(308 days)
Not Found
No
The summary describes image processing, registration, and tracking algorithms, but does not mention AI, ML, or related terms. The performance studies focus on accuracy of fusion and model illustration, not on AI/ML model performance metrics.
No
The device is an adjunctive tool for visualizing ultrasound images and assisting needle-targeted procedures; it does not directly treat or cure any condition.
No
The device is described as an "adjunctive tool" for visualization and assistance in needle-targeted procedures, such as biopsies. It does not provide a diagnosis itself, but rather aids physicians in performing diagnostic and interventional procedures by providing guidance and information visualization.
No
The device description explicitly states that the NaviGo™ Workstation "connects to the video output of the ultrasound system and by tracking the ultrasound probe's position". This indicates a physical connection and interaction with hardware (the ultrasound system and probe), which goes beyond being solely a software application.
Based on the provided information, the UC-CARE NaviGo Workstation is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostics are devices intended for use in the collection, preparation, and examination of specimens taken from the human body (such as blood, urine, or tissue) to provide information for the diagnosis, treatment, or prevention of disease.
- NaviGo Workstation's Function: The NaviGo Workstation is an adjunctive tool for ultrasound guided procedures. It processes and visualizes medical images (ultrasound and DICOM/MRI) to assist physicians in planning and performing procedures like prostate biopsies. It does not analyze biological specimens.
- Intended Use: The intended use clearly states it's for "2-D and 3-D visualization of ultrasound images of the prostate gland" and to "assist needle targeted procedures (e.g. biopsy)". This is focused on image guidance and procedural assistance, not the analysis of biological samples.
The device is a medical imaging and guidance system, falling under a different regulatory category than IVDs.
N/A
Intended Use / Indications for Use
The UC-CARE NaviGo Workstation is an adjunctive tool for ultrasound guided procedures and is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the other. Additional software features include patient data management, multi-planar reconstruction, segmentation, image measurement and 3-D image registration.
The device is specifically indicated to provide information within the prostate to assist needle targeted procedures (e.g. biopsy) regions of interest display, procedures planning and, reconstruction of a 3D rendered surface model of the prostate display needle locations that have been selected by the physician, as well as storage and future retrieval of this information.
Product codes (comma separated list FDA assigned to the subject device)
LLZ
Device Description
The NaviGo™ Workstation is an adjunctive tool in the management of prostate diagnostic and interventional procedures. The NaviGo™ Workstation allows prostate needle tracking, recording, and management solution. The NaviGo™ Workstation is designed to assist the physician in performing prostate diagnostic and interventional procedures by providing regional orientation information, displaying a 3D model with real-time tracking and recording of the needle location. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the . The device includes means to compensate for patient body and prostate motion at any time during the procedure.
The NaviGo™ Workstation is designed to work with standard trans-rectal ultrasound systems and biopsy setup without changing or interfering with the physician's existing workflow. The NaviGo™ Workstation connects to the video output of the ultrasound system and by tracking the ultrasound probe's position, the recorded 2D ultrasound images are transferred to the NaviGo™ Workstation for viewing and creation of a 3D model. As with any other procedure, the TRUS probe is used together with standard disposable cover sheath supplied by the user.
Two dimensional (2D) images and the 3D model of the prostate are displayed on the NaviGo™ Workstation's screen. The NaviGo™ Workstation is equipped with tools to manipulate (rotate, pan, zoom) the model, and to archive and retrieve the information for further use.
The tracking and recording enables display of an accurate 3D model of the prostate and to record needle locations on the model. Pathology diagnosis results may be updated on the 3D model and a color display representation provides a visual display of the biopsy results.
The NaviGo™ Workstation supports the display of ROIs on the 3D model and displays visual indication when the needle trajectory intersects with an ROI (region of interest) is defined by the physician by segmenting a portion of the prostate on a 2D image and displaying its location on the 3D model, thereby defining a portion of the prostate as a target to direct a needle within. The 2D model for segmentation of the ROI can be either an Ultrasound image frozen during the procedure or a DICOM complaint image from another imaging study completed prior to the Ultrasound procedure (e.g. MR) when selecting the fusion procedure to display the planned targets. The ROIs are clearly numbered and labeled by a letter symbolizing it origin (Ultrasound or DICOM), On each image, more than one ROI may be defined (segmented).
The live 2D ultrasound image is superimposed in real-time with the 2D ROI boundaries thus allowing the physician to see the real-time advancement of the needle on the live ultrasound image with reference to ROI boundaries. The 3D model is displayed on a separate window alongside the ultrasound's live continuous images display, allowing prostate 3D orientation.
