(416 days)
PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The recommended (and validated) sterilization cycle parameters for all models (Sterilization Pouches (FP) and Rolls(FL) with Chemical Indicator) are:
- For steam sterilization pre-vacuum cycle at 132 degrees C for 4 minutes. Validated minimum dry time is 20 minutes.
- For EO sterilization; 100%ethylene oxide(EO) with a concentration of 725mg/l at 55 degrees C and 50o 80%relative humidity for 60 minutes. Aeration time is 8 hours.
Chemical process indicators on the exterior of the pouch and roll indicate by color change that the pouch has undergone either a steam or ethylene oxide sterilization process.
Type1 steam indicator will change color from pink to brown after exposure to pre-vacuum steam sterilization process.
Type 1 ethylene oxide indicator will change from turquoise to ethylene oxide sterilization process.
As Worst Case Validated Load; general medical devices with stainless steel lumens with the following internal diameter (ID) and Maximum Lumen Lumen length can be processed in the validated steam and ETO sterilization cvcles as follows:
Table 1.Worst Case Validated Load and Lumen Claims.
| Pouch Size | ProductCode | Max.LoadWeight (Ibs.) | Max. lumen length(mm) | Min. internaldiameter (mm) | Max # Lumens |
|---|---|---|---|---|---|
| 50 x 200 mm | FP 0520 | 0.062 | 50.8 mm | 1.5 mm | 1 |
| 200 x 300 mm | FP 2030 | 0.516 | 200 mm | 3.0 mm | 1 |
| 250 x 300 mm | FP 2530 | 0.520 | 200 mm | 3.0 mm | 1 |
| 500 x 600 mm | FP 5060 | 3.344 | 400 mm | 3.0 mm | 1 |
PMSSteripack Flat Sterilization pouches (FP) and rolls (FL) with Chemical Indicator; are manufactured from medical grade paper/ laminated PET/PP film, by heat sealing on sides. Triple band seal provides three independent barriers to contamination, while reducing the risk of fiber tear.
PMSSteripack Flat Sterilization Pouches (FP) and Rolls (FL) with Chemical Indicator; are divided into 2 main groups as below;
-
Sterilization Pouches, Flat: These pouches are manufactured from medical grade paper/plastic film that are heat sealed on 3 sides, the remaining open side is heat sealed by the end user.
-
Sterilization Rolls, Flat: These rolls are manufactured from medical grade paper/plastic film that is heat sealed on opposite two sides. The required size is cut from the roll and the remaining open sides are heat sealed by the end user.
The medical grade paper conforms to recognized material standards and can be sterilized by steam or ethylene oxide gas.
After the sterilization process is completed, the sterility of the enclosed medical device is rnaintained for 30 days.
This document describes the PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator.
1. Table of Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Sterilant Penetration | Conforms to ANSI/AAMI/ISO 17665-1:2006/R2013, ISO 17665-2:2009, ANSI/AAMI/ISO 11135-2014, AAMI TIR 12 2010 | Passed |
| Drying Time | Conforms to AAMI TIR 12 | Passed |
| Aeration Time | Conforms to ANSI/AAMI/ISO 10993-7:2008/R 2012 | Passed |
| Biocompatibility | Conforms to ANSI/AAMI/ISO 10993-5:2009 | Passed (No cytotoxic effect) |
| Package Integrity | Conforms to ANSI/AAMI/ISO 11607-1:2006/A1 2014, ASTM F1140/F1140M - 13, ASTM F88-09, ASTM D645-97, ASTM F1886-09, ASTM F2096-11 | Passed |
| Material Compatibility | Conforms to ISO 10993-5 | Passed |
| Sterility Maintenance | Conforms to ANSI/AAMI/ISO 11607-1:2006/A1 2014, ANSI/AAMI/ISO 14937:2009/R2013 | Passed |
| Chemical Indicator Efficacy (Type 1) | Conforms to ANSI/AAMI/ISO 11140-1:2014 | Passed |
| Shelf Life (pre-sterilization use) | Not explicitly stated as acceptance criteria, but predicate was 3 years. | 5 years |
| Shelf Life (post-sterilization) | Not explicitly stated as acceptance criteria, but predicate was 3 years. | 30 days |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the specific sample sizes used for each individual test. However, it mentions a "Worst Case Validated Load" tested for different pouch sizes and lumen claims:
- Pouch Size 50 x 200 mm (FP 0520): Max. Load Weight 0.062 lbs., Max. lumen length 50.8 mm, Min. internal diameter 1.5 mm, Max # Lumens 1
- Pouch Size 200 x 300 mm (FP 2030): Max. Load Weight 0.516 lbs., Max. lumen length 200 mm, Min. internal diameter 3.0 mm, Max # Lumens 1
- Pouch Size 250 x 300 mm (FP 2530): Max. Load Weight 0.520 lbs., Max. lumen length 200 mm, Min. internal diameter 3.0 mm, Max # Lumens 1
- Pouch Size 500 x 600 mm (FP 5060): Max. Load Weight 3.344 lbs., Max. lumen length 400 mm, Min. internal diameter 3.0 mm, Max # Lumens 1
The document does not provide information on the country of origin of the data or whether the studies were retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The studies listed are performance tests against recognized standards rather than studies involving expert interpretation for ground truth.
