K Number
K161186

Validate with FDA (Live)

Date Cleared
2016-06-29

(63 days)

Product Code
Regulation Number
876.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Diathermic Slitter (FlushKnife) DK2618J and DK2623J are intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The devices are indicated for ablation, incision, dissection, avulsion, coagulation and hemostasis of tissue within the digestive tract.

Diathermic Slitter (ClutchCutter) DP2618DT is intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The device is indicated for ablation, incision, dissection, avulsion, cauterization, coagulation and hemostasis of tissue within the digestive tract.

Device Description

The Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter) are electrosurgical instruments intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The FlushKnife has needle or ball tip shapes and varying slitter lengths and working lengths. The ClutchCutter has a forceps type slitter shape and a working length of 1800mm. Both devices have a water feed function, are operated manually via a handle slider, use monopolar energy delivered from an electrosurgical generator, are single-use devices, and are biocompatible.

AI/ML Overview

The provided text is a 510(k) premarket notification for a medical device. This type of document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific performance criteria through a clinical study or effectiveness study with human readers.

Therefore, many of the requested categories for acceptance criteria and study details are not applicable or not provided in this document.

Here's an analysis based on the available information:

1. A table of acceptance criteria and the reported device performance

The document states that "In all cases, the device met the pre-defined acceptance criteria for the test." However, the specific acceptance criteria values are not detailed in this summary for each test. The tests performed were:

Test Performed (FlushKnife)Reported Device Performance
AirtightnessMet pre-defined acceptance criteria
Maximum diameter of insertion portionMet pre-defined acceptance criteria
Working lengthMet pre-defined acceptance criteria
Slitter Portion lengthMet pre-defined acceptance criteria
Electrical resistanceMet pre-defined acceptance criteria
Test Performed (ClutchCutter)Reported Device Performance
Working lengthMet pre-defined acceptance criteria
Maximum diameter of insertion portionMet pre-defined acceptance criteria
Distal end shapesMet pre-defined acceptance criteria
Electrical resistanceMet pre-defined acceptance criteria

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The studies mentioned are engineering/performance testing, not clinical studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the tests were performance testing of physical device attributes, not diagnostic performance requiring expert interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable for physical device performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This document describes a physical medical device (electrosurgical slitter), not an AI-powered diagnostic tool, so this type of study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable as the device is not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the performance tests, the "ground truth" would be established engineering specifications and measurement standards. For example, "Working length" would have a defined specification, and the device's measured length would be compared against it. This is not explicitly detailed in the summary, but it's inherent to engineering performance testing.

8. The sample size for the training set

This information is not applicable as there is no training set for an AI or algorithm in this context.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set for an AI or algorithm in this context.

Summary of the Device and its Review:

The document describes a Special 510(k) for the Fujifilm Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter). The purpose of this submission is to address a minor modification (change in material - epoxy resin) to an already cleared predicate device (K151474). The core argument for substantial equivalence is that the intended use, indications, technological characteristics, and principles of operation remain the same as the predicate device. The performance tests conducted (e.g., airtightness, working length, electrical resistance) were to ensure that the modified device performs equivalently to the predicate and met pre-defined acceptance criteria. The document concludes that these minor differences do not raise new questions of safety or efficacy.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, arranged in a stacked formation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 29, 2016

Fujifilm Medical Systems USA, Inc. Shraddha More Specialist, Regulatory Affairs and Quality Assurance 10 High Point Drive Wayne, NJ 07470

Re: K161186

Trade/Device Name: Fujifilm Diathermic Slitter (flushknife), Fujifilm Diathermic Slitter (clutchcutter) Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: Class II Product Code: KGE Dated: June 16, 2016 Received: June 17, 2016

Dear Shraddha More,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known)

K161186

Device Name

Fuiifilm Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter)

Indications for Use (Describe)

Diathermic Slitter (FlushKnife) DK2618J and DK2623J are intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The devices are indicated for ablation, incision, dissection, avulsion, coagulation and hemostasis of tissue within the digestive tract.

Diathermic Slitter (ClutchCutter) DP2618DT is intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The device is indicated for ablation, incision, dissection, avulsion, cauterization, coagulation and hemostasis of tissue within the digestive tract.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

XII. 510(K)

SUMMARY Submitter's

Information

FUJIFILM Medical Systems U.S.A., Inc. Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293

Contact Person:

Shraddha More Specialist, Regulatory Affairs and Quality Assurance Telephone: (973) 686-2627 Ext. 522627 Facsimile: (973) 633-8818 E-Mail smore@fujifilm.com

Date Prepared: April 26, 2016

Identification of the Proposed device:

Proprietary/Trade Name:Fujifilm Diathermic Slitter (FlushKnife) and Diathermic Slitter(ClutchCutter)
Common Name:Electrosurgical Instruments
Device Class:Class II
Review Panel:Gastroenterology/Urology

Classification Information:

Endoscopic electrosurgical unit and accessories 21 C.F.R. § 876.4300 Product Code: KGE

Predicate Device:

Fujifilm Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter) (K151474)

Purpose of the Special 510(k) notice:

The Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter) is a modification to Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter) (K151474).

