AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medisign GH81 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH82 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH83 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH81 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH82 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH83 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

Device Description

The Medisign GH81, GH82, GH83, GH81 BT, GH82 BT, and GH83 BT Blood Glucose Monitoring System quantitatively measures the glucose in the whole blood sample by using a harmless electrical current produced by a chemical reaction between the glucose in the blood and Flavin adenine dinucleotide-Glucose dehydrogenase (FAD-GDH) on the test strip. This current is proportionally converted to the amount of glucose in the blood sample to display as the blood glucose result. Glucose measurements from this system are reported as plasma equivalents.

The Medisign GH81, GH82, GH83, GH81 BT, GH82 BT, GH83 BT Blood Glucose Monitoring Systems are for over-the-counter use and systems' kit basically consist of a blood glucose meter, blood glucose test strips, a lancing device, sterile lancets, a carrying case, however some kits may not have test strips or a lancing device or sterile lancets. The Medisign GH81. GH82. GH83 Blood Glucose Monitoring System allows to transmit blood glucose test results to a PC via an external USB cable and Medisign GH81 BT, GH82 BT, GH83 BT Blood Glucose Monitoring System can transmit blood glucose test results to a PC via an external USB cable or a Bluetooth Low Energy transmitter in meter.

The Medisign GH81 Blood Glucose Test Strip is for the Medisign GH81 and GH81 BT Blood Glucose Meter. The Medisign GH82 Blood Glucose Test Strip is for the Medisign GH82 and GH82 BT Blood Glucose Meter. The Medisign GH83 Blood Glucose Test Strip is for the Medisign GH83 and GH83 BT Blood Glucose Meter.

Each of test strip box may contain 10 test strips or 25 test strips or 50 test strips with one vial or 50 test strips split into two vials. Also, test strip box may contain 25 or 50 test strips packed separately for each test strip.

The Medisign Link Diabetes Management Software is an optional software accessory and were previously cleared in K133260.

AI/ML Overview

The provided document is a 510(k) premarket notification for a Blood Glucose Monitoring System. It describes the device, its intended use, comparison to a predicate device, and a high-level summary of performance testing.

However, the document does not contain detailed information about specific acceptance criteria for the device performance beyond stating "the performance of the candidate devices were satisfied with all pre-determined acceptance criteria." It also lacks specifics on the study design, sample sizes for test sets (beyond implicitly mentioning clinical and non-clinical tests), data provenance, number or qualifications of experts, adjudication methods, multi-reader multi-case studies, standalone performance details, type of ground truth used for performance validation, training set details, or how ground truth for the training set was established.

Therefore, I cannot provide a table of acceptance criteria with reported device performance or many of the other requested details because that specific information is not present in the provided text.

Based on the available information, here's what can be extracted:

1. Table of Acceptance Criteria and Reported Device Performance

  • Acceptance Criteria: Not explicitly detailed in the provided text. The document generally states "the performance of the candidate devices were satisfied with all pre-determined acceptance criteria."
  • Reported Device Performance: Not explicitly detailed with specific metrics and thresholds in the provided text.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified. The document only mentions "clinical and non-clinical tests were conducted."
  • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not provided in the document.

4. Adjudication method for the test set

  • This information is not provided in the document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance.

  • This device is a Blood Glucose Monitoring System, which is an in-vitro diagnostic device, not an AI or imaging-based device involving human readers or interpretation of complex cases. Therefore, an MRMC study or assessment of human reader improvement with AI assistance is not applicable and not mentioned.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This device is a standalone blood glucose monitoring system. The performance described would inherently be the "standalone" performance of the device itself (meter + test strip). The document implies this by stating the device was "compared with the YSI 2300 STAT Plus as a reference instrument to evaluate the system reliability and accuracy performance."

7. The type of ground truth used

  • The document states: "The candidate devices were compared with the YSI 2300 STAT Plus as a reference instrument to evaluate the system reliability and accuracy performance."
    • Ground Truth Type: Comparison to a Reference Instrument (YSI 2300 STAT Plus). This is a common method for establishing ground truth in glucose measurement studies, where a highly accurate laboratory analyzer serves as the comparator.

8. The sample size for the training set

  • This information is not provided in the document, as the document describes validation testing against a reference, not the training of an algorithm in the AI sense.

