Intended Use

SinuSys Vent-Os® Sinus Dilation family is intended to provide a means to access the sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.

  • The Vent-Os Maxillary Dilation System is indicated for dilation of Maxillary sinus and associated spaces.
  • · The Vent-Os Frontal Dilation System is indicated for dilation of the Frontal recess and associated spaces.
  • · The Vent-Os Sphenoid Dilation System is indicated for dilation of Sphenoid sinus and associated spaces.
Device Description

The Vent-Os® Sinus Dilation family comprises a family of osmotically driven devices that access and dilate the sinus ostia and spaces associated with the paranasal sinus cavities. Each member of the Vent-Os® Sinus Dilation family is a system that includes a Dilation Device preloaded on a Placement Instrument. The Placement Instrument accesses the target site through the nasal passageway and delivers the Dilation Device. The Vent-Os® Sinus Dilation Device then expands and remodels the target tissues and then is removed.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided text.

Based on the provided text, the device in question is the "Vent-Os® Sinus Dilation Family."

Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for outcomes like "patent" rates for the clinical trial. Instead, it compares the device's performance to a literature control to demonstrate "substantial equivalence." The "acceptance criteria" can be inferred as achieving comparable safety and effectiveness to the predicate device and the literature control.

Attribute Tested/OutcomeAcceptance Criteria (Inferred: Comparable to Literature Control / Passed)Reported Device Performance (Vent-Os Study)
BiocompatibilityNon-cytotoxic, Non-sensitizing, Non-irritatingCytotoxicity (MEM Elution): Non-cytotoxicSensitization-Maximization: Non-sensitizingIntracutaneous Reactivity: Non-irritating
Bench TestingAll attributes PassedDevice weight: PassedPlacement System Working Length Profile: PassedPlacement System Working Length: PassedCrossing Profile: PassedDeployment Force: PassedDilation Device Diameter @ 1 hour: PassedDilation Device Working Length @ 1 hour: PassedDilation Device Integrity: PassedBond Strength: Passed
Transportation, Sterilization, & Shelf-Life TestingAll attributes Passed (Sterility Assurance Level (SAL) of 10⁻⁶)Sterilization: Achieved SAL of 10⁻⁶Transit Testing: PassedSterile Barrier Integrity (Bubble Test): PassedPackaging & Product Visual Inspection: PassedPouch Seal Testing: Passed
Cadaver TestingSuccessful delivery and placementEvaluators confirmed successful delivery and placement of the devices to the targeted sinus.
Clinical Outcomes (Frontal Recess - Acute)Comparable to Bolger Study (69% patent)Vent-Os Study: 33/33 (100%) patent (95% CI: 92.4% - 100%)Bolger Study (Literature Control): 85/124 (69%) patent (95% CI: 59.9% - 76.1%)
Clinical Outcomes (Frontal Recess - 3 Months)Comparable to Bolger Study (84% patent)Vent-Os Study: 28/30 (93%) patent (95% CI: 78.7% - 98.2%)Bolger Study (Literature Control): 82/98 (84%) patent (95% CI: 75.1% - 89.7%)
Clinical Outcomes (Sphenoid Sinus Ostium - Acute)Comparable to Bolger Study (55% patent)Vent-Os Study: 13/15 (87%) patent (95% CI: 62.1% - 96.3%)Bolger Study (Literature Control): 41/75 (55%) patent (95% CI: 43.5 - 65.4%)
Clinical Outcomes (Sphenoid Sinus Ostium - 3 Months)Comparable to Bolger Study (59% patent)Vent-Os Study: 10/15 (67%) patent (95% CI: 41.7% - 84.8%)Bolger Study (Literature Control): 32/54 (59%) patent (95% CI: 46.0% - 71.3%)

Study Details

  1. Sample sizes used for the test set and the data provenance:

    • Clinical Trial Test Set: 30 patients.
    • Data Provenance: The study was a "multi-center, prospective, single arm, non-randomized trial conducted at three (3) clinical sites." The country of origin is not explicitly stated, but given the FDA submission, it's typically understood to be within or recognized by the US regulatory framework.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document states "three (3) investigators" for the clinical trial. It implies these investigators were physicians involved in treating patients with chronic rhinosinusitis and evaluating the outcomes. Specific qualifications (e.g., years of experience, specific specialization beyond "physician users") are not detailed.
    • For cadaver testing, "multiple physician users" were involved in validating usability.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • The document does not describe any specific adjudication method for the clinical outcomes. It implies the "investigators" assessed the success of the treatment, but details on how agreement was reached for indeterminate cases or divergences in assessment are not provided.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is a surgical instrument, not an AI-assisted diagnostic tool for human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as the Vent-Os® Sinus Dilation Family is a surgical instrument and not an algorithm or AI system.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the clinical trial, the outcomes (patent, non-patent, indeterminate) appear to be based on clinical assessment by the physician investigators (outcomes data) at the 3-month follow-up. This type of assessment often relies on endoscopic evaluation and investigator judgment.
    • For cadaver testing, the ground truth was expert observation/confirmation of successful delivery and placement by "multiple physician users."
  7. The sample size for the training set:

