K Number
K160108

Validate with FDA (Live)

Device Name
Synapse PACS
Date Cleared
2016-02-03

(15 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FUJIFILM Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with FUJIFILM Synapse PACS Software (Server). The FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities.

The FUJIFILM Synapse PACS Software enables the display, comparison of 3D (MIP/MPR) of CT, MR, PET and SPECT studies. Typical users are radiologists and clinicians. These functions (MIPMPR/Fusion) are not intended for Mammography use.

The FUJIFILM Synapse PACS Software may be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA cleared, DICOM compatible displays for mammography.

Device Description

The proposed Synapse PACS Software is an implementation that combines capabilities of Synapse Workstation software (K112439) and Synapse MPR/Fusion software (K113244) in a single system, using the current Internet standards for Web clients and servers. The Synapse PACS Software and the predicate devices, i.e., Synapse Workstation Software (K112439) and Synapse MPR Fusion Software (K113244), are picture archiving and communication systems (as defined by 21 CFR 892.2050). Synapse PACS Software is the web based (client/server) application and implementation of the Synapse Workstation Software and the Synapse MPR Fusion Software. The Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with Synapse PACS Software (server). The Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities. Synapse PACS Software enables the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies (K113244).

In summary, this 510(k) submission introduces the Synapse PACS with the web based (client/server) application and the ability to the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies.

AI/ML Overview

The provided text is a 510(k) summary for the FUJIFILM Synapse PACS (K160108). This document primarily focuses on establishing substantial equivalence to previously cleared predicate devices (Synapse Workstation K112439 and Synapse MPR Fusion K113244) rather than presenting a performance study with detailed acceptance criteria and testing methodology for a novel diagnostic algorithm.

The document describes the device, its intended use, and provides comparisons of features, technology, and specifications with the predicate devices. It also mentions general testing activities. However, it does not detail a specific study proving the device meets acceptance criteria in the way one would for a new AI/CAD device. The Synapse PACS is a picture archiving and communication system, and its "performance" is more related to its functionality, speed, and ability to display and process images, rather than the diagnostic accuracy of an interpretation algorithm.

Therefore, many of the requested elements for describing a study proving the device meets acceptance criteria for a diagnostic AI/CAD device are not applicable or not detailed in this 510(k) summary.

Here's an attempt to answer the questions based on the provided text, indicating where information is not present:


Acceptance Criteria and Study for FUJIFILM Synapse PACS (K160108)

This 510(k) submission for the Synapse PACS focuses on demonstrating substantial equivalence to predicate devices for its functionalities as a PACS system, not on proving diagnostic performance of a new AI algorithm. Therefore, the "acceptance criteria" and "study" described below are interpreted in the context of a PACS system's performance and equivalence, rather than the accuracy of an automated diagnostic output.

1. Table of Acceptance Criteria and Reported Device Performance

For a PACS system, acceptance criteria are generally related to its functionality, performance (speed, responsiveness), and ability to display and process images as intended. The document states "Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests successfully passed." Specific quantitative criteria are not listed for individual functions.

Acceptance Criteria Category (Derived)Stated Performance / Verification
Functionality: Display, comparison, processing of various modalities (X-ray, CT, MR, US, NM, DICOM compliant images)"Synapse PACS Software enables the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies." "The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities."
User Interface: Serving as primary user interface for image processing and presentation."The Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed."
Mammography Image Handling: Processing, display, manipulation, and interpretation of mammography images."The FUJIFILM Synapse PACS Software may be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA cleared, DICOM compatible displays for mammography."
System Compatibility/Stability: Functioning on specified OS and browsers, networking, database.Verified through "system level functionality test, component testing, verification testing, integration testing, usability testing, installation/upgrade testing, labeling testing." (Specific metrics not provided). Assessed through direct comparison to predicate devices' features, technology, and specifications.
Safety and Effectiveness: No new safety/efficacy issues compared to predicates."Synapse PACS introduces no new safety or efficacy issues other than those already indentified with the predicate device." "The device is found to be safe and effective and substantially equivalent to the predicate devices."
Performance (General): Achieving expected accuracy performance (unspecified metric)."In addition, we conducted benchmark performance testing using actual clinical images to help demonstrate that the proposed device achieved the expected accuracy performance."