4. Adjudication method for the test set
This information is not applicable as the studies described are performance tests against standards, not studies requiring adjudication of output from a device or algorithm.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study was not done. The device in question is a sterilization pouch with a chemical indicator, not an AI-assisted diagnostic or interpretative device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable for this device type. The chemical indicator operates standalone as a chemical reaction, but it is not an algorithm, nor does it require human-in-the-loop performance in the sense of an AI system. Its function is to change color, indicating exposure to a sterilization process.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the tests performed is based on meeting the requirements of internationally recognized standards for sterilization packaging, such as ISO, AAMI, and ASTM. For example:
- Sterilant Penetration: Ground truth is effective penetration as defined by the aforementioned standards.
- Biocompatibility: Ground truth is the absence of cytotoxic effects as per ANSI/AAMI/ISO 10993-5.
- Chemical Indicator Efficacy: Ground truth is the reproducible color change according to ANSI/AAMI/ISO 11140-1 Class 1 indicator requirements.
8. The sample size for the training set
This is not applicable as the device is a physical product (sterilization pouch with a chemical indicator) and not an AI or machine learning model that requires a training set.
9. How the ground truth for the training set was established
This is not applicable as the device is not an AI or machine learning model that requires a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 13, 2017
Pms Tibbi Cihazlar Teknolojisi San Ve Tic As Derya Dikici Business Development Manager Karaduvar Mah.serbest Bolge, 11.cad No.46 Mersin Serbest Bolgesi, Akdeniz Mersin, 33020 TURKEY
Re: K162362 Trade/Device Name: PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: August 24, 2017 Received: September 11, 2017
Dear Derya Dikici:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) 162362
Device Name
PMSSteripack Flat Sterilization Pouch (FP) and Roll(FL) with Chemical Indicator
Indications for Use (Describe)
PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator, is a device intended to be used to enclose another medical device that is to be sterlized by a health care provider. It is intended to allow sterlization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The recommended (and validated) sterilization cvcle parameters for all models (Sterilization Pouches (FP) and Rolls(FL) with Chemical Indicator) are:
- For steam sterilization pre-vacuum cycle at 132 degrees C for 4 minutes. Validated minimum dry time is 20 minutes. ●
For EO sterilization; 100%ethylene oxide(EO) with a concentration of 725mg/l at 55 degrees C and 50o 80%relative humidity for 60 minutes. Aeration time is 8 hours.
Chemical process indicators on the exterior of the pouch and roll indicate by color change that the pouch has
undergone either a steam or ethylene oxide sterilization process.
Type1 steam indicator will change color from pink to brown after exposure to pre-vacuum steam sterilization process.
Type 1 ethylene oxide indicator will change from turquoise to ethylene oxide sterilization process.
As Worst Case Validated Load; general medical devices with stainless steel lumens with the following internal diameter (ID) and Maximum Lumen Lumen length can be processed in the validated steam and ETO sterilization cvcles as follows:
Table 1.Worst Case Validated Load and Lumen Claims.
| Pouch Size | ProductCode | Max.LoadWeight (Ibs.) | Max. lumen length(mm) | Min. internaldiameter (mm) | Max # Lumens |
|---|---|---|---|---|---|
| 50 x 200 mm | FP 0520 | 0.062 | 50.8 mm | 1.5 mm | 1 |
| 200 x 300 mm | FP 2030 | 0.516 | 200 mm | 3.0 mm | 1 |
| 250 x 300 mm | FP 2530 | 0.520 | 200 mm | 3.0 mm | 1 |
| 500 x 600 mm | FP 5060 | 3.344 | 400 mm | 3.0 mm | 1 |
{3}------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EFF
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for PMS. The logo consists of a red and gray abstract shape on the left, followed by the letters "PMS" in gray. The letters are large and bold, and they are spaced evenly apart.
K162362
510 (k) Summary For "PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator"
1.0. Name, address and contact
PMS TIBBI CIHAZLAR TEKNOLOJISI SANAYI VE TICARET ANONIM SIRKETI
Karaduvar Mahallesi, Serbest Bölge 11.cad. No: 46 Mersin Serbest Bölgesi Akdeniz 33020 Mersin –TURKEY
| Phone | : | 90 324 2387042 – 90 541 648 6312 |
|---|---|---|
| Fax | : | 90 324 2386549 |
| Prepared by | : | Taner ErsenQuality Management Representative |
| Contact | : | Derya DikiciBusiness Development ManagerPhone : 90 324 2387042- 90 541 648 6312Fax : 90 324 2386549E-mail : derya.dikici@pmsmedikal.com |
Date Prepared: October 13, 2017
{5}------------------------------------------------
2. Device Name
Trade Name: PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator
Common/Usual Name: Sterilization Pouch and Roll (Flat)
Device Classification Name: Wrap, Sterilization - Indicator, Physical/Chemical Sterilization Process
Product Code: FRG - JOJ
Product Classification: Class II
21 CFR 880.6850, General Hospital
3. Predicate Device
K102158, SIGMA Sterilization Pouch and Roll
4. Indications For Use
PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The recommended (and validated) sterilization cycle parameters for all models (Sterilization Pouches (FP) and Rolls(FL) with Chemical Indicator) are;
For steam sterilization pre-vacuum cycle at 132 degrees C for 4 minutes. Validated . minimum dry time is 20 minutes.
. For EO sterilization; 100%ethylene oxide(EO) with a concentration of 725mg/l at 55 degrees C and 50-80%relative humidity for 60 minutes. Aeration time is 8 hours.
