(58 days)
RELiZORB™ is indicated for use in adults to hydrolyze fats in enteral formula
The RELiZORB device is a point-of-care accessory designed to fit in line with currently used enteral feeding circuits. RELiZORB is designed to hydrolyze (break down) fats present in enteral formulas from triglycerides into fatty acids and monoglycerides to allow for their absorption and utilization by the body. This breakdown of fats by the RELiZORB is intended to mimic the function of the enzyme lipase in patients who do not excr ete sufficient levels of pancreatic lipase. RELiZORB is comprised of a cylindrical, hollow cartridge with a single inlet port and a single outlet port connection. Inside the cartridge, there are small white beads. The digestive enzyme, lipase, is covalently bound to the small white beads. The lipase-bead complex, iLipaseTM (immobilized lipase), is retained within the cartridge during use by filters on both ends of the cartridge. The fat in enteral formulas is hydrolyzed as it comes in contact with iLipase as the formula passes through the cartridge.
This document describes a 510(k) premarket notification for the RELiZORB™ device, an enzyme-packed cartridge designed to hydrolyze fats in enteral formula. The submission aims to demonstrate substantial equivalence to a previously cleared RELiZORB™ device (DEN150001).
Since this is a submission for an enzyme-packed cartridge and not a diagnostic AI/ML device, the detailed questions about acceptance criteria, test sets, ground truth establishment, expert adjudication, and comparative effectiveness studies as typically applied to AI/ML software are not directly applicable in the same manner. This submission focuses on the chemical and mechanical performance of the device rather than the diagnostic accuracy of an algorithm.
However, I can extract the relevant performance data and criteria mentioned in the document in the spirit of your request:
1. Table of acceptance criteria and the reported device performance:
Acceptance Criteria Category | Specific Criteria (Implicitly Met) | Reported Device Performance |
---|---|---|
Shelf Life | Product stability over time, retaining intended function. | "Shelf life testing of RELIZORB was completed and met the acceptance criteria in the protocol. The shelf life test results support the proposed edit to the labeling with respect to product stability." |
Device Performance | Device performs as intended under anticipated conditions of use. | "The RELIZORB device was previously tested to demonstrate that the device performs as intended under anticipated conditions of use, and safety, including: Tensile and mechanical strength, Flow rate, Demonstration of enzymatic effect on macronutrients, Impurities/degradants characterization, Animal testing to demonstrate hydrolysis effect, Human factors testing." (These tests were deemed sufficient for the predicate device and are implicitly accepted for the current device due to no changes in design, technology, or functionality). |
Safety | No significant changes to risk assessment due to proposed minor labeling change. | "The Risk management file was reviewed and there was no significant change to the risk assessment as a result of the proposed minor change to the labeling." |
Quality System Compliance | Compliance with medical device quality system regulations. | "Alcresta develops and manufactures RELIZORB in compliance with the Quality System regulations (21 CFR 820)." |
Biocompatibility | Biological safety of materials in contact with the body/formula. | Referenced standard: ISO 10993-1:2009/Cor 1:2010 Biological evaluation of medical devices. (Implicitly met through prior testing of predicate). |
Usability/Human Factors | Device is safe and effective for users. | Referenced standard: EN 62366:2008 Medical devices - Application of usability engineering. (Implicitly met through prior testing of predicate). |
Risk Management | Application of risk management processes. | Referenced standard: EN ISO 14971:2012 Medical devices. Application of risk management. (Implicitly met through prior testing of predicate and current review). |
Packaging/Shipping | Device integrity maintained during transport. | Referenced standard: ISTA 2A: Packaged-Products weighing 150 lb (68 kg) or Less. Basic Requirements: atmospheric conditioning, compression, fixed displacement or random vibration and shock testing. (Implicitly met through prior testing of predicate). |
Connectors/Adapters | Compatibility and safety of enteral feeding set adapters and connectors. | Referenced standard: AAMI/ANSI ID54: 1996/(R)2012: Enteral feeding set adapters and connectors. (Implicitly met through prior testing of predicate). |
Controlled Environments | Manufacturing in appropriate environments. | Referenced standard: ISO-14644: Cleanrooms and associated controlled environments. (Implicitly met through prior manufacturing of predicate). |
2. Sample size used for the test set and the data provenance:
- Test set sample size: Not explicitly stated for each specific performance test (tensile, flow rate, enzymatic effect, etc.). The document mentions "shelf life testing" was completed, but not the number of units or replicates. For prior testing of the predicate device, it indicates a range of tests were performed, but no numbers of devices tested are provided.
