K Number
K160762
Date Cleared
2016-11-21

(245 days)

Product Code
Regulation Number
864.7470
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Diazyme Direct HbA 1c Assay (Enzymatic, On-Board Lysis) test kit is intended for use in the quantitative determination of stable HbA in venous whole blood samples with on-board blood lysis application in a clinical laboratory. This test is not to be used to diagnose or screen for diabetes. The measurement of HbA 1c concentration is for use in monitoring longterm glucose control of persons with diabetes. For in-vitro diagnostic use only.

Diazyme Direct HbA1c Assay Calibrator Set is intended to be used for calibration of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis). For in-vitro diagnostic use only.

Diazyme Direct HbA1c Assay Control Set is intended to be used for quality control by monitoring accuracy and precision of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis). For in-vitro diagnostic use only.

Device Description

The Diazyme Direct HbA1c assay kit (Enzymatic, On-Board Lysis) is comprised of a Reagent 1, Reagent 2, Lysis Buffer.
Diazyme Direct HbA1c Assay Calibrator Set: 2 levels, human whole blood based in lyophilized form. It is intended to be used for calibration of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis).
Diazyme Direct HbA1c Assay Control Set: 2 levels, human whole blood based in lyophilized form. It is intended to be used for quality control by monitoring accuracy and precision of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis).

The verification and/or validation studies were performed on the automated chemistry analyzer Roche Modular P (K953239/005). Modular P analyzer is a high through-put automated chemistry analyzer and that can process 800 samples per hour. The predicate performance studies at submission were performed on Hitachi 917 which is an obsolete member of the Roche Mod P Roche family of instruments. The Control Unit of the Roche Modular P analyzer uses a graphical interface to control all instrument functions. The computer, keyboard, and touchscreen monitor allows users to navigate through the software, enter assay, calibrator, and control information, and make test selections. The Diazyme Direct HbA1c Assay Modular P application parameters provided are programmed into the Modular P analyzers. The reagents, calibrators and controls are loaded into the analyzer. The Roche Modular P stores the Diazyme Direct HbA1c Assay reagents in a refrigerated compartment. Reagent and sample pipettes automatically aspirate and dispense specified amounts of reagents or calibrators, controls, and samples into reaction cells. The change in absorbance is measured at specified wavelengths.

AI/ML Overview

The document describes the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) and its performance, comparing it to an earlier version of the device (K070734) and a predicate device (Tosoh G8 HPLC HbA1c method).

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" for all performance metrics in a single table, but the conclusion for each study indicates whether the criteria were met. Based on the "Conclusion" sections for Precision and Method Comparison, the implicit acceptance criteria appear to be:

Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
Internal Precision (CV%)Less than 2% for within-run, between-run, between-day, between-lot, and total CV%.Sample 1 (4.64%): Total CV: 1.7% Sample 2 (5.36%): Total CV: 1.2% Sample 3 (7.51%): Total CV: 0.9% Sample 4 (9.61%): Total CV: 0.9% Sample 5 (11.89%): Total CV: 1.0% Con 1 Lot 1 (6.22%): Total CV: 1.0% Con 2 Lot 1 (9.47%): Total CV: 0.8% Con 1 Lot 2 (5.70%): Total CV: 1.1% Con 2 Lot 2 (9.11%): Total CV: 0.7% Con 1 Lot 3 (6.04%): Total CV: 1.2% Con 2 Lot 3 (9.68%): Total CV: 0.7% Cal 1 Lot 1 (6.22%): Total CV: 0.9% Cal 2 Lot 1 (12.30%): Total CV: 0.7% Cal 1 Lot 2 (5.68%): Total CV: 1.2% Cal 2 Lot 2 (9.70%): Total CV: 0.8% Cal 1 Lot 3 (6.04%): Total CV: 1.1% Cal 2 Lot 3 (11.30%): Total CV: 0.6% (All met the < 2% criteria.)
Multi-Site Precision (CV%)Less than 2% for within-run, between-run, between-day, between-site, and total CV%.Sample 1 (4.67%): Total CV: 1.4% Sample 2 (5.37%): Total CV: 1.2% Sample 3 (7.52%): Total CV: 1.0% Sample 4 (9.67%): Total CV: 1.4% Sample 5 (11.92%): Total CV: 1.2% Con1 Lot 1 (6.21%): Total CV: 0.6% Con2 Lot 1 (9.48%): Total CV: 0.9% Con1 Lot 2 (5.63%): Total CV: 1.0% Con2 Lot 2 (9.11%): Total CV: 0.7% Con1 Lot 3 (6.01%): Total CV: 0.9% Con2 Lot 3 (9.65%): Total CV: 1.0% Call Lot 1 (6.21%): Total CV: 0.9% Cal2 Lot 1 (12.32%): Total CV: 0.8% Call Lot 2 (5.63%): Total CV: 1.1% Cal2 Lot 2 (9.74%): Total CV: 1.0% Call Lot 3 (6.02%): Total CV: 0.9% Cal2 Lot 3 (11.36%): Total CV: 1.0% (All met the < 2% criteria.)
LOD- (Not explicitly stated, but determined)0.5% (maximal value among three reagent lots)
LOB- (Not explicitly stated, but determined)0.2% (maximal value among three reagent lots)
LOQCV less than 20%0.8% (for all three reagent lots)
LinearityBias ≤ 10%Assay is linear from 4% to 12% HbA1c (Bias ≤ 10% was reported as met).
Method Comparison (Accuracy)Excellent correlation to predicate device (Tosoh G8 HPLC HbA1c method). The implicit criteria from the predicate comparison are: - Correlation Coefficient (R): High (e.g., > 0.98) - Slope: Close to 1.0 - Intercept: Close to 0.0Combined (Linear Regression): - Slope: 1.023 - Intercept: -0.090 - Correlation Coefficient (R): 0.9937 Combined (Deming Regression): - Slope: 1.030 - Intercept: -0.140 - Correlation Coefficient (R): 0.9937 (The conclusion states the results met acceptance criteria.)
Analytical Specificity (Interference)< ±10% deviation due to interfering substances.No significant interference (< ±10% deviation) for Ascorbic Acid, Bilirubin, Bilirubin Conjugated, Hemoglobin (trace), Triglycerides, Glucose, Uric Acid, Urea, Acetaminophen, Acetylsalicylic Acid, Metformin, Ibuprofen, Glyburide, Total Protein, Vitamin E, Rheumatoid Factor at specified concentrations. No significant interference for Hemoglobin C (≤ 38.2%), Hemoglobin D (≤ 43.1%), Hemoglobin E (≤ 21.1%), and Hemoglobin S (≤ 37.3%). (High HbF (>10%) may result in inaccurate values).

