copy workstation.Same forbiopsy
HardwareBiopsy unit slides onto objecttableBiopsy unit slides onto objecttableSame
PatientpositioningStereotactic biopsy can beperformed on seated andrecumbent patientsStereotactic biopsy can beperformed on seated andrecumbent patientsSame
Accuracy+/-1 mm in X, Y and Z axis+/-1 mm in X, Y and Z axisSame
Biopsy volume(vertical needleguidance)50 mm x 40 mm x 110 mm (2" x1.6" x 4.3")50 mm x 40 mm x 110 mm (2" x1.6" x 4.3")Same
Biopsy volume50 mm x 40 mm x 60 mm (2" x50 mm x 40 mm x 60 mm (2" xSame

Table 1: Comparison of the Subiect to the Primary Predicate

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and appear to be a logo or brand name.

AttributesSubject deviceMammomat Fusion withStereotactic BiopsyPrimary predicateK122286Mammomat InspirationRemarks
(lateral needleguidance)1.6" x 2.4"1.6" x 2.4"
Tube swivelrange in stereomode-15° and + 15°-15° and + 15°Same
Biopsycompressionplate withwindow96 mm x 100 mm (3.8" x 3.9")(window size 52 mm x 42 mm(2" x 1.7"), vertical needleguidance96 mm x 100 mm (3.8" x 3.9")(window size 52 mm x 42 mm(2" x 1.7"), vertical needleguidanceSame
Biopsycompressionplate withoutwindow96 mm x 100 mm (3.8" x 3.9")(lateral needle guidance)96 mm x 100 mm (3.8" x 3.9")(lateral needle guidance)Same
Software controlled functions for stereotactic biopsy
WorkflowAutomated workflowAutomated workflowSame
ImageprocessingOpView, the image impressionfor biopsy images andscreening/diagnostic images isidenticalOpView, the image impression forbiopsy images andscreening/diagnostic images isidenticalSame

Table 2: Comparison of the Subject to the Reference Predicate

AttributesSubject deviceMammomat Fusion withStereotactic BiopsyReference predicateK151645 Mammomat FusionRemarks
Indications foruseThe Mammomat Fusion systemgenerates digital mammographicimages that can be used forscreening and diagnosis, andstereotactic biopsies of the breastunder supervision of medicalprofessionals. The MammomatFusion system is intended to beused in the same clinicalapplications as traditional film-screen mammography systems.The Mammomat Fusion systemmay also be used for additionaldiagnostic workup of the breast.The Mammomat Fusion systemgenerates digital mammographicimages that can be used forscreening and diagnosis of thebreast under supervision ofmedical professionals. TheMammomat Fusion system isintended to be used in the sameclinical applications as traditionalfilm-screen mammographysystems. The Mammomat Fusionsystem may also be used foradditional diagnostic workup ofthe breast.Same exceptfor biopsy
System configuration
X-ray StandFloor mounted X-ray systemFloor mounted X-ray systemSame
X-ray5 kW5 kWSame
Generator30kV, 1s30kV, 1sSame
X-ray TubeSame tube but only Tungsten isusedSame tube but only Tungsten isusedSame
AttributesSubject deviceMammomat Fusion withStereotactic BiopsyReference predicateK151645 Mammomat FusionRemarks
Beam LimitingDeviceAutomatic for all sizesAutomatic for all sizesSame
CompressionunitAutomatic and manual operationAutomatic and manual operationSame
Object tableCarbon fiber mammographysupport systemCarbon fiber mammographysupport systemSame
GridReciprocating 5:1 ratioReciprocating 5:1 ratioSame
MagnificationtableMagnification 1.5 and 1.8Magnification 1.5Adding 1.8table
Biopsy unitYesNoReason for510(k)
Powerrequirements208V, 220V, 230V, 240V,277V, ± 10%, single-phase208V, 220V, 230V, 240V,277V, ± 10%, two-phase,50/60 Hz208V, 220V, 230V, 240V,277V, ± 10%, single-phase208V, 220V, 230V, 240V,277V, ± 10%, two-phase, 50/60HzSame
Monitor/Display19" and 21" TFT display19" and 21" TFT displaySame
Software controlled functions
AEC CalculationAEC calculation is done in theAEC calculation is done in theSame
Calculationacquisition workstationacquisition workstation
DetectorControllerSoftwaredetector controller softwaredetector controller software forSame
Softwarefor PaxScan 3023M detectorPaxScan 3023M detector
OperatingSystemWindows 7Windows 7Same
NetworkInterfacesDICOM SendYesYesSame
DICOM PrintYesYesSame
DICOM Query/ RetrieveYesYesSame
DICOMGet work list /MPPSYesYesSame
DICOM DoseStructuredReportYesYesSame
Detector
DetectormanufacturerVarianVarianSame
Detector TFTCesium Iodide (Csl) andamorphous Silicon (a-Si)Cesium Iodide (Csl) andamorphous Silicon (a-Si)Same
Detector size23 cm x 30 cm23 cm x 30 cmSame
AttributesSubject deviceMammomat Fusion withStereotactic BiopsyReference predicateK151645 Mammomat FusionRemarks
Array size2790 x 35802790 x 3580Same
Pixel size83 um x 83 um83 um x 83 umSame

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the word "SIEMENS" in a bold, teal-colored font. The letters are all capitalized and evenly spaced. The word appears to be a logo or brand name.

