(106 days)
The MAKO 7 is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the fallopian tube for cytological evaluation.
The MAKO 7 is a hysteroscope accessory intended to collect cell samples from the fallopian tube. The device is comprised of a catheter and a handle. The catheter includes a balloon, a shaft (which is made up of a stainless steel tube and a Nylon tube), a sheath (Nylon 12), and a sheath knob (Polycarbonate). The handle includes a drive wheel and an extension tube that is attached to a luer in the handle body. The extension tube attaches to a commercially available inflation device via a commercially available 3-way stopcock.
In summary, the physician inserts the MAKO 7 into the working channel of the hysteroscope until the distal tip of the catheter is positioned immediately proximal to the ostium of the fallopian tube. Then, the balloon is advanced into the fallopian tube and cells are collected on its surface.
The MAKO 7 is a sterile, single-use device. The device is terminally sterilized using ethylene oxide (EO).
Here's an analysis of the provided text regarding the acceptance criteria and study for the MAKO 7 device.
1. Table of Acceptance Criteria and Reported Device Performance
The FDA 510(k) summary provided does not explicitly present a formal "acceptance criteria" table with specific pass/fail thresholds. However, we can infer the key performance metrics and their reported outcomes from the "Clinical testing" section.
| Performance Metric | Implicit Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|---|
| Clinical - Device Navigate Fallopian Tube | High percentage of successful navigation, considering pre-existing anatomical factors. | Access achieved in 71/80 (89%) fallopian tubes. |
| Clinical - Collect Sample Adequate for Cytological Evaluation | High percentage of collected samples deemed adequate for cytological evaluation by a pathologist. | 70/71 (99%) of samples from accessed tubes were adequate. |
| Clinical - Adverse Events | No device-related adverse events. | No device-related adverse events reported. |
| Mechanical - Dimensional | Device dimensions meet specifications (assumed to ensure proper function and compatibility). | "Dimensional" testing performed (passed, implied). |
| Mechanical - Deployment | Device balloon deploys as intended (assumed to ensure effective sample collection). | "Deployment" testing performed (passed, implied). |
| Mechanical - Inflation Pressure | Device balloon inflates to specified pressure (assumed to ensure effective sample collection and prevent damage). | "Inflation Pressure" testing performed (passed, implied). |
| Ergonomics - Physician Feedback | Positive feedback from physicians on handle, drive wheel, hysteroscopic visibility, and adherence to IFU. | Confirmed handle ergonomics, drive wheel ergonomics, hysteroscopic visibility of balloon deployment, and ability to follow the IFU. |
Note: The "Implicit Acceptance Criteria (Inferred)" column is based on common regulatory expectations for medical device performance. The document doesn't explicitly state numerical targets for mechanical or ergonomic testing, but the inclusion of these tests implies successful completion.
2. Sample Size Used for the Test Set and Data Provenance
-
Clinical Test Set:
- Sample Size: 40 subjects (80 fallopian tubes)
- Data Provenance: Prospective, single-arm clinical study. Country of origin not explicitly stated, but typically assumed to be within the country of submission (USA for FDA).
-
Mechanical Test Set:
- Sample Size: n=15 for each of the following tests: Dimensional, Deployment, Inflation Pressure.
-
Ergonomics Test Set:
- Sample Size: 3 physicians per parameter.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
-
Clinical Study (Cytological Evaluation):
- Number of Experts: One study pathologist.
- Qualifications: "The study pathologist" is mentioned, implying a qualified individual in pathology, likely with expertise in cytological evaluation. Specific years of experience or board certifications are not provided.
-
Ergonomics:
- Number of Experts: 3 physicians.
- Qualifications: "Physician" is mentioned. Specific qualifications (e.g., gynecologist, experience with hysteroscopy) are not detailed but are implied by the nature of the device.
