(46 days)
The Spirit Hair Removal laser Family intended for dermatological use. The devices are specifically indicated for hair removal, permanent hair reduction by using selective laser energy.
The Spirit Hair Removal laser Family is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin. Permanent reduction in hair regrowth is defined as the longterm, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
The Spirit Hair Removal laser Family has two models of the same device (916, & 918) and several trade names. The devices apply photothermal energy to human skin tissue to effect a desired change in the structure of the tissue. The energy is transmitted from a 810nm laser diode to the target tissue by a Treatment Handpiece that is in contact with the skin.
The 918 and 916 models have fully identical HW. The difference in the optical output power (6 -90J/cm² for the 918 model and 4.5 to 10J/cm² for the 916 model) is preset.
For the hair removal treatment, the devices utilized laser energy at the spectrum of 810nm (IR) which is absorbed by the Melanin which is located in the hair follicles. Once the energy is absorbed, it is turned into heat. The heat coagulates the hair follicles and removes the unwanted hair.
The Spirit Hair Removal laser Family was designed to comply with international standards and is constructed from metal, plastic a diode laser and other electronic components.
The provided text is a 510(k) premarket notification for the "Spirit Hair Removal Laser Family" (K153718). It focuses on demonstrating substantial equivalence to a predicate device (Soprano XL K112031) rather than presenting a study to prove a device meets specific acceptance criteria based on its own clinical performance for effectiveness.
Therefore, many of the requested elements for describing a study proving the device meets acceptance criteria cannot be extracted directly from this document. The document primarily describes the justification for substantial equivalence through comparison with a previously cleared device and adherence to safety standards.
Here's a breakdown of the information that can be extracted, and where requested information is not available:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for clinical performance that the device itself must meet (e.g., a specific percentage of hair reduction to be considered effective). Instead, it establishes substantial equivalence by comparing its technological characteristics and intended use to a predicate device.
The tables provided compare features of the Spirit Hair Removal Laser Family to the predicate device, demonstrating they are technologically similar and perform comparably, which serves as the basis for substantial equivalence rather than fulfilling novel performance criteria.
| Subject | Spirit Hair Removal Laser Family (916 Data) | Spirit Hair Removal Laser Family (918 Data) | Predicate Device (Soprano XL-SHR mode) | Predicate Device (Soprano XL-LHR mode) |
|---|---|---|---|---|
| Intended Use | Hair removal, permanent hair reduction | Hair removal, permanent hair reduction | Hair removal, permanent hair reduction | Hair removal, permanent hair reduction |
| Severity and stage of the clinical condition | For all hair densities | For all hair densities | For all hair densities | For all hair densities |
| Site of application | Hairy Skin | Hairy Skin | Hairy Skin | Hairy Skin |
| Patient population | Fitzpatrick I-VI | Fitzpatrick I-VI | Fitzpatrick I-VI | Fitzpatrick I-VI |
| Energy intensity | 4.5-10 J/cm² | 6-90 J/cm² | 5-10 J/cm² | 1-120 J/cm² |
| Spectrum | 810nm | 810nm | 810nm | 810nm |
| Pulse train duration | 11-38ms | Up to 310ms | Up to 20ms | Up to 200ms |
| Pulse repetition rate | ≤10Hz | ≤10Hz | ≤10Hz | ≤3Hz |
| Treatment area size | 12*16mm | 12*16mm | 10*12mm | 10*12mm |
| Conditions of use | Used on Healthy skin | Used on Healthy skin | Used on Healthy skin | Used on Healthy skin |
| Biocompatibility of materials | Complies (ISO10993) | Complies (ISO10993) | Complies | Complies |
| Safety and Performance (IEC 60825-1, IEC 60601-2-22, IEC 60601-1, IEC 60601-1-2) | Complies | Complies | (Implied by prior clearance K112031) | (Implied by prior clearance K112031) |