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "actual clinical images" were used for "benchmark performance testing" but does not specify the sample size or their provenance (e.g., country of origin, retrospective/prospective collection). Given the nature of a PACS system, the "test set" would likely be a diverse range of images to ensure format compatibility and display accuracy across modalities, rather than a specific set for diagnostic accuracy assessment.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable. For a PACS system, "ground truth" in the diagnostic sense is not typically established by experts as the system's function is to display and process, not to provide diagnostic interpretations. The ground truth for its performance would be whether it accurately displays images, performs requested manipulations, and stores/retrieves data correctly. This would be verified through technical testing and potentially user acceptance testing by radiologists/clinicians, but not in the same way as establishing ground truth for an AI diagnostic algorithm.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used to resolve discrepancies among expert readers when establishing ground truth for diagnostic studies. This is not relevant to the functional validation of a PACS system.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. An MRMC study is typically performed to evaluate the impact of an AI algorithm on human reader performance (e.g., improved diagnostic accuracy, reduced reading time). This device is a PACS system, not a diagnostic AI algorithm, so such a study would not be applicable or expected.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a PACS system, which is a tool for human interpretation and management of medical images. It does not perform standalone diagnostic interpretations.

7. The Type of Ground Truth Used

The "ground truth" for a PACS system would be the technical specifications and clinical expectations for image display, processing, and storage. For example:

  • Technical Specifications: Does the system accurately render images according to DICOM standards? Are measurements correct?
  • Functional Expectations: Does it perform MIP/MPR as specified? Does it retrieve images within an acceptable timeframe?
    These are assessed through verification and validation testing, not typically through pathology, expert consensus, or outcomes data in the diagnostic sense. The document refers to "benchmark performance testing using actual clinical images," which suggests an evaluation against expected display and processing quality for real-world data.

8. The Sample Size for the Training Set

Not applicable. PACS software, like the Synapse PACS, is developed based on software engineering principles and medical imaging standards (e.g., DICOM). It does not use "training sets" in the machine learning sense to learn to perform a task. It is a rule-based system designed to display and manage images.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no "training set" for a PACS system in this context.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. Above the profiles is a stylized image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 3, 2016

FUJIFILM Medical Systems U.S.A., Inc. % Jyh-Shyan Lin Senior Manager, Regulatory, Ouality and Clinical Affairs 419 West Avenue STAMFORD CT 06902

Re: K160108 Trade/Device Name: Synapse PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 15, 2016 Received: January 19, 2016

Dear Jyh-Shyan Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ochs

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160108

Device Name Synapse PACS

Indications for Use (Describe)

FUJIFILM Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with FUJIFILM Synapse PACS Software (Server). The FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities.

The FUJIFILM Synapse PACS Software enables the display. comparison of 3D (MIP/MPR) of CT, MR, PET and SPECT studies. Typical users are radiologists and clinicians. These functions (MIPMPR/Fusion) are not intended for Mammography use.

The FUJIFILM Synapse PACS Software may be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA cleared, DICOM compatible displays for mammography.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

FUJ}FILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse PACS

510(k) Summary

510(k) Number: K160108

Date Prepared:

February 1, 2016

Submitter's Information:

FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (301) 251-1092 Fax: (203) 602-3785 Contact: Jyh-Shyan Lin

Device Trade Name:

Synapse PACS

Device Common Names:

Picture Archiving and Communications System (PACS)

Device Classification Name:

System, Image Processing, Radiological

Product Code:

LLZ

Regulation Number: 21 CFR 892.2050

Device Class:

Class II

Panel:

Radiology

Predicate Devices:

  • . FUJIFILM Synapse Workstation Software (K112439)
  • Synapse MPR Fusion (K113244) .

