(271 days)
VITEK® MS is a mass spectrometry system using matrix-assisted laser desorption/ionization – time of flight mass spectrometry (MALDI-TOF MS) for the identification of microorganisms cultured from human specimens.
The VITEK® MS is a qualitative in vitro diagnostic device in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial, yeast and mould infections.
VITEK® MS is mass spectrometry system using matrix-assisted laser desorption/ionization-time of flight mass spectrometry (MALDI-TOF MS) for the identification of microorganisms cultured from human specimens. The VITEK® MS is a qualitative in vitro diagnostic device indicated for use in coniunction with other clinical and laboratory findings to aid in the diagnosis of bacterial, yeast and mould infections.
The VITEK® MS is intended for laboratory use by professional users who are trained in microbiology and good laboratory practices.
This 510(k) is an update to the VITEK® MS (Mass Spectrometry) clinical knowledge base (KB v3.0.0) for the purposes of identifying Mycobacterium, Nocardia, and mould isolates. As the VITEK® MS KB v3.0.0 update includes new indications for use on the VITEK® MS system, new clinical data was required to establish safety and effectiveness. To account for the detection of higher mass peaks, relevant for some moulds and Mycobacterium, the VITEK® MS acquisition station software was optimized (in v1.5.0) .
Microorqanism identifications are are made via matrix-assisted laser desorption/ionization - time of flight mass spectrometry (MALDI-TOF MS) technology, which includes the three basic principles of ionization, separation, and detection. Depending on the isolate culture, the analyte sample may be directly spotted to a target slide, or for Mycobacterium, Nocardia and mould it must be processed/inactivated before adding to the target slide. Once spotted to the target slide, a matrix is added for the purpose of easy sublimation and strong absorbance in the laser wavelength employed by the instrument.
The slide is then loaded onto the instrument, where a laser targets the sample spot and pulses the isolate spot, resulting in vibrational excitation of matrix and analyte molecules. The matrix transfer protons to the analyte resulting in a positive charge. The ionized molecules are then accelerated in an electromagnetic field and a grid electrode in the ionization chamber. The velocity of the molecules depends on the mass-to-charge (m/z) ratio of the analyte, with heavier molecules having a higher moment of inertia resulting in a lower velocity.
The time of flight is measured precisely by the ions arrival at a particle detector. Based on the time of flight, the m/z ratio of each particle can be determined, and a mass spectrum of the analyte sample mixture is generated. The mass spectrum displays results as a series of peaks (spectrum) which correspond to the ionized proteins derived from the analyte sample. The mass spectra are sufficiently distinctive to allow taxonomic characterization at the genus and species.
The VITEK® MS system is a mass spectrometry system using matrix-assisted laser desorption/ionization – time of flight mass spectrometry (MALDI-TOF MS) for the identification of microorganisms cultured from human specimens. It is a qualitative in vitro diagnostic device used in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial, yeast, and mold infections. This summary specifically details the performance of the VITEK® MS with an updated clinical knowledge base (KB v3.0.0) for identifying Mycobacterium, Nocardia, and mold isolates.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the VITEK® MS updated with KB v3.0.0 are implied by the reported performance, aiming for high agreement rates and very low error rates when compared against a reference method (molecular sequencing).
| Metric | Acceptance Criteria (Implied) | Reported Device Performance (Clinical Strains, All Organism Groups Combined) |
|---|---|---|
| Agreement Rate (Single ID + Low Discrimination Correct Genus) | High (e.g., >90%) | 94.6% |
| Error Rate | Very low (e.g., <1%) | 0.7% (19/2695) |
| No Identification (No ID) Rate | Acceptable (e.g., <10%) | 4.7% (127/2695) |
| Agreement Rate (Excluding No ID, Single ID + Low Discrimination Correct Genus) | Very High (e.g., >95%) | 99.3% (2549/2568) |
Reported Device Performance for Challenge Strains (Mycobacterium, Nocardia, and Mold Combined):
| Metric | Acceptance Criteria (Implied) | Reported Device Performance (Challenge Strains) |
|---|---|---|
| Agreement Rate (Single ID + Low Discrimination Correct Genus) | High (e.g., >95%) | 96.7% (290/300) |
| Misidentification Rate | 0% desired | 0% |
| No Identification (No ID) Rate | Acceptable (e.g., <5%) | 3.3% (10/300) |
| Agreement Rate (Excluding No ID, Single ID + Low Discrimination Correct Genus) | Very High (e.g., >99%) | 100% (290/290) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Clinical Strains): A total of 2,695 isolates (Mycobacterium, Nocardia, and mold) were tested.
