K Number
K160639
Date Cleared
2016-05-04

(58 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Electrosurgical generator intended for tissue cutting and coagulation with electrosurgical accessories and ancillary equipment.

Device Description

The Olympus ESG-200 Electrosurgical Generators: (CELON Elite WA90001A, WA90002A and CELON Precision WA90008A, WA90009A), including its two variants the Celon Elite and Celon Precision are reusable, non-sterile electrosurgical generators that feature different mono- and bipolar cutting and coagulation modes. The maximum output power is 120 W. Both variants of the ESG-200 are intended for cutting and coagulation of tissue in open, laparoscopic and endoscopic surgery in conjunction with electrosurgical accessories and ancillary equipment. The front panels of the proposed ESG-200 features a touch screen GUI (graphical user interface) that displays the connection status of accessories and peripherals connected to the electrosurgical generator. Two contact quality indicators (one for split and one for non-split electrodes) are green illuminated if neutral electrodes are correctly connected and red if not. Three additional push buttons allow recalling a previously saved setting (Select Procedure), to assign a footswitch to a specific output socket (Footswitch), and to control several other functions (Menu), e.g. select language, touch tone control, output volume, or brightness. The touch screen is also used to show and modify the output settings (e.g. mode, output power, effect) as well as to control other functions (e.g. save settings). In addition, the proposed ESG-200 has a bipolar socket (with automatic instrument recognition of selected Olympus instruments), a monopolar socket (E- or B-Type depending on the model), and a neutral electrode socket.

AI/ML Overview

This document is a 510(k) summary for the Olympus ESG-200 Electrosurgical Generator, asserting its substantial equivalence to predicate devices. It describes the device, its intended use, technological characteristics, and performance data. However, it does not explicitly define acceptance criteria or detail a specific study proving the device meets them in the format requested.

Here's an analysis based on the provided text, addressing the points where information is available or can be inferred:

1. A table of acceptance criteria and the reported device performance

The document does not provide a formal table of acceptance criteria with corresponding performance metrics. Instead, it discusses the assessment of "substantial equivalence" based on comparing the ESG-200's performance with predicate devices.

The general performance criteria for electrosurgical generators would typically revolve around:

  • Tissue Effects: The ability to achieve comparable cutting and coagulation effects as the predicate device.
  • Electrical Waveforms: Generation of electrical waveforms that are comparable to those produced by the predicate device for various modes.
  • Safety Standards Compliance: Adherence to recognized international safety standards.

Based on the text:

Acceptance Criteria (Inferred)Reported Device Performance
Comparable Tissue Effects"The results demonstrate that comparable tissue effects and electrical waveforms are achieved with subject and predicate devices for all modes of operation." "Porcine muscle, kidney and liver tissue were used to demonstrated tissue effect equivalence between subject and predicate devices." "Testing confirmed that comparable tissue effects and electrical waveforms could be achieved for all modes of operation."
Comparable Electrical Waveforms"The electrical waveforms of the ESG-200 were in all cases comparable to the electrical waveforms of the predicate. This was confirmed for all output modes (waveforms) at rated load." "The purpose of the bench validation testing was to show equivalence of the electrical waveform between the subject and predicate devices." "Performance and validation tests incorporated the same range of waveform outputs and power levels." "During the validation testing the waveforms and test results were compared directly between the subject and predicate devices." "Waveform Validation reports for ESG-200 at all outputs and at rated load are provided in Appendix 12c."
Compliance with Recognized Safety and Performance StandardsThe document lists several FDA-recognized international standards to which the device complies, including: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-2, IEC 62304, IEC 60601-1-6, and ISO 14971. "The performance data support the safety of the device and demonstrate that the subject devices comply with the recognized standards as specified."
Software Validation (Moderate Level of Concern)"The software validation activities were performed in accordance with the FDA Guidance, 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices' (May 11, 2005)." "The device software is considered a 'Moderate Level of Concern'."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document mentions the use of "Porcine muscle, kidney and liver tissue" for tissue effect equivalence testing, but it does not specify the sample size (e.g., number of tissue samples or repetitions).
  • Data Provenance: Not explicitly stated, but bench testing on animal tissue (porcine) is indicated. The manufacturer, Olympus Winter & Ibe GmbH, is located in Hamburg, Germany, suggesting the testing likely occurred there or in a related facility. This would be prospective bench testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the study described is a bench performance test comparing physical characteristics (waveforms, tissue effects) between devices, not a study requiring human expert assessment for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable for the reasons stated above. The comparison was objective measurement of electrical waveforms and qualitative assessment of tissue effects.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device is an electrosurgical generator, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable as the device is not an algorithm or AI. The performance testing was of the electrosurgical generator itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the bench testing described:

