K Number
K160012
Date Cleared
2016-06-02

(150 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The O-pel laser and families of probe delivery devices are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including Arthroscopy, Urology, ENT, Dermatology, Plastic Surgery, General Surgery, Gastroenterology, Thoracic and Pulmonary and Gynecology.

Device Description

The O-pel laser is an infrared laser system with an output wavelength of approximately 2μm. The O-pel laser and families of probe delivery devices (O-pel laser system) have an Intended Use of cutting and coagulating tissue, specifically incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue. The O-pel laser system is comprised of the following components:
-a laser console
-control and display panel;
-a fiber port for delivery systems;
-a footswitch:
-a variety of fiber optic (probe) delivery devices

AI/ML Overview

The provided text is a 510(k) summary for the O-pel Laser and families of probe delivery devices. It primarily focuses on demonstrating substantial equivalence to predicate devices based on intended use, indications for use, design features, and functional features.

Crucially, the document explicitly states: "Substantial equivalence was not based upon clinical performance." This means that a clinical study designed to test the device's performance against pre-defined acceptance criteria was not conducted or presented in this submission. Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets them.

The sections you asked for, such as sample size, data provenance, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment, are all related to clinical performance studies, which were not part of this 510(k) submission.

The document does include information on non-clinical performance data, primarily related to safety and compatibility:

  • Electromagnetic Compatibility: Tested to IEC 60601-1-2: 2007 and found to comply.
  • Electrical Safety: Tested to IEC 60601-1: 2007 and found to comply.
  • Laser Safety: Tested to IEC 60601-2-22 and IEC 60825-1 and found to comply.
  • Coexistence: Tested to FDA Guidance, "Radio Frequency Wireless Technology in Medical Devices" (August 14, 2013) and found to comply.
  • Biocompatibility:
    • Tests performed: ISO 10993-5 (in vitro cytotoxicity) and ANSI/AAMI/ISO 10993-7 (ethylene oxide sterilization residues).
    • Evaluation conducted per FDA draft guidance document, "Use of International Standard ISO 10993. Biological Evaluation of medical devices Part 1: Evaluation and Testing" (April 23, 2013).

However, these are safety and engineering performance tests, not clinical efficacy or effectiveness studies with acceptance criteria in the sense of demonstrating how well the device performs a specific clinical task in a patient population.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three wavy lines that also suggest human profiles.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 2, 2016

Precise Light Surgical Mr. Ken Arnold President And CEO 310 W. Hamilton Ave., Suite 210 Campbell, California 95008

Re: K160012

Trade/Device Name: O-pel Laser And Families Of Probe Delivery Devices Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: Mav 2, 2016 Received: May 4, 2016

Dear Mr. Arnold:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160012

Device Name

O-pel Laser And Families Of Probe Delivery Devices

Indications for Use (Describe)

The O-pel laser and families of probe delivery devices are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including Arthroscopy, Urology, ENT, Dermatology, Plastic Surgery, General Surgery, Gastroenterology, Thoracic and Pulmonary and Gynecology.

ENT

Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including:

  • · Endonasal/sinus Surgery
  • · Partial Turbinectomy
  • · Polypectomy
  • · Dacryocystorhinostomy
  • · Frontal Sinusotomy
  • · Ethmoidectomy
  • Maxillary antrostomy
  • · Functional endoscopic sinus surgery
  • · Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal
  • · Tonsillectomy
  • · Adenoidectomy

Dermatology and Plastic Surgery

Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:

  • · Basal cell carcinomas
  • · Lesions of skin and subcutaneous tissue
  • · Skin tags
  • Plantar warts

General Surgery

Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:

  • · Cholecystectomy
  • · Lysis of adhesion
  • · Appendectomy
  • Biopsy
  • · Skin incision
  • · Tissue dissection
  • · Excision of external tumors and lesions
  • · Complete or partial resection of internal organs, tumors and lesions
  • · Mastectomy
  • · Hepatectomy
  • · Pancreatectomy

{3}------------------------------------------------

  • · Splenectomy
  • · Thyroidectomy
  • · Parathyroidectomy
  • · Herniorrhaphy
  • · Tonsillectomy
  • Lymphadenectomy
  • · Partial Nephrectomy
  • · Pilonidal Cystectomy
  • · Resection of lipoma
  • Debridement of Decubitus Ulcer
  • Hemorrhoids
  • · Debridement of Statis Ulcer
  • · Biopsy
  • Thoracic and Pulmonary

Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue

  • · Laryngeallesions
  • · Airway obstructions including carcinoma
  • Polyps and Granulomas
  • · Palliation of obstructing carcinomas of the tracheobronchial tree

Urology

Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:

  • · Urethral strictures
  • · Bladder neck incisions(BNI)
  • · Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors
  • · Ablation of Benign Prostatic Hypertrophy (BPH)
  • · Transurethral incision of the prostrate (TUIP)
  • · Laser resection of the prostate (HoLRP)
  • · Laser enucleation of the prostate (HoLEP)
  • · Laser ablation of the prostate (HoLAP)
  • · Condylomas
  • · Lesions of external genitalia

Gastroenterology

Open and endoscopic gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:

  • · Appendectomy
  • Polyps
  • Biopsy
  • Gall Bladder calculi
  • Biliary/Bile duct calculi
  • Ulcers
  • · Gastric ulcers
  • · Duodenal ulcers
  • · Non bleeding ulcers
  • · Pancreatitas
  • · Hemorrhoids
  • · Cholecystectomy
  • · Benign and Malignant Neoplasm
  • · Angiodysplasia
  • · Colorectal cancer
  • · Telangiectasias

{4}------------------------------------------------

· Telangiectasias of the Osler-Weber-Renu disease

  • Vascular Malformation
  • Gastritis
  • · Esophagitis
  • · Esophageal ulcers
  • · Varices
  • · Colitis
  • · Mallory-Weiss tear
  • · Gastric erosions

Gynecology

Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis)

  • · Intra-uterine treatment of submucous fibroids, benign endometrial polyps, and uterine septum by incision, excision, ablation, and or vessel coagulation
  • · Soft tissue excision procedures such as excisional conization of the cervix

Arthroscopy

Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)

  • · Ablation of soft and cartilaginous tissue in Minimal Invasive Spinal Surgery including:
  • · Percutaneous Laser Disc Decompression/Discectomy
  • · Foraminoplasty
  • · Ablation and coagulation of soft vascular and non vascular tissue in minimally invasive

spinal surgery

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary

Precise Light Surgical O-pel Laser and families of probe delivery devices

I SUBMITTER

Precise Light Surgical 310 W. Hamilton Ave Suite 210 Campbell. CA 95008 Phone: 844-669-1845 e-mail: karnold@preciselightsurgical.com

Contact person: Ken Arnold, President Date prepared: December 30, 2015

Establishment Registration Number: TBD

II DEVICE

Name of Device: O-pel laser and families of Probe delivery devices

Common or Usual Name: 2 micron Laser System

Classification Name: Powered Laser Surgical Instrument and Accessories (21CFR 878.4810)

Regulatory Code: II

Product Code: GEX

III PREDICATE DEVICE(S)

K051167 Allmed RevoLix and RevoLix Jr K011703 Modified Lumenis Versapulse PowerSuite Holmium (Ho:YAG) and Dual Wavelength (Ho:YAG/Nd:YAG) Surgical Lasers & Delivery Devices with Accessories

IV DEVICE DESCRIPTION

The O-pel laser is an infrared laser system with an output wavelength of approximately 2μm. The O-pel laser and families of probe delivery devices (O-pel laser system) have an Intended Use of cutting and coagulating tissue, specifically incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue. The O-pel laser system is comprised of the following components:

-a laser console -control and display panel; -a fiber port for delivery systems;

{6}------------------------------------------------

-a footswitch:

-a variety of fiber optic (probe) delivery devices

V INDICATIONS FOR USE

The O-pel laser and families of probe delivery devices are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including Arthroscopy, Urology, ENT, Dermatology, Plastic Surgery, General Surgery, Gastroenterology, Thoracic and Pulmonary and Gynecology.

ENT

Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including:

  • . Endonasal/sinus Surgery
  • . Partial Turbinectomy
  • . Polypectomy
  • . Dacryocystorhinostomy
  • . Frontal Sinusotomy
  • . Ethmoidectomy
  • . Maxillary antrostomy
  • . Functional endoscopic sinus surgery
  • . Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal
  • Tonsillectomy
  • . Adenoidectomy

Dermatology and Plastic Surgery

Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:

  • . Basal cell carcinomas
  • . Lesions of skin and subcutaneous tissue
  • . Skin tags
  • . Plantar warts

General Surgerv

Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:

  • Cholecystectomy
  • . Lysis of adhesion
  • . Appendectomy
  • . Biopsy
  • . Skin incision
  • . Tissue dissection
  • . Excision of external tumors and lesions