{4}------------------------------------------------

Synapse PACS

1. Description of the Device

The proposed Synapse PACS Software is an implementation that combines capabilities of Synapse Workstation software (K112439) and Synapse MPR/Fusion software (K113244) in a single system, using the current Internet standards for Web clients and servers. The Synapse PACS Software and the predicate devices, i.e., Synapse Workstation Software (K112439) and Synapse MPR Fusion Software (K113244), are picture archiving and communication systems (as defined by 21 CFR 892.2050). Synapse PACS Software is the web based (client/server) application and implementation of the Synapse Workstation Software and the Synapse MPR Fusion Software. The Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with Synapse PACS Software (server). The Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities. Synapse PACS Software enables the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies (K113244).

In summary, this 510(k) submission introduces the Synapse PACS with the web based (client/server) application and the ability to the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies.

2. Indication for Use

FUJIFILM Synapse PACS Software (client) is intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with FUJIFILM Synapse PACS Software (Server). The FUJIFILM Synapse PACS Software is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The Synapse PACS Software can process medical images from the following modality types: plane X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine and images from other DICOM compliant modalities.

The FUJIFILM Synapse PACS Software enables the display, comparison and fusion of 3D (MIP/MPR) of CT, MR, PET and SPECT studies. Typical users are radiologists and clinicians. These functions (MIP/MPR/Fusion) are not intended for Mammography use.

The FUJIFILM Synapse PACS Software may be used to process FUJIFILM's DICOM MG "For Processing" images and also for the display, manipulation, and interpretation of lossless compressed or non-compressed mammography images that have been received in the DICOM For Presentation format and displayed on FDA cleared, DICOM compatible displays for mammography.

{5}------------------------------------------------

FUJIYFILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse PACS

3. Substantial Equivalence Comparison

3.1 Device features and technological characteristics comparison

Table 1. Synapse PACS 5.1.0 vs. Synapse Workstation 3.3.0 (K112439)

FeaturesSynapse PACS 5.1.0(This submission)Synapse Workstation 3.3.0(K112439)
Product availabilityProvided as Software onlyProvided as Software only
Operating SystemsWindows 7/8.1Windows 2000/XP
Web BrowserInternet Explorer Version: 10 or 11or Google ChromeInternet Explorer
Database and file access methodWeb based (http(s))Web based (http(s))
Display of Digital Mammography ImagesYesYes
Process of Digital Mammography (Non-Processed or For Process) ImagesYesYes
On-demand access to database and imagesYesYes
Viewing study listsYesYes
Decompression for compressed images beforedisplayYesYes
Display imagesYesYes
Viewing ReportsYesYes
Hanging ProtocolYesYes
Spine LabelingYesYes
Reference line displayYesYes
Key images identificationYesYes
Link multiple seriesYesYes
Measurement ToolsYesYes
Annotation ToolsYesYes
Standard Image Manipulation Tools (Windowwidth/level, Zoom, Pan, etc.)YesYes
FCR IPSS for Fuji-CR imagesYesYes
CT IPSS for CT imagesYesYes

{6}------------------------------------------------

Synapse PACS

FeaturesSynapse PACS 5.1.0(This submission)Synapse MPR Fusion 2.5(K113244)
Product availabilityProvided as Software onlyProvided as Software only
Operating SystemsWindows XP/Vista/Windows 7Windows 7, Windows 8.1 64bit
Web BrowserInternet Explorer Version: 10 or 11 or GoogleChromeInternet Explorer
Database and file access methodWeb based (http(s))Web based (http(s))
Orthogonal and Oblique Multi PlanarReconstruction (MPR)YesYes
Maximum, Average, and Minimum IntensityProjection (MIP, MPVR, MinIP)YesYes
Curved Multi Planar Reconstruction (CPR)NoYes
Independent Window/Level of upper and lowerimages in Fusion StudyYesYes
Save Preferences by User ProfileNoYes
Save Window/Level PreferencesYesYes
Multi-Modal Fusion to include MR/PET/NM/andCTYesYes
Compare Current Study to previous study inFusion ViewYesYes
Compare Current Study to Prior Study inCompare ModeYesYes
Manual Fusion Correction with RigidRegistration in Blend and overlay modesNoYes
Save series to Synapse PACSYesYes