- Sample Size (Challenge Strains): 100 well-characterized challenge strains were included (50 molds, 35 Mycobacterium, and 15 Nocardia).
- Data Provenance: The study was conducted at five clinical sites. The document does not specify the country of origin of the data, but it is implied to be clinical data from human specimens. The data is prospective in the sense that it was collected specifically for this performance testing, but the isolates themselves would have been collected as part of routine clinical practice.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number or qualifications of experts used to establish the ground truth beyond stating that the reference method was "molecular sequencing methods." This suggests that the ground truth was established through laboratory testing rather than human expert consensus on morphological characteristics, for example.
4. Adjudication Method for the Test Set
The document does not explicitly state an adjudication method. The performance evaluation was a direct comparison of the VITEK MS identification results to molecular sequencing methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device identifies microorganisms and does not involve human readers interpreting images or data directly in a way that an MRMC study would typically assess. The performance is assessed on the accuracy of the device's identification against a reference standard.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
Yes, the performance study described is a standalone performance assessment of the VITEK® MS system. The device automatically generates an identification based on MALDI-TOF MS data, and its results are compared directly to a molecular sequencing reference method. While professional users trained in microbiology operate the device and interpret its output in a laboratory setting, the core performance reported (agreement, error, no-ID rates) reflects the algorithm's capability without direct human interpretation influencing the primary identification result itself.
7. The Type of Ground Truth Used
The ground truth used was molecular sequencing methods. This is considered a highly accurate and definitive method for identifying microorganisms at the species level. Clinical strains included samples from both solid and liquid culture media.
8. The Sample Size for the Training Set
The document does not explicitly state the sample size for the training set. It refers to the "VITEK® MS clinical knowledge base (KB v3.0.0)," implying a database of known mass spectra used for identification. The development of this knowledge base would have involved a large and diverse collection of characterized organisms, but the specific numbers for the training set are not provided in this summary.
9. How the Ground Truth for the Training Set was Established
The document does not explicitly detail how the ground truth for the training set (clinical knowledge base) was established. However, for such diagnostic platforms, the ground truth for training data is typically established through a combination of:
- Reference methods: Including molecular sequencing (e.g., 16S rRNA gene sequencing for bacteria, ITS sequencing for fungi), conventional biochemical tests, and expert microbiological characterization.
- Pure, well-characterized strains: Often sourced from culture collections (e.g., ATCC).
- Clinical isolates: Confirmed by multiple reliable methods.
The knowledge base would be built by acquiring mass spectral profiles from these ground-truthed organisms.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2017
BIOMERIEUX, INC. NATHAN HARDESTY MANAGER, REGULATORY AFFAIRS MICROBIOLOGY 595 ANGLUM RD. HAZELWOOD MO 63042
Re: K162950
Trade/Device Name: VITEK MS Regulation Number: 21 CFR 866.3361 Regulation Name: Mass spectrometer system for clinical use for the identification of microorganisms
Regulatory Class: II Product Code: PEX Dated: June 16, 2017 Received: June 19, 2017
Dear Mr. Hardesty:
This letter corrects our substantially equivalent letter of July 22, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
{1}------------------------------------------------
as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kristian M. Roth - $S_{\text{For:}}$
Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162950
Device Name VITEK® MS
Indications for Use (Describe)
VITEK® MS is a mass spectrometry system using matrix-assisted laser desorption/ionization – time of flight mass spectrometry (MALDI-TOF MS) for the identification of microorganisms cultured from human specimens.
The VITEK® MS is a qualitative in vitro diagnostic device in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial, yeast and mould infections.