Regardless of the type of ultrasound probe used for the procedure (side-fire or end-fire) a cannula for a needle is built within the outer shell of the probe and therefore a needle can be inserted through the designated cannula. The designated cannula defines the needle path along its trajectory. The needle trajectory is displayed on the 3D model representing the potential needle route to allow the physician to direct the needle in real time into the target. When the projected path intersects with an RQL the color of the ROI will change indicating the needle can be directed inside the ROI if the physician ensures that the needle is inserted deep enough.
During the procedure, the physician may change some of the parameters on the US system or perform different actions that require the NaviGo™ Workstation to adjust. A physician may wish to freeze an image and perform measurements, label the image and save it to the report, switch between transversal and longitudinal view, or change the view of the ultrasound image. The NaviGo™ Workstation is equipped with image state algorithms to automatically detect the change in parameters and adjust itself to the new parameters. For example, if the physician freezes the US image, the NaviGo will automatically present additional menu option for a frozen image like measurements tools and labeling. The image state algorithm was developed to support generic ultrasound parameters and is specifically tested for each new ultrasound the NaviGo™ is required to support.
In off-line mode, the NaviGo™ software further enables the physician to analyze previous procedures using the prostate model, update the 3D model if necessary, update the recorded biopsies' locations, generate report, and provides a DICOM (e.g. MRI) interface for ROI definition, 3D model display, and data communication.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Ultrasound, DICOM images (e.g. MRI)
Anatomical Site
Prostate gland
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Physicians in the clinic or hospital
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Fusion accuracy performance test:
The accuracy of targeting DICOM (MRI) derived regions of interest (ROI) by using the NaviGo Workstation was evaluated. Four phantoms each with 3 isoechoic lesions (CIRS Model 778-05, containing 3 randomly located isoechoic lesions of 0.5 cc) were targeted 2-3 times each using the NaviGo MRI fusion algorithm. Two phantoms were targeted using side fire ultrasound probes and two phantoms were targeted using end fire ultrasound probes. Needle location were defined relative to the ROI center line and accuracy is defined as the difference in needle location between the needle trajectory (targeted location in lesion) as marked on the real time US image, to the actual result as seen in post procedure CT scan. A total of 34 ultrasound guided biopsies were targeted into 12 isoechoic lesions. All 34 biopsies (100%) successfully hit the target lesion with an average error of 1.32±0.84mm. 32.3% of the biopsies had accuracy of
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2017
UC-CARE Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW WASHINGTON DC 20004
Re: K160934
Trade/Device Name: NaviGo™ Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 10, 2017 Received: January 10, 2017
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K160934
Device Name NaviGo™ Workstation
Indications for Use (Describe)
The UC-CARE NaviGo Workstation is an adjunctive tool for ultrasound guided procedures and is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the other. Additional software features include patient data management, multi-planar reconstruction, segmentation, image measurement and 3-D image registration.
The device is specifically indicated to provide information within the prostate to assist needle targeted procedures (e.g. biopsy) regions of interest display, procedures planning and, reconstruction of a 3D rendered surface model of the prostate display needle locations that have been selected by the physician, as well as storage and future retrieval of this information.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | |
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo is in blue, with the exception of the "A" in CARE, which is in green. Above the text are five blue ovals that are progressively lighter in color.
510(k) SUMMARY
NaviGo™ Workstation
510(k) Number K160934
Applicant's Name:
UC-CARE Ltd. Apollo Bldg., New Industrial Zone P.O.B 67 Yokneam 20692, Israel Tel: +972-4-909-7427 ext. 656 Fax: +972-4-993-7323 Email: tomers@uc-care.com Contact Person: Tomer Schatzberger
Contact Person:
Stephen Bellomo H.M.Da Consulting and Development Ltd. 40 Geulim Stree Zichron-Yaakov 3091847 Israel Tel: +972-54-227-4831 Email: stephen.bellomo@gmail.com
Date Prepared:
January 10, 2017
Trade Name:
NaviGo™ Workstation
Classification Name:
Picture archiving and communications system.
Classification:
Picture archiving and communications systems have been classified as Class II devices, product code: LLZ, Regulation Number: 892.2050, reviewed by the Division of Radiology.
Predicate Devices:
Primary: NaviGo™ Workstation (UC-CARE Ltd.), cleared under K100784, concurrence date November 1, 2011.
Secondary: Urostation - 3D Prostate Suite (Koelis), cleared under K131448, concurrence date July 26, 2013.
UC-CARE Ltd.
4
Image /page/4/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo is in blue, with the exception of the 'A' in CARE, which is in green. Above the text are five blue ovals that are progressively lighter in color from left to right.
