K Number
K163106
Date Cleared
2016-12-01

(24 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CODMAN Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.

Device Description

The CODMAN Integrated Bipolar Cord and Tubing Set consists of an inlet spike with protective cap, an on/off clamp, silicone tubing and fittings, a male luer connector, two plugs, and a female socket connector. The set is packaged sterile for single use with a nonpyrogenic fluid pathway.

AI/ML Overview

The provided documentation is a 510(k) Premarket Notification for a medical device called the "Codman Integrated Bipolar Cord and Tubing Set." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy through clinical studies, especially for Class II devices.

Therefore, the information regarding acceptance criteria and a "study that proves the device meets the acceptance criteria" will be focused on the performance testing conducted to demonstrate substantial equivalence, rather than a clinical effectiveness study. The "acceptance criteria" here refer to the passing criteria for the various technical and safety tests performed.

Here's the breakdown of the information requested, based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Bench Testing
Electrical Functionality Test - Continuity and ResistancePass
Irrigation Functionality Test - Leak (Dry)Pass
Mating Part - Part 1: PinsPass
Mating Part - Part 2: SocketPass
Mating Part - Part 3: LuerPass
Mating Part - Part 4: SpikePass
Pull TestingPass
Irrigation Simulated Use TestPass
Simulated Use Pull TestPass
Electrical Safety and EMC per IEC 60601
High Frequency Leakage Current - 60601-1 and 60601-2-2Pass
High Frequency Dielectric Strength - 60601-1 and 60601-2-2Pass
Main Frequency Dielectric Strength - 60601-1 and 60601-2-2Pass
Cable Flexing - 60601-1 and 60601-2-2Pass
Electromagnetic Compatibility - 60601-1-2Pass
Sterilization
Bioburden ValidationPass
Bioburden TestPass
EO/ECH ResidualsPass
LAL ValidationPass
Sterility Assurance Level (SAL)10⁻⁶ (Pass)
Shelf Life - Functionality Testing Complete after One Year Accelerated Aging
Electrical Functionality Test - Continuity and ResistancePass
Irrigation Functionality Test - Leak (Dry)Pass
Mating Part - Part 1: PinsPass
Mating Part - Part 2: SocketPass
Mating Part - Part 3: LuerPass
Mating Part - Part 4: SpikePass
Pull TestingPass
Device and packaging remain functional and maintain sterility for 1 yearAchieved through accelerated aging
Biocompatibility
Cytotoxicity - ISO MEM ElutionPass
Delayed Type Sensitization - ISO Maximization in Guinea PigsPass
Intracutaneous Reactivity - ISO Intracutaneous Study in RabbitsPass
Acute Systemic Toxicity - ISO Systemic Toxicity in MicePass
Pyrogenicity Testing - USP Rabbit Pyrogen Study, Material MediatedPass
Hemolysis - ASTM F756 (Extract only)Pass
Chemical Characterization of Extractables (ISO 10993-18)Low level of extractables
Toxicology Risk AssessmentExtractables do not represent a significant toxicological risk to patients

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document consistently refers to "final sterile devices" being tested across various performance categories. However, it does not specify the exact sample sizes (e.g., number of units, batches) used for each individual test (Bench, Electrical Safety, Sterilization, Shelf Life, Biocompatibility).

The data provenance is internal testing performed by the manufacturer (Medos International SARL, based in Le Locle, Switzerland, or its affiliate Codman & Shurtleff, Inc. in Raynham, MA, USA). The testing is prospective as it was conducted specifically to support this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and therefore not provided in the document. This is a medical device for electrosurgical cutting and coagulation, and the testing performed is technical and safety-related (e.g., electrical, mechanical, biological). It does not involve human interpretation or expert evaluation to establish a "ground truth" in the way a diagnostic AI device would.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and therefore not provided in the document. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations, typically in diagnostic studies. As mentioned above, the testing involved in this 510(k) is technical and objective.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an accessory (bipolar cord and tubing set) for electrosurgical equipment. Its evaluation for substantial equivalence relies on physical, electrical, and biological safety and performance testing, not on human reader performance with or without AI assistance. The document explicitly states: "No clinical studies were performed as appropriate verification and validation of the modified device was achieved from the results of the bench testing."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This concept is not applicable to this device. This is a physical medical accessory, not an AI algorithm. Therefore, there is no "standalone algorithm only" performance to evaluate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is established by objective engineering specifications, material standards, and compliance with recognized industry standards (e.g., ISO, IEC, ASTM, ANSI/AAMI, USP). For example:

