(16 days)
The SIDEKICK® and USHER® Support Catheters are single lumen catheters intended to create a pathway for other devices in the peripheral vasculature.
The SIDEKICK® and USHER® Support Catheters are single lumen catheters with a standard luer fitting hub and separate attachable hemostatic valve to support the CROSSER® CTO Recanalization Catheters 14S and S6. The product hub identifies SD for SIDEKICK® Catheter, USH for USHER® Catheter, A for Angled and T for Tapered in addition to the sheath profile and working length in centimeters. A guidewire introducer is provided to facilitate the guidewire passage through the optional hemostatic valve. The guidewire introducer shaft color matches the shaft color of the recommended support catheter.
The SIDEKICK® Catheter is available in straight, angled, tapered and non-tapered configurations in 70cm and 110cm working lengths. The USHER® Catheter is tapered and is available in straight and angled configurations in 83cm and 130cm working lengths. The SIDEKICK® and straight USHER® Catheters have a single radiopaque marker 1mm from the distal tip. The angled USHER® Catheter configurations have three radiopaque markers at the distal tip for enhanced visualization of the catheter tip and angle under fluoroscopy. The third proximal radiopaque marker is located 15mm from the distal tip.
The GEOALIGN® Marking System has been added to the device. The GEOALIGN® Marking System is a non-radiopaque ruler on the catheter shaft measured from the distal tip. The GEOALIGN® Markings are designated on the catheter shaft by 1cm increment bands with an accuracy within ±1mm. The distance from the distal catheter tip is labeled in 10cm increments. Thicker bands denote the midway point (5cm) between the labeled distances. The GEOALIGN® Marking System is designed to be used as a tool to externally measure the intravascular advancement and/or retraction of the catheter. This can provide an intravascular reference regarding the location of the distal tip of the catheter or an approximate intravascular length measurement between two points. The GEOALIGN® Marking System may also facilitate geographic alignment of an adjunctive therapy that includes the same GEOALIGN® Marking System.
The provided text describes a 510(k) premarket notification for the SIDEKICK® and USHER® Support Catheters, focusing on the addition of the GEOALIGN® Marking System. This is a medical device submission, and the "acceptance criteria" and "device performance" mentioned in the request refer to the performance testing of the device for regulatory approval, not typically the performance of an AI algorithm or a clinical study in the traditional sense.
Based on the document, here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly present a table with acceptance criteria thresholds and discrete performance values in the way one might expect for an AI algorithm. Instead, it describes general performance testing and concludes that the device is "comparable" and "substantially equivalent" to the predicate device.
Test Performed | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
GEOALIGN® Marking Legibility | Markings are legible. | Evaluated and found satisfactory to ensure visibility. The results demonstrate that the technological characteristics and performance criteria... are comparable to the predicate device and that it performs substantially equivalent to the legally marketed predicate device. |
Dimensional Analysis: | Markings are within specified tolerances for location and spacing. | GEOALIGN® Marking Location (Distal Position): Within ±1mm accuracy for 1cm increment bands. GEOALIGN® Marking Location (Proximal Position): Within ±1mm accuracy for 1cm increment bands. GEOALIGN® Marking Spacing: Within ±1mm accuracy for 1cm increment bands. |
Catheter Outer Diameter (OD) | Within specified tolerances. | Evaluated and found satisfactory. The results demonstrate that the technological characteristics and performance criteria... are comparable to the predicate device and that it performs substantially equivalent to the legally marketed predicate device. |
GEOALIGN® Marking durability with Introducer Sheaths | Markings maintain integrity after interaction with introducer sheaths. | Evaluated and found satisfactory. The results demonstrate that the technological characteristics and performance criteria... are comparable to the predicate device and that it performs substantially equivalent to the legally marketed predicate device. |
GEOALIGN® Marking compatibility with Introducer Sheaths | Markings function correctly and safely with introducer sheaths. | Evaluated and found satisfactory. The results demonstrate that the technological characteristics and performance criteria... are comparable to the predicate device and that it performs substantially equivalent to the legally marketed predicate device. |
Biocompatibility and Chemical Characterization | Device is biocompatible and does not elicit substances at concerning levels. | Per ISO 10993-1:2009 CORR 1, biocompatibility and chemical characterization demonstrate the subject device is biocompatible and does not elicit any substances at levels of concern as result of this change. |
2. Sample size used for the test set and the data provenance:
The document does not specify the exact sample sizes used for each of the performance tests (e.g., how many catheters were tested for marking legibility, durability, or dimensions). It refers generally to "nonclinical tests." The data provenance is internal testing performed by the manufacturer, Bard Peripheral Vascular, Inc., in the USA (Tempe, Arizona). These are laboratory and engineering tests, not clinical data from patients. The study is not a clinical study; it's a non-clinical, benchtop performance evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable to the type of testing described in the document. The "ground truth" here is defined by engineering specifications, measurement standards, and regulatory requirements (e.g., ISO 10993 for biocompatibility), not by expert consensus on clinical data. No "experts" were used in the sense of clinical reviewers to establish ground truth for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This is not applicable. The tests performed are objective, measurable non-clinical tests (e.g., dimensional measurements, durability tests, biocompatibility assays). There is no "adjudication" in the sense of conflicting expert opinions on a clinical outcome.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This document is for a physical medical device (support catheters), not an AI algorithm. Therefore, no MRMC study or assessment of human readers with/without AI assistance was performed or described.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This document is for a physical medical device (support catheters), not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for the performance tests is based on:
- Engineering specifications and design requirements: For dimensional accuracy, legibility, and durability.
- Established scientific and regulatory standards: Such as ISO 10993-1:2009 CORR 1 for biocompatibility.
8. The sample size for the training set:
Not applicable. This document describes the testing of a physical medical device, not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established:
Not applicable, as there is no AI algorithm training set.
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).