{7}------------------------------------------------

  • . Complete or partial resection of internal organs, tumors and lesions
  • Mastectomy .
  • . Hepatectomy
  • . Pancreatectomy
  • . Splenectomv
  • . Thyroidectomy
  • . Parathyroidectomy
  • . Herniorrhaphy
  • . Tonsillectomy
  • . Lymphadenectomy
  • . Partial Nephrectomy
  • Pilonidal Cystectomy
  • . Resection of lipoma
  • . Debridement of Decubitus Ulcer
  • . Hemorrhoids
  • . Debridement of Statis Ulcer
  • . Biopsy

Thoracic and Pulmonary

Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection. ablation, vaporization, coagulation and hemostasis) of soft tissue

  • . Laryngeal lesions
  • . Airway obstructions including carcinoma
  • . Polyps and Granulomas
  • . Palliation of obstructing carcinomas of the tracheobronchial tree

Urology

Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:

  • Urethral strictures
  • Bladder neck incisions(BNI)
  • Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral . Tumors
  • . Ablation of Benign Prostatic Hypertrophy (BPH)
  • . Transurethral incision of the prostrate (TUIP)
  • . Laser resection of the prostate (HoLRP)
  • . Laser enucleation of the prostate (HoLEP)
  • Laser ablation of the prostate (HoLAP) .
  • . Condylomas
  • . Lesions of external genitalia

Gastroenterology

Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:

{8}------------------------------------------------

  • Appendectomy .
  • . Polyps
  • . Biopsv
  • . Gall Bladder calculi
  • . Biliary/Bile duct calculi
  • . Ulcers
  • . Gastric ulcers
  • . Duodenal ulcers
  • . Non bleeding ulcers
  • . Pancreatitas
  • . Hemorrhoids
  • . Cholecystectomy
  • . Benign and Malignant Neoplasm
  • . Angiodysplasia
  • . Colorectal cancer
  • . Telangiectasias
  • . Telangiectasias of the Osler-Weber-Renu disease
  • Vascular Malformation
  • . Gastritis
  • . Esophagitis
  • . Esophageal ulcers
  • . Varices
  • . Colitis
  • Mallory-Weiss tear ●
  • . Gastric erosions

Gynecology

Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis)

  • Intra-uterine treatment of submucous fibroids, benign endometrial polyps, and uterine septum by incision, excision, ablation, and or vessel coagulation
  • . Soft tissue excision procedures such as excisional conization of the cervix

Arthroscopy

Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)

  • Ablation of soft and cartilaginous tissue in Minimal Invasive Spinal Surgery ● including:
  • Percutaneous Laser Disc Decompression/Discectomy
  • . Foraminoplasty
  • . Ablation and coagulation of soft vascular and non vascular tissue in minimally invasive spinal surgery

8-4

{9}------------------------------------------------

VI RATIONAL FOR SUBSTANTIAL EQUIVALENCE

The O-pel laser system with fiber optic delivery devices share the same intended use, indications for use, similar design features and functional features and therefore are substantially equivalent to the Allmed Revolix and the Lumenis VersaPulse systems.

PERFORMANCE DATA VII

Electromagnetic Compatibility. The laser console was tested to IEC 60601-1-2: 2007 by a third party and was found to comply.

Electrical Safety: The laser console was tested to IEC 60601-1: 2007 by a third party and was found to comply.

Laser Safety: The laser console was tested to IEC 60601-2-22 and IEC 60825-1: by a third party and was found to comply.

Coexistence: The laser console was tested to the FDA Guidance, "Radio Frequency Wireless Technology in Medical Devices" issued August 14, 2013 and was found to comply.

Biocompatibility. The biocompatibility evaluation of the Probe delivery device was conducted per FDA draft guidance document, "Use of International Standard ISO 10993. "Biological Evaluation of medical devices Part 1: Evaluation and Testing "" dated April 23, 2013. The following tests were performed:

ISO 10993-5: Biological Evaluation of medical devices Part 5- Tests for in vitro cytotoxicity ANSI /AAMI/ISO 10993-7:2008 (R) 2012 Biological evaluation of medical device - Part 7 Ethylene Oxide sterilization residues.

VIII CONCLUSIONS

The O-pel laser and families of Probe delivery devices share indications for use, similar design features, and functional features and therefore are substantially equivalent to the predicate devices. Any differences in technological characteristics do not raise any new issues of safety or efficacy. Substantial equivalence was not based upon clinical performance.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.