K Number
K160038
Device Name
UniStrip1 Generic Blood Glucose Test Strips
Date Cleared
2016-07-29

(204 days)

Product Code
Regulation Number
862.1345
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The UniStrip1™ Generic Blood Glucose Test Strips are used with the OneTouch® Ultra@2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The strips are meant for self-testing of blood glucose as an aid to monitor the effectiveness of diabetes control. They are for single patient use only and should not be shared. They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm. Testing is done outside the body (in vitro diagnostic use). They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control. Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. The UniStrip1™ Generic Blood Glucose Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch® Ultra® meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly). The UniStrip™ Control Solutions are for use with the UniStrip1™ Generic Blood Glucose Test Strips and OneTouch® Ultra® meters to check that the meters and test strip are working together properly and that the test is performing correctly.
Device Description
UniStrip1 Generic Blood Glucose Test Strips are used with the OneTouch® Ultra, OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters. The test strips are intended for use outside the body (in vitro diagnostic use) by people with diabetes using the OneTouch® Ultra®, OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® blood glucose monitoring systems as an aid to monitor the effectiveness of diabetes control. UniStrip1 Generic Blood Glucose Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch* Ultra* meter). UniStrip1 Generic Blood Glucose Test Strips can only be used with the OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012. UniStrip1 Generic Blood Glucose Test Strips are only for use with calibration code 49.
More Information

Not Found

No
The summary describes a standard blood glucose test strip and its performance characteristics, with no mention of AI or ML in the device description, intended use, or performance studies.

No.
This device is a diagnostic tool that measures glucose levels, which aids in monitoring diabetes control. It does not directly treat or provide therapy.

Yes
The device is described as measuring glucose (sugar) in whole capillary blood as an aid to monitor the effectiveness of diabetes control, which is a diagnostic purpose. It explicitly states "Testing is done outside the body (in vitro diagnostic use)." and "Not intended for the diagnosis of or screening for diabetes mellitus", further clarifying its role in monitoring, which is a diagnostic activity post-diagnosis.

No

The device is a blood glucose test strip, which is a physical component used in conjunction with a blood glucose meter. It is an in vitro diagnostic device that requires a physical sample of blood and chemical reactions on the strip to measure glucose levels. This is not a software-only device.

Yes, this device is an IVD (In Vitro Diagnostic).

The document explicitly states:

  • "Testing is done outside the body (in vitro diagnostic use)."
  • "The test strips are intended for use outside the body (in vitro diagnostic use)..."

N/A

Intended Use / Indications for Use

The UniStrip1™ Generic Blood Glucose Test Strips are used with the OneTouch® Ultra@2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The strips are meant for self-testing of blood glucose as an aid to monitor the effectiveness of diabetes control.

They are for single patient use only and should not be shared.

They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm. Testing is done outside the body (in vitro diagnostic use).

They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control. Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

The UniStrip1™ Generic Blood Glucose Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch® Ultra® meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly).

The UniStrip™ Control Solutions are for use with the UniStrip1™ Generic Blood Glucose Test Strips and OneTouch® Ultra® meters to check that the meters and test strip are working together properly and that the test is performing correctly.

Product codes (comma separated list FDA assigned to the subject device)

NBW, CGA, JJX

Device Description

UniStrip1 Generic Blood Glucose Test Strips are used with the OneTouch® Ultra, OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters. The test strips are intended for use outside the body (in vitro diagnostic use) by people with diabetes using the OneTouch® Ultra®, OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® blood glucose monitoring systems as an aid to monitor the effectiveness of diabetes control. UniStrip1 Generic Blood Glucose Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch* Ultra* meter). UniStrip1 Generic Blood Glucose Test Strips can only be used with the OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012. UniStrip1 Generic Blood Glucose Test Strips are only for use with calibration code 49.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

finger, palm, or forearm

Indicated Patient Age Range

Not intended for use on neonates.

Intended User / Care Setting

self-testing by people with diabetes in their home

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

  1. Sample volume study per ISO 15197:2013: The glucose measurements of sample volume from 1.0 to 1.5 µL met acceptance criteria, requiring at least 1.0 µL for normal testing.
  2. Operation condition study per EN 23640:2013: Evaluated performance in normal and extreme environments (42-111 °F (6-44°C), 10-90% R.H.). Individual bias was within ±10 mg/dL for glucose 75 mg/dL compared to Glucose Analyzer YSI 2300.
  3. Precision study per NCCLS/CLSI EP05-A3:2015: Pooled and maximum SD less than 5.0 mg/dL for glucose 0.95; bias of individual glucose results within ± 10 mg/dL (glucose 75 mg/dL).
  • Precision Study: pooled and maximum SD

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that overlap and create a sense of movement.

