AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Powder Free Latex Patient Examination Gloves, with Protein Content labeling Claim (Contains 50 Micrograms per dm- of Glove or Less of Water Extractable Protein), meets all of the requirements of ASTM D3578:05-2010

AI/ML Overview

The provided text describes the acceptance criteria and the performance of an A1 Brand Powder Free Patient Examination Glove. This document is a 510(k) summary for a medical device (patient examination glove), which is often supported by non-clinical performance data rather than extensive clinical studies as seen with AI/software as a medical device (SaMD). Therefore, many of the requested points, particularly those related to AI algorithm performance, ground truth establishment, expert adjudication, and multi-reader multi-case studies, are not applicable to this type of device.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and the Reported Device Performance

CHARACTERISTICSSTANDARDSSPECIFICATION (Acceptance Criteria)DEVICE PERFORMANCE/ACTUAL SPECIFICATION (Reported Performance)
DimensionsASTM D3578:05-2010Thickness Palm: min 0.08mm Finger: min 0.08mmMeet. Palm Thickness: 0.09 mm Finger Thickness: 0.11 mm
Length (by size) S: min 220 M: min 220 L: min 230Length (by size) S: 240 mm M: 240 mm L: 240 mm
Physical PropertiesASTM D3578:05-2010
ASTM D412:06
ASTM D573:04-2010Tensile Strength (MPa) Before aging: Min 18 MPa After aging: Min 14 MPaMeet.
Tensile Strength (MPa) Before aging: 22.91 MPa After Aging: 20.98 MPa
Elongation at Break (%) Before aging: Min 650mm After aging: Min 500mmElongation at Break (%) Before aging: 850 mm After aging: 820 mm
Freedom from pin-holesASTM D5151:06-2011AQL 1.5Meet
Powder Free ResidueASTM D6124:06-2011Below 2mg/gloveMeet
Protein ContentASTM D5712:10-2010Below 50ug/dm²Meet
BiocompatibilityAnimal Irritation Test (ISO 10993-10:2010)Under the conditions tested, the subject glove was not an irritant.Under the conditions tested, the subject glove was not an irritant.
Dermal Sensitization (ISO 10993-10:2010)Under the condition tested, the subject glove was not a sensitizer.Under the condition tested, the subject glove was not a sensitizer.

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (e.g., number of gloves tested for each characteristic) used for the non-clinical tests. It refers to compliance with ASTM and ISO standards, which typically include sampling plans. The data provenance is Malaysia, where A1 Glove Sdn. Bhd. is located. These are non-clinical (laboratory/bench) tests, not retrospective or prospective human subject data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is not an AI/SaMD and does not use "ground truth" derived from expert consensus on medical images or diagnoses. The "ground truth" for glove performance is defined by the objective measurements and chemical analyses specified by the ASTM and ISO standards.

4. Adjudication method for the test set

Not applicable. There is no adjudication method described as this is not a study involving human interpretation or subjective assessment of medical data. The tests are objective measurements against defined standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI device.

7. The type of ground truth used

The "ground truth" for this device's performance is based on established international standards and objective laboratory measurements (e.g., precise measurements of thickness, tensile strength, chemical residue, and biological reactivity according to ASTM and ISO protocols). There is no pathology, outcomes data, or expert consensus used in the typical sense for medical imaging or diagnostic algorithms.

8. The sample size for the training set

Not applicable. This is not an AI device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI device that requires a training set.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.