2. Sample Size Used for the Test Set and the Data Provenance

  • Internal Precision:

    • Sample Size: 5 unaltered whole blood specimens (4.6%, 5.4%, 7.5%, 9.7%, 11.9% HbA1c), 3 lots of calibrators (2 levels each), and 3 lots of controls (2 levels each). Each sample/control/calibrator was tested 2 runs per day, 2 replicates per run over 20 working days. This results in N=240 data points per sample/control/calibrator (2 replicates * 2 runs/day * 20 days * 3 lots).
    • Data Provenance: The study was performed at Diazyme Laboratories. The whole blood samples were "IRB approved." Specific country of origin is not mentioned but Diazyme Laboratories is located in Poway, CA, USA. This appears to be prospective data collection, specifically for this study.
  • Multiple Site Precision:

    • Sample Size: Same 5 whole blood samples (4.6%, 5.4%, 7.5%, 9.7%, 11.9% HbA1c), 3 lots of calibrators and 3 lots of controls. Tested in duplicates per run, 2 runs per day for 5 working days with the same lot of reagent. N=60 data points per sample/control/calibrator (2 replicates * 2 runs/day * 5 days * 3 sites).
    • Data Provenance: Performed at Diazyme (internal site) and two external sites. Specific country of origin for external sites is not mentioned, but the overall study is assumed to be US-based given the FDA submission. This is prospective data collection.
  • LOD/LOB/LOQ: Based on three lots of reagents and analysis using CLSI EP17-A2, but specific sample numbers are not detailed beyond "three lots of reagents" tested with "5 concentrations" for LOQ.

  • Linearity:

    • Sample Size: 11 levels of whole blood linearity set, prepared by dilution of two whole blood samples (Low 3.8%, High 12.3% HbA1c). Tested in triplicate.
    • Data Provenance: Not explicitly stated, but likely from Diazyme Laboratories and prospective.
  • Method Comparison:

    • Sample Size:
      • Site 1: 124 unique K2 EDTA whole blood samples
      • Site 2: 132 unique K2 EDTA whole blood samples
      • Site 3: 120 unique K2 EDTA whole blood samples
      • Total unique samples: 376 (124+132+120).
    • Data Provenance: Not explicitly stated, but samples were "IRB approved" K2 EDTA whole blood samples, suggesting human subject data. Conducted at three different sites, likely US-based, and prospective for the purpose of this study.
  • Analytical Specificity (Interference): Three whole blood samples ("low", "medium", and "high" HbA1c concentrations), spiked with interfering substances and tested in triplicates. Data provenance not explicitly stated but likely internal and prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable in the context of this device. The device is an in-vitro diagnostic assay for measuring HbA1c concentration. The "ground truth" for these studies is typically established by:

  • Reference Methods: Such as the Tosoh G8 HPLC HbA1c method, which is an NGSP certification method.
  • Known Concentrations: For controls, calibrators, and linearity samples.
  • Clinical Laboratory Standards Institute (CLSI) guidelines: For study design and statistical analysis.