  • Electrical Waveforms: The "ground truth" for electrical waveforms would be their direct measurement using an oscilloscope and high voltage probe and comparison against the established characteristics of the predicate device (which is itself a legally marketed, validated device).
  • Tissue Effects: The "ground truth" for tissue effects would be the observed impact on porcine tissue. The assessment was a direct comparison to the effects produced by the predicate device, implying the predicate's tissue effects serve as the reference standard.

8. The sample size for the training set

This section is not applicable as the device is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

This section is not applicable as the device is not an AI/machine learning device that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 4, 2016

Olympus Winter & Ibe GmbH % Mr. Graham Baillie Manager, Regulatory Affairs Gyrus ACMI. Inc. 136 Turnpike Rd. Southborough, Massachusetts 01772

Re: K160639

Trade/Device Name: Olympus ESG-200 Electrosurgical Generator: (CELON Elite WA90001A, WA90002A and CELON Precision WA90008A, WA90009A) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 2, 2016 Received: March 7, 2016

Dear Mr. Baillie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160639

Device Name

Olympus ESG-200 Electrosurgical Generator: (CELON Elite WA90002A and CELON Precision WA90008A, WA 90009A)

Indications for Use (Describe)

Electrosurgical generator intended for tissue cutting and coagulation with electrosurgical accessories and ancillary equipment.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary May 4, 2016 K160639

General Information

Manufacturer:Olympus Winter & Ibe GmbHKuehnstr. 6122045 Hamburg, GermanyEstablishment Registration Number: 9610773
Official Correspondent:Graham A.L. Baillie, MSManager, Regulatory AffairsGyrus ACMI, Inc.136 Turnpike Rd.Southborough, MA 01772-2104Phone: 508.804.2738Fax: 508. 804.2624Email: Graham.baillie@olympus-osta.comEstablishment Registration Number: 300379030-

Device Identification

Proprietary name:Olympus ESG-200 Electrosurgical Generators:(CELON Elite WA90001A, WA90002A andCELON Precision WA90008A, WA90009A)
Device Classification name:Electrosurgical cutting and coagulation device andaccessories
Regulation Medical Specialty:General & Plastic Surgery
Regulations Number:21 CFR 878.4400
Regulatory class:Class II
Product code:GEI

Predicate Device

Both variants (CELON PRECISION / CELON ELITE) of the proposed ESG-200 are considered substantially equivalent to the following legally marketed device:

Olympus ESG-400 (Primary Predicate, K141225) .

Both ESG-200 variants also have two secondary predicates for specific features as identified below:

Subject device variantSecondary predicate510(k) No
CELON PRECISIONESG-200VNUS RadiofrequencyGenerator, Model RFG2K040638
CELON ELITE ESG-200Celon Lab ENTK032838

Table 2.1: secondary predicate devices for a few specific features

{4}------------------------------------------------

Device Description

The Olympus ESG-200 Electrosurgical Generators: (CELON Elite WA90001A, WA90002A and CELON Precision WA90008A, WA90009A), including its two variants the Celon Elite and Celon Precision are reusable, non-sterile electrosurgical generators that feature different mono- and bipolar cutting and coagulation modes. The maximum output power is 120 W. Both variants of the ESG-200 are intended for cutting and coagulation of tissue in open, laparoscopic and endoscopic surgery in conjunction with electrosurgical accessories and ancillary equipment.

