K Number
K161966

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-09-08

(52 days)

Product Code
Regulation Number
892.1680
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The XRpad2 4336 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.

Device Description

The XRpad2 4336 HWC-M is a lightweight, cassette-sized, flat panel X-ray detector for digital radiography. The X-ray detector consists of an amorphous silicon flat panel with a directly deposited CsI:T1 scintillator and dedicated read-out, scan, and control electronics, all packaged in a carbon-fiber and aluminum enclosure. The outside dimensions of the detector are 460.0 mm × 383.6 mm × 15.5 mm, which fits into a standard X-ray cassette Bucky.

The detector can be integrated into a fixed room X-ray system to enable digital radiography.

AI/ML Overview

This document describes the 510(k) summary for the PerkinElmer XRpad2 4336 HWC-M Flat Panel Detector, which seeks to establish substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a quantitative table format with corresponding "reported device performance" values for the XRpad2 4336 HWC-M. Instead, it relies on demonstrating equivalence to a predicate device (PerkinElmer XRpad 4336 MED, K140551) through non-clinical testing and stating that the physical values are "comparable."

However, a comparison chart for device characteristics is provided (Page 5), which implicitly indicates the performance metrics considered for demonstrating equivalence. The "acceptance criteria" for the new device are essentially that its performance is equivalent to the predicate.

CharacteristicPredicate Device Performance (PerkinElmer XRpad 4336 MED, K140551)Proposed Device Performance (PerkinElmer XRpad2 4336 HWC-M)Reference/Comment (Implicit Acceptance Criteria)
Intended Use / Indications for UseAs described for predicateSameMust be the same as the predicate.
PanelSingle substrate amorphous silicon active TFT/diode arraySameMust be the same.
ScintillatorDirect deposition CsI:TlSameMust be the same.
Pixel pitch100 µmSameMust be the same.
Pixel matrix3556 × 4320 pixels3524 × 4288 pixelsSlightly different, but not noted as impacting image quality or substantial equivalence.
Limiting resolution5 lp/mmSameMust be the same.
Binning capable2 × 2 binning for 200 µmSameMust be the same.
Data transmit area355 mm × 432 mm352 mm × 429 mmSlightly different, corresponding to pixel matrix change, but not noted to impact image quality.
External dimensions (w×l×h)384 mm × 460 mm × 15 mm384 mm × 460 mm × 15.5 mmMinimal difference, attributed to housing update.
Weight3.8 kg3.1 kgDifferent, due to housing update, and stated not to impact image quality.
Housing materialAluminum with carbon-fiberSameMust be the same.
FinishMatte carbon-fiber front with white silk-screen active area markersSameMust be the same.
Communication interfaceGb Ethernet or 802.11n WiFiSameMust be the same.
WiFi band5.1-5.3 GHz5.1-5.9 GHzExpanded, but considered equivalent for networking.
X-ray synchronization interfaceDedicated trigger in/out signal lines or Automatic Exposure DetectionSameMust be the same.
PowerExternal power supply or batterySameMust be the same.
Software libraryWindows OSWindows and Linux OSExpanded OS support, but functionality considered equivalent.
DQE and MTFNot explicitly stated, but established for predicateComparable to predicate deviceMust be comparable to the predicate device to demonstrate substantial equivalence (non-clinical test).

2. Sample Size Used for the Test Set and Data Provenance

A new clinical study was not conducted for the XRpad2 4336 HWC-M device. The clinical data supporting the predicate device, XRpad 4336 MED, was derived from testing for K122495 (XRD 1622 AP3 MED). Therefore, no specific "test set" sample size or data provenance is provided for the current device's clinical performance. The current device relies on demonstrating that its differences "do not invalidate the applicability of the clinical study data submitted in K122495."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

Since a new clinical study was not conducted for this device, and it relies on previous clinical data, this information is not provided in the document. The document refers to the predicate clearance (K140551) which itself references K122495 for clinical data. Details on experts and ground truth for K122495 are not in this document.

4. Adjudication Method for the Test Set

As no new clinical study test set was used for this device, a specific adjudication method is not described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done for the XRpad2 4336 HWC-M. The submission focuses on non-clinical equivalence to a predicate device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This device is a flat panel X-ray detector, not an AI algorithm. Therefore, a standalone (algorithm only) performance study as typically understood for AI/CAD systems is not applicable or performed. Its performance is measured by image quality metrics (like DQE, MTF) and compared to a predicate device.