Chemical process indicators on the exterior of the pouch and roll indicate by color change that the pouch has undergone
either a steam or ethylene oxide sterilization process.
Type 1 steam indicator will change color from pink to brown after exposure to pre-vacuum steam sterilization process.
Type 1 ethylene oxide indicator will change from turquoise to yellow after exposure to ethylene oxide sterilization process.
As Worst Case Validated Load; general medical devices of stainless steel and devices with stainless steel lumens with the following internal diameter (ID) and Maximum Load weight and Maximum Lumen length can be processed in the validated steam and ETO sterilization cycles as follows;
{6}------------------------------------------------
Table 1.Worst Case Validated Load and Lumen Claims.
| Pouch Size | ProductCode | Max.LoadWeight(Ibs.) | Max. lumenlength(mm) | Min. internaldiameter(mm) | Max #Lumens |
|---|---|---|---|---|---|
| 50 x 200 mm | FP 0520 | 0.062 | 50.8 mm | 1.5 mm | 1 |
| 200 x 300 mm | FP 2030 | 0.516 | 200 mm | 3.0 mm | 1 |
| 250 x 300 mm | FP 2530 | 0.520 | 200 mm | 3.0 mm | 1 |
| 500 x 600 mm | FP 5060 | 3.344 | 400 mm | 3.0 mm | 1 |
5. Device Description
PMSSteripack Flat Sterilization pouches (FP) and rolls (FL) with Chemical Indicator; are manufactured from medical grade paper/ laminated PET/PP film, by heat sealing on sides. Triple band seal provides three independent barriers to contamination, while reducing the risk of fiber tear.
PMSSteripack Flat Sterilization Pouches (FP) and Rolls (FL) with Chemical Indicator; are divided into 2 main groups as below;
-
Sterilization Pouches, Flat: These pouches are manufactured from medical grade paper/plastic film that are heat sealed on 3 sides, the remaining open side is heat sealed by the end user.
-
Sterilization Rolls, Flat: These rolls are manufactured from medical grade paper/plastic film that is heat sealed on opposite two sides. The required size is cut from the roll and the remaining open sides are heat sealed by the end user.
The medical grade paper conforms to recognized material standards and can be sterilized by steam or ethylene oxide gas.
After the sterilization process is completed, the sterility of the enclosed medical device is rnaintained for 30 days.
-List of Model Numbers are given in Table 2 and Table 3.
| Table 2. Product List of PMSSteripack Flat Sterilization Pouch (FP) with Chemical |
|---|
| Indicator. |
| PRODUCT LIST (FP) | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| ProductName | (A)TotalWidth(mm)±2 | (B)TotalHeight(mm)±5 | (A)TotalWidth(in.)±0.08 | (B)TotalHeight(in.)±0.2 | (L)SealWidth(mm)±2 | (V)SingleSealWidth(mm)min.2 | (P)DistanceBetween2 Seals(mm)±0.1 | (S)Diameter(mm)min. | (T)Thickness(μm) |
| FP0520 | 50 | 200 | 2 | 8 | 10 | 2 | 2 | 10 | 153±18 |
| FP0525 | 50 | 250 | 2 | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP0530 | 50 | 300 | 2 | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP0532 | 50 | 320 | 2 | 12¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP0535 | 50 | 350 | 2 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP7515 | 75 | 150 | 3 | 6 | 10 | 2 | 2 | 10 | 153±18 |
| FP7518 | 75 | 180 | 3 | 7 | 10 | 2 | 2 | 10 | 153±18 |
| FP7520 | 75 | 200 | 3 | 8 | 10 | 2 | 2 | 10 | 153±18 |
| FP7523 | 75 | 230 | 3 | 9 | 10 | 2 | 2 | 10 | 153±18 |
| FP7525 | 75 | 250 | 3 | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP7528 | 75 | 280 | 3 | 11 | 10 | 2 | 2 | 10 | 153±18 |
| FP7530 | 75 | 300 | 3 | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP7535 | 75 | 350 | 3 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP7540 | 75 | 400 | 3 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP7545 | 75 | 450 | 3 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP0815 | 80 | 150 | 3¼ | 6 | 10 | 2 | 2 | 10 | 153±18 |
| FP0817 | 80 | 170 | 3¼ | 6¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP0818 | 80 | 180 | 3¼ | 7 | 10 | 2 | 2 | 10 | 153±18 |
| FP0820 | 80 | 200 | 3¼ | 8 | 10 | 2 | 2 | 10 | 153±18 |
| FP0827 | 80 | 270 | 3¼ | 10¼ | 10 | 2 | 2 | 10 | 153±18 |
| FP0860 | 80 | 600 | 3¼ | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP0956 | 90 | 560 | 3½ | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP0960 | 90 | 600 | 3½ | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP0961 | 90 | 610 | 3½ | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP1015 | 100 | 150 | 4 | 6 | 10 | 2 | 2 | 10 | 153±18 |
| FP1018 | 100 | 180 | 4 | 7 | 10 | 2 | 2 | 10 | 153±18 |
| FP1019 | 100 | 190 | 4 | 7½ | 10 | 2 | 2 | 10 | 153±18 |
| FP1020 | 100 | 200 | 4 | 8 | 10 | 2 | 2 | 10 | 153±18 |
| FP1023 | 100 | 230 | 4 | 9 | 10 | 2 | 2 | 10 | 153±18 |
| FP1025 | 100 | 250 | 4 | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP1026 | 100 | 260 | 4 | 10¼ | 10 | 2 | 2 | 10 | 153±18 |
| FP1027 | 100 | 270 | 4 | 10¼ | 10 | 2 | 2 | 10 | 153±18 |
| FP1028 | 100 | 280 | 4 | 11 | 10 | 2 | 2 | 10 | 153±18 |
| FP1030 | 100 | 300 | 4 | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP1033 | 100 | 330 | 4 | 13 | 10 | 2 | 2 | 10 | 153±18 |
