K Number
K162203
Date Cleared
2017-04-06

(244 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The POGO Automatic Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The POGO Automatic Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The POGO Automatic Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The POGO Automatic Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The POGO Automatic Blood Glucose Monitoring System is indicated for use in adults and adolescents (13 and up).

POGO Automatic Test Cartridges are for use with the POGO Automatic Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

Device Description

The POGO Automatic Blood Glucose Monitoring System (POGO BGMS) is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip. The system includes multiple glucose-oxidase-based dry-reagent test strips housed in a cartridge and a photometer to read the glucose-dependent color change from the in-use strip. The meter uses calibration information from a barcode on the disposable cartridge to convert the reflectance information into a plasma-equivalent glucose value.

The POGO BGMS automates finger lancing, blood sample collection and placement onto the test strip, and calculation of the blood glucose result, and so requires significantly fewer steps than existing BGMSs to obtain a glucose result. The POGO BGMS accomplishes this via a 10-test cartridge where each foil-sealed test includes a hollow lancet, spring, and test strip. The user does not need a separate lancing device since the lancing mechanism is built into individual test cells of the cartridge. The blood sample is drawn into the test through capillary action and the meter, by monitoring the time and degree of blood spreading on the reagent pad, can detect when the test is under-filled and allow the user a limited amount of time to add blood to the test trip to obtain a glucose result rather than an error. Additionally, used tests are retained within the cartridge for added user convenience. Control solutions are available to confirm correct system performance.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria (from FDA Guidance for Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use, October 2016)Reported Device Performance (POGO BGMS Self-Test Glucose Results)
Accuracy relative to YSI reference:
For glucose concentrations < 75 mg/dL:
All data within ±15 mg/dL bias to the plasma YSI reference results.100% of POGO BGMS self-test glucose results were within ±15 mg/dL of the reference YSI results at a glucose concentration of <75 mg/dL for both Automated and Reapplication results.
For glucose concentrations ≥ 75 mg/dL:
At least 95% of results within ±15% of the YSI reference.At glucose concentrations of ≥75 mg/dL, 95.0% of POGO BGMS self-test glucose results were within ±15% of the reference YSI (95.8% Automated and 91.5% Reapplication).
(Implicitly, better than 95% within ±20% is also typically expected)99.2% of POGO BGMS results were within ±20% of the reference YSI (99.1% Automated and 100% Reapplication).
Reapplication Feature Accuracy (implied by additional studies):The overall results from each study demonstrated at least 95% of results within 15% (for glucose > 75mg/dL) or 10 mg/dL (for glucose < 75mg/dL) of the YSI when used in Reapplication Mode.

The document states, "These results met the predetermined accuracy specifications."

Study Details

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Clinical Lay User Evaluation (initial study):
    • Sample Size: 285 subjects with diabetes.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates it was a "multi-site" study, suggesting it was prospective clinical data.
  • Additional Reapplication Studies:
    • Sample Size: 215 data points were obtained in total.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates these were "clinical studies" where "lay users tested," suggesting prospective clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The ground truth was established using a YSI reference. YSI is a laboratory-grade glucose analyzer, not an expert panel. Therefore, no human experts were involved in establishing the ground truth for comparison.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

There was no human adjudication as the ground truth was established by a YSI laboratory reference.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed. This device is a blood glucose monitoring system, not an AI-assisted diagnostic imaging tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, the studies evaluated the standalone performance of the POGO Automatic Blood Glucose Monitoring System. The results provided are for the device (algorithm and hardware) directly compared to the YSI reference, representing its intrinsic accuracy without human interpretation influencing the numerical glucose value. The "self-test data" implies users operated the device themselves, but the performance metrics are about the device's accuracy.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used was plasma YSI reference results. YSI (Yellow Springs Instruments) analyzers are commonly used as "gold standard" or laboratory reference methods for glucose measurement in clinical studies due to their high accuracy and precision.