{7}------------------------------------------------

FUJIFILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse PACS

3.2 Technology Similarities and Differences

TechnologySynapse PACS 5.1.0(This Submission)Synapse Workstation 3.3.0(K112439)
Image and Data ProcessingServer SideClient Side
Technology Platform (Synapse Client)Platform Independent (Browser based UI)Windows Active-X
Technology Platform (Synapse Server)Windows ServerWindows Server
System SecurityWindows AuthenticationWindows Authentication
Programming Language(s) (Synapse Client)Javascript (TypeScript), HTML5, CSSC++, Active-X

Table 3. Synapse PACS 5.1.0 vs. Synapse Workstation 3.3.0 (K112439)

Table 4. Synapse PACS 5.1.0 vs. Synapse MPR Fusion 2.5 (K113244)

TechnologySynapse PACS 5.1.0(This Submission)Synapse MPR Fusion 2.5(K113244)
Image and Data ProcessingServer SideClient Side
Technology Platform (Synapse Client)Platform Independent (Browser based UI)Windows Native Application
Technology Platform (Synapse Server)Windows ServerWindows Server
System SecurityWindows AuthenticationWindows Authentication
Programming Language(s) (Synapse Client)Javascript (TypeScript), HTML5, CSSC++

{8}------------------------------------------------

FUJ}{F||LM|MEDICAL SYSTEMS U.S.A., Inc.

Synapse PACS

3.3 Device Specifications Comparison

SpecificationsSynapse PACS 5.1.0(This Submission)Synapse Workstation 3.3.0(K112439)
Client O/SWindows 7/8.1Windows 2000/XP
BrowserInternet Explorer Version: 10/11 or Google ChromeInternet Explorer
Server O/SWindows 2008 R2, or 2012 R2 Server OSWindows 2003, or 2008 Server OS
DatabaseOracleOracle

Table 5. Synapse PACS 5.1.0 vs. Synapse Workstation 3.3.0 (K112439)

Table 6. Synapse PACS 5.1.0 vs. Synapse MPR Fusion 2.5 (K113244)

SpecificationsSynapse PACS 5.1.0(This Submission)Synapse MPR Fusion 2.5(K113244)
Client O/SWindows 7/8.1Windows 2000/XP
BrowserInternet Explorer Version: 10 or 11 or Google ChromeInternet Explorer
Server O/SWindows 2008 R2, or 2012 Server R2 OSWindows 2003, or 2008 Server OS
DatabaseOracleN/A

{9}------------------------------------------------

FUJIFILM MEDICAL SYSTEMS U.S.A., Inc.

Synapse PACS

4. Safety Information

Synapse PACS introduces no new safety or efficacy issues other than those already indentified with the predicate device. The Risk Management and the results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The Synapse PACS labeling contains instructions for use and necessary cautions, warnings and notes to provide the safe and effective use of the device.

5. Testing and Performance Information

Synapse PACS is tested successfully with reference to its product requirements, as well as design verification and validation documents and traceability matrix document. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the Synapse PACS software, which is found to be safe and effective and substantially equivalent to the predicate devices.

Testing involved system level functionality test, component testing, verification testing, integration testing, usability testing, installation/upgrade testing, labeling testing, as well as the testing for risk mitigations associated with the risk management process. In addition, we conducted benchmark performance testing using actual clinical images to help demonstrate that the proposed device achieved the expected accuracy performance.

Pass/Fail criteria were based on the requirements and intended use of the product. Test results showed that all tests successfully passed.

6. Conclusion

This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).