- Data provenance: Not explicitly stated (e.g., country of origin). The testing seems to have been conducted by or for Alcresta Therapeutics, Inc. (Newton, MA, USA). The studies appear to be prospective as they involve testing the device under various conditions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in the typical sense for this device. This device is not a diagnostic tool requiring expert interpretation of results for ground truth. The "ground truth" for this device would be established by objective measurements of its physical properties (tensile strength, flow rate) and chemical activity (enzymatic effect on macronutrients), likely performed by laboratory technicians or engineers following validated protocols, rather than expert interpretation of a diagnostic outcome.
4. Adjudication method for the test set:
- Not applicable. Adjudication methods like 2+1 or 3+1 are used to reconcile disagreements among human readers in diagnostic studies. For this device, performance is evaluated against objective, measurable criteria and laboratory standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI/ML diagnostic device and does not involve human readers interpreting cases with or without AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used:
- For the enzymatic effect: Objective chemical measurements (e.g., measuring the breakdown of triglycerides into fatty acids and monoglycerides).
- For mechanical/physical properties: Engineering measurements against established standards (e.g., tensile strength, flow rate).
- For biocompatibility: Laboratory testing results compared to biological safety standards.
8. The sample size for the training set:
- Not applicable. This device does not use an algorithm that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable. As there is no training set for an algorithm, there is no ground truth to be established for it.
§ 876.5985 Enzyme packed cartridge.
(a)
Identification. An enzyme packed cartridge is anex vivo prescription device that is used in enzymatic hydrolysis of macronutrients into their essential nutrient forms at the time of delivery. The device consists of an outer casing containing an inert polymer with a covalently bound enzyme through which nutritional formula is directed. The device fits in line with enteral feeding systems.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient contacting components of the device must be demonstrated to be biocompatible.
(2)
In vivo testing must be performed and must demonstrate that the device causes neither an adverse tissue response nor adverse performance.(3) Non-clinical testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
(i) Mechanical testing to demonstrate that the device can withstand clinical forces;
(ii) Flow rate and leakage testing to demonstrate that the device does not impede the flow of enteral formula;
(iii) Demonstration of enzymatic effect on intended macronutrient;
(iv) The amount of enzyme that exits the cartridge must be characterized;
(v) Validation that the device does not adversely impact the nutritional composition of enteral formula; and
(vi) Validation that the device does not impede flow alarms on enteral feeding pumps.
(4) Human factors testing must be performed to characterize use error risks.
(5) Performance data must support shelf life by demonstrating package integrity and device functionality over the identified shelf life.
(6) Labeling must include the following:
(i) A detailed summary of
in vivo testing pertinent to use of the device, including device-related adverse events;(ii) A detailed summary of compatible formulas that is supported by non-clinical testing, including the expected enzymatic conversion as a percentage;
(iii) Detailed instructions on how to place the device into an enteral feeding circuit;
(iv) A warning regarding the possibility for misconnections; and
(v) Expiration date or shelf life.
(7) Patient labeling must be provided and must include:
(i) Relevant warnings, precautions, adverse effects, and complications;
(ii) A description of the device and how it operates;
(iii) Instructions on how to correctly use the device; and
(iv) The benefits and risks associated with the use of the device.