(See Attached for 'List of Claimed Organisms')
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Indications for Use Attachment
Abiotrophia defectiva Achromobacter denitrificans 1 Achromobacter xylosoxidans ¹ Acinetobacter baumannii complex Acinetobacter haemolyticus Acinetobacter junii Acinetobacter lwoffii Actinomyces meyeri Actinomyces neuii Actinomyces odontolyticus Aerococcus viridans Aeromonas hydrophila/caviae 2 Aeromonas sobria 2 3 Aggregatibacter actinomycetemcomitans Aggregatibacter aphrophilus Aggregatibacter segnis Alcaligenes faecalis ssp faecalis Bacteroides caccae Bacteroides fragilis Bacteroides ovatus 4 Bacteroides thetaiotaomicron Bacteroides uniformis Bacteroides vulgatus Bordetella parapertussis Bordetella pertussis Brevundimonas diminuta Burkholderia multivorans Campylobacter coli Campylobacter jejuni Candida albicans Candida dubliniensis Candida famata Candida glabrata Candida guilliermondii Candida haemulonii Candida inconspicua Candida intermedia Candida kefyr Candida krusei Candida lambica Candida lipolytica Candida lusitaniae Candida norvegensis
Candida parapsilosis Candida pelliculosa Candida rugosa Candida tropicalis Candida utilis Candida zeylanoides Chryseobacterium indologenes Citrobacter amalonaticus Citrobacter braaki 5 Citrobacter freundii 5 Citrobacter koseri Citrobacter youngae 5 Clostridium clostridioforme Clostridium difficile Clostridium perfringens Clostridium ramosum Corynebacterium jeikeium Cronobacter sakazakii Cryptococcus neoformans Edwardsiella hoshinae Edwardsiella tarda Eikenella corrodens Elizabethkingia meningoseptica Enterobacter aerogenes Enterobacter asburiae 6 Enterobacter cancerogenus Enterobacter cloacae 6 Enterococcus avium Enterococcus casseliflavus Enterococcus durans Enterococcus faecalis Enterococcus faecium Enterococcus gallinarum Escherichia coli ' Escherichia fergusonii Escherichia hermannii Ewingella americana Finegoldia magna Fusobacterium necrophorum Fusobacterium nucleatum Gardnerella vaginalis Gemella haemolysans Gemella morbillorum
{4}------------------------------------------------
Granulicatella adiacens Haemophilus influenzae Haemophilus parahaemolyticus Haemophilus parainfluenzae Hafnia alvei Kingella denitrificans Kingella kingae Klebsiella oxytoca Klebsiella pneumoniae Kodamaea ohmeri Lactococcus garvieae Lactococcus lactis ssp lactis 8 Leclercia adecarboxylata Legionella pneumophila Leuconostoc mesenteroides Leuconostoc pseudomesenteroides Listeria monocytogenes Malassezia furfur Malassezia pachydermatis Micrococcus luteus Mobiluncus curtisii Moraxella (Branhamella) catarrhalis Morganella morganii Neisseria cinerea Neisseria gonorrhoeae 9 Neisseria meningitidis Neisseria mucosa 10 Ochrobactrum anthropi Oligella ureolytica Oligella urethralis Pantoea agglomerans Parvimonas micra Pasteurella multocida Pediococcus acidilactici Peptoniphilus asaccharolyticus Peptostreptococcus anaerobius Pluralibacter gergoviae Prevotella bivia Prevotella buccae Prevotella denticola Prevotella intermedia Prevotella melaninogenica Propionibacterium acnes
Proteus mirabilis Proteus penneri 11 Proteus vulgaris 11 Providencia rettgeri Providencia stuartii Pseudomonas aeruginosa Pseudomonas fluorescens Pseudomonas putida Pseudomonas stutzeri Ralstonia pickettii Raoultella ornithinolytica Raoultella planticola Rhizobium radiobacter Rhodotorula mucilaginosa Rothia mucilaginosa Saccharomyces cerevisiae Salmonella group 9 Saprochaete capitata Serratia fonticola Serratia liquefaciens Serratia marcescens Serratia odorifera Sphingobacterium multivorum Sphingobacterium spiritivorum Sphingomonas paucimobilis Staphylococcus aureus Staphylococcus capitis Staphylococcus cohnii ssp cohnii Staphylococcus cohnii ssp urealyticus Staphylococcus epidermidis Staphylococcus haemolyticus Staphylococcus hominis ssp hominis 12 Staphylococcus lugdunensis Staphylococcus saprophyticus Staphylococcus schleiferi Staphylococcus sciuri Staphylococcus simulans Staphylococcus warneri Stenotrophomonas maltophilia Streptococcus agalactiae Streptococcus anginosus Streptococcus constellatus Streptococcus dysgalactiae Streptococcus gallolyticus ssp gallolyticus
{5}------------------------------------------------