| FP1035 | 100 | 350 | 4 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP1036 | 100 | 360 | 4 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP1040 | 100 | 400 | 4 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP1045 | 100 | 450 | 4 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP1050 | 100 | 500 | 4 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP1056 | 100 | 560 | 4 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP1057 | 100 | 570 | 4 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP1060 | 100 | 600 | 4 | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP1061 | 100 | 610 | 4 | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP1128 | 110 | 280 | 4¼ | 11 | 10 | 2 | 2 | 10 | 153±18 |
| FP1225 | 120 | 250 | 4¾ | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP1230 | 120 | 300 | 4¾ | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP1240 | 120 | 400 | 4¾ | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP1250 | 120 | 500 | 4¾ | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP1325 | 130 | 250 | 5 | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP1338 | 130 | 380 | 5 | 15 | 10 | 2 | 2 | 10 | 153±18 |
| FP1520 | 150 | 200 | 6 | 8 | 10 | 2 | 2 | 10 | 153±18 |
| FP1525 | 150 | 250 | 6 | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP1527 | 150 | 270 | 6 | 10¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP1528 | 150 | 280 | 6 | 11 | 10 | 2 | 2 | 10 | 153±18 |
| FP1530 | 150 | 300 | 6 | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP1532 | 150 | 320 | 6 | 12¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP1535 | 150 | 350 | 6 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP1538 | 150 | 380 | 6 | 15 | 10 | 2 | 2 | 10 | 153±18 |
| FP1540 | 150 | 400 | 6 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP1543 | 150 | 430 | 6 | 17 | 10 | 2 | 2 | 10 | 153±18 |
| FP1545 | 150 | 450 | 6 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP1550 | 150 | 500 | 6 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP1552 | 150 | 520 | 6 | 20¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP1555 | 150 | 550 | 6 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP1556 | 150 | 560 | 6 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP1640 | 160 | 400 | 6¼ | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP1658 | 160 | 580 | 6¼ | 23 | 10 | 2 | 2 | 10 | 153±18 |
| FP1722 | 170 | 220 | 6¾ | 8¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP1735 | 170 | 350 | 6¾ | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP1830 | 180 | 300 | 7 | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP1832 | 180 | 320 | 7 | 12¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP1846 | 180 | 460 | 7 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP1933 | 190 | 330 | 7½ | 13 | 10 | 2 | 2 | 10 | 153±18 |
| FP2020 | 200 | 200 | 8 | 8 | 10 | 2 | 2 | 10 | 153±18 |
| FP2025 | 200 | 250 | 8 | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP2027 | 200 | 270 | 8 | 10¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP2030 | 200 | 300 | 8 | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP2033 | 200 | 330 | 8 | 13 | 10 | 2 | 2 | 10 | 153±18 |
| FP2035 | 200 | 350 | 8 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP2040 | 200 | 400 | 8 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP2041 | 200 | 410 | 8 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP2045 | 200 | 450 | 8 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP2050 | 200 | 500 | 8 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP2060 | 200 | 600 | 8 | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP2128 | 210 | 280 | 8¼ | 11 | 10 | 2 | 2 | 10 | 153±18 |
| FP2135 | 210 | 350 | 8¾ | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP2140 | 210 | 400 | 8¼ | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP2142 | 210 | 420 | 8¼ | 16¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP2228 | 220 | 280 | 8¾ | 11 | 10 | 2 | 2 | 10 | 153±18 |
| FP2235 | 220 | 350 | 8¾ | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP2242 | 220 | 420 | 8¾ | 16¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP2525 | 250 | 250 | 10 | 10 | 10 | 2 | 2 | 10 | 153±18 |
| FP2530 | 250 | 300 | 10 | 12 | 10 | 2 | 2 | 10 | 153±18 |
| FP2535 | 250 | 350 | 10 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP2538 | 250 | 380 | 10 | 15 | 10 | 2 | 2 | 10 | 153±18 |
| FP2540 | 250 | 400 | 10 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP2545 | 250 | 450 | 10 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP2550 | 250 | 500 | 10 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP2735 | 270 | 350 | 10¼ | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP2745 | 270 | 450 | 10¼ | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP2835 | 280 | 350 | 11 | 14 | 10 | 2 | 2 | 10 | 153±18 |
| FP2840 | 280 | 400 | 11 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP2845 | 280 | 450 | 11 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP2850 | 280 | 500 | 11 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP2857 | 280 | 570 | 11 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP3038 | 300 | 380 | 12 | 15 | 10 | 2 | 2 | 10 | 153±18 |