Performance Standards:
No performance standards have been established for such a device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
Device Description:
The NaviGo™ Workstation is an adjunctive tool in the management of prostate diagnostic and interventional procedures. The NaviGo™ Workstation allows prostate needle tracking, recording, and management solution. The NaviGo™ Workstation is designed to assist the physician in performing prostate diagnostic and interventional procedures by providing regional orientation information, displaying a 3D model with real-time tracking and recording of the needle location. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the . The device includes means to compensate for patient body and prostate motion at any time during the procedure.
The NaviGo™ Workstation is designed to work with standard trans-rectal ultrasound systems and biopsy setup without changing or interfering with the physician's existing workflow. The NaviGo™ Workstation connects to the video output of the ultrasound system and by tracking the ultrasound probe's position, the recorded 2D ultrasound images are transferred to the NaviGo™ Workstation for viewing and creation of a 3D model. As with any other procedure, the TRUS probe is used together with standard disposable cover sheath supplied by the user.
Two dimensional (2D) images and the 3D model of the prostate are displayed on the NaviGo™ Workstation's screen. The NaviGo™ Workstation is equipped with tools to manipulate (rotate, pan, zoom) the model, and to archive and retrieve the information for further use.
The tracking and recording enables display of an accurate 3D model of the prostate and to record needle locations on the model. Pathology diagnosis results may be updated on the 3D model and a color display representation provides a visual display of the biopsy results.
The NaviGo™ Workstation supports the display of ROIs on the 3D model and displays visual indication when the needle trajectory intersects with an ROI (region of interest) is defined by the physician by segmenting a portion of the prostate on a 2D image and displaying its location on the 3D model, thereby defining a portion of the prostate as a target to direct a needle within. The 2D model for segmentation of the ROI can be either an Ultrasound image frozen during the procedure or a DICOM complaint image from another imaging study completed prior to the Ultrasound procedure (e.g. MR) when selecting the fusion procedure to display the planned targets. The ROIs are clearly numbered and labeled by a letter symbolizing it origin (Ultrasound or DICOM), On each image, more than one ROI may be defined (segmented).
The live 2D ultrasound image is superimposed in real-time with the 2D ROI boundaries thus allowing the physician to see the real-time advancement of the needle on the live ultrasound image with reference to ROI boundaries. The 3D model is displayed on a separate window alongside the ultrasound's live continuous images display, allowing prostate 3D orientation.
Reqardless of the type of ultrasound probe used for the procedure (side-fire or end-fire) a cannula for a needle is built within the outer shell of the probe and therefore a needle can be inserted through the designated cannula. The designated cannula defines the needle path along its trajectory. The needle trajectory is displayed on the 3D model representing the potential needle route to allow the physician to direct the needle in real time into the target. When the projected path
UC-CARE Ltd.
5
Image /page/5/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo is primarily blue, with the letters "UC" and "CARE" in a bold, sans-serif font. The "A" in "CARE" is colored green. Above the text are a series of blue ovals that appear to be fading from dark to light.
intersects with an RQL the color of the ROI will change indicating the needle can be directed inside the ROI if the physician ensures that the needle is inserted deep enough.
During the procedure, the physician may change some of the parameters on the US system or perform different actions that require the NaviGo™ Workstation to adjust. A physician may wish to freeze an image and perform measurements, label the image and save it to the report, switch between transversal and longitudinal view, or change the view of the ultrasound image. The NaviGo™ Workstation is equipped with image state algorithms to automatically detect the change in parameters and adjust itself to the new parameters. For example, if the physician freezes the US image, the NaviGo will automatically present additional menu option for a frozen image like measurements tools and labeling. The image state algorithm was developed to support generic ultrasound parameters and is specifically tested for each new ultrasound the NaviGo™ is required to support.
In off-line mode, the NaviGo™ software further enables the physician to analyze previous procedures using the prostate model, update the 3D model if necessary, update the recorded biopsies' locations, generate report, and provides a DICOM (e.g. MRI) interface for ROI definition, 3D model display, and data communication.
Intended Use:
The UC-CARE NaviGo Workstation is an adjunctive tool for ultrasound guided procedures and is intended to be used by physicians in the clinic or hospital for 2-D and 3-D visualization of ultrasound images of the prostate gland. The NaviGo™ Workstation offers the ability to fuse DICOM originated information (e.g. MRI) with the ultrasound images and thus superimpose information from one modality onto the other. Additional software features include patient data management, multiplanar reconstruction, segmentation, image measurement and 3-D image registration.
The device is specifically indicated to provide information on regional orientation within the prostate to assist needle targeted procedures (e.g. biopsy) regions of interest display, procedures planning and, reconstruction of a 3D rendered surface model of the prostate display needle locations that have been selected by the physician, as well as storage and future retrieval of this information.