There are no human experts subjectively interpreting results or establishing a ground truth in the way a radiologist would for an imaging device.

4. Adjudication Method for the Test Set

Not applicable. As described above, the ground truth is established by reference methods and known concentrations, not by human adjudication of qualitative results. The studies involve quantitative measurements and statistical comparisons.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

Not applicable. This device is an automated in-vitro diagnostic assay, not an AI-assisted diagnostic tool that aids human readers in making a diagnosis. There is no human-in-the-loop performance component or "human readers" in the context of this specific regulatory submission.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

Yes, the studies presented (Precision, LOD/LOB/LOQ, Linearity, Method Comparison, Analytical Specificity) represent the standalone performance of the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) on the Roche Modular P analyzer. The device operates automatically without human intervention during the measurement process, aside from loading samples and reagents.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the performance studies was established using:

  • Reference Method: For the method comparison study, the Tosoh G8 HPLC HbA1c method was used as the comparator. This method is noted as an NGSP certification method used in NGSP reference laboratories, making it a highly reliable and recognized standard for HbA1c measurement.
  • Known Concentrations: For precision, LOD/LOQ, and linearity studies, samples, calibrators, and controls were prepared with known or target HbA1c concentrations. The calibrators are traceable to NGSP/DCCT and IFCC.
  • CLSI Guidelines: The studies followed established CLSI guidelines (EP5-A2, EP17-A2, EP6-A, EP9-A2), which specify rigorous methods for determining performance characteristics against accepted laboratory standards.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of a machine learning or AI model development. This device is a chemical assay, not an AI algorithm that requires training data in the traditional sense. The data provided describes the validation and verification of the assay's analytical performance.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for an AI model in this context, this question is not applicable. The device's performance characteristics are validated against established analytical methods and known standards (as described in point 7).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, with the head tilted upwards. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 21, 2016

DIAZYME LABORATORIES ABHIJIT DATTA DIRECTOR, TECHNICAL OPERATIONS 12889 GREGG COURT POWAY CA 92064

Re: K160762

Trade/Device Name: Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis), Diazyme Direct HbA1c Assay Calibrator Set, Diazyme Direct HbA1c Assay Control Set Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated hemoglobin assay Regulatory Class: II Product Code: LCP, JIT, JJX Dated: October 5, 2016 Received: October 6, 2016

Dear Abhijit Datta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160762

Device Name

Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis), Diazyme Direct HbA1c Assay Calbrator Set, Diazyme Direct HbA1c Assay Control Set

Indications for Use (Describe)

Diazyme Direct HbA 1c Assay (Enzymatic, On-Board Lysis) test kit is intended for use in the quantitative determination of stable HbA in venous whole blood samples with on-board blood lysis application in a clinical laboratory. This test is not to be used to diagnose or screen for diabetes. The measurement of HbA 1c concentration is for use in monitoring longterm glucose control of persons with diabetes. For in-vitro diagnostic use only.

Diazyme Direct HbA1c Assay Calibrator Set is intended to be used for calibration of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis). For in-vitro diagnostic use only.

Diazyme Direct HbA1c Assay Control Set is intended to be used for quality control by monitoring accuracy and precision of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis). For in-vitro diagnostic use only.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K160762

510k Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's name:Diazyme Laboratories
Submitter's address:12889 Gregg CourtPoway, CA 92064USA
Name of Contact Person:Dr. Abhijit DattaDiazyme Laboratories12889 Gregg CourtPoway, CA 92064Phone: 858-455-4762Fax: 858-455-2120
Date the Summary was Prepared:November 15, 2016
Name of the DeviceDiazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis)Diazyme Direct HbA1c Assay Calibrator SetDiazyme Direct HbA1c Assay Control Set
Trade Name:Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis)Diazyme Direct HbA1c Assay Calibrator SetDiazyme Direct HbA1c Assay Control Set
Common Name:Direct Enzymatic Assay, HbA1c
Classification Name:Glycosylated Hemoglobin Assay

Establishment Registration

The Device Establishment Registration Number is 2032900.

{4}------------------------------------------------

Regulatory Information:

Regulation DescriptionProduct CodeDevice ClassRegulationPanel
GlycosylatedHemoglobin AssayLCPII21 CFR 864.7470Hematology, 81
CalibratorJITII21 CFR §862.1150Chemistry, 75
Quality ControlMaterialJJXI21 CFR §862.1660Chemistry, 75

Submission type: Traditional 510k

Predicate Device

Diazyme Laboratories claims substantial equivalence to the currently marketed Diazyme Direct HbA1c Enzymatic Assay. The reference numbers are K070743.

Performance Standards

At this time, no special controls under Section 513 or performance standard under Section 514 have been issued for in vitro diagnostic products.