| FP3040 | 300 | 400 | 12 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP3042 | 300 | 420 | 12 | 16¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP3045 | 300 | 450 | 12 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP3046 | 300 | 460 | 12 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP3050 | 300 | 500 | 12 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP3055 | 300 | 550 | 12 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP3060 | 300 | 600 | 12 | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP3245 | 320 | 450 | 12¾ | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP3250 | 320 | 500 | 12¾ | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP3255 | 320 | 550 | 12¾ | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP3346 | 330 | 460 | 13 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP3540 | 350 | 400 | 14 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP3545 | 350 | 450 | 14 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP3550 | 350 | 500 | 14 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP3552 | 350 | 520 | 14 | 20¾ | 10 | 2 | 2 | 10 | 153±18 |
| FP3555 | 350 | 550 | 14 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP3660 | 360 | 600 | 14 | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP4040 | 400 | 400 | 16 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP4045 | 400 | 450 | 16 | 18 | 10 | 2 | 2 | 10 | 153±18 |
| FP4055 | 400 | 550 | 16 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP4060 | 400 | 600 | 16 | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP4141 | 410 | 410 | 16 | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP4150 | 410 | 500 | 16 | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP4241 | 420 | 410 | 16¾ | 16 | 10 | 2 | 2 | 10 | 153±18 |
| FP4250 | 420 | 500 | 16¾ | 20 | 10 | 2 | 2 | 10 | 153±18 |
| FP4255 | 420 | 550 | 16¾ | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP4260 | 420 | 600 | 16¾ | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP4656 | 460 | 560 | 18 | 22 | 10 | 2 | 2 | 10 | 153±18 |
| FP4661 | 460 | 610 | 18 | 24 | 10 | 2 | 2 | 10 | 153±18 |
| FP4858 | 480 | 580 | 19 | 23 | 10 | 2 | 2 | 10 | 153±18 |
| FP5060 | 500 | 600 | 20 | 24 | 10 | 2 | 2 | 10 | 153±18 |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| Table 3. Product List of PMSSteripack Flat Sterilization Roll (FL) with Chemical | |
|---|---|
| Indicator. |
| PRODUCT LIST (FL) | ||||||||
|---|---|---|---|---|---|---|---|---|
| ProductName | (A)TotalWidth(cm)$\pm$ 0.2 | (B)TotalLength(m)$\pm$ %1 | (A)TotalWidth(in)$\pm$ 0.008 | (B)TotalLength(ft)$\pm$ %1 | (L)SealWidth(mm)$\pm$ 2 | (V)SingleSealWidth(mm)min.2 | (P)DistanceBetween2 Seals(mm)$\pm$ 0.1 | (T)Thickness(μm) |
| FL 0530 | 5 | 30 | 2 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 7530 | 7.5 | 30 | 3 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 1030 | 10 | 30 | 4 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 12530 | 12.5 | 30 | 5 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 1530 | 15 | 30 | 6 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 2030 | 20 | 30 | 8 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 2330 | 23 | 30 | 9 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 2530 | 25 | 30 | 10 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 3030 | 30 | 30 | 12 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 3530 | 35 | 30 | 14 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 3830 | 38 | 30 | 15 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 4030 | 40 | 30 | 16 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 4530 | 45 | 30 | 18 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 5030 | 50 | 30 | 20 | 100 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 0560 | 5 | 60 | 2 | 200 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 7560 | 7.5 | 60 | 3 | 200 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 1060 | 10 | 60 | 4 | 200 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 12560 | 12.5 | 60 | 5 | 200 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 1560 | 15 | 60 | 6 | 200 | 10 | 2 | 2 | 153 $\pm$ 18 |
| FL 2060 | 20 | 60 | 8 | 200 | 10 | 2 | 2 | 153±18 |
| FL 2360 | 23 | 60 | 9 | 200 | 10 | 2 | 2 | 153±18 |
| FL 2560 | 25 | 60 | 10 | 200 | 10 | 2 | 2 | 153±18 |
| FL 3060 | 30 | 60 | 12 | 200 | 10 | 2 | 2 | 153±18 |
| FL 3560 | 35 | 60 | 14 | 200 | 10 | 2 | 2 | 153±18 |
| FL 3860 | 38 | 60 | 15 | 200 | 10 | 2 | 2 | 153±18 |
| FL 4060 | 40 | 60 | 16 | 200 | 10 | 2 | 2 | 153±18 |
| FL 4560 | 45 | 60 | 18 | 200 | 10 | 2 | 2 | 153±18 |
| FL 5060 | 50 | 60 | 20 | 200 | 10 | 2 | 2 | 153±18 |
| FL 05100 | 5 | 100 | 2 | 330 | 10 | 2 | 2 | 153±18 |
| FL 75100 | 7.5 | 100 | 3 | 330 | 10 | 2 | 2 | 153±18 |
| FL 10100 | 10 | 100 | 4 | 330 | 10 | 2 | 2 | 153±18 |
| FL 125100 | 12.5 | 100 | 5 | 330 | 10 | 2 | 2 | 153±18 |
| FL 15100 | 15 | 100 | 6 | 330 | 10 | 2 | 2 | 153±18 |
| FL 20100 | 20 | 100 | 8 | 330 | 10 | 2 | 2 | 153±18 |
| FL 23100 | 23 | 100 | 9 | 330 | 10 | 2 | 2 | 153±18 |
| FL 25100 | 25 | 100 | 10 | 330 | 10 | 2 | 2 | 153±18 |
| FL 30100 | 30 | 100 | 12 | 330 | 10 | 2 | 2 | 153±18 |
| FL 35100 | 35 | 100 | 14 | 330 | 10 | 2 | 2 | 153±18 |
| FL 38100 | 38 | 100 | 15 | 330 | 10 | 2 | 2 | 153±18 |
| FL 40100 | 40 | 100 | 16 | 330 | 10 | 2 | 2 | 153±18 |
| FL 45100 | 45 | 100 | 18 | 330 | 10 | 2 | 2 | 153±18 |
| FL 50100 | 50 | 100 | 20 | 330 | 10 | 2 | 2 | 153±18 |
| FL 05200 | 5 | 200 | 2 | 656 | 10 | 2 | 2 | 153±18 |
| FL 75200 | 7.5 | 200 | 3 | 656 | 10 | 2 | 2 | 153±18 |
| FL 10200 | 10 | 200 | 4 | 656 | 10 | 2 | 2 | 153±18 |
| FL 125200 | 12.5 | 200 | 5 | 656 | 10 | 2 | 2 | 153±18 |
| FL 15200 | 15 | 200 | 6 | 656 | 10 | 2 | 2 | 153±18 |
| FL 20200 | 20 | 200 | 8 | 656 | 10 | 2 | 2 | 153±18 |
| FL 23200 | 23 | 200 | 9 | 656 | 10 | 2 | 2 | 153±18 |
| FL 25200 | 25 | 200 | 10 | 656 | 10 | 2 | 2 | 153±18 |