Streptococcus infantarius ssp coli Streptococcus infantarius ssp infantarius Streptococcus intermedius Streptococcus mitis/Streptococcus oralis Streptococcus mutans Streptococcus pneumoniae Streptococcus pyogenes Streptococcus salivarius ssp salivarius 13 Streptococcus sanguinis Trichosporon asahii Trichosporon inkin Trichosporon mucoides Vibrio cholerae Vibrio parahaemolyticus Vibrio vulnificus Yersinia enterocolitica Yersinia frederiksenii Yersinia intermedia Yersinia kristensenii Yersinia pseudotuberculosis 14
MYCOBACTERIUM
Mycobacterium abscessus Mycobacterium avium Mycobacterium chelonae Mycobacterium fortuitum group: 15 Mycobacterium alvei Mycobacterium farcinogenes Mycobacterium fortuitum Mycobacterium houstonense Mycobacterium peregrinum Mycobacterium porcinum Mycobacterium senegalense Mycobacterium gordonae Mycobacterium haemophilum Mycobacterium immunogenum Mycobacterium intracellulare Mycobacterium kansasii Mycobacterium lentiflavum Mycobacterium malmoense Mycobacterium marinum Mycobacterium mucogenicum Mycobacterium scrofulaceum Mycobacterium simiae Mycobacterium smegmatis Mycobacterium szulgai
Mycobacterium tuberculosis complex: 16 Mycobacterium africanum Mycobacterium bovis Mycobacterium canettii Mycobacterium microti Mycobacterium pinnipedii Mycobacterium tuberculosis Mycobacterium xenopi
NOCARDIA
Nocardia abscessus Nocardia asteroides Nocardia brasiliensis Nocardia cyriacigeorgica Nocardia farcinica Nocardia nova 17 Nocardia otitidiscaviarum Nocardia paucivorans Nocardia pseudobrasiliensis Nocardia transvalensis Nocardia veterana Nocardia wallacei
MOULDS
Acremonium sclerotigenum Alternaria alternata Aspergillus brasiliensis Aspergillus flavus/oryzae Aspergillus fumigatus Aspergillus lentulus Aspergillus nidulans Aspergillus niger complex Aspergillus sydowii Aspergillus terreus complex 18 Aspergillus calidoustus Aspergillus versicolor Blastomyces dermatitidis Cladophialophora bantiana Coccidioides immitis/posadasii Curvularia hawaiiensis Curvularia spicifera Epidermophyton floccosum Exophiala dermatitidis
{6}------------------------------------------------
Exophiala xenobiotica 19 Exserohilum rostratum 19 Fusarium oxysporum complex Fusarium proliferatum Fusarium solani complex Histoplasma capsulatum Lecythophora hoffmannii Lichtheimia corymbifera Microsporum audouinii Microsporum canis Microsporum gypseum Mucor racemosus complex 20 Paecilomyces variotii complex Penicillium chrysogenum
Pseudallescheria boydii Purpureocillium lilacinum Rasamsonia argillacea complex 21 Rhizopus arrhizus complex Rhizopus microsporus complex Sarocladium kiliense Scedosporium apiospermum Scedosporium prolificans Sporothrix schenckii complex Trichophyton interdigitale Trichophyton rubrum Trichophyton tonsurans Trichophyton verrucosum Trichophyton violaceum 22
-
Achromobacter denitrificans and Achromobacter xylosoxidans identifications should be considered as a slashline result, Achromobacter denitrificans/ Achromobacter xylosoxidans. The final identification should be confirmed before sending the final selection to the VITEK® 2 for AST results
-
Aeromonas hydrophila/caviae and Aeromonas sobria should be considered as an Aeromonas species group identification.
-
In KB V3.0.0, Aeromonas sobria is displayed as a low discrimination result with Aeromonas veronii but only Aeromonas sobria has been clinically validated.
-
In KB V3.0.0. Bacteroides ovatus is grouped in a slashline with Bacteroides xylanisolvens. It is not possible to distinguish between the two species. BB. ovatus is more commonly associated with human infection.
-
Citrobacter freundii, Citrobacter braakii and Citrobacter youngae should be considered as Citrobacter freundii complex.
-
Enterobacter cloacae and Enterobacter asburiae identifications should be considered as a slashline result, Enterobacter cloacae/ Enterobacter asburiae. The final identification should be confirmed before sending the final selection to the LIS or to the VITEK® 2 for AST results.
-
Shigella species and E. coli O157 are identified as Escherichia coli. Confirmatory tests are required to differentiate Escherichia coli from Shigella species or E. coli O157.