Manufacturing Address

The Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) will be manufactured and distributed by Diazyme Laboratories.

Manufacturing Site Address Diazyme Laboratories 12889 Gregg Court Poway, CA 92064

Description of the Device

Clinical Significance

Hemoglobin A1c is an important test recommended by the American Diabetes Association (ADA) and its usefulness was clarified by the United Kingdom Prospective Diabetes Study (UKPDS) and Diabetes Control and Complications Trial (DCCT). Currently, as part of the diabetes management program, HbA1c testing should be performed at least biannually in all patients and quarterly for patients whose therapy has changed or who are not meeting treatment goals. Glycohemoglobin is produced by non-enzymatic addition of glucose to amino groups in hemoglobin. HbA1c refers to glucose modified hemoglobin A (HbA) specifically at N-terminal

{5}------------------------------------------------

valine residues of hemoglobin beta chains. HbA1c test is used both as an index of mean glycemia and as a measure of risk for the development of diabetes complications. Therefore, the HbA1c test is a good indicator of glycemic control in the preceding 2-3 months.

Assay Principle

Diazyme Direct HbA1c test is an enzymatic assay in which lysed whole blood samples are subjected to extensive protease digestion with Bacillus sp protease. This process releases amino acids including glycated valines from the hemoglobin beta chains. Glycated valines then serve as substrates for specific recombinant fructosyl valine oxidase (FVO) enzyme. The recombinant FVO specifically cleaves N-terminal valines and produces hydrogen peroxide. This, in turn, is measured using a horseradish peroxidase (POD) catalyzed reaction and a suitable chromogen. No separate measurement for total Hemoglobin (Hb) is needed in this Direct HbA1c Assay.

Detailer Device Description

    1. The Diazyme Direct HbA1c assay kit (Enzymatic, On-Board Lysis) is comprised of a Reagent 1, Reagent 2, Lysis Buffer.
    1. Diazyme Direct HbA1c Assay Calibrator Set: 2 levels, human whole blood based in lyophilized form. It is intended to be used for calibration of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis).
    1. Diazyme Direct HbA1c Assay Control Set: 2 levels, human whole blood based in lyophilized form. It is intended to be used for quality control by monitoring accuracy and precision of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis).

The verification and/or validation studies were performed on the automated chemistry analyzer Roche Modular P (K953239/005). Modular P analyzer is a high through-put automated chemistry analyzer and that can process 800 samples per hour. The predicate performance studies at submission were performed on Hitachi 917 which is an obsolete member of the Roche Mod P Roche family of instruments. The Control Unit of the Roche Modular P analyzer uses a graphical interface to control all instrument functions. The computer, keyboard, and touchscreen monitor allows users to navigate through the software, enter assay, calibrator, and control information, and make test selections. The Diazyme Direct HbA1c Assay Modular P application parameters provided are programmed into the Modular P analyzers. The reagents, calibrators and controls are loaded into the analyzer. The Roche Modular P stores the Diazyme Direct HbA1c Assay reagents in a refrigerated compartment. Reagent and sample pipettes automatically aspirate and dispense specified amounts of reagents or calibrators, controls, and samples into reaction cells. The change in absorbance is measured at specified wavelengths.

Assay Scheme for Roche Modular P Analyzer

Use the following scheme as a guideline for analyzer application. Note: Diazyme Direct HbA1c Assay is an end-point assay and the first reading point A1 is right before the addition of reagent R2.

{6}------------------------------------------------

Image /page/6/Figure/0 description: This image shows a timeline of a process. At 0 minutes, 10 uL of blood, 125 uL of lysis buffer, and 128 uL of R1 are added. At 5 minutes, 56 uL of R2 is added, and A1 is measured at 700 nm. At 8 minutes, A2 is measured.

After a 2-point calibration, linear-fitting is used to fit calibration curve through mean values of the response for calibrators of known concentrations. The Roche Modular P calculates the HbA1c% concentration of a patient sample by interpolation of the obtained signal to a stored 2point calibration curve.

#: The Modular P analyzer performs automatic on-board 1: 12.5 dilution using lysis buffer (10 uL sample + 125 uL lysis buffer) for rapid blood sample lysis.

The following key parameters are input into the Mod P user defined channel interface using the provided Diazyme Direct HbA1c Assay (DZ168C) Parameter sheet.

Modular P Applications
Measuring mode:Absorbance
Abs. calculation mode:2 Point END
Reaction mode:S-R1/R3
Reaction direction:Increase
Primary/Secondary Wavelength:700nm/800nm
Calc. first/last:16/28
Unit:%
Pipetting parameters
R1:128 μL
Sample:10 µL; Sample Diluent (lysis buffer) 125 µL (1:12.5 dilution)
Diluted Sample:20 μL
R2:56 µL
Total volume:204 µL

The Design control verification and/or validation studies were performed on Roche Mod P automated analyzer.