| FL 30200 | 30 | 200 | 12 | 656 | 10 | 2 | 2 | 153±18 |
| FL 35200 | 35 | 200 | 14 | 656 | 10 | 2 | 2 | 153±18 |
| FL 38200 | 38 | 200 | 15 | 656 | 10 | 2 | 2 | 153±18 |
| FL 40200 | 40 | 200 | 16 | 656 | 10 | 2 | 2 | 153±18 |
| FL 45200 | 45 | 200 | 18 | 656 | 10 | 2 | 2 | 153±18 |
| FL 50200 | 50 | 200 | 20 | 656 | 10 | 2 | 2 | 153±18 |
{11}------------------------------------------------
{12}------------------------------------------------
6. Description of the Principle of Operation
PMSSteripack Sterilization Pouches (FP) and Rolls (FL) with Chemical Indicator, are used to enclose medical devices that are to be sterilized by health care provider via sterilization methods. Medical device to be sterilized is put into roll or pouch and the open parts of the packages are closed by heat sealing. Sterilization packages then are subjected to sterilization operation in related sterilization devices (steam sterilizers, EO sterilization devices). Sterilant penetration is carried out through the medical grade paper into the packages and microorqanisms on the surface of the medical devices are destroyed with the effect of the sterilant vapors. Other parameters of the sterilization process are temperature. pressure; humidity and time are deteminded according to the sterilization type.
The process indicator on the pouches and rolls are intended to be used by a health care provider with sterilization pouches to distinguish between processed and unprocessed units by changing color. Chemical process indicators on the pouch and roll indicate that the pouch has been exposed to sterilization process by changing color.
Chemical process indicators are printed on the pouch or roll (printed on medical grade paper) changes color when exposed to sterilant vapor during processing. After the sterilization is completed, the sterility of the enclosed medical device is maintained for 30 days. Chemical indicator bars printed on to the medical grade paper component of the pouches are 100 mm² size, having dimensions of 5 mmx20 mm.
The indicators used on the pouch and roll are classified as Class 1 type Indicator according to the ISO 11140-1 standard. The steam and ETO chemical indicator inks used in the PMS FL and FP sterilization pouches have been commercially available for many years and both indicator inks were cleared by the FDA in a prior PMS 510k pouch submission under No. K152669.
7. Comparison of Submission Device and Predicate Device and Substantial Equivalence
Technological characteristics of the submission device and the comparison with predicate devices are given in table 7.1 as a summary.
TABLE-7.1: Comparison of Submission Device and Predicate Device (Characteristics)
{13}------------------------------------------------
| DEVICE NAME | SUBMISSION DEVICE | PREDICATE DEVICE | COMPARISON | |||
|---|---|---|---|---|---|---|
| CHARACTERISTICS | PMSSteripack Flat Sterilization Pouch(FP) and Roll (FL) with Chemical Indicator | SIGMA Sterilization Pouch and Roll (K102158) | ||||
| DESIGN ANDCONSTRUCTION&MATERIALCOMPOSITION | Medical grade paper and heat sealed laminated PET/PP plastic film. Externally printed Steam and EO process indicator ink.The pouches are manufactured from medical grade paper and heat sealed laminated PET/PP plastic film that is heat sealed on 3 sides. The remaining open side is heat sealed by the end user.Rolls are manufactured from medical grade paper and heat sealed laminated PET/PP plastic film that is heat sealed on opposite two sides. The required size is cut from the roll and the remaining open two sides are heat sealed by the end user. | Medical Grade Paper and heat sealed plastic film. EO and Steam Process Indicator Print ink. Pouches are made from a medical grade plastic film that is heat sealed on three sides. The forth side is left opened and is heat-sealed when using. In gusseted pouches, plastic film is folded on both longest sides. The medical grade paper conforms to recognized material standards and can be sterilized by steam or ethylene oxide gas. | Similar | |||
| INTENDED USE | PMSSteripack Flat Sterilization Pouch (FP) and Roll (FL) with Chemical Indicator, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.The indicated sterilization parameters for either steam or EO are the only validated sterilization parameters to be used/applied.The recommended (and validated) sterilization cycle parameters for all models (Sterilization Pouches (FP) and Rolls(FL) with Chemical Indicator) are;• For steam sterilization pre-vacuum cycle at 132 degrees C for 4 minutes. Validated minimum dry time is 20 minutes.• For EO sterilization; 100%ethylene oxide(EO) with a concentration of 725mg/l at 55 degrees C and 50-80%relative humidity for 60 minutes. Aeration time is 8 hours.Chemical process indicators on the exterior of the pouch and roll indicate by color change that the pouch has undergone either a steam or ethylene oxide sterilization process.Class 1 steam indicator will change color from pink to brown after exposure to pre-vacuum steam sterilization process.Class 1 ethylene oxide indicator will change from turquoise to yellow after exposure to ethylene oxide sterilization process.As Worst Case Validated Load; general medical devices of stainless steel and devices with stainless steel lumens with the following internal diameter (ID) and Maximum Load weight and Maximum Lumen length can be processed in the validated steam and ETO sterilization cycles as follows; | Single use devices, to enclose other medical devices that are to be sterilized by a health provider. Sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam auto claves and via Ethylene Oxide (EO).The recommended steam sterilization cycle parameters are 30 minutes at 121 °C.The recommended EO sterilization cycle is 4 hours at 55 °C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L.Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post sterilization.Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. | The intended use differences are minor and does not impact substantial equivalence | |||