    • As this is a surgical device and not an AI or machine learning model, there is no "training set" in the conventional sense. The development of the device would involve engineering, material science, and iterative testing, but not data-driven training of an algorithm.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth. The graphic is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 29, 2016

SinuSys Corporation Mr. Lloyd Griese Vice President, Quality and Regulatory 2468 Embarcadero Wav Palo Alto, CA 94303

Re: K160770

Trade/Device Name: Vent-Os® Sinus Dilation Family Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: May 31, 2016 Received: June 1, 2016

Dear Mr. Griese:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160770

Device Name Vent-Os Sinus Dilation Family

Indications for Use (Describe)

SinuSys Vent-Os® Sinus Dilation family is intended to provide a means to access the sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.

  • The Vent-Os Maxillary Dilation System is indicated for dilation of Maxillary sinus and associated spaces.
  • · The Vent-Os Frontal Dilation System is indicated for dilation of the Frontal recess and associated spaces.
  • · The Vent-Os Sphenoid Dilation System is indicated for dilation of Sphenoid sinus and associated spaces.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

8 Premarket Notification 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number: __K160770

Applicant Information:

Date Prepared:June 25, 2016
Name:SinuSys Corporation
Address:2468 Embarcadero WayPalo Alto, CA 94303
Phone:650-213-9988
Fax:650-213-9688

Contact Person: Lloyd H. Griese, Vice President of Quality & Regulatory

Phone Number (Office):650-213-9988
Phone Number (Mobile):925-786-8617
Email:lgriese@sinusys.com
Back-up Contact Person:Michael A. Daniel, Daniel & Daniel Consulting, LLC
Phone Number (Mobile):415-407-0223
Email:madaniel@clinregconsult.com

Device Information:

Device Trade Name:Vent-Os® Sinus Dilation Family
Common Name:Instrument, ENT Manual Surgical
Classification Name(s):Ear, nose, and throat manual surgical instrument.
Product Code/ Regulation:LRC / 874.4420
Classification:Class I, Exempt

Predicate Device:

XprESS Multi-Sinus Dilation Tool (K121943)

Reference Devices:

Vent-Os® Sinus Dilation System (K133016) Relieva Spin Sinus Dilation System (K111875)

{4}------------------------------------------------

Premarket Notification 510(k) Summary - Continued

Device Description:

The Vent-Os® Sinus Dilation family comprises a family of osmotically driven devices that access and dilate the sinus ostia and spaces associated with the paranasal sinus cavities. Each member of the Vent-Os® Sinus Dilation family is a system that includes a Dilation Device preloaded on a Placement Instrument. The Placement Instrument accesses the target site through the nasal passageway and delivers the Dilation Device. The Vent-Os® Sinus Dilation Device then expands and remodels the target tissues and then is removed.

Indications for Use:

SinuSys Vent-Os® Sinus Dilation family is intended to provide a means to access the sinus space and to dilate sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.

  • . The Vent-Os Maxillary Dilation System is indicated for dilation of Maxillary sinus and associated spaces.
  • The Vent-Os Frontal Dilation System is indicated for dilation of the Frontal recess and associated spaces.
  • . The Vent-Os Sphenoid Dilation System is indicated for dilation of Sphenoid sinus and associated spaces.

Comparison to Predicate Device:

The Vent-Os® Sinus Dilation family has the same intended use and similar indication, mechanism of action, and procedural outcomes as the XprESS Multi-Sinus Dilation Tool (the predicate device). The Vent-Os® Sinus Dilation family also shares similar technological features and mechanism of action with the cleared Vent-Os Sinus Dilation System (for the maxillary sinus). The Vent-Os Sinus Dilation family dilates sinus ostia and spaces associated with paranasal sinus cavities via pressure applied radially in a substantially equivalent manner to the predicate device. The technological differences between the Vent-Os® Sinus Dilation family and its predicate do not raise new questions of safety or effectiveness. Performance data demonstrate substantial equivalence in terms of safety and effectiveness to the predicate device.