Intended Use of the Device:

Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) test kit is intended for use in the quantitative determination of stable HbA1c in venous whole blood samples with on-board blood lysis application in a clinical laboratory. This test is not to be used to diagnose or screen for

{7}------------------------------------------------

diabetes. The measurement of HbA1c concentration is for use in monitoring long-term glucose control of persons with diabetes. For in-vitro diagnostic use only.

Diazyme Direct HbA1c Assay Calibrator Set is intended to be used for calibration of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis). For in-vitro diagnostic use only.

Diazyme Direct HbA1c Assay Control Set is intended to be used for quality control by monitoring accuracy and precision of Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis). For in-vitro diagnostic use only.

Performance Characteristics of Device

Precision Summary

  1. Internal Precision

Internal precision of the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) was evaluated according to CLSI EP5-A2 guideline. This assessment was performed on the Roche Modular P analyzer at Diazyme Laboratories.

In the study, three lots of reagents were used. For each lot of reagent, 5 unaltered whole blood specimens containing 4.6%, 5.4%, 7.5%, 9.7% and 11.9% HbA1c spanning AMR were tested. Three lots of HbA1c calibrators and three lots of HbA1c controls, two levels each, were tested as samples to evaluate precision. All samples were tested 2 runs per day, 2 replicates per run over 20 working days according to CLSI EP5-A2 protocol. Whole blood samples used in the study were IRB approved.

Precision data for the three lots of reagents tested are given below. Data analyzed with EP Evaluator (Release 11).

SampleMean(N=240)Within-RunBetween-RunBetween-dayBetweenlotTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
Sample 14.640.040.8%0.071.5%0.000.0%0.081.6%0.081.7%
Sample 25.360.050.9%0.050.9%0.000.0%0.071.2%0.071.2%
Sample 37.510.050.6%0.050.7%0.000.0%0.070.9%0.070.9%
Sample 49.610.060.6%0.050.5%0.030.3%0.080.9%0.080.9%
Sample 511.890.090.7%0.080.6%0.040.4%0.121.0%0.121.0%
Con 1 Lot 16.220.050.8%0.030.5%0.010.1%0.061.0%0.061.0%
Con 2 Lot 19.470.060.6%0.040.4%0.020.2%0.070.8%0.070.8%
Con 1 Lot 25.700.040.8%0.040.7%0.020.4%0.061.1%0.061.1%
Con 2 Lot 29.110.050.5%0.040.4%0.030.3%0.070.7%0.070.7%
Con 1 Lot 36.040.050.7%0.050.9%0.000.0%0.071.1%0.071.2%
Con 2 Lot 39.680.050.6%0.040.4%0.000.0%0.070.7%0.070.7%
Cal 1 Lot 16.220.040.6%0.040.6%0.000.0%0.050.8%0.050.9%
Cal 2 Lot 112.300.070.6%0.040.3%0.030.2%0.090.7%0.090.7%
Cal 1 Lot 25.680.050.9%0.040.7%0.020.4%0.071.1%0.071.2%
Cal 2 Lot 29.700.060.6%0.040.4%0.020.2%0.070.8%0.070.8%

{8}------------------------------------------------

Cal 1 Lot 36.040.050.8%0.040.7%0.000.0%0.061.0%0.061.1%
Cal 2 Lot 311.300.060.5%0.040.3%0.010.1%0.070.6%0.070.6%

Conclusion: 20-day reproducibility data for all samples tested showed that the within-run, between-run, between-day, between-lot and total CV% were less than 2%, meeting acceptance criteria.

  1. Multiple Site Precision Study (Three sites)

Multiple site precision study was performed at two external sites and the Diazyme site. In this study, same sets of samples were used at all sites. Multiple site precision of the Diazyme Direct HbA1c Assay was evaluated according to a modified CLSI EP5-A2 guideline on Roche Modular P analyzer. Five whole blood samples containing 4.6%, 5.4%, 7.5%, 9.7% and 11.9% HbA1c spanning assay AMR, three lots of calibrators and three lots of controls were tested in duplicates per run, 2 runs per day for 5 working days with the same lot of the reagent to evaluate between site assay precision. Testing was conducted at 3 different sites, by 3 different operators, on 3 different Modular P instruments. Within-run, between-day, between-site, and total precision were calculated.