8. The sample size for the training set

The document does not provide specific details about a training set for an algorithm. It discusses "nonclinical data" where "IMI obtained 60 POGO BGMS whole blood results via intermittent sampling at nominal glucose levels from 60 to 225 mg/dL. All data were within ±15% bias to the plasma YSI reference results." This sounds more like a verification or early validation step rather than a formal, large-scale training set for a machine learning model. The device's "photometric" detection method and "glucose oxidase" enzyme suggest a more traditional biochemical assay rather than a complex AI algorithm that typically requires a large training dataset in the context of AI/ML devices.

9. How the ground truth for the training set was established

As mentioned above, the available information describes a nonclinical dataset where results were compared against "plasma YSI reference results." This indicates the YSI system was also used as the ground truth for this nonclinical testing, similar to the clinical studies.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 6, 2017

INTUITY MEDICAL, INC. C/O CINDY DOMECUS, R.A.C. PRINCIPAL, DOMECUS CONSULTING SERVICES, LLC 1171 BARROILHET DRIVE HILLSBOROUGH, CA 94010

Re: K162203

Trade/Device Name: POGO Automatic Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. CGA Dated: February 28, 2017 Received: March 1, 2017

Dear Cindy Domecus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162203

Device Name

POGO Automatic Blood Glucose Monitoring System

Indications for Use (Describe)

The POGO Automatic Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The POGO Automatic Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The POGO Automatic Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The POGO Automatic Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The POGO Automatic Blood Glucose Monitoring System is indicated for use in adults and adolescents (13 and up).

POGO Automatic Test Cartridges are for use with the POGO Automatic Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Intuity Medical. The logo features a stylized "i" inside a circle, followed by the word "intuity" in a sans-serif font. Below "intuity" is the word "MEDICAL" in a smaller font. To the right of "intuity" is the phrase "Simplifying Diabetes".

510(k) Summary

1 510(k) Owner

Owner:Intuity Medical, Inc.
Address:526 Almanor AvenueSunnyvale, CA 94085
Phone:(408) 530-1700 x275
Fax:(408) 530-1717
Contact:Robb Hesley

2 Submission Correspondent:

Cindy Domecus, R.A.C. (US & EU) Principal, Domecus Consulting Services LLC (650) 343-4813

3 Date Summary Prepared: February 28, 2017

4 Device name - trade name and common name, and classification

Trade NamePOGO® Automatic™ Blood Glucose Monitoring System
Common NameGlucose Test System
ClassificationClass II
Regulations21 CFR 862.1345
Product CodesNBW, CGA
PanelClinical Chemistry

5 Identification of the legally marketed predicate device

The POGO® Automatic™ Blood Glucose Monitoring System is substantially equivalent to the Contour Next EZ Blood Glucose Monitoring System (Bayer HealthCare LLC, Mishawaka, IN), cleared under K130265 on June 23, 2014.

{4}------------------------------------------------

6 Device Description

The POGO Automatic Blood Glucose Monitoring System (POGO BGMS) is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip. The system includes multiple glucose-oxidase-based dry-reagent test strips housed in a cartridge and a photometer to read the glucose-dependent color change from the in-use strip. The meter uses calibration information from a barcode on the disposable cartridge to convert the reflectance information into a plasma-equivalent glucose value.

The POGO BGMS automates finger lancing, blood sample collection and placement onto the test strip, and calculation of the blood glucose result, and so requires significantly fewer steps than existing BGMSs to obtain a glucose result. The POGO BGMS accomplishes this via a 10-test cartridge where each foil-sealed test includes a hollow lancet, spring, and test strip. The user does not need a separate lancing device since the lancing mechanism is built into individual test cells of the cartridge. The blood sample is drawn into the test through capillary action and the meter, by monitoring the time and degree of blood spreading on the reagent pad, can detect when the test is under-filled and allow the user a limited amount of time to add blood to the test trip to obtain a glucose result rather than an error. Additionally, used tests are retained within the cartridge for added user convenience. Control solutions are available to confirm correct system performance.

7 Indications for Use

The POGO Automatic Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The POGO Automatic Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The POGO Automatic Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The POGO Automatic Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. The POGO Automatic Blood Glucose Monitoring System is indicated for use in adults and adolescents (13 and up).