Performance Testing

Biocompatibility in conformance with applicable requirements of ISO 10993 was completed. This testing included:

Item #TestStandard ReferenceResults
1Cytotoxicity (MEM Elution)ISO 10993-5:2009Non-cytotoxic
2Sensitization-MaximizationISO 10993-10: 2010Non-sensitizing
3Intracutaneous ReactivityISO 10993-10: 2010Non-irritating

{5}------------------------------------------------

Premarket Notification 510(k) Summary - Continued

Bench testing to evaluate conformance with required device attributes and performance was completed. This testing included the following evaluations:

Attribute TestedPass/Fail
Device weightPassed
Placement System Working Length ProfilePassed
Placement System Working LengthPassed
Crossing ProfilePassed
Deployment ForcePassed
Dilation Device Diameter @ 1 hourPassed
Dilation Device Working Length @ 1 hourPassed
Dilation Device IntegrityPassed
Bond StrengthPassed

Transportation, sterilization, and shelf-life testing was completed to evaluate conformance with required attributes and performance. Sterilization was performed in conformance with all applicable requirements of ANSI/AAMI/ISO 11137-2: Sterilization of Health Care Products -Radiation. Results demonstrate achievement of a sterility assurance level (SAL) of 10°.

Transportation and packaging validation included the following:

Attribute TestedPass/Fail
Transit TestingPassed
Sterile Barrier Integrity (Bubble Test)Passed
Packaging & Product Visual InspectionPassed
Pouch Seal TestingPassed

{6}------------------------------------------------

Premarket Notification 510(k) Summary - Continued

Cadaver testing was completed by multiple physician users to validate usability. The cadaver evaluations included use of cadaveric head specimens with the devices placed multiple times into each targeted sinus site. Evaluators confirmed successful delivery and placement of the devices to the targeted sinus.

A clinical trial was also performed by SinuSys Corporation to demonstrate the substantial equivalence, in terms of the safety and effectiveness, of the Vent-Os® Sinus Dilation family to the predicate device cited. A multi-center, prospective, single arm, non-randomized trial was conducted at three (3) clinical sites with three (3) investigators to evaluate basic usability and confirm safety and effectiveness of the Vent-Os® Sinus Dilation family to dilate the frontal recess and sphenoid sinus ostia in patients with chronic rhinosinusitis. Use of the Vent-Os® Sinus Dilation family for maxillary sinus ostia dilation was the subject of cleared pre-market notification K133016. A total of 30 patients were enrolled and treated in the study. In the 30 treated subjects, 33/33 frontal recesses and 15/15 sphenoid sinus ostia were successfully treated. Clinical trial results achieved for frontal recess and sphenoid sinus ostia were comparable to the literature control as summarized in the following table:

SinuSys StudyBolger Study*
AnalysisPer Protocol95% CI**Per Protocol95% CI**
Frontal Recess
Acute1 week
N33124
Patent33 (100%)92.4% - 100%85 (69%)59.9% - 76.1%
Non-Patent07 (6%)2.8% - 11.2%
Indeterminate032 (26%)18.9% - 34.2%
3 Months3 Months
N3098
Patent28 (93%)78.7% - 98.2%82 (84%)75.1% - 89.7%
Non-Patent03 (3%)1.1% - 8.6%
Indeterminate2 (7%)1.9% - 21.3%13 (13%)7.9% - 21.4%
Sphenoid Sinus Ostium
Acute1 week
N1575

Comparison to Literature Control

{7}------------------------------------------------

Patent13 (87%)62.1% - 96.3%41 (55%)43.5 - 65.4%
Non-Patent0-0-
Indeterminate2 (13%)3.7% - 37.9%34 (45%)34.6% - 56.6%
3 Months3 Months
N1554
Patent10 (67%)41.7% - 84.8%32 (59%)46.0% - 71.3%
Non-Patent0-0-
Indeterminate5 (33%)15.2% - 58.3%22 (41%)28.7% - 54.0%
  • Literature control Bolger, et al. Safety and outcomes of balloon catheter sinusotomy: A multicenter 24-week analysis in 115 patients. Otolaryngol Head and Neck Surg 2007, 137: 10-20.

**Wilson score confidence intervals

Clinical trial results substantiated the Vent-Os® devices perform safely, effectively, and in a manner substantially equivalent to cited predicate and reference devices.

Summary:

Based upon the device description and test data provided in this submission the Vent-Os® Sinus Dilation family is substantially equivalent to the predicate device cited.

§ 874.4420 Ear, nose, and throat manual surgical instrument.

(a)
Identification. An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.(b)
Classification Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 874.9.