  • Bench Testing: Mechanical integrity, electrical continuity, leak-free operation, proper mating, pull strength are measured against predetermined engineering specifications.
  • Electrical Safety: Compliance with specific limits defined in IEC 60601 standards.
  • Sterilization: Achievement of a specified Sterility Assurance Level (SAL) of 10⁻⁶ validated against microbiological standards (ISO 11135, ISO 11737-1).
  • Biocompatibility: Absence of toxic effects and compliance with ISO 10993 standards and specific biological response criteria in animal models (e.g., no sensitization, no systemic toxicity).

8. The sample size for the training set

This information is not applicable. This is a physical medical device, not an AI or machine learning model. Therefore, there is no "training set" in the context of data used to train an algorithm.

9. How the ground truth for the training set was established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 1, 2016

Medos International SARL Ms. Megan Palumbo Senior Regulatory Affairs Specialist Chemin-blanc 38 2400 Le Locle, Switzerland

Re: K163106

Trade/Device Name: Codman Integrated Bipolar Cord and Tubing Set Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 4, 2016 Received: November 7, 2016

Dear Ms. Palumbo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

Sincerely,

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K163106

Device Name

CODMAN Integrated Bipolar Cord and Tubing Set

Indications for Use (Describe)

The CODMAN Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 – 510(k) Summary

I. Submitter Codman & Shurtleff, Inc. 325 Paramount Drive Raynham, MA 02767

Establishment Registration Number: 1226348

On behalf of: Medos International SARL Chemin-Blanc 38 2400 LeLocle, Switzerland

Establishment Registration Number: 3008114965

Contact: Megan Palumbo Phone: (508) 828-3571 Fax: (508) 977-6979

Date of Submission: November 4, 2016

II. Device

II. DeviceDevice Proprietary NameCodman Integrated Bipolar Cord and Tubing Set
Common NameIrrigation Tubing and Bipolar Cord
Classification NameElectrosurgical, Cutting & Coagulation & Accessories(21 CFR 878.4400)
Regulatory ClassificationII
Product CodeGEI
III. PredicateDeviceThe predicate device for this submission is the Codman Integrated IrrigationTubing and Bipolar Cord Set (K052449), which was cleared on October 17,2005.
IV. DeviceDescriptionThe CODMAN Integrated Bipolar Cord and Tubing Set consists of an inletspike with protective cap, an on/off clamp, silicone tubing and fittings, a maleluer connector, two plugs, and a female socket connector. The set is packaged

V. Indications The CODMAN Integrated Bipolar Cord and Tubing Sets are intended to for Use provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.

sterile for single use with a nonpyrogenic fluid pathway.

{4}------------------------------------------------

  • the same intended use, ●
  • use the same fundamental scientific technology, .
  • incorporates the same basic cord and tube set design, ●
  • . incorporates similar materials,
  • packaged and sterilized using the same packaging design, materials, . and processes.

The minor differences between the predicate and subject device are:

  • a slight difference in device design, .
  • minor formulation changes to the existing materials (material types are . remaining the same), and
  • . change in shelf life.

The table below details the comparison of the predicate and subject devices.