The front panels of the proposed ESG-200 features a touch screen GUI (graphical user interface) that displays the connection status of accessories and peripherals connected to the electrosurgical generator. Two contact quality indicators (one for split and one for non-split electrodes) are green illuminated if neutral electrodes are correctly connected and red if not. Three additional push buttons allow recalling a previously saved setting (Select Procedure), to assign a footswitch to a specific output socket (Footswitch), and to control several other functions (Menu), e.g. select language, touch tone control, output volume, or brightness. The touch screen is also used to show and modify the output settings (e.g. mode, output power, effect) as well as to control other functions (e.g. save settings). In addition, the proposed ESG-200 has a bipolar socket (with automatic instrument recognition of selected Olympus instruments), a monopolar socket (E- or B-Type depending on the model), and a neutral electrode socket.Indications for Use

Intended Use

Olympus ESG-200 Electrosurgical Generators intended for tissue cutting and coagulation in conjunction with electrosurgical accessories and ancillary equipment.

Technological Characteristics

The ESG-200 generator has the same intended use and technological characteristics as the primary predicate ESG-400 device. Various instruments can be connected to the monopolar and bipolar sockets. In addition, dedicated Olympus instruments can be connected to the bipolar socket with integrated self-recognition.

The PK modes of the primary predicate device (K141225) are not integrated in the ESG-200. The basic design philosophy of the User Interface (UI) and GUI flow chart concept is identical, except for the special ESG-400 amendment in regards to the PK instruments.

In comparison to the primary predicate device the following output modes are available:

Subject ESG-200 Model"CELON PRECISION"Subject ESG-200 Model"CELON ELITE"Predicate ESG-400
FineCutFineCutFineCut (Effect 1)
PureCutPureCutPureCut (Effect 2)
StrongCutStrongCutPureCut (Effect 3)
N/APulseCutPulseCut slow (Effect 2)

Output modes in comparison to the primary predicate device ESG-400

{5}------------------------------------------------

N/AN/ABlend Cut
N/AN/APulseCut fast

Monopolar Cut Modes

Subject ESG-200Model"CELON PRECISION"Subject ESG-200Model"CELON ELITE"Predicate ESG-400
SoftCoagSoftCoagSoftCoag (Effect 3)
ForcedCoagForcedCoagForcedCoag (Effect 3)
N/AN/APowerCoag
N/AN/ASprayCoag

Monopolar Coagulation Modes

Subject ESG-200 Model"CELON PRECISION"Subject ESG-200 Model"CELON ELITE"Predicate ESG-400
FineCutFineCutBipolarCut (Effect 1)
PureCutPureCutBipolarCut (Effect 2)
StrongCutStrongCutBipolarCut (Effect 3)
N/AN/ASalineCut
N/AN/APK PureCut
N/AN/APK SoftCut
N/AN/APK LoopCut
N/AN/APK MorceCut

Bipolar Cut Modes

Subject ESG-200 Model"CELON PRECISION"Subject ESG-200Model "CELON ELITE"Predicate ESG-400
SoftCoagSoftCoagBiSoftCoag (Effect 3)
N/AStrong RFITTRFCoag (w/o RCAP)
N/AStrong RFITT + RCAPRFCoag (w/ RCAP)
N/AN/AAutoCoag
N/AN/ASalineCoag
N/AN/AHardCoag
N/AN/AFineCoag
N/AN/APK Coag
N/AN/APK Softcoag
N/AN/APK AutoCoag

Bipolar Coagulation Modes

The range of output waveforms is identical but the power levels are decreased in comparison to the FDA cleared ESG-400 electrosurgical generator, K141225.

RFITT output modes in comparison to the secondary predicate devices

Additional bipolar coagulation modes were implemented exclusively for the RFITT (radiofrequency induced thermotherapy) functionality in conjunction with dedicated Olympus bipolar applicators.

Subject ESG-200 Model"CELON PRECISION"Secondary predicate VNUS Radiofrequency Generator,Model RFG2 (K040638)
RFITTonly RF power available
Pulse RFITT(no further modes)
Power RFITT

CELON PRECISION RFITT - Bipolar Coagulation Modes

{6}------------------------------------------------

Subject ESG-200 Model"CELON ELITE"Secondary predicate Celon Lab ENT (K032838)
Fine RFITTRFITT mode
Pure RFITT(no further modes)
CELON ELITE RFITT - Bipolar Coagulation Modes