July 29, 2016

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

OK BIOTECH CO., LTD. DR. KE-MIN JEN OFFICIAL CORRESPONDENT NO. 91, SEC. 2, GONGDAO 5TH ROAD HSINCHU CITY 30070 TAIWAN

Re: K160038

Trade/Device Name: UniStrip1 Generic Blood Glucose Test Strips Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JJX Dated: June 3, 2016 Received: June 14, 2016

Dear Dr. KeMin Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

1

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K160038

Device Name

UniStrip1 Generic Blood Glucose Test Strips

Indications for Use (Describe)

The UniStrip1™ Generic Blood Glucose Test Strips are used with the OneTouch® Ultra@2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The strips are meant for self-testing of blood glucose as an aid to monitor the effectiveness of diabetes control.

They are for single patient use only and should not be shared.

They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm. Testing is done outside the body (in vitro diagnostic use).

They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control. Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

The UniStrip1™ Generic Blood Glucose Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch® Ultra® meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly).

The UniStrip™ Control Solutions are for use with the UniStrip1™ Generic Blood Glucose Test Strips and OneTouch® Ultra® meters to check that the meters and test strip are working together properly and that the test is performing correctly.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/0 description: The image shows a logo with the letter K inside of two circles. To the right of the K is the word "biotech" in a bold, sans-serif font. Below the logo is text in a different language, possibly the company's name in another language. The logo appears to be for a company called "Kbiotech."

Taiwan : TEL : 886-3-5160258 FAX : 886-3-5160028 China: TEL:86-791-3899362 FAX:86-791-3880131

http://www.okbiotech.com E-mail:service@okbiotech.com

5. 510(k) Summary of Safety and Effectiveness

(Per 21 CFR 807.92)

Type Of 510(K) SubmissionTraditional
Basis for the submissionAdditional or Expanded Indications
Common Name Of The Proposed
DeviceBLOOD GLUCOSE TEST STRIPS
Trade nameUniStrip1 Generic Blood Glucose Test Strips
510(k) SubmitterOK BIOTECH CO., LTD.
No. 91, Sec. 2, Gongdao 5th Road, 30070,
Hsinchu City, Taiwan
Telephone: +886-3-516-0258
Fax:+886-3-516-0028
Email: service@okbiotech.com
Owner Number9090860
Date preparedJuly 29, 2016
Official CorrespondentDr. JEN, KE-MIN
TEL: 886-3-5208829
FAX: 886-3-5209783
Email: ceirs.jen@msa.hinet.net
Preference For Continued
Confidentiality (21 CFR 807.95)510(k) Summary
Device Classification NameSystem, Test, Blood Glucose, Over The Counter
Regulation DescriptionGlucose test system
Review PanelClinical Chemistry
Product CodeNBW, CGA, JJX
Regulation number21 CFR 862.1345
Class2
Primary Predicate DeviceUniStrip Technologies, LLC
UniStrip1TM Test Strips
K113135
Secondary Predicate Device
(for control solution)DIAGNOSTIC DEVICES, INC
PRODIGY BLOOD GLUCOSE TEST SYSTEM
K060467

4

Image /page/4/Picture/0 description: The image shows a logo with the letter K inside of a circle. Next to the logo is the word "biotech" in a stylized font. Below the word "biotech" is some text in a different language.

http://www.okbiotech.com E-mail:service a okbiotech.com

. Intended Use:

The UniStrip1™ Generic Blood Glucose Test Strips are used with the OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The strips are meant for self-testing of blood glucose as an aid to monitor the effectiveness of diabetes control.

They are for single patient use only and should not be shared.

They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm.

Testing is done outside the body (in vitro diagnostic use).

They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control.

Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

The UniStrip1™ Generic Blood Glucose Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch® Ultra® meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly).

The UniStrip™ Control Solutions are for use with the UniStrip1TM Generic Blood Glucose Test Strips and OneTouch® Ultra® meters to check that the meters and test strip are working together properly and that the test is performing correctly.

● Device Description:

UniStrip1 Generic Blood Glucose Test Strips are used with the OneTouch® Ultra, OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters. The test strips are intended for use outside the body (in vitro diagnostic use) by people with diabetes using the OneTouch® Ultra®, OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® blood glucose monitoring systems as an aid to monitor the effectiveness of diabetes control. UniStrip1 Generic Blood Glucose Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch* Ultra* meter). UniStrip1 Generic Blood Glucose Test Strips can only be used with the OneTouch® Ultra®2, OneTouch® UltraMini® and OneTouch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012. UniStrip1 Generic Blood Glucose Test Strips are only for use with calibration code 49.