POGO Automatic Test Cartridges are for use with the POGO Automatic Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

{5}------------------------------------------------

CharacteristicPOGO BGMSContour Next EZK130265
Intended UseQuantitative measurement ofblood glucoseSame
Indications For UseThe POGO Automatic BloodGlucose Monitoring System isintended to be used for thequantitative measurement ofglucose (sugar) in freshcapillary whole blood samplesdrawn from the fingertips. ThePOGO Automatic BloodGlucose Monitoring System isintended to be used by a singleperson and should not beshared.The POGO Automatic BloodGlucose Monitoring System isintended for self testingoutside the body (in vitrodiagnostic use) by people withdiabetes at home as an aid tomonitor the effectiveness ofdiabetes control. The POGOAutomatic Blood GlucoseMonitoring System should notbe used for the diagnosis of orscreening of diabetes or forneonatal use. The POGOAutomatic Blood GlucoseMonitoring System is indicatedfor use in adults andadolescents (13 and up).POGO Automatic TestCartridges are for use with thePOGO Automatic Meter toquantitatively measure glucose(sugar) in fresh capillary wholeblood samples drawn from thefingertips.The Contour Next EZ bloodglucose monitoring system isan over the counter (OTC)device utilized for self-testingby persons with diabetes athome for the quantitativemeasurement of glucose inwhole blood, is for single-patient use only, and shouldnot be shared. The ContourNext EZ blood glucosemonitoring system is indicatedfor use with fresh fingertipcapillary whole blood samples.The clinical utility of thisdevice is to aid in monitoringthe effectiveness of yourdiabetes control program. TheContour Next EZ bloodglucose monitoring system isnot intended for use for thediagnosis of or screening fordiabetes mellitus and is notintended for use on neonates.The Contour Next test stripsare intended for self-testing bypersons with diabetes for thequantitative measurement ofglucose in whole bloodsamples from 20 to 600mg/dL.
Rx/OTCOTCSame
CharacteristicPOGO BGMSContour Next EZK130265
ClassificationRegulation862.1345Same
Product CodesNBW, CGANBW, LFR
Product DesignBattery-powered handheldmeterSame
Delivery of Blood toReagent Pad viaCapillary ActionYesSame
Automatic BloodSample CollectionYesNo
Detects Under-filledReagent PadYesSame
Allows BloodReapplicationto Complete TestYes, provides glucose result orerror code after blood additionto under-filled stripSame
Time period allowedfor patient to addmore blood whenreagent pad is under-filled15 seconds20 seconds
Time period allowedfor patient to applyblood when blooddoes not reachreagent pad90 seconds180 seconds
Meter enforces limitson strip exposure?Yes, device design preventsstrip use after 5 minuteexposure (open cell timewindow)No, device design does notprevent strip re-insertionunless blood or seriousmishandling is detected
Patient able to re-uselancet?No, device design preventsre-useYes, device design does notprevent re-use
Cal CodingAutomatedSame
Detection MethodPhotometric (LED)Electrochemical
EnzymeGlucose oxidaseGlucose dehydrogenase
Hematocrit Range20% to 60%15% to 65%
Humidity Range10% to 90% RH10% to 93% RH
CharacteristicPOGO BGMSContour Next EZ K130265
Lancing DeviceIntegrated in Meter withStrip/Lancet in CartridgeSeparate from Meter
Measurement Range20 - 500 mg/dL20 - 600 mg/dL
Minimum Sample Size0.25 µL0.6 µL
Monitor Memory500 results480 results
Operating Temperature Range50 to 104° F41 to 113° F
Principle of OperationGlucose is oxidized by theenzyme glucose oxidase andelectrons are transferred toform the intermediatehydrogen peroxide.Horseradish peroxidasecatalyzes the transfer ofelectrons between hydrogenperoxide and precursor dyemolecules to form a blueoxidation product, theconcentration of which isproportional to theconcentration of glucose in thesample. The amount of colorformed is read by a photometerin the meter.Glucose is oxidized by theenzyme glucosedehydrogenase-FAD andelectrons are transferred toform a reduced mediatorintermediate. The reducedmediator transfers electrons toan electrode to produce anelectric current, which isproportional to theconcentration of glucose in thesample. The current is read bya potentiometer in the meter.