CELON ELITE RFITT - Bipolar Coagulation Modes

Substantial Equivalence

Substantial equivalence is demonstrated by acknowledged verification/validation methodologies. The subject devices have equivalent technology, performance, dimensions and materials. The differences to the primary predicate ESG-400 are:

  • only one monopolar socket, no universal socket, no docking station interface . on the bottom
  • to be used only in conjunction with a double pedal footswitch ●
  • new enclosure
  • no PK instruments incorporation ●
  • RFITT Radiofrequency induced thermotherapy in conjunction with . dedicated Olympus instruments

Regarding the additionally implemented bipolar coagulation modes, exclusively for RFITT (radiofrequency induced thermotherapy) functionality, two secondary predicate devices were chosen because of their specific output modes. Substantial equivalence has been demonstrated by acknowledged verification/validation methodologies. The subject devices have equivalent technology and performance in respect to the compared modes.

Performance Data

The following performance data were provided in support of the substantial equivalence determination. All standards applied were FDA recognized international standards. The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005). The device software is considered a "Moderate Level of Concern".

Performance Testing - Bench

In general, the evaluations compared the function of the ESG-200 against the performance characteristics defined by the Design Specification and in comparison to the performance characteristics of the predicate devices. To demonstrate substantial equivalence the following aspects were considered within the validation versus the predicate devices. The results demonstrate that comparable tissue effects and electrical waveforms are achieved with subject and predicate devices for all modes of operation.

    1. Performance and validation tests incorporated the same range of waveform outputs and power levels.
    1. During the validation testing the waveforms and test results were compared directly between the subject and predicate devices.

{7}------------------------------------------------

The purpose of the bench validation testing was to show equivalence of the electrical waveform between the subject and predicate devices. Appendices in section 12 detail the validation testing. An oscilloscope and high voltage probe were used as measuring and test equipment. The electrical waveforms of the ESG-200 were in all cases comparable to the electrical waveforms of the predicate. This was confirmed for all output modes (waveforms) at rated load.

Porcine muscle, kidney and liver tissue were used to demonstrated tissue effect equivalence between subject and predicate devices (Appendix 12b). Waveform Validation reports for ESG-200 at all outputs and at rated load are provided in Appendix 12c

Comprehensive validation bench tests demonstrated and confirmed substantial equivalence to the predicate devices. Testing confirmed that comparable tissue effects and electrical waveforms could be achieved for all modes of operation. Clinical and animal studies were not deemed necessary to support substantial equivalence.

Standard No.Standard TitleFDA-Recognitionno + date
AAMI/ANSI ES60601-1:2005/(R)2012and C1:2009/(R)2012and, A2:2010/(R)2012Medical electrical equipment - Part 1: Generalrequirements for basic safety and essential performance(IEC 60601-1:2005, MOD)19-507/09/2014
IEC 60601-1-2Ed. 3.0 b:2007Medical electrical equipment - Part 1-2: Generalrequirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances -Requirements and tests19-110/17/2014
IEC 60601-1-8 Ed. 2.1:2012Medical electrical equipment - Part 1-8: Generalrequirements for basic safety and essential performance -Collateral Standard: General requirements, tests andguidance for alarm systems in medical electricalequipment and medical electrical systems5-7608/05/2013
IEC 60601-2-2 Ed. 5.0:2009Medical electrical equipment - Part 2-2: Particularrequirements for the basic safety and essentialperformance of high frequency surgical equipment andhigh frequency surgical accessories [Including:Technical Corrigendum 1 (2014)]6-33601/27/2015
IEC 62304 Ed. 1.0 b:2006Medical device software - Software life cycle processes13-808/20/2012
IEC 60601-1-6 Ed. 3.0:2010Medical electrical equipment Part 1-6: Generalrequirements for basic safety and essential performance -Collateral standard: Usability5-8507/09/2014
ISO 14971Secondedition 2007Medical devices - Application of risk management tomedical devices5-4008/20/2012

Applied standards

{8}------------------------------------------------

Conclusion

The performance data support the safety of the device and demonstrate that the subject devices comply with the recognized standards as specified.

In summary, we believe the ESG-200 is substantially equivalent with the predicate devices with respect to the general design approach, function, and the intended use. The ESG-200 raises no new concerns of safety or effectiveness when compared to the predicate devices.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.