5

Image /page/5/Picture/0 description: The image shows a logo with the letter K inside of a circle. To the right of the logo is the word "biotech" in a stylized font. Below the logo and the word is a line of text in a different language. The logo appears to be for a company or organization related to biotechnology.

http://www.okbiotech.com E-mail:service@okbiotech.com

● Test Principle

The OneTouch® Ultra® Family Meters are plasma-calibrated to allow easy comparison of results with laboratory methods. Glucose in the blood sample mixes with special chemicals on the UniStrip1 Generic Blood Glucose Test Strips and a small electrical current is produced. This current is measured by the OneTouch® Ultra® Family Meters and displayed as your blood glucose result. The strength of the current changes with the amount of glucose in the blood sample.

● Comparison Table

ItemPredicate deviceSubject device
ManufacturerUniStrip Technologies, LLCOK Biotech Co., Ltd.
Device trade
nameUniStrip1™ Test StripsUniStrip1 Generic Blood Glucose
Test Strips
510(k) numberK113135K160038
Common nameBlood Glucose Test StripsBlood Glucose Test Strips
Classification
nameSystem, Test, Blood Glucose, Over
The CounterSystem, Test, Blood Glucose, Over
The Counter
Regulation
DescriptionGlucose, Oxidase, GlucoseGlucose, Oxidase, Glucose
Product codeNBW, CGANBW, CGA, JJX
Regulation
number21 CFR 862.134521 CFR 862.1345
Device classIIII
Review panelClinical ChemistryClinical Chemistry
Similarities
Test PrincipleThe OneTouch® Ultra® Family Meters
are plasma-calibrated to allow easy
comparison of results with laboratory
methods. Glucose in the blood sample
mixes with special chemicals on the
UniStrip1 Test Strips and a smallSame test principle
electrical current is produced. This
current is measured by the OneTouch®
Ultra® Family Meters and displayed as
your blood glucose result. The strength
of the current changes with the amount
of glucose in the blood sample.
EnzymeGlucose oxidase (Aspergillus niger)Same enzyme
Specimen TypeCapillary whole blood from fingertip
and alternative sites palm, forearm.
(Use of palm AST is not to be done
with OneTouch® Ultra® meter).Same specimen type
Sample Volume1.0 µLSame sample volume
Operating
Temperature42 - 111 °F (6 - 44 °C)
10 - 90% R.H.Same operating temperature
HCT Range30 - 55%Same HCT range
CalibrationCode 49 onlySame calibration code
Detecting Limit20 - 600 mg/dLSame detecting limit
Measuring time5 secondsSame measuring time
Open vial shelf
life90 days after openingSame open vial shelf life
Chemical
makeup of test
strips1. Glucose oxidase (Aspergillus niger):
20 IU
  1. Potassium ferricyanide: 0.12 mg
  2. Non-reactive ingredients: 1.8 mg | Same chemical makeup of test strips |
    | | 510(k) cleared as K060467 | |
    | Chemical
    makeup of
    control solution | 1. D-Glucose
  3. Polyvinyl acetate (aqueous
    emulsion): 10%
  4. Antifoaming agent (Polyethylene
    Glycol 4000): 0.02%
  5. Disodium EDTA: 0.1%
  6. Fumed silica: 0.2%
  7. Food Pigment Red No.6: 0.05%
  8. Sodium Benzoate: 0.2% | Same chemical makeup of control
    solution as K060467 |
    | Differences: | | |
    | Item | Predicate device | Subject device |
    | Indications for
    use | The UniStrip1 Test Strips are used
    with the OneTouch® Ultra®,
    OneTouch® Ultra®2, OneTouch®
    UltraMini® and One-Touch®
    UltraSmart® meters
    purchased before October 2012, set at
    calibration code 49, for measuring
    glucose (sugar) in whole capillary
    blood. The Unistrip1 TM is meant for
    self-testing of blood glucose as an aid
    to monitor the effectiveness of
    diabetes control.
    They are for single patient use only
    and should not be shared.
    They are used to quantitatively
    measure glucose in fresh capillary
    whole blood samples taken from the
    finger, palm, or forearm.
    Testing is done outside the body (in
    vitro diagnostic use).
    They are indicated for use by people
    with diabetes in their home as an aid
    to monitor the effectiveness of
    diabetes control.
    Not intended for the diagnosis of or
    screening for diabetes mellitus and is
    not intended for use on neonates.
    UniStrip1 Test Strips allow alternate
    site testing (AST) from the fingertip,
    palm and/or the forearm. (Use of palm
    AST is not to be done with
    OneTouch® Ultra® meter). Alternative
    site testing (AST) should only be done
    during steady-state times (when
    glucose is not changing rapidly). | The UniStrip1™ Generic Blood
    Glucose Test Strips are used with the
    OneTouch® Ultra®2, OneTouch®
    UltraMini® and OneTouch®
    UltraSmart® meters purchased before
    April 2016, and OneTouch® Ultra®
    purchased before October 2012, set at
    calibration code 49, for measuring
    glucose (sugar) in whole capillary
    blood. The strips are meant for
    self-testing of blood glucose as an aid
    to monitor the effectiveness of diabetes
    control.
    They are for single patient use only and
    should not be shared.
    They are used to quantitatively measure
    glucose in fresh capillary whole blood
    samples taken from the finger, palm, or
    forearm.
    Testing is done outside the body (in
    vitro diagnostic use).
    They are indicated for use by people
    with diabetes in their home as an aid to
    monitor the effectiveness of diabetes
    control.
    Not intended for the diagnosis of or
    screening for diabetes mellitus and is
    not intended for use on neonates.
    The UniStrip1™ Generic Blood
    Glucose Test Strips allow alternate site
    testing (AST) from the fingertip, palm
    and/or the forearm. (Use of palm AST
    is not to be done with OneTouch®
    Ultra® meter). Alternative site testing
    should only be done during steady-state
    times (when glucose is not changing
    rapidly).
    The UniStrip™ Control Solutions
    are for use with the UniStrip1™
    Generic Blood Glucose Test Strips
    and OneTouch® Ultra® meters to
    check that the meters and test strip
    are working together properly and
    that the test is performing correctly. |
    | Strip Storage
    Temperature | 39 °F – 86 °F
    (4°C – 30 °C) | 39 °F – 104 °F
    (4 °C – 40 °C) |
    | Brand name of
    Control solution | Prodigy Control Solution (K060467) | UniStrip Control Solution
    (same as K060467) |

