K Number
K161255

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-01-19

(261 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The meso-relle needles are intended to inject fluids intradermally.

Device Description

The meso-relle® needles are single lumen needle intended to inject fluids intradermally. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe. Needles have different length, which make them suitable to inject fluids intradermally. The meso-relle are not intended for injection in the bloodstream. The devices do not contact the central nervous system. The meso-relle® needles are suitable for administration of fluids intradermally. Each needle is provided with a protective cap (cover is rigid and not colored. The dimensions are suitable to accommodate the corresponding needle and vary according to the dimensions of the needle. The meso-relle® needles are disposable single use devices, sold sterilized by ethylene oxide.

AI/ML Overview

The provided document describes the acceptance criteria and the study conducted for the meso-relle needles (AAL34, AAL36, AM30G), a hypodermic single lumen needle intended to inject fluids intradermally.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present acceptance criteria in a dedicated table format with corresponding numerical performance values. Instead, it states that the meso-relle needles were tested according to specific ISO standards and "met the acceptance criteria." The device's performance is reported indirectly by stating its compliance with these standards and demonstrating substantial equivalence to predicate devices.

However, based on the non-clinical performance data section and the comparison table, we can infer the acceptance criteria are met by conforming to the specified ISO standards and achieving substantial equivalence to the predicate devices in terms of materials, design, and functional characteristics.

Acceptance Criteria (Inferred from Standards & Predicate Comparison)Reported Device Performance (Compliance Statement)
Biocompatibility: Meets requirements of ISO 10993-1.Passed tests for Cytotoxicity, Sensitization, Intracutaneous reactivity, Systemic toxicity, and Haemolysis.
Sterility: Achieves SAL 10-6 via Ethylene Oxide (EO) sterilization.Sterilization method complies with AAMI / ANSI / ISO 11135-1:2007. EO and ECH residuals met acceptance criteria per UNI EN ISO 10993-7:2008.
Functional Performance (Hypodermic Needles): Meets requirements of ISO 7864.Tested according to ISO 7864:2016 and deemed to show substantial equivalence.
Functional Performance (Stainless Steel Needle Tubing): Meets requirements of ISO 9626.Tested according to ISO 9626:2016 and deemed to show substantial equivalence.
Functional Performance (Conical Fittings - Luer taper): Meets requirements of ISO 594-1.Tested according to ISO 594-1:1986 and deemed to show substantial equivalence.
Material Composition: AISI 304 Stainless Steel cannula, Polypropylene hub.Same or similar to predicate devices.
Tip Configuration: Triple sharpened, non-coring.Same as the MV pilot needle and reference predicate device.
Connection to Syringe: Luer taper.Same as predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for each non-clinical performance test. It mentions that "tests were performed on the final finished device," suggesting a representative sample was used for each test (biocompatibility, sterility, functional testing).

The data provenance is retrospective, as the tests were conducted on the finished devices to demonstrate compliance with existing standards and equivalence to already marketed predicate devices. The country of origin of the data is not explicitly stated, but the manufacturer (Biotekne S.r.l.) and contact person (Enrico Bisson) are located in ITALY.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable to this type of device and study. The ground truth for hypodermic needles is established by compliance with international engineering and medical device standards (e.g., ISO standards for materials, sterility, and functional performance) and comparison to established predicate devices, not by expert consensus on clinical images or outcomes.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation of medical data (e.g., radiology images) where there may be disagreement among experts. This study focuses on objective engineering and performance testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not the focus of this device (a hypodermic needle).

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No, a standalone algorithm performance study was not done. This device is a physical medical instrument, not an AI algorithm.

7. Type of Ground Truth Used

The ground truth used for this device is based on:

  • International Standards: Compliance with ISO 10993-1 (Biocompatibility), AAMI / ANSI / ISO 11135-1:2007 and UNI EN ISO 10993-7:2008 (Sterility), ISO 7864:2016 (Hypodermic Needles), ISO 9626:2016 (Stainless Steel Needle Tubing), and ISO 594-1:1986 (Luer Taper Fittings).
  • Performance of Predicate Devices: The established safe and effective performance of the legally marketed predicate devices (K110606, MV Intradermic Needles and K051783, Artsana Hypodermic Needles). The subject device demonstrates "substantial equivalence" to these predicates.