8 In KB V3.0.0 Lactococcus lactis is integrated in a new group Lactococcus lactis (with two other subspecies Lactococcus lactis ssp cremoris and Lactococcus lactis ssp hordniae). In the Lactoccus lactis group only subspecies Lactococcus lactis ssp lactis has been clinically validated.
Confirmatory tests recommended for Neisseria gonorrhea and Salmonella species. 9.
-
In KB V3.0.0, Neisseria mucosa is grouped in a slashline with Neisseria sicca. It is not possible to distinguish between the two species. Both species have been associated with human infection.
-
Proteus penneri and Proteus vulgaris identifications should be considered as a slashline result. Proteus penneri/ Proteus vulgaris.
-
In KB V3.0.0 Staphylococcus hominis is integrated in a new group Staphylococus.hominis (with another subspecies S. hominis ssp novobiosepticus). In the Staphylococus.hominis group only subspecies Staphylococcus hominis ssp hominis has been clinically validated.
-
In KB V3.0.0, Streptococcus salivarius ssp salivarius is displayed as a low discrimination result with Streptococcus salivarius ssp thermophilus and Streptococcus vestibularis but only Streptococcus salivarius ssp salivarius has been clinically validated.
{7}------------------------------------------------
-
There is a possibility of cross-identification between Yersinia similis and Yersinia pseudotuberculosis
-
The Mycobacterium fortuitum complex or group displayed in VITEK® MS includes the seven most prominent and closely related species. The VITEK® MS group differs from the ones reported in the literature that may include up to 13 species.
-
The Mycobacterium tuberculosis complex displayed in VITEK® MS includes the four most prevalent species based on different worldwide geographic regions. It does not include two additional species Mycobacterium microti and Mycobacterium pinnipedii as reported in the literature.
-
In KB V3.0.0, Nocardia nova is displayed as a low discrimination result with Nocardia africana but only Nocardia nova has been clinically validated.
-
Both Aspergillus alabamensis and Aspergillus niveus are not identified as Aspergillus terreus complex. No identification is expected with either species.
-
All of the no identifications for this organism were from multiple replicates of the same isolate.
-
Mucor racemosus f. sphaerosporus is not identified as Mucor racemosus complex. No identification is expected. The Mucor racemosus complex is comprised of M. racemosus f. brunneus, M. racemosus f. chibinensis, M. racemosus f. racemosus, and M. racemosus f. sphaerosporus.
-
3 out of the 5 no identifications for this organism were from multiple replicates of the same isolate.
-
All of the discordant identifications for this organism were from multiple replicates of the same isolate.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for bioMérieux, a French multinational biotechnology company. The logo features the company name in a stylized font below a circular graphic. The graphic is bisected with one half solid and the other half containing a series of curved lines. A curved line extends from the top of the image, bisecting the circle, and continuing to the bottom of the image.
510(k) SUMMARY
VITEK® MS
B.
C.
510(k) Submission Information:
| Submitter's Name: | bioMérieux, SA |
|---|---|
| Address: | 3 Route de Port MichaudLa Balme les Grottes, 38390 (France) |
| Contact Person: | Nathan HardestyManager, Regulatory Affairs Microbiology |
| Phone Number: | 314-731-8666 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | October 14, 2016 |
| Device Name: | |
| Formal/Trade Name: | VITEK® MS |
| Classification Name: | 21 CFR 866.3361, Mass spectrometer system forclinical use for the identification of microorganisms |
| Common Name: | VITEK® MS, VITEK MS |
| Predicate Device: | VITEK® MS (DEN130013 / K124067) |
D. 510(k) Summary:
VITEK® MS is mass spectrometry system using matrix-assisted laser desorption/ionization-time of flight mass spectrometry (MALDI-TOF MS) for the identification of microorganisms cultured from human specimens. The VITEK® MS is a qualitative in vitro diagnostic device indicated for use in coniunction with other clinical and laboratory findings to aid in the diagnosis of bacterial, yeast and mould infections.
The VITEK® MS is intended for laboratory use by professional users who are trained in microbiology and good laboratory practices.
This 510(k) is an update to the VITEK® MS (Mass Spectrometry) clinical knowledge base (KB v3.0.0) for the purposes of identifying Mycobacterium, Nocardia, and mould isolates. As the VITEK® MS KB v3.0.0 update includes new indications for use on the VITEK® MS system, new clinical data was required to establish safety and effectiveness. To account for the detection of higher mass peaks, relevant for some
bioMérieux, Inc.