{8}------------------------------------------------

The Mammomat Fusion with Stereotactic Biopsy has the same intended use, fundamental scientific technology and performance characteristics as the predicates. Documentation is provided to support a claim of substantial equivalence to Siemens' predicate devices the Mammomat Inspiration and the Mammomat Fusion (K122286 and K151645).

8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Devices:

Mammomat Fusion with Stereotactic Biopsy is based on the same mechanical stand as the primary. X-ray generation and control are the same. The Collimator, Compression unit and AEC technology are the same. The image processing algorithms are identical to those of the predicates. Mammomat Fusion with Stereotactic Biopsy will now feature the same biopsy accessory as the primary predicate Mammomat Inspiration.

9. Summary of Non-Clinical Tests:

The Siemens Mammomat Fusion with Stereotactic Biopsy was tested and complies with the voluntary standards listed in the table below:

Table 3: Conformance to Standards

Reference Number, Date and Title of Standard

AAMI ANSI, ES 60601-1: 2005 /( R ) 2012, Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2: 2007. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests

IEC 60601-1-3 2008 + A1: 2013 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment

IEC 62366: 2007, Medical devices - Application of usability engineering to medical devices

ISO 14971:2012, Medical devices - application of risk management to medical devices

IEC 62304: 2006, Medical device software - Software life cycle processes

ISO 10993-1:2009, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process

IEC 60601-2-28: 2010. Medical electrical equipment - Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis -Ed. 1.0

IEC 61223-3-2:2007, Evaluation and routine testing in medical imaging departments -

{9}------------------------------------------------

Healthcare

Reference Number, Date and Title of Standard

Part 3-2: Acceptance tests - Imaging performance of mammographic X-ray equipment

IEC 60601-2-45: 2011, Medical electrical equipment - Part 2-45: Particular requirements for the basic safety and essential performance of mammographic X-ray equipment and mammographic stereotactic devices

NEMA PS 3.1 - 3.20: 2016, Digital Imaging and Communications in Medicine (DICOM) Set

IEC 60336: 2005, Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Characteristics of focal spots

In addition, the following bench tests were also conducted and passed:

TestObjectiveTest MethodAcceptance CriteriaResults
TargetingaccuracyEnsureaccuracyof thebiopsydeviceRemove plastic tray, attachphantom and compress manually.The needletip must beno morethan +/-1mm in x, y,z directionfrom theselectedtarget point.Within 1mm oftarget
Take a scout exposure and a stereopair and check reference points.
Set an accurate target mark thenselect needle. Select target mark
with corresponding needle andneedle holder. Transmit the target
mark to the biopsy unit. Mount the
calibration needle and check the
needle tip position of the
calibration needle.
BiopsyneedleCorrectbiopsyneedlewill beusedPosition biopsy phantom on objecttable and compress. Acquire scoutA messagewill appearthat thetargetcannot bereachedwith thisneedle andanappropriateneedlelength issuggested.Themessagewasreceivedthat thetargetcould notbe reachedwith thatneedle andanappropriateneedlelength wassuggested.
and check that target for biopsy is
within the possible biopsy volume.
Acquire stereo pair and check and
adapt reference marks if
applicable. Set target in both
stereo images and select
calibration needle. Check that
target coordinates and calculated
distance between needle tip and
biopsy table are displayed. Setfurther target at the border of thebiopsy volume. Select a needlethat won't reach the target.

Table 4: Summary of Bench Tests

{10}------------------------------------------------

Healthcare

The following quality assurance measures were applied to the development of the system:

  • Risk Analysis
  • Requirement Specification Reviews
  • Design Reviews
  • Integration testing (System verification)

General Safety and Effectiveness Concerns:

Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Mammomat Fusion with Stereotactic Biopsy is continually monitored and if an error occurs the system functions will be blocked and an error message will be displayed.

Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards. Siemens adheres to recognized and established industry practice and all equipment is subject to final performance testing.

10. Conclusion as to Substantial Equivalence:

The Mammomat Fusion with Stereotactic Biopsy has the same intended use, fundamental scientific technology and performance characteristics as the predicate devices. Therefore the Mammomat Fusion with Stereotactic Biopsy is substantially equivalent to the primary predicate the Mammomat Inspiration and the reference predicate the Mammomat Fusion.

§ 892.1715 Full-field digital mammography system.

(a)
Identification. A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”See § 892.1(e) for the availability of this guidance document.