4. Adjudication Method for the Test Set
- For the cytological evaluation ground truth in the clinical study, the text states, "The study pathologist determined that 70/71 (99 %) of the samples were adequate for cytological evaluation." This indicates a single expert's determination, with no mention of a formal multi-reader adjudication method (e.g., 2+1, 3+1).
- For access success/failure, the determination appears to be based on direct observation during the procedure and confirmed by methylene blue dye injection in cases of non-access. This is a procedural outcome, not typically subject to expert adjudication in the same way as image interpretation.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The provided text does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss human readers' improvement with or without AI assistance. This device is a physical sample collection tool, not an AI-driven diagnostic or imaging interpretation system.
6. Standalone Performance
Yes. The clinical study of the MAKO 7 evaluated the standalone performance of the device in its intended use: "to demonstrate the safety and effectiveness of the subject device in navigating the fallopian tube and obtaining cell samples." This was an algorithm-only (device-only) evaluation without human-in-the-loop performance being a separate measured outcome. Physician action is required to use the device, but the device's performance in navigating and collecting samples was the primary endpoint.
7. Type of Ground Truth Used
-
Clinical Study:
- Clinical outcomes/observations (ability to navigate, successful sample collection).
- Pathology (assessment of sample adequacy for cytological evaluation by a study pathologist).
- Procedural observation (adverse events, methylene blue dye injection for tubal patency).
-
Mechanical Testing:
- Engineering specifications and physical measurements.
-
Ergonomics:
- Expert user (physician) qualitative feedback.
8. Sample Size for the Training Set
The document does not specify a separate training set. This is not an AI/machine learning device, so the concept of a "training set" as it relates to model development is not applicable here. The clinical and non-clinical data described are primarily for verification and validation of the manufactured product.
9. How the Ground Truth for the Training Set Was Established
Since there is no training set mentioned or implied for an AI/ML context, this question is not applicable to the MAKO 7 device as described. The ground truth mentioned above relates to the evaluation of the device's performance directly.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 9, 2016
nVision Medical Corporation % Cindy Domecus Principal Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsbrough, CA 94010
Re: K160510
Trade/Device Name: Mako 7 Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: Class II Product Code: HIH Dated: May 10, 2016 Received: May 13, 2016
Dear Cindy Domecus,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K160510
Device Name MAKO 7
Indications for Use (Describe)
The MAKO 7 is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the fallopian tube for cytological evaluation.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
SUBMITTER I.
510(k) Owner: nVision Medical Corporation 2500 Grant Road, G01 Mountain View, CA 94040 Surbhi Sarna, CEO Telephone: (408) 655-3577 surbhi@nvisionmedical.com
Contact Person: Cindy Domecus, R.A.C. (US &EU) Principal, Domecus Consulting Services LLC (650) 343-4813 (phone) (650) 343-7822 (fax) DomecusConsulting@comcast.net
Date Prepared: May 31, 2016
II. DEVICE
| Name of Device: | MAKO 7 |
|---|---|
| Common or Usual Name: | Hysteroscope Accessory |
| Classification Name: | Hysteroscope and accessories (884.1690) |
| Regulatory Class: | II |
| Product Code: | HIH |
PREDICATE DEVICE III.
Substantial equivalence was claimed to the following predicate device:
| 510(k) Number, Company | Trade Name | Model Number |
|---|---|---|
| K151275,nVision Medical Corporation | MAKO Device | MAKO 2.5 |
{4}------------------------------------------------
IV. DEVICE DESCRIPTION
The MAKO 7 is a hysteroscope accessory intended to collect cell samples from the fallopian tube. The device is comprised of a catheter and a handle. The catheter includes a balloon, a shaft (which is made up of a stainless steel tube and a Nylon tube), a sheath (Nylon 12), and a sheath knob (Polycarbonate). The handle includes a drive wheel and an extension tube that is attached to a luer in the handle body. The extension tube attaches to a commercially available inflation device via a commercially available 3-way stopcock.