The acceptance criterion for this submission is "substantial equivalence" to the predicate device. The performance shown is that the Spirit Hair Removal Laser Family's characteristics and safety comply with relevant standards and are comparable to those of the predicate.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document mentions "Bench and clinical data demonstrate that the Spirit Hair Removal laser Family meet the required specifications." However, no sample size or specific details about a clinical test set are provided. The focus is on demonstrating equivalence through technical specifications and safety standards, rather than direct new clinical efficacy data for this specific device. The clinical data referred to might be from the predicate device or a general understanding, but details are absent for the Spirit Hair Removal Laser Family.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. This document does not describe a study involving expert assessment to establish a "ground truth" for a test set.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. No such study involving adjudication is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a laser for hair removal, not an AI-assisted diagnostic device that would involve human readers or AI assistance in interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This refers to an algorithm's performance, which is not relevant for a physical laser device. The "performance" of the laser is its ability to meet documented technical specs and operate safely.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no specific clinical efficacy study is detailed for this device where a ground truth would be established. The document relies on the "ground truth" that the predicate device is safe and effective for its indicated use, and the new device is substantially equivalent. The definition of "Permanent reduction in hair regrowth" as "the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime" could be considered an "outcomes data" ground truth for hair removal efficacy generally, but it's not a parameter measured in a new clinical study for this submission.
8. The sample size for the training set
Not applicable. This device is a physical laser system, not an algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This device is a physical laser system, not an algorithm that requires a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 12, 2016
Active Optical Systems, Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K153718 Trade/Device Name: Spirit Hair Removal Laser Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: January 18, 2016 Received: February 2, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K153718
Device Name
Spirit Hair Removal laser Family.
Trade / Prorietary Device Name: Spirit - 916, Spirit - 918, cFactor- 918, mFactor- 916, mFactor- 918
Indications for Use (Describe)
The Spirit Hair Removal laser Family intended for dermatological use. The devices are specifically indicated for hair removal, permanent hair reduction by using selective laser energy.
The Spirit Hair Removal laser Family is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin. Permanent reduction in hair regrowth is defined as the longterm, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | ☑ |
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 005 - 510(k) Summary of Safety and Effectiveness (In accordance to 21 CFR 807.87(h))
Date of preparation: 18/01/16
Device Name
Device Name: Spirit Hair Removal laser Family. Trade / Proprietary Device Name: Spirit - 916, Spirit - 918 cFactor - 916, cFactor - 918 mFactor - 916, mFactor - 918
Establishment Name and Registration Number of Submitter
Submitter Name: Active Optical Systems Ltd. 510(k) Owner: Active Optical Systems Ltd. Registration Number: 3005180774 Contact person: Gil Bidas Address: 11 Shoham St. Petach Tikva, Israel Tel: +972-3-9236610 Fax: +972-3-9236620
Device Classification
| Product Code: | GEX |
|---|---|
| Regulation Number: | 878.4810 |
| Common Name: | Spirit Hair Removal laser Family |
| Classification Name: | Laser surgical instrument for use in general andplastic surgery and in dermatology. (21 CFR 878.4810) |
| Regulatory class: | Class II |
| Panel | 79 General and Plastic Surgery |
Reason for 510(k) Submission
Traditional 510(k) Submission
Identification of Legally Marketed Equivalent Devices
Modified Diode Laser with SHR Treatment Mode for use with the Family of Soprano XL K112031
Device Description
The Spirit Hair Removal laser Family has two models of the same device (916, & 918) and several trade names. The devices apply photothermal energy to human skin tissue to effect a desired change in the structure of the tissue. The energy is transmitted from a 810nm laser diode to the target tissue by a Treatment Handpiece that is in contact with the skin.
The 918 and 916 models have fully identical HW. The difference in the optical output power (6 -90J/cm² for the 918 model and 4.5 to 10J/cm² for the 916 model) is preset.
{4}------------------------------------------------
For the hair removal treatment, the devices utilized laser energy at the spectrum of 810nm (IR) which is absorbed by the Melanin which is located in the hair follicles. Once the energy is absorbed, it is turned into heat. The heat coagulates the hair follicles and removes the unwanted hair.