SampleMeanWithin-RunBetween-RunBetween-dayTotal
(n=60)SD%CVSD%CVSD%CVSD%CV
Sample 14.670.051.0%0.040.8%0.020.5%0.071.4%
Sample 25.370.040.8%0.051.0%0.000.0%0.071.2%
Sample 37.520.050.7%0.060.8%0.000.0%0.081.0%
Sample 49.670.070.8%0.111.1%0.000.0%0.131.4%
Sample 511.920.090.8%0.090.8%0.060.5%0.141.2%
Con1 Lot 16.210.030.6%0.000.0%0.000.0%0.030.6%
Con2 Lot 19.480.050.6%0.050.6%0.030.3%0.080.9%
Con1 Lot 25.630.040.7%0.020.4%0.030.6%0.061.0%
Con2 Lot 29.110.050.5%0.040.4%0.030.3%0.070.7%
Con1 Lot 36.010.050.8%0.010.2%0.020.3%0.060.9%
Con2 Lot 39.650.060.7%0.050.5%0.040.4%0.091.0%
Call Lot 16.210.040.7%0.030.5%0.000.0%0.060.9%
Cal2 Lot 112.320.070.6%0.050.4%0.050.4%0.100.8%
Call Lot 25.630.061.0%0.000.0%0.020.3%0.061.1%
Cal2 Lot 29.740.060.6%0.060.7%0.030.4%0.101.0%
Call Lot 36.020.050.8%0.030.4%0.010.1%0.050.9%
Cal2 Lot 311.360.060.6%0.080.7%0.050.5%0.121.0%

Multiple site precision data is summarized in the following table:

MeanBetween-Site
Sample(n=60)SD%CV
Sample 14.70.071.4
Sample 25.40.061.2
Sample 37.50.070.9
Sample 49.70.121.3

{9}------------------------------------------------

Sample 511.90.141.2
Control 1 Lot 16.20.030.5
Control 2 Lot 19.50.080.9
Control 1 Lot 25.60.061.0
Control 2 Lot 29.10.070.7
Control 1 Lot 36.00.060.9
Control 2 Lot 39.60.090.9
Calibrator 1 Lot 16.20.050.8
Calibrator 2 Lot 112.30.100.8
Calibrator 1 Lot 25.60.051.0
Calibrator 2 Lot 29.70.101.0
Calibrator 1 Lot 36.00.050.9
Calibrator 2 Lot 311.40.121.0

Conclusion: All samples tested showed that the within-run precision, between-run, between-day, between-site, and total CV% were less than 2% meeting the acceptance criteria.

LOD/LOB/LOQ

The LOB, LOD, and LOQ of the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis Assay were determined according to CLSI EP17-A2 with three lots of the reagents.

The LOB was determined as 0.0%, 0.1% and 0.2% for reagent lot 1, lot 2, and lot 3 respectively. The HbA1c Assay LOB is obtained as the maximal value between all three reagents lots, 0.2%.

The LOD was determined as 0.38%. 0.43%, and 0.5% for reagent lot 1. lot 2. and lot 3 respectively. The Direct HbA1c Assay LOD is obtained as the maximal value among the three lots, 0.5%.

Using the calculated CV's from the 5 concentrations, curve fit analysis was performed to determine the upper 95% confidence level of point with CV less than 20%. EP Evaluator (Release 11) analysis showed LOQ = 0.8% for Reagent Lots 1, 2, and 3, respectively.

Linearity

11 levels of whole blood linearity set were prepared by dilution of two whole blood samples, Low and high samples with known HbA1c of 3.8% and 12.3% respectively were used to generate the test levels according to CLSI EP6-A and tested in triplicate, again, per guideline. Data analysis showed bias ≤10%. The assay is linear from 4% to12% HbA1c.

Calibrator Traceability

Diazyme Direct HbA1c Assay is traceable to NGSP/DCCT. The assay methodology is traceable to IFCC per NGSP correlation guideline (www.ngsp.org).

Method Comparison Study

{10}------------------------------------------------

Following the CLSI: EP9-A2 guideline, the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) was evaluated on the Roche Modular P testing individual K2 EDTA whole blood samples (IRB approved) in comparison with Tosoh G8 HPLC HbA1c method, which is used as the NGSP certification method in NGSP reference laboratories (www.NGSP.org). A total of 124 unique samples at site 1, total of 132 unique samples at site 2, and a total of 120 unique samples at site 3 were tested using the same lot of the reagents at three different sites on three different Modular P analyzers by three different operators.

Summary of Regression Analysis: Regular Linear

ParameterSite 1Site 2Site 3Combined
Slope1.0061.0331.0261.023
95% CI0.983 - 1.0301.015 - 1.0511.005 - 1.0471.011 - 1.035
Intercept0.04-0.126-0.18-0.090
95% CI-0.13 to 0.21-0.268 to 0.015-0.34 to -0.01-0.180 to 0.000
CorrelationCoefficient(R)0.99180.99500.99380.9937
Sample Range(Diazyme)4.4 - 12.04.2 - 11.94.4 - 12.04.2 - 12.0

Summary of Regression Analysis: Deming

ParameterSite 1Site 2Site 3Combined
Slope1.0151.0391.0331.030
95% CI0.992 – 1.0381.021 - 1.0571.012 - 1.0541.018 – 1.042
Intercept-0.02-0.167-0.23-0.140
95% CI-0.19 to 0.15-0.309 to -0.026-0.40 to -0.06-0.230 to -0.050
CorrelationCoefficient(R)0.99180.99500.99380.9937
Sample Range(Diazyme)4.4 - 12.04.2 - 11.94.4 - 12.04.2 - 12.0

{11}------------------------------------------------

Conclusion: Method comparison data analysis of the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) versus the predicate Tosoh HPLC Assay testing K2 EDTA whole blood at three different sites on three different Modular P analyzers by three different operators demonstrated the results met the acceptance criteria.

