(12 days)
This product is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.
RadiForce GX550 is a monochrome LCD monitor for viewing medical images including those of mammography. The monochrome panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 2,048 x 2,560 pixels (5MP) with a pixel pitch of 0.165 mm.
Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.
There are two model variations, GX550 and GX550-AR. The difference of the two variations is the surface treatment of the display screens; the surface treatment of the GX550 is Anti-Glare (AG) treatment and that of the GX550-AR is Anti-Reflection (AR) coating.
RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce GX550 based on the QC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce GX550.
The document describes a 510(k) premarket notification for the EIZO RadiForce GX550 and GX550-AR 5MP Monochrome LCD Monitors, intended for displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis, and diagnosis by trained medical practitioners, and specially designed for breast tomosynthesis applications.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The submission refers to "pre-defined criteria" but does not explicitly list them. However, it indicates that the device meets these criteria. The performance testing section outlines the measurements taken, implicitly suggesting these are the aspects against which performance was assessed. Since specific acceptance values are not provided, the table below lists the measured parameters and the statement that the device meets criteria.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Conformance to DICOM GSDF (AAPM TG18 guideline) | Verified conformance to DICOM GSDF |
| Angular dependency of luminance response (horizontal, vertical, diagonal) | Measured and meets pre-defined criteria (stated in conclusion) |
| Luminance non-uniformity characteristics (TG18 guideline) | Measured and meets pre-defined criteria (stated in conclusion) |
| Chromaticity non-uniformity characteristics (TG18 guideline) | Measured and meets pre-defined criteria (stated in conclusion) |
| Chromaticity at center (5%, 50%, 95% maximum luminance) | Measured and meets pre-defined criteria (stated in conclusion) |
| Display reflections (specular, diffuse, haze) | Measured and meets pre-defined criteria (stated in conclusion) |
| Small-spot contrast ratio | Measured and meets pre-defined criteria (stated in conclusion) |
| Spatial resolution (Modulation Transfer Function - MTF) | Measured and meets pre-defined criteria (stated in conclusion) |
| Noise (Noise Power Spectrum - NPS) | Measured and meets pre-defined criteria (stated in conclusion) |
| Pixel aperture ratio | Measured and meets pre-defined criteria (stated in conclusion) |
| Absence of miscellaneous artifacts (TG18 guideline) | Visual check performed and meets pre-defined criteria (stated in conclusion) |
| Temporal response | Measured and meets pre-defined criteria (stated in conclusion) |
| Luminance stability | Performance data provided and meets pre-defined criteria (stated in conclusion) |
| Maximum allowed pixel defects/faults | Measured and meets pre-defined criteria (stated in conclusion) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document describes bench tests performed on the RadiForce GX550. This indicates a sample size of one device (or a small number of devices directly tested).
- Data Provenance: The tests are prospective bench tests performed on the manufactured device. The location of the testing is not explicitly stated but implies to be conducted by EIZO Corporation, based in Japan.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is a technical performance verification for a medical display, not a diagnostic device requiring expert interpretation for ground truth establishment. The "ground truth" here is the established technical specifications and performance standards (e.g., DICOM GSDF, AAPM TG18 guidelines).
4. Adjudication Method for the Test Set
Not applicable. See point 3. Bench tests are objective measurements against technical standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, what was the effect size of how much human readers improve with AI vs without AI assistance
No. This submission is for a medical display monitor, not an AI-powered diagnostic tool. Therefore, an MRMC study assessing human reader performance with or without AI assistance is not relevant and was not performed.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This is a medical display, not an algorithm. Therefore, "standalone" algorithm performance is not applicable.
7. The Type of Ground Truth Used
The "ground truth" for the performance testing is based on established technical standards and guidelines for medical displays. Specifically, the document mentions:
- DICOM GSDF (Grayscale Standard Display Function): A standardized luminance response curve for medical images.
- AAPM Task Group 18 (TG18) guideline: "Assessment of Display Performance for Medical Imaging Systems," which provides methodologies and criteria for evaluating medical display monitors.