6

Image /page/6/Picture/0 description: The image shows the logo for "Kbiotech". The logo features a stylized letter "K" enclosed in a circular design, followed by the word "biotech" in a bold, italicized font. Below the word "biotech" are Chinese characters, likely representing the company's name or a related description.

http://www.okbiotech.com

E-mail:service@okbiotech.com

7

Image /page/7/Picture/0 description: The image shows a logo with the letters 'K' and 'biotech' stacked on top of each other. The 'K' is stylized with a circular design around it. Below the word 'biotech' is a line of text in a different language, possibly Chinese. The logo appears to be for a company named 'Kbiotech'.

Taiwan : TEL : 886-3-5160258 FAX : 886-3-5160028 China: TEL:86-791-3899362 FAX:86-791-3880131

http://www.okbiotech.com E-mail:service@okbiotech.com

8

Image /page/8/Picture/0 description: The image contains a logo with the letter K inside of a circular design. Next to the logo is the word "biotech" in bold, italicized letters. Below the word biotech is a line of text in a different language.

Taiwan : TEL : 886-3-5160258 FAX : 886-3-5160028 TEL : 86-791-3899362 FAX : 86-791-388013 China :

http://www.okbiote E-mail: service a okbio

● Safety and Effectiveness Tests and Non-Clinical Studies

The following tests and studies were conducted to ensure the UniStrip1 Generic Blood Glucose Test Strips used with the following blood glucose meters: OneTouch Ultra 2, OneTouch® UltraMini® and One-Touch® UltraSmart® meters purchased before April 2016, and OneTouch® Ultra® purchased before October 2012 were safe and effective in measuring blood glucose concentration.

    1. Sample volume study per ISO 15197:2013 In vitro diagnostic test systems -Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus
      As the test results shown, the glucose measurements of sample volume from 1.0 to 1.5 µL met acceptance criteria. The sample volume of the System was required at least 1.0 µL to obtain the normal testing results.
    1. Operation condition study per EN 23640:2013 In vitro diagnostic medical devices. Evaluation of stability of in vitro diagnostic reagents
      The performance of the UniStrip1 Generic Blood Glucose Test Strips with OneTouch Series Blood Glucose Meters was evaluated in the normal and extreme environments. According to the test results, the individual bias was within ±10 mg/dL at glucose concentration 75 mg/dL compared with Glucose Analyzer YSI 2300. The test results met the acceptance criteria. So the UniStrip1 Generic Blood Glucose Test Strips used with the One Touch Ultra2, One Touch UltraMini and OneTouch UltraSmart purchased before April 2016 pass the linearity study.
MeterTest SiteLinearityR2
One Touch®
Ultra 2®FingertipY=0.9783X + 1.49380.9829
PalmY=0.9718X + 2.36210.9876
ForearmY=0.9945X - 0.8910.9865
One Touch®
UltraMini®FingertipY=0.9759X + 1.6860.9849
PalmY=0.9928X - 0.72680.9848
ForearmY=0.9773X + 1.2990.9856
One Touch®
UltraSmart®FingertipY=0.9806X + 1.33150.9853
PalmY=0.9924X - 0.84470.9849
ForearmY=1.0152X - 2.60140.9884

Linear regression analysis:

    1. Precision study per NCCLS/CLSI EP05-A3:2015 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline- Third Edition (FDA recognition number 7-251)
      According to the test results, the pooled and maximum SD were less than 5.0 mg/dL at glucose concentration