7. Type of Ground Truth Used

For the current device, direct clinical ground truth (e.g., pathology, outcomes data) was not established in a new study. The device's "ground truth" for clinical performance is indirectly established by demonstrating its equivalence to a predicate device whose clinical effectiveness was previously established, likely through methods typical for X-ray detectors (e.g., image quality assessment, possibly expert consensus on diagnostic images from K122495).

8. Sample Size for the Training Set

This device is an X-ray detector, not a machine learning algorithm that requires a "training set" in the conventional sense. The document describes engineering, design, and performance validation, not algorithmic training.

9. How the Ground Truth for the Training Set Was Established

As this is not an AI/ML device requiring a training set, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 8, 2016

PerkinElmer, Inc. % Ms. Dawn Spooner Associate Director, Regulatory Affairs 940 Winter Street WALTHAM MA 02451

Re: K161966

Trade/Device Name: PerkinElmer XRpad2 4336 HWC-M Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: August 15, 2016 Received: August 17, 2016

Dear Ms. Spooner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K161966

Device Name PerkinElmer XRpad2 4336 HWC-M Flat Panel Detector

Indications for Use (Describe)

The XRpad2 4336 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

14

{3}------------------------------------------------

510(k) Summary XRpad2 4336 HWC-M Flat Panel Detector

This summary of 510(k) safety and effectiveness information is supplied in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K161966

August 15, 2016

Submitted By:PerkinElmer Medical Imaging 2175 Mission College Blvd. Santa Clara, CA 95054 U.S.A.
Contact Person:Dawn Spooner Associate Director, Regulatory Affairs Tel. 781-663-6071 Fax. 781-663-5969
Device Name:PerkinElmer XRpad2 4336 HWC-M Flat Panel Detector
Classification:Product Code:MQBClassification Name:Stationary X-ray SystemClassification Regulation:21 CFR 892.1680Product Code:MQBClassification Name:Stationary X-ray SystemClassification Regulation:21 CFR 892.1680
Product Code:MQB
Classification Name:Stationary X-ray System
Classification Regulation:21 CFR 892.1680
Predicate Device:PerkinElmer XRpad 4336 MED Flat Panel Detector
510(k) Clearance:K140551; August 1, 2014Product Code:MQBClassification Name:Stationary X-ray SystemClassification Regulation:21 CFR 892.1680510(k) Clearance:K140551; August 1, 2014Product Code:MQBClassification Name:Stationary X-ray SystemClassification Regulation:21 CFR 892.1680
510(k) Clearance:K140551; August 1, 2014
Product Code:MQB
Classification Name:Stationary X-ray System
Classification Regulation:21 CFR 892.1680

{4}------------------------------------------------

Device Description:

The XRpad2 4336 HWC-M is a lightweight, cassette-sized, flat panel X-ray detector for digital radiography. The X-ray detector consists of an amorphous silicon flat panel with a directly deposited CsI:T1 scintillator and dedicated read-out, scan, and control electronics, all packaged in a carbon-fiber and aluminum enclosure. The outside dimensions of the detector are 460.0 mm × 383.6 mm × 15.5 mm, which fits into a standard X-ray cassette Bucky.

The detector can be integrated into a fixed room X-ray system to enable digital radiography. The following accessories are available for the XRpad2 4336 HWC-M:

  • XRpad LBC-2 (Lithium Battery Charger) ●
  • XRpad IPU-2 (Interface and Power Unit) .
  • XRpad LPT2 Detector Cable ●
  • XRpad Protective Insert 4336 ●
  • AC Cable IEC 60320 C13 DE
  • AC Cable IEC 60320 C13 US
  • Trigger Cable 5 m/16.5 ft.
  • Trigger Cable 20 m/65.5 ft. ●
  • GigE Interface Cable 7.6 m/25 ft. ●
  • GigE Interface Cable 15.25 m/50 ft. ●
  • GigE Interface Cable 30.5 m/100 ft. .

Indications For Use:

The XRpad2 4336 HWC-M, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.