8. The Sample Size for the Training Set
Not applicable. This device is a display monitor, not an algorithm that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable. See point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three heads, representing the department's focus on health, human services, and science. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2016
EIZO Corporation % Mr. Hiroaki Hashimoto Manager 153 Shimokashiwano Hakusan, Ishikawa 924-8566 JAPAN
Re: K162497
Trade/Device Name: 5MP Monochrome LCD Monitor, Radiforce GX550, GX550-AR Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: September 2, 2016 Received: September 7, 2016
Dear Mr. Hashimoto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K162497
Device Name
5MP Monochrome LCD Monitor, RadiForce GX550, GX550-AR
Indications for Use (Describe)
This product is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.
Type of Use (Select one or both, as applicable)
| X Prescription Use (Part 21 CER 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CER 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the EIZO logo. The logo consists of a stylized square with a checkered pattern on the left and the word "EIZO" in bold, sans-serif font on the right. A registered trademark symbol is located to the upper right of the "O" in "EIZO".
EIZO Corporation, 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Janan
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
Name Department
Hiroaki Hashimoto Medical System Standards
Telephone Fax E-Mail
+81 (76) 274-2468 +81 (76) 274-2484 hiroaki.hashimoto@eizo.com
510(k) Summary (in accordance with 21 CFR 807.92)
1. Company
EIZO Corporation 153 Shimokashiwano, Hakusan Ishikawa 924-8566 Japan Tel: +81 (76) 274-2468 Fax: +81 (76) 274-2484
2. Contact Person
Hiroaki Hashimoto
-
- Date of Summary September 2nd, 2016
4. Device Information
- Trade Name/Model: RadiForce GX550, GX550-AR ●
- Common Name: 5MP Monochrome LCD Monitor
- Classification Name: Display, Diagnostic Radiology ●
- 21 CFR 892.2050, Product Code PGY ● Regulation Number:
5. Predicate Device
- 5MP Monochrome LCD Monitor, RadiForce GX540 (K151883) ●
{4}------------------------------------------------
6. Device Description
RadiForce GX550 is a monochrome LCD monitor for viewing medical images including those of mammography. The monochrome panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 2,048 x 2,560 pixels (5MP) with a pixel pitch of 0.165 mm.
Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.
There are two model variations, GX550 and GX550-AR. The difference of the two variations is the surface treatment of the display screens; the surface treatment of the GX550 is Anti-Glare (AG) treatment and that of the GX550-AR is Anti-Reflection (AR) coating.
RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce GX550 based on the QC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce GX550.
7. Indications for use
This product is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.
{5}------------------------------------------------
8. Comparison of Technological Characteristics
The comparison table below enumerates information derived from the product brochure of the each device and different technological characteristics are discussed in it:
| Attributes | RadiForce GX550 | RadiForce GX540 | Explanation ofDifferences |
|---|---|---|---|
| Display Performance/Specifications | |||
| Screentechnology | TFT MonochromeLCD Panel (IPS) | TFT MonochromeLCD Panel (IPS) | - |
| Viewing angle(H, V) | H: 178°, V: 178° | H: 176°, V: 176° | - |
| Resolution | 5MP (2,048 x 2,560) | 5MP (2,048 x 2,560) | - |
| Aspect ratio | 4 : 5 | 4 : 5 | - |
| Active screensize | 337.9 mm x 422.4 mm | 337.9 mm x 422.4 mm | - |
| Pixel pitch | 0.165 mm x 0.165 mm | 0.165 mm x 0.165 mm | - |
| Maximumluminance | 2,000 cd/m² | 1,200 cd/m² | - |
| DICOMcalibratedluminance | 600 cd/m² | 500 cd/m² | - |
| Contrast ratio | 1500 : 1 | 1200:1 | - |
| ResponseTime (typical) | 25 ms (On/Off) | 25 ms (On/Off) | - |
| Backlighting | LED | LED | - |
