(22 days)
Galil Medical's IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures. The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are indicated for use as a cryosurgical tool in the fields of general surgery, dermatology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, oncology, and urology. The systems are designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
The Galil Medical IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles have the following specific indications:
- · Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")
- · Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention)
- · Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.)
- · Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia)
- · General surgery (palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma)
- · ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth)
- · Thoracic surgery (with the exception of cardiac tissue)
- · Proctology (ablation of benign or malignant growths of the anus or rectum, and hemorrhoids)
Galil Medical's IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are sterile, single use, disposable components, when used in conjunction with Galil Medical's Visual-ICE Cryoablation System to perform cryoablative destruction of tissue. The needles are intended to convert high-pressure gas to either a very cold freezing application or to a warm thawing application. The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are disposable 2.1 mm needles that have a sharp cutting tip, a color-coded handle, a gas tube, and a connector. The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles offer handles configured in a 90° angled configuration.
The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles were designed to provide the same cryoablation functionality with the same iceball shapes as compared to Galil Medical's predicate needles, IcePearl 2.1 CX and IceFORCE 2.1 CX (K152133).
The new devices differ from the predicate devices in that the needle shaft is 55 mm longer than the needle shaft of the predicate devices. The longer shaft length was incorporated into the needles to treat deep-seated tumors in patients with a larger body habitus, while still allowing for CT gantry clearance using a 2.1 CX type needle.
The provided text describes the 510(k) summary for the IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles. This document primarily focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a study to prove performance against those criteria in the way one might expect for a novel device or a standalone AI algorithm.
Based on the provided text, the device itself is a physical medical instrument (cryoablation needles), not an AI-powered diagnostic or treatment planning system. Therefore, many of the requested points regarding AI device studies (e.g., sample size for test set, data provenance, ground truth experts, MRMC studies, standalone performance, training set details) are not directly applicable in the context of this 510(k) submission.
However, I can extract information related to the device and its performance testing as presented in the document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of numerical acceptance criteria with corresponding performance values in the way you might see for an AI algorithm (e.g., sensitivity, specificity thresholds). Instead, for a medical device like a cryoablation needle, acceptance is demonstrated by showing "same as predicate" or "meets defined specifications" through performance testing.
| Acceptance Criteria Category (as implied by the 510(k)) | Reported Device Performance (IcePearl 2.1 CX L / IceFORCE 2.1 CX L) |
|---|---|
| Design and Construction | |
| Needle Tip | Same as predicate (IcePearl 2.1 CX and IceFORCE 2.1 CX) |
| Needle Shaft | 55 mm longer than predicate (This is the primary difference) |
| Gas Pathway Tubing | Same as predicate |
| Handle | Same as predicate |
| Needle Connector | Same as predicate |
| Performance and Function | |
| Freezing/Thawing Technology | Same as predicate |
| Function | Same as predicate |
| Freezing Parameters | Same as predicate |
| Thaw Parameters | Same as predicate |
| Track Ablation | Same as predicate |
| Iceball Shapes | Same as predicate |
| Safety and Performance Characteristics | Meets defined specifications and does not raise any new safety or effectiveness issues |
| Indications for Use | Same as predicate |
Study Proving Device Meets Acceptance Criteria:
The study proving the device meets the acceptance criteria is detailed as "Performance testing... with the previously cleared Visual-ICE System Software to verify safety and performance characteristics and to establish substantial equivalence." This testing included "freezing and in vivo performance." The outcome was that "Test results demonstrated that the IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles meet defined specifications and do not raise any new safety or effectiveness issues."
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not specified in the provided text. The testing conducted included "freezing and in vivo performance," implying physical or biological testing, but the number of units tested or subjects in vivo is not given.
- Data Provenance: Not explicitly stated. The testing was described as "in-house requirements" and "according to protocols based on international standards," which suggests laboratory and potentially animal or human (if "in vivo" refers to human trials, which is unlikely for a 510(k) for a device modification of this nature) testing. Given Galil Medical Ltd. is based in Israel, some testing might have been conducted there.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This question is not applicable as the device is a physical cryoablation needle, not a diagnostic or AI device that requires expert ground truth for interpretation. Ground truth for its performance would be physical measurements of iceball size, temperature, consistency, and in vivo effects, not expert medical opinion on images or data.