Comparison of Technological Characteristics:
The subject NaviGo Workstation device and the primary predicate NaviGo Workstation device have the same principles of operation and basic design. Both devices have the same following technological characteristics:
- assist the physician by transfer and display of ultrasound images on the workstation screen
- provide regional orientation information during prostate biopsy procedures ●
- build display and manipulate a 3D model of the prostate on screen ●
- archive procedure data and report generating
- provide data management solutions ●
- track, display and record the biopsy needle trajectory location retrieved from the ultrasound probe
- . display the scanning history, including pathology analyses
Several modifications have been made from this predicate device, including the following: addition of illustration model (side-fire first procedure only), needle-fire recognition, image-state correction offline configuration, and end-fire ultrasound probe support. These modifications are considered minor and do not raise new issues of safety and efficacy.
UC-CARE Ltd.
6
Image /page/6/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo has a series of blue ovals in the top portion of the image. The text "UC-CARE" is in blue, with the "C" in "CARE" being green. Below this text is the text "Medical Systems" in blue.
Fusion functionality has been added to the subject NaviGo device, including fusion of DICOM originated information, ROI display, and procedure planning. The indications for use statement of the subject device has been modified from the predicate device to include the addition of fusion functionality. The fusion functionality of the subject NaviGo device is substantially equivalent to the fusion functionality of the secondary predicate Urostation device, including the stated accuracy of the system. Therefore, the addition of fusion functionality does not raise any safety or effectiveness concern in comparison with the cleared device.
| Product | NaviGo (K100784)
[Primary Predicate
Device] | Urostation
(K131448)
[Secondary Predicate
Device] | Subject NaviGo
Device
(K160934) |
|----------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Product code | LLZ | LLZ | LLZ |
| Class | II | II | II |
| Device Classification Name | System, Image
Processing,
Radiological | System, Image
Processing,
Radiological | System, Image
Processing,
Radiological |
| Visualization Means | Side-fire Ultrasound | End-fire Ultrasound | Side -fire & End-fire
Ultrasound |
| Supported Modalities | Ultrasound | Ultrasound, DICOM
images (e.g. MRI) | Ultrasound, DICOM
images (e.g. MRI) |
| Accuracy | +/-5mm | +/-3mm | +/-3mm |
| Segmentation (Contouring) | Yes | Yes, only for fusion | Yes |
| 3D model reconstruction | Yes, contouring only | Yes, only for fusion | Yes, illustration &
contouring |
| Needle trajectory display | Real-time | Upon request | Real-time |
| Tracking technology | Electro-magnetic
(mattress and cube) | Image processing | Electro-magnetic
(cube only) |
| Offline | Post procedure data
analysis, pathology
results | Post procedure data
analysis, pathology
results, planning (on
DICOM images) | Post procedure data
analysis, pathology
results, planning (on
DICOM images) |
| Regions of interest | No | Originated from
DICOM images (e.g.
MRI) | Originated from US
images + DICOM
images (e.g. MRI) |
| Registration | US - US | US - US, and MR - US
(fusion) | US - US, and MR - US
(fusion) |
| DICOM connectivity | No | Yes | Yes |
Substantial Equivalence Comparison Table:
7
Image /page/7/Picture/0 description: The image contains the logo for UC-CARE Medical Systems. The logo consists of the text "UC-CARE" in a bold, blue font, with the "A" in "CARE" being green. Above the text are five blue ovals, arranged horizontally and decreasing in size from left to right. Below the main text, "Medical Systems" is written in a smaller, blue, italicized font.
Non-clinical Performance Data:
The performance characteristics and operation/usability of the modified NaviGo™ Workstation were evaluated in the following non-clinical (bench) testing.
Fusion accuracy performance test:
The accuracy of targeting DICOM (MRI) derived regions of interest (ROI) by using the NaviGo Workstation was evaluated. . Four phantoms each with 3 isoechoic lesions (CIRS Model 778-05, containing 3 randomly located isoechoic lesions of 0.5 cc) were targeted 2-3 times each using the NaviGo MRI fusion algorithm. Two phantoms were targeted using side fire ultrasound probes and two phantoms were targeted using end fire ultrasound probes. Needle location were defined relative to the ROI center line and accuracy is defined as the difference in needle location between the needle trajectory (targeted location in lesion) as marked on the real time US image, to the actual result as seen in post procedure CT scan. A total of 34 ultrasound guided biopsies were targeted into 12 isoechoic lesions. All 34 biopsies (100%) successfully hit the target lesion with an average error of 1.32±0.84mm. 32.3% of the biopsies had accuracy of