Intended Use:

Diathermic Slitter (FlushKnife) DK2618J and DK2623J are intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The devices are indicated for ablation, incision, dissection, avulsion, coaqulation and hemostasis of tissue within the digestive tract.

Diathermic Slitter (ClutchCutter) DP2618DT is intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The device is indicated for ablation, incision, dissection, avulsion, coagulation and hemostasis of tissue within the digestive tract.

{4}------------------------------------------------

Technological Characteristics:

Tables comparing the technological characteristics between FlushKnife and ClutchCutter and their respective predicate devices are shown below.

Predicate DeviceProposed Device
Device NameFujifilm Diathermic Slitter(FlushKnife)Fujifilm Diathermic Slitter(FlushKnife)
Device ModelsDIATHERMIC SLITTER, MODELDK2618J -N10-, DK2618J -N15-,DK2618J -N20-, DK2618J -N25-,DK2618J -N30-, DK2618J -B15-,DK2618J -B20-, DK2618J -B25-,DK2618J -B30-, DK2623J -N15-,DK2623J -N20-, DK2623J -B15-,DK2623J -B20-Same as predicate device

Summary of differences between modified FlushKnife from its predicate device

510(k) NumberK151474Pending
Intended UseThese instruments have beendesigned to be used with specifiedendoscopes to cut tissue usinghigh-frequency current within thedigestive tract. Both the ball tiptype and the needle typeinstruments are indicated forablation, incision, dissection,avulsion, cauterization,coagulation and hemostasis oftissue within the digestive tract.Same as predicate device
TechnologicalCharacteristicsSlitter Length1.0/1.5/2.0/2.5/3.0mm (forDK2618 series)1.5/2.0mm (for DK2623 series)Same as predicate device
Slitter ShapeNeedle Type(With Ball Tip:-BXX-)(Without Ball Tip:-NXX-)Same as predicate device
Maximum Diameter ofInsertion Portionφ2.7Same as predicate device
Working Length1800mm/2300mmSame as predicate device
Water Feed FunctionYesSame as predicate device
Method of OperationManually (handle slider)Same as predicate device
Energy used or deliveredenergy delivered from aelectrosurgical generatorSame as predicate device
Monopolar/BipolarMonopolarSame as predicate device
SterilizationYes (Single Use Device)Same as predicate device
BiocompatibilityYesSame as predicate device
AccessoriesNoneSame as predicate device
Combination equipmentEndoscope, Electrosurgicalgenerator, A CordSame as predicate device

{5}------------------------------------------------

Summary of differences between modified ClutchCutter from its predicate device

Predicate deviceProposed Device
Device NameFujifilm Diathermic Slitter(ClutchCutter)Fujifilm Diathermic Slitter(ClutchCutter)
Device ModelsDP2618DT -35-, DP2618DT -50-Same as predicate device
510(k) NumberK151474Pending
Intended UseThese instruments have beendesigned to be used withspecified endoscopes to cuttissue using high-frequencycurrent within the digestive tract.These instruments are indicatedfor ablation, incision, dissection,avulsion, cauterization,coagulation and hemostasis oftissue within the digestive tract.Same as predicate device
TechnologicalCharacteristicsSlitter ShapeForceps TypeSame as predicate device
Maximum Diameterof Insertion Portionφ2.7Same as predicate device
Working Length1800mmSame as predicate device
Slitter PortionRotation FunctionYesSame as predicate device
Method of OperationManually (handle slider)Same as predicate device
Energy used or deliveredenergy delivered from aelectrosurgical generatorSame as predicate device
Monopolar/BipolarMonopolarSame as predicate device
SterilizationYes (Single Use Device)Same as predicate device
BiocompatibilityYesSame as predicate device
AccessoriesNoneSame as predicate device
Combination equipmentEndoscope, Electrosurgicalgenerator, A CordSame as predicate device

Performance Data

Fujifilm conducted the following performance testing on the proposed device, FlushKnife and ClutchCutter to ensure that the modified device performs equivalently to the predicate device:

FlushKnife

  • . Airtightness
  • Maximum diameter of insertion portion .
  • Working length .
  • Slitter Portion length
  • . Electrical resistance

{6}------------------------------------------------

ClutchCutter

  • . Working length:
  • Maximum diameter of insertion portion ●
  • . Distal end shapes
  • . Electrical resistance

In all cases, the device met the pre-defined acceptance criteria for the test.

Substantial Equivalence

The company's proposed device, FlushKnife and ClutchCutter have the same intended use as the previously cleared predicate FlushKnife and ClutchCutter (K151474). In addition, the proposed device, FlushKnife and ClutchCutter has same indications, technological characteristics, and principles of operation as its predicate. Although there are minor differences between the proposed device and its predicate device, namely modification in material (epoxy resin), those differences do not raise new questions of safety or efficacy. The minor change done in the material would not raise any new or additional questions in of safety or effectiveness of the proposed device. Thus, the proposed device, FlushKnife and ClutchCutter, is substantially equivalent to its predicate device.

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).