Comparison of the Predicate and Subject Device
Predicate Device:Codman Integrated Irrigation Tubingand Bipolar Cord Set (K052449)Subject Device:Codman Integrated Bipolar Cord and TubingSet (This Submission)
FDA Regulatory Information
ManufacturerCodman & Shurtleff, Inc.Same as predicate
FDA ProductCodeGEISame as predicate
ClassificationClass II - 21 CFR 878.4400Same as predicate
ClassificationNameElectrosurgical, Cutting & Coagulation &AccessoriesSame as predicate
Indications for Use
Indications forUseThe CODMAN Integrated Bipolar Cordand Tubing Sets are intended to provideirrigation and energy simultaneously tobipolar forceps specifically designed forirrigation.[They are intended for use with theCodman/Malis CMC-II and theCodman/Malis CMC III I.E.C. IrrigationModules and the Codman/Malis BipolarCoagulators.]Same as predicate
Device Design
Single UseYesSame as predicate
Device DesignIntegrated disposable bipolar cord with anirrigation tube.Same as predicate
Comprised of irrigation inlet spike withprotective cap, male luer connector, femaleSame as predicate
Comparison of the Predicate and Subject Device
Predicate Device:Subject Device:
Codman Integrated Irrigation Tubingand Bipolar Cord Set (K052449)Codman Integrated Bipolar Cord and TubingSet (This Submission)
Devices for whichthis Accessory isto be used withsocket connector, on/off clamp, siliconetubing and fittings, and two bipolargenerator plug components.They are intended for use with theCodman/Malis CMC-II and theCodman/Malis CMC III I.E.C. IrrigationModules and the Codman/Malis BipolarCoagulators.Same as predicate
Device ColorBipolar CordTransparent tubing surrounding coppercord with white plugs and socketIrrigation TubingTransparent tubing and connectors withwhite inlet spikeSame as predicate
Device Materials
SpikeABSABS
PVC TubingPVCPVC
Tubing AdaptorPVCGravity Feed:PVCRotary Pump:ABS
Silicone TubingSiliconeSilicone
Extruded PVC/Copper CablePVC and CopperPVC and Copper
Male Slip LuerPVCPVC
BushingPVCPVC
Spike CapPolyethylenePolyethylene
Tubing ClampPolyethylene, Polysterene, PolypropyleneCelanese and Polyacetal
Male Slip LuerCapPolyethylenePolyethylene
Banana Plug Pinnickel plate brassnickel plate brass
Banana PlugOvermoldPVCPVC
Socketnickel plated brasstin plated brass
Socket OvermoldPVCPVC
Other
Electrical SafetyTestingTesting in accordance with:• ANSI/AAMI HF18:2001Testing in accordance with• IEC 60601-1• IEC 60601-1-2• IEC 60601-2-2
PackagingConfiguration(Sterile Barrier)Peel PouchSame as predicate
Comparison of the Predicate and Subject Device
Predicate Device:Codman Integrated Irrigation Tubingand Bipolar Cord Set (K052449)Subject Device:Codman Integrated Bipolar Cord and TubingSet (This Submission)
PackagingMaterials(Sterile Barrier)Transparent bottom stock formable film-thermoform and 1073B Tyvek top stockthermoformSame as predicate
Packaging Unit10 units per boxSame as predicate
SterilityAssurance Level(SAL)10-6Same as predicate
SterilizationMethodEthylene oxideSame as predicate
Shelf Life5 years1 year

{5}------------------------------------------------

{6}------------------------------------------------

VII. Performance Data

The following performance testing has been conducted in support of the substantial equivalence determination. All testing was performed on final sterile devices unless otherwise specified.

Please see the Summary of Testing table below for a summary of testing and testing results. Additional information regarding the Bench, Electrical Safety, Sterilization, Shelf Life and Biocompatibility Testing follows.

Summary of Testing
TestResult
Bench Testing
Electrical Functionality Test - Continuity and ResistancePass
Irrigation Functionality Test - Leak (Dry)Pass
Mating Part - Part 1: PinsPass
Mating Part - Part 2: SocketPass
Mating Part - Part 3: LuerPass
Mating Part - Part 4: SpikePass
Pull TestingPass
Irrigation Simulated Use TestPass
Simulated Use Pull TestPass
Electrical Safety and EMC per IEC 60601
High Frequency Leakage Current - 60601-1 and 60601-2-2Pass
High Frequency Dielectric Strength - 60601-1 and 60601-2-2Pass
Main Frequency Dielectric Strength - 60601-1 and 60601-2-2Pass
Cable Flexing - 60601-1 and 60601-2-2Pass
Electromagnetic Compatibility - 60601-1-2Pass
Sterilization
Bioburden ValidationPass
Bioburden TestPass
EO/ECH ResidualsPass
LAL ValidationPass