9. How the ground truth for the training set was established

  • This information is not provided in the document, as the document describes validation testing against a reference, not the training of an algorithm in the AI sense.

In summary, the provided text offers a high-level overview of the device and its regulatory submission but lacks the detailed performance study results and methodology typically found in a full study report or technical documentation.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is black, and the text is in a simple, sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 7, 2017

TIANJIN EMPECS MEDICAL DEVICE CO., LTD. C/O PRISCILLA CHUNG LK CONSULTING GROUP USA, INC. 690 ROOSEVELT IRVINE, CA 92620

Re: K162564

Trade/Device Name: Medisign GH81 Blood Glucose Monitoring System Medisign GH82 Blood Glucose Monitoring System Medisign GH83 Blood Glucose Monitoring System Medisign GH81 BT Blood Glucose Monitoring System Medisign GH82 BT Blood Glucose Monitoring System Medisign GH83 BT Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345

Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR Dated: May 4, 2017 Received: May 5, 2017

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K162564

Device Name

Medisign GH81 Blood Glucose Monitoring System

Indications for Use (Describe)

The Medisign GH81 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH81 Blood Glucose Monitoring System is comprised of the Medisign GH81 Blood Glucose Meter and Medisign GH81 Blood Glucose Test Strips.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K162564

Device Name

Medisign GH82 Blood Glucose Monitoring System

Indications for Use (Describe)

The Medisign GH82 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH82 Blood Glucose Monitoring System is comprised of the Medisign GH82 Blood Glucose Meter and Medisign GH82 Blood Glucose Test Strips.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K162564

Device Name

Medisign GH83 Blood Glucose Monitoring System

Indications for Use (Describe)

The Medisign GH83 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH83 Blood Glucose Monitoring System is comprised of the Medisign GH83 Blood Glucose Meter and Medisign GH83 Blood Glucose Test Strips.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{5}------------------------------------------------

Indications for Use

510(k) Number (if known) K162564

Device Name

Medisign GH81 BT Blood Glucose Monitoring System

Indications for Use (Describe)

The Medisign GH81 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH81 BT Blood Glucose Monitoring System is comprised of the Medisign GH81 BT Blood Glucose Meter and Medisign GH81 Blood Glucose Test Strips.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

Indications for Use

510(k) Number (if known) K162564

Device Name

Medisign GH82 BT Blood Glucose Monitoring System

Indications for Use (Describe)

The Medisign GH82 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH82 BT Blood Glucose Monitoring System is comprised of the Medisign GH82 BT Blood Glucose Meter and Medisign GH82 Blood Glucose Test Strips.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{7}------------------------------------------------

Indications for Use

510(k) Number (if known) K162564

Device Name

Medisign GH83 BT Blood Glucose Monitoring System

Indications for Use (Describe)

The Medisign GH83 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (invitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steadystate times (when glucose is not changing rapidly).

The Medisign GH83 BT Blood Glucose Monitoring System is comprised of the Medisign GH83 BT Blood Glucose Meter and Medisign GH83 Blood Glucose Test Strips.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{8}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92

1.510(k) Number

K162564

2. Date Prepared

June 06, 2017

3. Applicant

Tianjin Empecs Medical Device Co., Ltd. No.35 and 37, Yingcheng street, Hangu, Binhai New Area, Tianjin, 300480 China Contact: Eric.Chae, Senior Regulatory Affairs Manager Phone: +86-22-2569-5007 Fax: +86-22-2569-4461 e-mail: eric@empecs.com

4. Correspondent

Priscilla Chung LK Consulting Group USA, Inc. 690 Roosevelt, Irvine CA 92606 Tel. 714-202-5789 Email: juhee.c@lkconsultinggroup.com

5. Measurement

Glucose in fresh capillary whole blood

6. Type of Test

Quantitative, Amperometric detection, Flavin adenine dinucleotide-Glucose dehydrogenase (FAD-GDH)

7. Device Name

Medisign® GH81 Blood Glucose Monitoring System Medisign® GH82 Blood Glucose Monitoring System Medisign® GH83 Blood Glucose Monitoring System

Medisign® GH81 BT Blood Glucose Monitoring System Medisign® GH82 BT Blood Glucose Monitoring System Medisign® GH83 BT Blood Glucose Monitoring System

{9}------------------------------------------------

8. Common Classification Name

Blood Glucose Test System

9. Classification

Product CodeClassificationRegulation SectionPanel
NBW - System, Test, Blood Glucose,Over The CounterClass II21 CFR § 862.134575-Chemistry
LFR - Glucose dehydrogenase, glucosetest systemClass II21 CFR § 862.134575-Chemistry

10. Predicate Device Information

Device Name: OneTouch Verio Flex Blood Glucose Monitoring System 510(k) Number: K150214

11. Indications for Use

Medisign®GH81 Blood Glucose Monitoring System

The Medisign GH81 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (in-vitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steady-state times (when glucose is not changing rapidly).