Analytical specificity

Endogenous substances:

To determine the level of interference from substances present whole blood samples, the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) was used to test three whole blood samples, which contained "low", "medium", and "high" HbA1c concentrations, following CLSI EP7-A2. To evaluate interference, each whole blood samples were spiked with potential endogenous interference substances per protocol and tested in triplicates on the Roche Modular P analyzer.

Interfering SubstancesConcentration
Ascorbic Acid12 mg/dL
Bilirubin15 mg/dL
Bilirubin Conjugated13 mg/dL
Hemoglobin21 mg/dL
Triglycerides4000 mg/dL
Glucose4000 mg/dL
Uric Acid30 mg/dL
Urea80 mg/dL
Acetaminophen20 mg/dL
Acetylsalicylic Acid65.2 mg/dL
Metformin4 mg/dL
Ibuprofen50 mg/dL
Glyburide0.19 mg/dL
Total Protein21000 mg/dL
Vitamin E13.6 mg/dL
Rheumatoid Factor375 IU/mL

The following substances showed no significant interference (< ±10% deviation) up to the concentrations summarized below.

Additionally acetylated, carbamylated and labile HbA1c does not adversely affect the assay. Hemoglobinopathies may interfere with glycated hemoglobin analysis. Testing results indicate that there is no significant interference for Hemoglobin C (≤ 38.2%), Hemoglobin D (≤ 43.1%), Hemoglobin E (≤ 21.1%), and Hemoglobin S (≤ 37.3%). High HbF (>10%) may result in inaccurate HbA1c values.

{12}------------------------------------------------

Substantial Equivalence

Device Modifications

The intended use of the modified device and fundamental scientific technology used in the test has not changed as a result of the modification.

Specific modifications:

    1. Change in number of reagents supplied with on-board lysis application
    1. Change in concentration of reagent components
    1. Change in product labeling to reflect the modification

Summary of Assay Kit Components

Predicate device K070734Subject Device
Lysis bufferLysis buffer
100mM CHES pH 8.7, 1% Triton X 100, 0.45% SDS, 0.5mM redox agent>100mM Tris pH >8.0, 1% Triton X 100,>1.5% nonionic and ionic detergents,>4KU/mL proteases
Reagent1aReagent 1
5mM MES pH 7.0, >4KU/mL proteases,>10µM redox agent5mM MES pH >6.0, <3mM redox agent
Reagent 1bN/A
1mM MES pH 6.3, <3mM redox agent
Reagent 2Reagent 2
15mM Tris pH8.0, >10U/mL FVO enzyme,90U/mL POD, 0.8mM chromagen>5mM bis-Tris pH>7.0, >10U/mL FVOenzyme, 90U/mL POD, >50μΜ chromagen
NoneNone
HbA1c Calibrator set(linear mode, lyophilized whole blood based)HbA1c Calibrator set(linear mode, lyophilized whole blood based)
1 x 0.5mL Level 11 x 0.5mL Level 1
1 x 0.5mL Level 21 x 0.5mL Level 2
HbA1c Control set(lyophilized whole blood based)HbA1c Control set(lyophilized whole blood based)
1 x 0.5mL Level 11 x 0.5mL Level 1

Substantial Equivalence Comparison Table

Indications for Use

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis) (modified device; special 510k submission)Equivalency
Diazyme Direct Enzymatic HemoglobinDiazyme Direct HbA1c AssaySame

{13}------------------------------------------------

A1c (glycated hemoglobin A1c; A1c;HbA1c) reagents are intended for use inthe quantitative determination of stableHbA1c in human whole blood samples.Measurement of hemoglobin A1c is avaluable indicator for long-term diabeticcontrol. For in-vitro diagnostic use only.(Enzymatic, On-Board Lysis) test kit isintended for use in the quantitativedetermination of stable HbA1c in venouswhole blood samples with on-boardblood lysis application in a clinicallaboratory. This test is not to be used todiagnose or screen for diabetes. Themeasurement of HbA1c concentration isfor use in monitoring long-term glucosecontrol of persons with diabetes. For in-vitro diagnostic use only.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Principle