8. Sample Size for the Training Set

This information is not applicable. There is no "training set" in the context of validating a physical medical device like a hypodermic needle. Training sets are used for machine learning algorithms.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 19, 2017

Biotekne S.r.1. % Enrico Bisson President Studio di Ingegneria Enrico Bisson Via Marzia 9 Abano Terme, Padova 35031 ITALY

Re: K161255

Trade/Device Name: meso-relle (AAL34, AAL36, AM30G) Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: December 21, 2016 Received: December 22, 2016

Dear Enrico Bisson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/1/Picture/8 description: The image contains a signature and the name "Tina Kiang-S". The signature is a stylized, looping design to the left of the name. The name is written in a simple, sans-serif font, with "Tina" stacked above "Kiang-S".

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161255

Device Name meso-relle AAL34, AAL36, AM30G

Indications for Use (Describe) The meso-relle needles are intended to inject fluids intradermally.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Representation Use (Part 21 CFR 601 Subpart D) ☐ Compassionate Use (21 CFR 601 Subpart E)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 05 510(k) Summary

510 (k) Summary

K161255

APPLICANT

Company Name:Biotekne S.r.l.
Company Address:Via della Bastia, 9
40033 Casalecchio di Reno (BO) - ITALY
Company Phone:+39 051565211
Company Fax:+39 0516182524
Company e-mailinfo@biotekne.it
CONTACT PERSON:Enrico Bisson - Studio ingegneria Enrico Bisson
Via Marzia n. 9
Abano Terme (PD) - ITALY
Contact Phone:+39 0498630080
Contact Fax+39 0498630080
Contact E-mailenrico.bisson@gmail.com

Date Summary Prepared: January 17, 2017

DEVICE IDENTIFICATION

  • A. Trade name: meso-relle AAL34, AAL36, AM30G
  • B. Generic/ Common Name: Hypodermic Needle
  • ். Classification name: Hypodermic single lumen needle, 21 CFR 880.5570, Class II
  • D. Product Code: FMI

LEGALLY MARKETED DEVICES (PREDICATE DEVICES)

K110606, MV Intradermic Needles, MV S.R.L.

Reference Predicate Device: K051783, Artsana Hypodermic Needles, ARTSANA SPA.

{4}------------------------------------------------

INDICATIONS FOR USE

The meso-relle needles are intended to inject fluids intradermally.

DEVICE DESCRIPTION

The meso-relle® needles are single lumen needle intended to inject fluids intradermally. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe. Needles have different length, which make them suitable to inject fluids intradermally. The

meso-relle are not intended for injection in the bloodstream. The devices do not contact the central nervous system. The meso-relle® needles are suitable for administration of fluids intradermally.

Each needle is provided with a protective cap (cover is rigid and not colored. The dimensions are suitable to accommodate the corresponding needle and vary according to the dimensions of the needle.

The meso-relle® needles are disposable single use devices, sold sterilized by ethylene oxide.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The meso-relle Needles are same or similar in materials, design and intended use to the predicate devices. Both the subject and the predicate devices are intended to inject fluids intradermally; they have the same indications for use. Each of the technical attributes of the meso-relle Needles are present in the predicate devices. The materials, tip configuration and other fundamental design characteristics are all the same. Comparison elements are as follows:

{5}------------------------------------------------

ATTRIBUTE /meso-relle NEEDLESMV Intradermic NeedlesArtsana HypodermicComparison
CHARACTERISTICS(Submitted Product)(Legally MarketedNeedles (Reference
Predicate Device)Predicate Device)
'K" numbersK161255K110606K051783
CFR Section880.5570880.5570880.5570same
Pro-codeFMIFMIFMIsame
Classification nameHypodermic single lumen needleHypodermic single lumen needleHypodermic single lumen needlesame
Intended / IndicationsFor UseThe meso-relle needles areintended to inject fluidsintradermally.The MV intradermic needlesare intended to inject fluidsintradermally.Artsana disposable sterilehypodermic needles areintended for use to inject fluidsinto, or withdraw fluids from,parts of the body below thesurface of the skin.same as the primarypredicate
Cannula materialAISI 304 Stainless SteelAISI 304 Stainless SteelAISI 304 Stainless Steelsame
Hub materialPolypropylene (PolypropyleneMG03MA)Polypropylene (PolypropyleneMG03MA)Polypropylenesame
Hub colorYellowCAS No.22094-93-5,Dispersant (CAS No.557-04-0)internal material code:SCP-19906color coded ISO 6009YellowCAS No.22094-93-5,Dispersant (CAS No.557-04-0)internal material code:SCP-19906color coded ISO 6009Not availablesame as the primarypredicate
Coverrigid cover, polypropylene orpropylene\ethylenecopolymerrigid cover, polyethylenerigid cover, non toxicpropylenesame or similar
Adhesiveepoxy type adhesiveepoxy type adhesiveepoxy type adhesivesame
Lubricantmedical grade silicone oilmedical grade silicone oilmedical grade silicone oilsame
ATTRIBUTE /CHARACTERISTICSmeso-relle NEEDLES(Submitted Product)MV Intradermic Needles(Legally MarketedPredicate Device)Artsana HypodermicNeedles (ReferencePredicate Device)Comparison
'K" numbersK161255K110606K051783
Needle diameter (gauge)30G22G, 25G, 26G, 27G, 30G /pilot needle: 21G, 23G, 26G,27G27G, 30GThe needle diameter of thesubject device is includedwithin the configuration ofthe predicate devices
Needle length (mm)4, 6, 1225, 27,35, 37, 40, 50, 57, 70 /pilot needle: 13, 254, 6, 13The differences in the needlelength between the subjectand the primary predicatedevice do not raise newquestions of safety oreffectiveness for the subjectdevice. Shorter lengthneedles are more appropriatefor intradermal use.
Tip configurationtriple sharpened, non-coringInjection needle: closed blunttip, lateral openingPilot needle: triple sharpened,non-coringtriple sharpened, non-coringsame as the MV pilot needleand as the referencepredicate device
Connection to syringe orinjection deviceLuer taperLuer taperLuer tapersame
SterilizationEthylene Oxide (EO)Per ISO 11135-1:2007Ethylene Oxide (EO)Per ISO 11135-1:2007Not availablesame as the primarypredicate
SAL LevelSAL 10-6SAL 10-6Not availablesame as the primarypredicate
BiocompatibilityISO 10993-1ISO 10993-1Not availablesame as the primarypredicate
ATTRIBUTE /CHARACTERISTICSmeso-relle NEEDLES(Submitted Product)MV Intradermic Needles(Legally MarketedPredicate Device)Artsana HypodermicNeedles (ReferencePredicate Device)Comparison
"K" numbersK161255K110606K051783
Functional testingISO 7864:2016ISO 9626:2016ISO 594-1:1986ISO 7864:1993ISO 9626:1991ISO 594-1:1986Not availablesame as the primarypredicate

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

The shorter length of the subject device do not raise new issues of safety or effectiveness; shorter lengths are even more suitable for intradermal use as they are less invasive. This difference does not affect the substantial equivalence of the subject device.

The test results and comparison results show that the subject device is substantial equivalent to the predicate in performance.

NON-CLINICAL PERFORMANCE DATA

All materials used for the meso-relle needles have a long history of safe use for the same or equivalent intended use. Biocompatibility has been tested according to the requirements of ISO 10993-1. In consideration of the International Standard ISO 10993-1, Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing, following biocompatibility tests were performed on the final finished device to evaluate: Cytotoxicity, Sensitization, Intracutaneous reactivity, Systemic toxicity and Haemolysis.

The sterility of the meso-relle Needles is assured by using a validated sterilization method which complies with the requirements of the Recognized Consensus Standard: AAMI / ANSI / ISO 11135-1:2007, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices. Ethylene oxide (EtO) and Ethylene Chlorohydrin (ECH) residuals were tested according to UNI EN ISO 10993-7: 2008 - Method K.4.3.+ HRGC/MS Detection for EtO and Method K.4.8+ HRGC/MS Detection for ECH and met the acceptance criteria.

Meso-relle Needles were tested according to ISO 7864:2016 "Requirements and test methods Sterile hypodermic needles for single use", ISO 9626:2016 "Stainless steel needle tubing for the manufacture of medical devices. Requirements and test methods" and ISO 594-1.1986 "Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 1: General requirements" in order to demonstrate that the proposed device is substantial equivalent to the predicate in safety and performance.

CONCLUSION:

The subject and the predicate devices have the same indications for use and they have same technological characteristics. The test results and comparison results show that the proposed device is substantial equivalent to the predicate in performance.

Based on the intended use, technological characteristics and performance testing, the proposed product meso-relle Needles is considered to be substantially equivalent to the predicate device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).