The Medisign GH81 Blood Glucose Monitoring System is comprised of the Medisign GH81 Blood Glucose Meter and Medisign GH81 Blood Glucose Test Strips.

Medisign®GH82 Blood Glucose Monitoring System

The Medisign GH82 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and It is intended for self-testing outside the body (in-vitro diagnostics use only) by people forearm. with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steady-state times (when glucose is not changing rapidly).

The Medisign GH82 Blood Glucose Monitoring System is comprised of the Medisign GH82 Blood Glucose Meter and Medisign GH82 Blood Glucose Test Strips.

Medisign®GH83 Blood Glucose Monitoring System

The Medisign GH83 Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (in-vitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis

{10}------------------------------------------------

of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steady-state times (when glucose is not changing rapidly).

The Medisign GH83 Blood Glucose Monitoring System is comprised of the Medisign GH83 Blood Glucose Meter and Medisign GH83 Blood Glucose Test Strips.

Medisign® GH81 BT Blood Glucose Monitoring System

The Medisign GH81 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (in-vitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steady-state times (when glucose is not changing rapidly).

The Medisign GH81 BT Blood Glucose Monitoring System is comprised of the Medisign GH81 BT Blood Glucose Meter and Medisign GH81 Blood Glucose Test Strips.

Medisign® GH82 BT Blood Glucose Monitoring System

The Medisign GH82 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (in-vitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steady-state times (when glucose is not changing rapidly).

The Medisign GH82 BT Blood Glucose Monitoring System is comprised of the Medisign GH82 BT Blood Glucose Meter and Medisign GH82 Blood Glucose Test Strips.

Medisign® GH83 BT Blood Glucose Monitoring System

The Medisign GH83 BT Blood Glucose Monitoring System is intended for the quantitative measurement of the glucose in fresh capillary whole blood drawn from fingertip, palm and forearm. It is intended for self-testing outside the body (in-vitro diagnostics use only) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system is intended for single patient use only and should not be shared. It is not to be used for the diagnosis of or screening of diabetes, or for neonatal use. Alternative site testing (palm and forearm) should be done only during steady-state times (when glucose is not changing rapidly).

The Medisign GH83 BT Blood Glucose Monitoring System is comprised of the Medisign GH83 BT Blood Glucose Meter and Medisign GH83 Blood Glucose Test Strips.

12. Device Description

The Medisign GH81, GH82, GH83, GH81 BT, GH82 BT, and GH83 BT Blood Glucose Monitoring System quantitatively measures the glucose in the whole blood sample by using a harmless electrical current produced by a chemical reaction between the glucose in the blood and Flavin adenine dinucleotide-Glucose dehydrogenase (FAD-GDH) on the test strip. This current is

{11}------------------------------------------------

proportionally converted to the amount of glucose in the blood sample to display as the blood glucose result. Glucose measurements from this system are reported as plasma equivalents.

The Medisign GH81, GH82, GH83, GH81 BT, GH82 BT, GH83 BT Blood Glucose Monitoring Systems are for over-the-counter use and systems' kit basically consist of a blood glucose meter, blood glucose test strips, a lancing device, sterile lancets, a carrying case, however some kits may not have test strips or a lancing device or sterile lancets. The Medisign GH81. GH82. GH83 Blood Glucose Monitoring System allows to transmit blood glucose test results to a PC via an external USB cable and Medisign GH81 BT, GH82 BT, GH83 BT Blood Glucose Monitoring System can transmit blood glucose test results to a PC via an external USB cable or a Bluetooth Low Energy transmitter in meter.

The Medisign GH81 Blood Glucose Test Strip is for the Medisign GH81 and GH81 BT Blood Glucose Meter. The Medisign GH82 Blood Glucose Test Strip is for the Medisign GH82 and GH82 BT Blood Glucose Meter. The Medisign GH83 Blood Glucose Test Strip is for the Medisign GH83 and GH83 BT Blood Glucose Meter.