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis)Equivalency
Direct Enzymatic HbA1c test is anenzymatic assay in which lysed wholeblood samples are subjected to extensiveprotease digestion with Bacillus spprotease. This process releases aminoacids including glycated valines from thehemoglobin beta chains. Glycated valinesthen serve as substrates for specificrecombinant fructosyl valine oxidase(FVO) enzyme, produced in E. coli. Therecombinant FVO specifically cleaves N-terminal valines and produces hydrogenperoxide. This, in turn, is measured usinga horse radish peroxidase (POD) catalyzedreaction and a suitable chromagen.Diazyme Direct HbA1c Assay is anenzymatic assay in which lysed wholeblood samples are subjected to extensiveprotease digestion with Bacillus spprotease. This process releases aminoacids including glycated valines from thehemoglobin beta chains. Glycated valinesthen serve as substrates for specificrecombinant fructosyl valine oxidase(FVO) enzyme. The recombinant FVOspecifically cleaves N-terminal valinesand produces hydrogen peroxide. This, inturn, is measured using a horseradishperoxidase (POD) catalyzed reaction anda suitable chromogen.Same
The HbA1c concentration is expresseddirectly as %HbA1c by use of a suitablecalibration curve in which the calibratorshave values for each level in %HbA1c.The HbA1c concentration is expresseddirectly as %HbA1c by use of a suitablecalibration curve in which thecalibrators have values for each level in%HbA1c.

Test Objective

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis)Equivalency
For the in vitro quantitativedetermination of % HbA1c in humanwhole blood.For the in vitro quantitativedetermination of % HbA1c in humanwhole blood.Same

Type of Test

{14}------------------------------------------------

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis)Equivalency
QuantitativeQuantitativeSame

Specimen Type

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis)Equivalency
EDTA Human whole blood.K2 EDTA Human whole blood.Same

Specimen Lysis

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis)Equivalency
Human whole blood Specimens arelysed manually off-boardHuman whole blood specimens arelysed on-board on analyzerDifferent

Product Type

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis)Equivalency
3 Reagents plus lysis buffer, Calibrators,Controls2 Reagents plus lysis buffer, Calibrators,ControlsDifferent

Performance Comparison

Direct HbA1c Enzymatic Assay(K070734)Direct HbA1c Assay (Enzymatic, On-Board Lysis)Equivalency
Linear Range: 4-16% HbA1c1Total Precision: 1.8%Method Comparison:Linear Range: 4-12 % HbA1c1Total Precision: 0.6-1.9%Method Comparison:Substantiallysimilar
Correlation Coefficient: 0.9874Slope/Intercept: y = 1.0212x + 0.0135Correlation Coefficient: 0.9937Slope/Intercept: y = 1.023x - 0.09

¹Linear Range limited to 4-12% HbA1c so that the upper limit of AMR is not too far from NGSP protocol for certification sample range.

{15}------------------------------------------------

Summary of Rationale for Considering the Device Substantially Equivalent to Devices Approved for Interstate Commerce

In summary, the Intended Use of the modified device has not changed. The assay principle and fundamental scientific technology of the modified device has not changed. The major changes are in the number of reagents supplied (from 3 reagent plus lysis buffer to 2 reagents plus lysis buffer), Concentration of components in the reagents and labeling changes reflecting the modification. Design control activities including precision, limit of quantitation, linearity, accuracy and interference studies demonstrates that the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) is substantially similar to the predicate device (K070734) and is safe and effective. There is no significant deviation between performance of the modified device Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) and the predicate device. The differences in the assay steps should not affect the safety and effectiveness of the Diazyme Direct Hbalc assay (Enzymatic, On-Board Lysis). Diazyme Direct HbA1c assay calibrator is traceable to NGSP certification for the modified device. The legally marketed device Tosoh HPLC HbA1c Test is used as standard NGSP certification method in NGSP reference laboratories and was thus selected for accuracy study. The Tosoh HPLC HbA1c Test is also the device used in accuracy studies for the predicate Diazyme HbA1c Enzymatic assay device (K070734). The accuracy studies with whole blood samples tested with the modified device unequivocally demonstrated excellent correlation. In conclusion, the Diazyme Direct HbA1c Assay (Enzymatic, On-Board Lysis) for measurement of HbA1c in patient whole blood samples is substantially equivalent to legally marketed devices. The differences in the assay steps should not affect the safety and effectiveness of the Diazyme Direct HbA1c assay (Enzymatic, On-Board Lysis) and risk analysis performed indicates no change in risk stance. The Diazyme Direct HbA1c assay gives similar diagnostic information as the other legally marketed HbA1c tests and offers users a rapid and convenient device for measuring HbA1c in human blood.

§ 864.7470 Glycosylated hemoglobin assay.

(a)
Identification. A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.(b)
Classification. Class II (performance standards).