4. Adjudication method for the test set:
Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth from multiple human readers for AI diagnostic studies. The performance of a physical device like a cryoablation needle is evaluated by direct measurement and observation, not by adjudicating expert opinions.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an AI device.
7. The type of ground truth used:
For a physical medical device like this, the "ground truth" would be the direct physical and biological effects observed during testing. This would include:
- Physical measurements of iceball dimensions, temperatures, and freezing/thawing parameters in simulated environments.
- Histopathological (pathology) assessment of ablated tissue in in vivo studies.
- Direct observation of device function and integrity.
The text vaguely refers to "freezing and in vivo performance," which would be assessed against established physical or biological metrics.
8. The sample size for the training set:
Not applicable. This is not an AI device that requires a training set in the sense of machine learning.
9. How the ground truth for the training set was established:
Not applicable. This is not an AI device.
In summary, the document describes a 510(k) clearance for cryoablation needles that are substantially equivalent to a predicate device, with the primary difference being a longer needle shaft. The "study"
referred to is performance testing (including freezing and in vivo performance) to confirm that the longer needles maintain the same functional characteristics and safety profile as the predicate. The criteria are largely based on equivalence to the predicate and meeting defined internal specifications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure with three faces in profile, stacked on top of each other. The figure is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 11, 2016
Galil Medical Ltd. Ms. Lynne Davies Sr. Manager, Regulatory Affairs 4364 Round Lake Road Arden Hills, Minnesota 55112
Re: K162599
Trade/Device Name: IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: September 15, 2016 Received: September 19, 2016
Dear Ms. Davies:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162599
Device Name
IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles
Indications for Use (Describe)
Galil Medical's IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures. The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are indicated for use as a cryosurgical tool in the fields of general surgery, dermatology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, oncology, and urology. The systems are designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
The Galil Medical IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles have the following specific indications:
- · Urology (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")
- · Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention)
- · Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions,
angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors,
dermatofibromas small hemanglomas, mucocele cysts, multiple warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.)
-
· Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia)
· General surgery (palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma) -
· ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth)
-
· Thoracic surgery (with the exception of cardiac tissue)
-
· Proctology (ablation of benign or malignant growths of the anus or rectum, and hemorrhoids)
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Galil Medical. The logo consists of a stylized letter "G" with a green dot inside, followed by the text "GALILMEDICAL" in a bold, sans-serif font. Below the logo, there is a tagline that reads "Leading Through Innovation" in a smaller, lighter font.
K162599 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Galil Medical Ltd. |
|---|---|
| Tavor Building 1 | |
| Industrial Park, PO Box 224 | |
| Yokneam Industrial Park 2069203 | |
| Israel | |
| Company Contact Person: | Ms. Lynne A Davies |
| Sr. Manager, Regulatory Affairs | |
| Galil Medical Ltd. | |
| Phone: | 651-287-5098 |
| Fax: | 651-287-5099 |
| Email: | lynne.davies@btgplc.com |
| Alternate Contact Person: | Ms. Amy E McKinney |
| VP, Regulatory Affairs and Quality | |
| Galil Medical Ltd. | |
| Phone: | 651-287-5096 |
| Fax: | 651-287-5097 |
| Email: | amy.mckinney@btgplc.com |
| Device Name: | IcePearl 2.1 CX L and IceFORCE 2.1 CX L |
| Cryoablation Needles | |
| Device Classification Name: | Cryosurgical unit and accessories (GEH) |
| 21 CFR 878.4350 | |
| Predicate Devices / Reference | IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation |
| 510(k): | Needles (K152133) |
| Date of Preparation: | September 15, 2016 |
Device Description:
Galil Medical's IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are sterile, single use, disposable components, when used in conjunction with Galil Medical's Visual-ICE Cryoablation System to perform cryoablative destruction of tissue. The needles are intended to convert high-pressure gas to either a very cold freezing application or to a warm thawing application. The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are disposable 2.1 mm needles that have a sharp cutting tip, a color-coded handle, a gas tube, and a connector. The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles offer handles configured in a 90° angled configuration.