Each of test strip box may contain 10 test strips or 25 test strips or 50 test strips with one vial or 50 test strips split into two vials. Also, test strip box may contain 25 or 50 test strips packed separately for each test strip.

The Medisign Link Diabetes Management Software is an optional software accessory and were previously cleared in K133260.

{12}------------------------------------------------

13. Comparison with the Predicate DeviceSimilarities
FeaturesPredicate Device (K150214)OneTouch Verio Flex BloodGlucose Monitoring SystemSubject Device (K162564)Medisign GH81, GH82, GH83
Intended UseIntended to quantitativelymeasure glucose (sugar) in freshcapillary whole blood samplesdrawn from fingertip, palm orforearm and in venous wholeblood as an aid in monitoring theeffectiveness of glucose control.Refer to the aforementionedIntended Use
EnzymeFAD-GDHSame
Test PrincipleAmperometrySame
CalibrationPlasma equivalentSame
Test SampleCapillary whole bloodSame
Operating Temperature$50 ~ 104°F$Same
Operating Humidity$10 ~ 90%RH$Same
Hematocrit Range$20 ~ 60%$Same
Measuring Range$20 ~ 600 mg/dL$Same
Memory Capacity500 test resultSame
Differences
Measuring Time5 seconds6 seconds
BatteryOne 3.0V Lithiumbatteries(CR2032)Two 3.0V Lithiumbatteries(CR2032)
Coding of Test StripNon-CodingAuto coding
Sample Volume$0.4 \u03bcl$$0.5 \u03bcl$
Averaging Results7,14,30 and 90 days7,14 and 30 days
Alternative SiteNot AvailablePalm and forearm
Pre/Post-meal flaggingNot availableAvailable
Test Strip EjectorNot availableAvailable
Range IndicatorAvailableNot available
Data DownloadUSB or Bluetooth Low EnergyUSB

13. Comparison with the Predicate Device

{13}------------------------------------------------

Similarities
FeaturesPredicate Device (K150214)OneTouch Verio Flex BloodGlucose Monitoring SystemSubject Device (K162564)Medisign GH81 BT, GH82 BT,GH83 BT
Intended UseIntended to quantitativelymeasure glucose (sugar) in freshcapillary whole blood samplesdrawn from fingertip, palm orforearm and in venous wholeblood as an aid in monitoring theeffectiveness of glucose control.Refer to the aforementionedIntended Use
EnzymeFAD-GDHSame
Test PrincipleAmperometrySame
CalibrationPlasma equivalentSame
Test SampleCapillary whole bloodSame
Operating Temperature50 ~ 104°FSame
Operating Humidity10 ~ 90%RHSame
Hematocrit Range20 ~ 60%Same
Measuring Range20 ~ 600 mg/dLSame
Memory Capacity500 test resultSame
Data DownloadUSB or Bluetooth Low EnergySame
Differences
Measuring Time5 seconds6 seconds
BatteryOne 3.0V Lithiumbatteries(CR2032)Two 3.0V Lithiumbatteries(CR2032)
Coding of Test StripNon-CodingAuto coding
Sample Volume0.4 $\mu$ l0.5 $\mu$ l
Averaging Results7,14,30 and 90 days7,14 and 30 days
Alternative SiteNot AvailablePalm and forearm
Pre/Post-meal flaggingNot availableAvailable
Test Strip EjectorNot availableAvailable

{14}------------------------------------------------

Range IndicatorAvailableNot available
-------------------------------------------

{15}------------------------------------------------

14. Performance Test Summary

The clinical and non-clinical tests were conducted and demonstrated the performance of the candidate devices were satisfied with all pre-determined acceptance criteria. The candidate devices were compared with the YSI 2300 STAT Plus as a reference instrument to evaluate the system reliability and accuracy performance. No adverse events pertaining to the safety and effectiveness occurred during the tests.

15. Performance Test Summary

Design verification and validation tests for the candidate devices having the same intended use, fundamental scientific technology and performance characteristics as the predicate device have demonstrated and confirmed that the performance, safety and effectiveness are satisfied with those of the predicate device. Therefore, the candidate devices can be considered substantially equivalent to the predicate device cleared under K150214.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.