Technological Characteristics Comparison 8

{6}------------------------------------------------

{7}------------------------------------------------

Comparison of Intended Use

POGO BGMS and the listed predicate device have the same intended use; namely, the quantitative measurement of blood glucose. Both devices use fingerstick capillary whole blood as the sample. Both devices are for single-patient use in the home.

Comparison of Technological Characteristics

The technology used in the POGO BGMS is similar to that used in the predicate device. Similar elements include: a capillary channel to automate blood delivery from the fingerstick wound to the reagent section of the test strip, a means to detect when an insufficient amount of blood has been applied to the test strip, and the ability to add blood to an in-process test until either the meter has sufficient sample or the time allowed to add blood has expired. The test strips are all dry-reagent chemistries specific for glucose in samples of whole blood delivered via capillary action.

{8}------------------------------------------------

Differences between the POGO BGMS and the predicate device include: a different enzyme in the test strip (glucose oxidase in POGO BGMS and glucose dehydrogenase in the predicate), a different signal transduction system (photometric in POGO BGMS, electrochemical in predicate), a new, sterile, single-use and auto-disabling lancet is provided with every test in the POGO BGMS versus user-controlled lancet replacement or re-use in the predicate, smaller minimum sample volume in POGO BGMS (0.25 uL versus 0.6 uL in the predicate), and an automatic storage of used test strips and lancets within the cartridge for POGO BGMS versus user-required disposal of at least the test strip after every test in the predicate. The POGO BGMS also prevents use of a test strip that has been exposed to the environment for more than five minutes (open cell time window), whereas the predicate device has no means to measure strip exposure time and does not prevent strip re-insertion until the strip contains blood or has been seriously degraded by moisture.

9 Brief Discussion of Nonclinical Data

The software change that is the subject of this 510(k) was validated.

In accordance with FDA's Guidance for Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use, issued in October, 2016, IMI obtained 60 POGO BGMS whole blood results via intermittent sampling at nominal glucose levels from 60 to 225 mg/dL. All data were within ±15% bias to the plasma YSI reference results.

10 Brief Description of Clinical Data

In support of the Reapplication feature of the POGO BGMS. Intuity Medical, Inc. (IMI) has conducted three clinical studies. The initial study was a multi-site Clinical Lay User Evaluation where some of the system accuracy data was from users who completed the test using Reapplication. The two subsequent studies focused on collecting additional data specifically from the system in the Reapplication Mode using native fingerstick capillary whole blood samples.

In the Clinical Lay User Evaluation, self-test data from 285 subjects with diabetes, from a total of four cartridge lots, were evaluated for accuracy versus plasma from a fingerstick sample analyzed on a YSI. A total of 100% of POGO BGMS self-test glucose results were within ±15 mg/dL of the reference YSI results at a glucose concentration of <75 mg/dL for both Automated and Reapplication results. At glucose concentrations of ≥75 mg/dL, 95.0% of POGO BGMS self-test glucose results were within ±15% of the reference YSI (95.8% Automated and 91.5% Reapplication) and 99.2% of POGO BGMS results were within ±20 % of the reference YSI (99.1% Automated and 100% Reapplication). These results met the predetermined accuracy specifications.

Following the Clinical Lay User Evaluation, additional clinical studies were conducted where lay users tested with the POGO BGMS in Reapplication Mode to evaluate the accuracy of the POGO BGMS glucose result versus a YSI reference. A total of 215 data points were obtained over a range of native glucose levels from 48 to 396 mg/dL. The overall results from each study demonstrated at least 95% of results within 15%

{9}------------------------------------------------

(for glucose > 75mg/dL) or 10 mg/dL (for glucose < 75mg/dL) of the YSI. These studies confirmed that the POGO BGMS provides accurate results when used in the Reapplication Mode.

11 Conclusions from Nonclinical and Clinical Testing

Results from the nonclinical and clinical testing using the POGO BGMS demonstrate that the POGO BGMS is substantially equivalent to the predicate device. The POGO BGMS is safe and effective for its intended use and performs as well as or better than the predicate device.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.