| Pouch Size | Product Code | Max. Load Weight (lbs.) | Max. lumen length (mm) | Min. internal diameter (mm) | Max # Lumens | |
| 50 x 200 mm | FP 0520 | 0.062 | 50.8 mm | 1.5 mm | 1 | |
| 200 x 300 mm | FP 2030 | 0.516 | 200 mm | 3.0 mm | 1 | |
| 250 x 300 mm | FP 2530 | 0.520 | 200 mm | 3.0 mm | 1 | |
| 500 x 600 mm | FP 5060 | 3.344 | 400 mm | 3.0 mm | 1 | |
| Table 1.Worst Case Validated Load and Lumen Claims. |
{14}------------------------------------------------
| STERILIZATIONPROPERTIES | The recommended(validated) sterilization cycle parametersare:For Steam sterilization; pre-vacuum cycle at 132 °C for 4minutes.For EO Sterilization; 100% ethylene oxide (EO) with aconcentration of 725 mg/L at 55 °C and 50-80% relativehumidity for 60 minutes. Aeration time is 8 hours. | Therecommended steamsterilization cycle parameters are30 minutes at 121 °C.The recommended EOsterilization cycle is 4 hours at 55°C with a relative humiditybetween 50%/85% and asterilant concentration of 600mg/L. | Similar |
|---|---|---|---|
| PRINCIPLE OFOPERATION | Medical device to be sterilized is put into pouch or roll andthe open parts of the packages are closed by heat sealing.Sterilization packages then are subjected to validatedsterilization operation of steam or EO. Sterilant penetrationis carried out through the medical grade paper into thepackage and microorganisms on the surface of the medicaldevice are destroyed with the effect of the sterilant vapors.Other parameters of the sterilization process aretemperature, pressure, humidity, time and are determinedaccording to the sterilization type. Chemical processindicator is printed exterior on the pouch/roll (printed onmedical grade paper) changes color when exposed tosterilant vapor during processing. After the sterilization iscompleted, the sterility of the enclosed medical device ismaintained for 30 days | Medical device to be sterilized isput into roll or pouch and theopen parts of the packages areclosed by heat sealing.Sterilization packages then aresubjected to sterilizationoperation in related sterilizationdevices (steam autoclaves, EOsterilization).The process Indicators Inkprinted on the medical gradepaper will exhibit a color changeafter the pouch is exposed tosteam or ethylene oxide gas.The SIGMA sterilization pouchand roll is offered in thefollowing 5 types:* Self-sealing sterilizationpouches* Sterilization pouches, Flat* Sterilization pouches,Gusseted* Sterilization rolls, Flat* Sterilization rolls, Gusseted | Similar |
| PRINCIPLES OFOPERATION FORCHEMICALINDICATORS | The Process Indicators Ink printed on the medical gradepaper will exhibit a color change after the pouch is exposedto steam or ethylene oxide gas. In steam sterilization, printedindicator bar changes from pink to brown when exposed tosteam. In EO sterilization, printed indicator bar changes fromturquoise to yellow when exposed to EO gas. | The Process Indicators Inkprinted on the medical gradepaper will exhibit a color changeafter the pouch is exposed tosteam or ethylene oxide gas. | Similar |
| SHELF LIFE (pre-sterilization use) | 5 years | 3 years | The testingsupportssubstantialequivalence. |
| SHELF LIFE (poststerilization) | 30 days | 3 years | The testingsupportssubstantialequivalence. |
| CONFIGURATIONS& DIMENSIONS | Various sizes (width and height/length) | Various sizes (width,height/length and gusset) | Similar |
| DEVICE NAME | PROPOSED DEVICE | PREDICATE DEVICE | COMPARISON |
| PERFROMANCE(Test Standard) | PMSSteripack Flat SterilizationPouch (FP) and Roll (FL) withChemical Indicator | SIGMA Sterilization Pouch andRoll (K102158) | Similar |
| Sterilant Penetration(AAMI TIR 12) | passed | passed | Similar |
| Microbial BarrierProperties(ANSI/AAMI/ISO14937) | passed | passed | Similar |
| MaterialCompatibility (ISO10993-5) | passed | passed | Similar |
| Biocompatibility(ISO 10993-5) | passed | passed | Similar |
| Shelf Life (poststerilization)(ANSI/AAMI/ISO11607-1) | passed | passed | Similar |
| Drying Time(AAMI TIR 12) | passed | passed | Similar |
| Aeration Time(ISO 10993-7) | passed | passed | Similar |
| Package Integrity(ANSI/AAMI/ISO11607-1) | passed | passed | Similar |
{15}------------------------------------------------
TABLE-7.2: Comparison of Submission Device and Predicate Device (Non-Clinical Performance Testing)
8. Discussion - Safety and Effectiveness
The PMSSteripack Flat Sterilization pouch (FP) and roll (FL) has the same intended use and substantially equivalent indication for use as the predicate devices. Substantial equivalence was established by testing sterilant penetration, drying time, aeration, biocompatibility, package integrity, materials compatibility, sterility maintenance and chemical indicator efficacy.