{7}------------------------------------------------

Shelf Life - Functionality Testing Complete after One Year Accelerated Aging
Electrical Functionality Test - Continuity and ResistancePass
Irrigation Functionality Test - Leak (Dry)Pass
Mating Part - Part 1: PinsPass
Mating Part - Part 2: SocketPass
Mating Part - Part 3: LuerPass
Mating Part - Part 4: SpikePass
Pull TestingPass
Biocompatibility
Cytotoxicity - ISO MEM ElutionPass
Delayed Type Sensitization - ISO Maximization in GuineaPigsPass
Intracutaneous Reactivity - ISO Intracutaneous Study inRabbitsPass
Acute Systemic Toxicity - ISO Systemic Toxicity in MicePass
Pyrogenicity Testing - USP Rabbit Pyrogen Study, MaterialMediatedPass
Hemolysis - ASTM F756 (Extract only)Pass
Chemical Characterization of Extractables (ISO 10993-18)Low level of extractables
Toxicology Risk AssessmentExtractables do not representa significant toxicological riskto patients

Bench Testing

Bench testing data demonstrated that the Codman Integrated Bipolar Cord and Tubing Set performed according to its description and intended use, and established the performance characteristics of this device. The Codman Integrated Bipolar Cord and Tubing Set passed each bench test conducted, as compared to the device characteristics and performance specifications of the predicate device. Clinical testing was not required to establish substantial equivalence.

Results of verification and validation testing conducted on Codman Integrated Bipolar Cord and Tubing Set demonstrated that the proposed device performed as designed, is suitable for the intended use, and is substantially equivalent to the predicate device.

Performance bench testing was conducted in alignment with FDA's guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements.

Electrical Safety and Electromagnetic Compatibility Testing

Electrical safety testing was conducted in accordance with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and IEC 60601-2-2:2009/C1:2014 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. Electromagnetic Compatibility Testing

{8}------------------------------------------------

(EMC) was conducted in accordance with IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests. The Codman Integrated Bipolar Cord and Tubing Set passed all electrical safety and EMC testing.

Sterilization

The Codman Integrated Bipolar Cord and Tubing Sets are sterilized via ethylene oxide. The Cord and Tubing Sets have been validated to ensure a sterility assurance level (SAL) of 10t in accordance with ISO 11135:2014 Sterilization of health care products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices and ISO 11737-1:2006, "Sterilization of medical devices -Microbiological methods - Part 1: Determination of a population of microorganisms on products." Sterilization testing also demonstrated that ethylene oxide residuals can be reduced to an acceptable level in accordance with ISO 10993-7:2008 Biological Evaluation of Medical Devices - Ethylene Oxide Sterilization Residuals and the proposed device can be successfully adopted into Codman's existing sterilization cycle.

Shelf-Life Testing

Shelf-life testing was conducted in accordance with FDA's guidance document Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016 and internal requirements. The Codman Integrated Bipolar Cord and Tubing Set was subjected to accelerated aging. The aging studies established that the device and packaging remain functional and maintain sterility for 1 year.

Biocompatibility Testing

Biocompatibility with ISO 10993-1:2009/AC:2010, Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process, and FDA's guidance documents, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued June 16, 2016 and Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued August 15, 2016.

Animal Studies

No animal studies were performed as appropriate verification and validation of the modified device was achieved from the results of the bench testing, biocompatibility evaluation, and electrical/safety testing.

Clinical Studies

No clinical studies were performed as appropriate verification and validation of the modified device was achieved from the results of the bench testing.

{9}------------------------------------------------

biocompatibility evaluation, and electrical/safety testing.

VIII. Based upon the same indications for use, intended use, fundamental scientific Conclusion technology, material types, comparison to the predicate device, and testing conducted, it is concluded that the subject device, the Codman Bipolar Cord and Tubing Set, is substantially equivalent to the predicate device, the Codman Integrated Irrigation Tubing and Bipolar Cord Set (K052449), and therefore does not raise any new issues of safety and effectiveness.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.