| GrayscaleTones | 10-bit (DisplayPort): 1,024from a palette of 16,369tones8-bit: 256 from a palette of16,369 tones | 10-bit (DisplayPort): 1,024from a palette of 16,369tones8-bit: 256 from a palette of16,369 tones | - |
| Luminancenon-uniformitycompensation | Digital UniformityEqualizer | Digital UniformityEqualizer | - |
| Video Signals | |||
| Input videosignals | DVI-D (dual link) x 1,DisplayPort x 1 | DVI-D (dual link) x 1,DisplayPort x 1 | - |
| Outputvideosignals | DisplayPort x 1 (daisychain) | - | - |
| ScanningFrequency (H,V) | 31 - 135 kHz / 23 - 61 HzFrame synchronous mode:23.5 - 25.5 Hz, 47 - 51 Hz | 31 - 135 kHz / 24 - 61 HzFrame synchronous mode:24.5 - 25.5 Hz, 49 - 51 Hz | - |
| Power Related Specifications | |||
| PowerRequirements | AC 100 - 240 V:50 / 60 Hz | AC 100 - 120 V,200 - 240 V: 50 / 60 Hz | - |
| PowerConsumption /Save Mode | 95 W / Less than 1 W | 108 W / Less than 0.7 W | Compared with thepredicate device, theproposed deviceconsumes less power inthe operating mode andmore power in the powersaving mode. |
| PowerManagement | DVI DMPM,DisplayPort 1.2a | DVI DMPM,DisplayPort 1.1a | - |
| Miscellaneous Features/Specifications | |||
| QC software | RadiCS | RadiCS | - |
| Sensors | Backlight Sensor,Integrated Front Sensor,Presence Sensor,Ambient Light Sensor | Backlight Sensor,Integrated Front Sensor,Presence Sensor,Ambient Light Sensor | - |
| USB Ports /Standard | 1 upstream,2 downstream / Rev. 2.0 | 1 upstream,2 downstream / Rev. 2.0 | - |
| Dimensionsw/o stand(W x H x D) | 367 x 452 x 78 mm | 388 x 496 x 99 mm | Different housing design. |
{6}------------------------------------------------
It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the GX550.
{7}------------------------------------------------
9. Performance Testing
The bench tests below were performed on the RadiForce GX550 following the instructions in Guidance for Industry and FDA Staff: Display Accessories for Full-Field Digital Mammography Systems-Premarket Notification (510(k)) Submissions:
- . Verification of the conformance to DICOM GSDF as specified in Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
- . Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directions
- . Measurement of the luminance non-uniformity characteristics of the display screen as specified in TG18 guideline
- Measurement of the chromaticity non-uniformity characteristics of the display screen . as specified in TG18 guideline
- . Measurement of the chromaticity at the center of the display screen at 5%, 50% and 95% of the maximum luminance
- Measurement of display reflections including specular, diffuse and haze components ●
- . Measurement of small-spot contrast ratio
- . Measurement of spatial resolution expressed as modulation transfer function (MTF)
- Measurement of noise expressed as noise power spectrum (NPS)
- Measurement of pixel aperture ratio ●
- Visual check of presence or absence of miscellaneous artifacts on the display screen ● as specified in TG18 guideline
- Measurement of temporal response
- Performance data on luminance stability
- . The maximum number allowed for each type of pixel defects/faults
The test results showed that the RadiForce GX550 has display characteristics equivalent to those of the predicate device, RadiForce GX540.
Besides, the display characteristics of the RadiForce GX550 meet the pre-defined criteria when criteria are set.
No animal or clinical testing was performed on the RadiForce GX550.
10. Conclusion
The RadiForce GX550 was determined to be substantially equivalent to the predicate device due to the following reasons:
{8}------------------------------------------------
- . The stated intended use is completely the same as that of the predicate device.
- . It was confirmed that the technological characteristics differences from those of the predicate device do not affect the safety or the effectiveness.
The bench tests demonstrated that the display characteristics are equivalent to those of the predicate device.
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).