The PMSSteripack Flat Sterilization pouch (FP) and roll (FL) validates its effectiveness and safety using recommended practice, standards and quidelines developed by independent orqanizations, such as the Association for the advancement of Medical Instrumentation (AAMI). International Organization for Standardization (ISO), and American Society for Testing and Materials (ASTM). The PMSSteripack Flat Sterilization pouch (FP) and roll (FL) was validated to meet the requirements of ANSI/AAMIIISO 11607-1:2006/A1 2014.
The results of the PMSSteripack Flat Sterilization pouch (FP) and roll (FL) validation studies demonstrate that the sterilization pouches perform as intended. The results are summarized as follows:
- . The Sterilant Penetration, Drying Time, Aeration, testing performed as described in ANSI/AAMI/ISO 17665-1:2006/R2013, ISO 17665-2:2009, ANSI/AAMI/ISO 11135-2014, AAMI TIR 12 2010, ANSI/AAMI/ISO 10993-7:2008/R 2012 and ISO 17025:2005. The testing results demonstrate the ability of the PMSSteripack Flat Sterilization pouch (FP) and roll (FL) to effectively allow adequate sterilant penetration to the most difficult areas to reach inside the packaging. The results confirm that the sterilant is able to penetrate the PMSSteripack Flat Sterilization pouch (FP) and roll (FL) and sustain direct contact with the medical instrument inside the package. And the result of aeration time validation test meets the
{16}------------------------------------------------
requirements ANSI/AAMI/ISO 10993-7:2008/R 2012.
- . The Biocompatibility testing performed as described in ANSI/AAMI/ISO 10993-5:2009 and ISO 17025:2005. The testing results demonstrate the PMSSteripack Flat Sterilization pouch (FP) and roll (FL) did not elicit any cytotoxic effect.
- . The Package Integrity, Material Compatibility, Sterility Maintenance testing performed as described in ANSI/AAMI/ISO 11607-1:2006/A1 2014, ASTM F1140/F1140M - 13, ASTM F88-09, ASTM D882-12, ISO 1924-2 , ASTM F1929-12, ASTM D645-97, ASTM F2251-13, ISO 1974-12, ASTM D1922-09, ASTM F1886-09, ASTM F2096-11, ANSI/AAMI/ISO 14937:2009/R2013 and ISO 17025:2005. The testing results demonstrate the ability of the PMSSteripack Flat Sterilization pouch (FP) and roll (FL) to effectively withstand the sterilization process, while allowing sterilization to occur and to retain sterility (Package Integrity) of the sterilized contents till point of use.
- . The Chemical Indicator Efficacy testing was performed as described in ANSI/AAMI/ISO 11140-1:2014 for a class 1 process indicator. The testing demonstrates the ability of the steam and EtO process indicators on the PMSSteripack Flat Sterilization pouch (FP) and roll (FL) to reproducibly meet their performance specifications throughout the shelf life of the pouch and roll. The PMSSteripack Flat Sterilization pouch (FP) and roll (FL) meets the requirements ISO 11140-1:2014.
Any differences between the predicate and proposed device concerning materials, manufacturing or intended use are insignificant and they do not impact the safety or effectiveness of the subject device.
Any differences between the predicate and proposed device performance testing are insignificant and they do not impact the safety or effectiveness of the subject device. The main difference noted is that the PMSSteripack Flat Sterilization pouch (FL) were tested with the latest versions of the same standards.
PMSSteripack Flat Sterilization pouch (FP) and roll (FL) with Chemical Indicator (subject device) and SIGMA Sterilization pouch and roll (predicate device) are both single use devices that are used to enclose another medical device to be sterilized in required sterilization methods by the end user.
PMSSteripack Flat Sterilization pouch (FP) and roll (FL) with Chemical Indicator and predicate device apply substantially equivalent technological characteristics. SIGMA Sterilization pouch and rolls are made from medical grade paper and laminated plastic film by heat sealing. PMSSteripack Sterilization pouch and SIGMA Sterilization pouch and roll have the equivalent material and design features and they all have various size.
PMSSteripack Flat Sterilization pouch (FP) and roll (FL) with Chemical Indicator and predicate device have equivalent sterilization methods (Steam and ETO).
9. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as the legally marketed predicate device (K102158 ), Class II 21 CFR 880.6850 (FRG-JOJ).
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).