{9}------------------------------------------------
moulds and Mycobacterium, the VITEK® MS acquisition station software was optimized (in v1.5.0) .
Microorqanism identifications are are made via matrix-assisted laser desorption/ionization - time of flight mass spectrometry (MALDI-TOF MS) technology, which includes the three basic principles of ionization, separation, and detection. Depending on the isolate culture, the analyte sample may be directly spotted to a target slide, or for Mycobacterium, Nocardia and mould it must be processed/inactivated before adding to the target slide. Once spotted to the target slide, a matrix is added for the purpose of easy sublimation and strong absorbance in the laser wavelength employed by the instrument.
The slide is then loaded onto the instrument, where a laser targets the sample spot and pulses the isolate spot, resulting in vibrational excitation of matrix and analyte molecules. The matrix transfer protons to the analyte resulting in a positive charge. The ionized molecules are then accelerated in an electromagnetic field and a grid electrode in the ionization chamber. The velocity of the molecules depends on the mass-to-charge (m/z) ratio of the analyte, with heavier molecules having a higher moment of inertia resulting in a lower velocity.
The time of flight is measured precisely by the ions arrival at a particle detector. Based on the time of flight, the m/z ratio of each particle can be determined, and a mass spectrum of the analyte sample mixture is generated. The mass spectrum displays results as a series of peaks (spectrum) which correspond to the ionized proteins derived from the analyte sample. The mass spectra are sufficiently distinctive to allow taxonomic characterization at the genus and species.
Clinical performance testing for he VITEK® MS v3 / KB v3.0.0 was conducted at five sites, for the purpose of an expanded intended use for Mycobacterium, Nocardia, and moulds. Performance testing included a total of 2.695 Mycobacteria. Nocardia and mould isolates, and evaluation of performance was a comparison of the VITEK MS identification results to molecular sequencing methods (i.e. the reference method). Testing Mycobacteria isolates included samples from both solid and liguid culture media.
Clinical strains for all isolates tested from solid and liquid media at all sites combined show an acceptable agreement rate of 94.6% (correct single identification plus a low discrimination correct genus result), as well as with each organism group [moulds (92.7%), mycobacteria (96.5%) and Nocardia (97.9%). A combined very low error rate of 0.7% (19/2695) was obtained with all isolates tested, and the combined no identification rate was acceptable at 4.7% (127/2695).
When excluding No ID results, clinical strains for all isolates tested from solid and liguid media at all sites combined show a high agreement rate of 99.3% (correct single choice identification plus a low discrimination correct genus result), as well as high agreement
{10}------------------------------------------------
with each organism group [moulds (99.1%), mycobacteria (99.6%) and Nocardia (99.2%)]. A very low error rate of 0.7% (19/2568) was obtained with all isolates combined.
One hundred well-characterized challenge strains included 50 moulds, 35 mycobacteria and 15 Nocardia strains. Three trial sites tested the mould challenge strains, and three trial sites tested the mycobacteria and Nocardia challenge strains. 95.7% (287/300) of Challenge results were correct one choice, and 1.0% (3/300) of results were low discrimination correct genus, for a combined agreement of 96.7% (290/300), for the mould, mycobacteria and Nocardia challenqe isolates tested at all sites combined. No misidentifications were obtained, and no identification was obtained with 10 (3.3%) of all challenge isolates tested at all sites combined.
When excluding the No ID Challenge results, 99.0% (287/290) of results were correct one choice, and 1.0% (3/290) of results were low discrimination correct genus, for a combined agreement of 100% (290/290) for the mould, mycobacteria and Nocardia challenge isolates tested at all clinical trial sites combined.
Quality control strains Mycobacterium smegmatis, ATCC 19420, and Nocardia farcinica, ATCC 3308. showed 100% agreement. the quality control strain Aspergillus brasilienss. ATCC 16404, showed 97.5% agreement, and the negative control showed 99.7% agreement.
§ 866.3361 Mass spectrometer system for clinical use for the identification of microorganisms.
(a)
Identification. A mass spectrometer system for clinical use for the identification of microorganisms is a qualitative in vitro diagnostic device intended for the identification of microorganisms cultured from human specimens. The device is comprised of an ionization source, a mass analyzer, and a spectral database. The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(2) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols.
(3) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(4) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(5) Premarket notification submissions must include details on the appropriate end user device training program that will be offered while marketing the device.