{5}------------------------------------------------

Comparison Chart

CharacteristicsModelPerkinElmerXRpad 4336 MED(K140551)Proposed ModelPerkinElmerXRpad2 4336 HWC-M
Intended Use /Indications forUseThe XRpad 4336 MED when usedwith a radiographic imaging system,is indicated for use in generatingradiographic images of humananatomy for diagnostic X-rayprocedures, wherever conventionalscreen-film (SF), digital radiography(DR), or computed radiography(CR) systems may be used. It is notintended for mammographic use.Same
PanelSingle substrate amorphous siliconactive TFT/diode arraySame
ScintillatorDirect deposition CsI:TlSame
Pixel pitch100 μmSame
Pixel matrix3556 × 4320 pixels3524 × 4288
Limitingresolution5 lp/mmSame
Binning capable2 × 2 binning for 200 μmSame
Data transmitarea355 mm × 432 mm352 mm × 429 mm
Externaldimensions(w×l×h)384 mm × 460 mm × 15 mm384 mm × 460 mm × 15.5 mm
Weight3.8 kg3.1 kg
HousingmaterialAluminum with carbon-fiberSame
FinishMatte carbon-fiber front with whitesilk-screen active area markersSame
CommunicationinterfaceGb Ethernet or 802.11n WiFiSame
WiFi band5.1-5.3 GHz5.1-5.9 GHz
X-raysynchronizationinterfaceDedicated trigger in/out signal linesor Automatic Exposure DetectionSame
PowerExternal power supply or batterySame
SoftwarelibraryWindows OSWindows and Linux OS

{6}------------------------------------------------

Summary of Studies:

The proposed new device and the XRpad predicate device utilize similar technology and materials, are similar in design and construction, and have the same intended use. The construction and physical characteristics of the PerkinElmer XRpad2 4336 HWC-M compared to the predicate PerkinElmer XRpad 4336 MED are similar. The difference is related to updates to the housing design which results in a lighter detector and does not impact image quality.

The PerkinElmer XRpad2 4336 HWC-M flat panel detector has successfully completed internal non-clinical testing, complies with standards and regulations such as UL and IEC. A new clinical study was not required for the XRpad2 4336 HWC-M device. The predicate device, XRpad 4336 MED, was cleared using clinical data derived from testing to support K122495 (XRD 1622 AP3 MED) and applicable to the predicate device. The XRpad2 4336 HWC-M device differences do not invalidate the applicability of the clinical study data submitted in K122495. Through non-clinical testing, we have demonstrated the clinical data collected with the (XRD 1622 AP3 MED) in K122495 is also applicable to the XRpad2 4336 HWC-M device as it was applicable for the XRpad 4336 MED device.

Summary of Design Control and Risk Management:

The XRpad2 4336 HWC-M flat panel X-ray detector is a modification of the XRpad 4336 MED (K140551). The modifications are related to the decrease in dimensions and do not impact image quality.

The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted.

Summary of Non-Clinical Data:

To demonstrate the equivalence of the PerkinElmer XRpad2 4336 HWC-M to the predicate device. PerkinElmer has performed internal non-clinical testing and demonstrated compliance with accepted standards and regulations such as UL and IEC following the non-clinical considerations outlined in the Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, August, 1999. The non-clinical testing was conducted and demonstrated the main physical values of the XRpad2 4336 HWC-M such as DOE and MTF are comparable to the predicate device. The non-clinical performance testing conducted and resulting data demonstrate substantial equivalence.

{7}------------------------------------------------

Substantial Equivalency:

The proposed device and predicate device (XRpad 4336 MED flat panel detector) both utilize similar technology and materials, are similar in design and construction, and have been shown to produce images of equivalent diagnostic quality. Both devices are intended for use in generating radiographic images of human anatomy for diagnostic Xray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. The devices are not intended for mammographic use. Both devices produce digital images which can be transmitted to imaging software of the X-ray unit.

Conclusion:

Similar to the predicate device, the XRpad2 4336 HWC-M has comparable performance and operational standards. Potential hazards have been studied and controlled by a Risk Management Plan. The non-clinical software verification and validation test results demonstrate that the PerkinElmer XRpad2 4336 HWC-M complies with international and FDA recognized consensus standards and meets the acceptance criteria and is adequate for its intended use.

Based on the information supplied in this 510(k) PerkinElmer concludes, the XRpad2 4336 HWC-M is substantially equivalent to the currently marketed device, XRpad 4336 MED (K140551) in terms of safety and effectiveness.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.