In summary, the physician inserts the MAKO 7 into the working channel of the hysteroscope until the distal tip of the catheter is positioned immediately proximal to the ostium of the fallopian tube. Then, the balloon is advanced into the fallopian tube and cells are collected on its surface.
The MAKO 7 is a sterile, single-use device. The device is terminally sterilized using ethylene oxide (EO).
The device is intended for use in the hospital/clinic/physician's office.
The duration and type of contact is classified as follows:
- Category: External Communicating Device ●
- Contact: Mucosal membrane
- Duration: A - Less than 24 hours
INTENDED USE/INDICATIONS FOR USE V.
The indications for use for the subject device are as follows:
The MAKO 7 is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the fallopian tube for cytological evaluation.
The indications for use differ from that of the predicate, in that the indication for use for the subject device refers to the collection of samples from the fallopian tube instead of the proximal fallopian tube only. This difference does not represent a new intended use, since 1) device use is in the same anatomical structure – the fallopian tube, and 21 the difference in location within the fallopian tube does not alter the general purpose of the device to collect cells for cytological evaluation. As described in FDA's July 2014 guidance regarding evaluating substantial equivalence, the proposed indication is a "tool type" indication, which is not affected by the change in location within the fallopian tube.
{5}------------------------------------------------
Also, reference to device use in the fallopian tube does not raise different questions of safety and effectiveness as compared to use within the proximal portion of the fallopian tube.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The mechanism of action and principles of operation of the subject device and the predicate device are identical. Both devices use an identical handle, catheter body, and, most distally, a balloon that collects the sample. For both devices, this distal portion is actuated by a drive wheel on the handle and a hysteroscopic view is required.
There is only one difference in design between the predicate device and the subject device. This difference is that the balloon of the subject device extends 7cm into the fallopian tube while the balloon of the predicate device extends 2.5cm into the fallopian tube.
Since the subject and predicate devices collect cells by the same mechanism and in the same anatomical structure, and because the handle body, catheter body, and cell collection portion are the same, the subject device does not raise different questions of safety or effectiveness. There are also no new risks for collection of cells in the fallopian tube as compared to cell collection in the proximal fallopian tube. Indeed, in a clinical study of the subject device, no adverse events were reported.
VII. PERFORMANCE DATA
Nonclinical and clinical testing included the following:
Mechanical testing
- Dimensional (sample size of n=15 was used for each test) ●
- Deployment (sample size of n=15 was used for each test) ●
- Inflation Pressure (sample size of n=15 for each test)
Ergonomics
- Physician feedback on MAKO 7 confirmed handle ergonomics, drive wheel ● ergonomics, hysteroscopic visibility of balloon deployment, and ability to follow the IFU.
- . A sample size of 3 physicians per parameter was used.
Clinical testing
A prospective, single-arm clinical study was conducted to demonstrate the safety and effectiveness of the subject device in navigating the fallopian tube and obtaining cell samples. The MAKO 7 was evaluated in 40 subjects (80 fallopian tubes) who were already scheduled to undergo a laparoscopic tubal ligation.
{6}------------------------------------------------
Study endpoints included 1) ability of the device to navigate the fallopian tube, 2) ability of the device to collect a sample adequate for cytological evaluation, and 3) adverse events. Access was achieved in 71/80 (89%) fallopian tubes. The remaining 9 tubes were determined to have pre-existing tubal occlusion, as determined by methylene blue dye injection (in two of the tubes not initially accessed a second access and navigation was successful after incidental clearing of tubal blockage after methylene blue dye injection).
The study pathologist determined that 70/71 (99 %) of the samples were adequate for cytological evaluation.
There were no device-related adverse events reported.
VIII. CONCLUSION
The bench and clinical performance data provided in support of the 510(k) confirm that the MAKO 7 subject device is as safe and effective and substantially equivalent to the MAKO 2.5 predicate device.
§ 884.1690 Hysteroscope and accessories.
(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.