The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles were designed to provide the same cryoablation functionality with the same iceball shapes as compared to Galil Medical's predicate needles, IcePearl 2.1 CX and IceFORCE 2.1 CX (K152133).
{4}------------------------------------------------
The new devices differ from the predicate devices in that the needle shaft is 55 mm longer than the needle shaft of the predicate devices. The longer shaft length was incorporated into the needles to treat deep-seated tumors in patients with a larger body habitus, while still allowing for CT gantry clearance using a 2.1 CX type needle. Galil Medical is using the alpha letter "L" identifier as part of the needle name as well as a different needle part numbers to assist users in distinguishing the different shaft length from the predicate devices.
The table below provides a summary comparison of the submitted devices compared to the predicate devices.
| Description of Submitted Device: | Comments related to Predicates: |
|---|---|
| IcePearl 2.1 CX L and IceFORCE 2.1 CX LCryoablation Needles | IcePearl 2.1 CX and IceFORCE 2.1 CX Cryoablation Needles(K152133) |
| Design and Construction | |
| Needle Tip | Same as predicate |
| Needle Shaft | Shaft length is 55 mm longer than predicate |
| Gas Pathway Tubing | Same as predicate |
| Handle | Same as predicate |
| Needle Connector | Same as predicate |
| Performance and Function | |
| Freezing/Thawing Technology | Same as predicate |
| Function | Same as predicate |
| Freezing Parameters | Same as predicate |
| Thaw Parameters | Same as predicate |
| Track Ablation | Same as predicate |
| Indications for Use | Same as predicate |
In summary, the submitted IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles have the same technology, principle of operation and indications for use as the predicate devices.
There are no changes to the Visual-ICE System Software. The previously cleared software is applicable for use with submitted IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles.
Intended Use:
Galil Medical's IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures. The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurology (including cryoanalgesia), thoracic surgery (with the exception of cardiac tissue), ENT, gynecology, proctology, and urology. The systems are designed to destroy tissue by the application of extreme cold temperatures including prostate and kidney tissue, liver metastases, tumors, skin lesions, and warts.
{5}------------------------------------------------
The Galil Medical IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles have the following specific indications:
- . Uroloqy (ablation of prostate tissue in cases of prostate cancer and Benign Prostate Hyperplasia "BPH")
- Oncology (ablation of cancerous or malignant tissue and benign tumors, and palliative intervention)
- Dermatology (ablation or freezing of skin cancers and other cutaneous disorders. Destruction of warts or lesions, angiomas, sebaceous hyperplasia, basal cell tumors of the evelid or canthus area, ulcerated basal cell tumors, dermatofibromas small hemanglomas, mucocele cysts, multiple warts, plantar warts, actinic and seborrheic keratoses, cavernous hemanglomas, perianal condylomata, and palliation of tumors of the skin.)
- Gynecology (ablation of malignant neoplasia or benign dysplasia of the female genitalia)
- General surgery (palliation of tumors of the rectum, hemorrhoids, anal fissures, pilonidal cysts, and recurrent cancerous lesions, ablation of breast fibroadenoma)
- ENT (Palliation of tumors of the oral cavity and ablation of leukoplakia of the mouth)
- Thoracic surgery (with the exception of cardiac tissue)
- . Proctology (ablation of benign or malignant growths of the anus or rectum, and hemorrhoids)
Summary of Performance Data and Substantial Equivalence:
The IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles use the same technology and have the same intended use and method of operation as compared to the predicate devices.
Performance testing was conducted on the IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles with the previously cleared Visual-ICE System Software to verify safety and performance characteristics and to establish substantial equivalence. Testing was conducted according to protocols based on international standards and in-house requirements and included freezing and in vivo performance. Test results demonstrated that the IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles meet defined specifications and do not raise any new safety or effectiveness issues.
Conclusion:
The information and data provided in this Special 510(k) Notification establish that the IcePearl 2.1 CX L and IceFORCE 2.1 CX L Cryoablation Needles are substantially equivalent to the legally marketed predicate devices.
§ 878.4350 Cryosurgical unit and accessories.
(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.