The Spirit Hair Removal laser Family was designed to comply with international standards and is constructed from metal, plastic a diode laser and other electronic components.
Indications for use
The Spirit Hair Removal laser Family is generally intended for dermatological use. The devices are specifically indicated for hair removal, permanent hair reduction by using selective laser energy.
The Spirit Hair Removal laser Family is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin.
Permanent reduction in hair regrowth is defined as the longterm, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Rationale for Substantial Equivalence
The Spirit Hair Removal laser Family shares the same indications for use, energy source and technological characteristics with the predicated device mentioned above (K112031) - therefore it is substantially equivalent to the predicated device.
The following table (table1) will summarize the main technological characteristics of the Spirit Hair Removal laser Family and the predicated device:
| No. | Subject | 916 Data | SopranoXL-SHRmode | 918 Data | SopranoXL-LHRmode |
|---|---|---|---|---|---|
| 1 | Manufacturer | Active | Almalasers | Active | Almalasers |
| 2 | intended use | indicatedfor hairremoval,permanenthairreduction | indicatedfor hairremoval,permanenthairreduction | indicatedfor hairremoval,permanenthairreduction | indicatedfor hairremoval,permanenthairreduction |
| 3 | severity and stage of theclinical condition | For allhairdensities | For allhairdensities | For allhairdensities | For allhairdensities |
| 4 | site of application | Hairy Skin | Hairy Skin | Hairy Skin | Hairy Skin |
{5}------------------------------------------------
| No. | Subject | 916 Data | SopranoXL-SHRmode | 918 Data | SopranoXL-LHRmode |
|---|---|---|---|---|---|
| 5 | patient population | FitzpatrickI-VI | FitzpatrickI-VI | FitzpatrickI-VI | FitzpatrickI-VI |
| 6 | energy intensity | 4.5-10 J/ $cm^2$ | 5-10 J/ $cm^2$ | 6-90 J/ $cm^2$ | 1-120 J/ $cm^2$ |
| 7 | Spectrum | 810nm | 810nm | 810nm | 810nm |
| 8 | Pulse train duration | 11-38ms | Up to20ms | Up to310ms | Up to200ms |
| 9 | Pulse repetition rate | ≤10Hz | ≤10Hz | ≤10Hz | ≤3Hz |
| 10 | Treatment area size | 12*16mm | 10*12mm | 12*16mm | 10*12mm |
| 11 | conditions of use | Used onHealthyskin | Used onHealthyskin | Used onHealthyskin | Used onHealthyskin |
| 12 | biocompatibility ofmaterials | Complies | Complies | Complies | Complies |
| 13 | FDA approval No. | K112031 | K112031 |
Table 1 - device substantial main points equivalence comparison
Safety & Effectiveness
The device has been designed, verified and validated complying to 21CFR 820.30 regulations. Bench and clinical data demonstrate that the Spirit Hair Removal laser Family meet the required specifications. No adverse affects have been detected.
High level summary of the tests that were used to demonstrate substantial equivalence
In order to compare the Spirit Hair Removal laser Family with the predicated device various tests were used.
The first tests were the ones that compared the intended usage and patient population.
Once it was clear that both devices are intended for the same usage and same patient population it was checked if the method of operation is equivalent.
In order to verify the data regarding the Spirit Hair Removal laser Family's method of operation, the spirit was checked under IEC 60825-1 and IEC 60601-2-22. These tests were performed by SII. Their results are summarized in the original submission Sec. 18 and show us that both the Spirit Hair Removal laser Family and the predicated device share an equivalent method of operation in terms of wavelength used, energy used, accuracy and other technological characteristics.
Once it was established that both devices share the same intended usage and method of operation a safety comparison was made by applying ISO14971 requirements on the
{6}------------------------------------------------
Spirit Hair Removal laser Family. In addition the applied part was tested under ISO10993 for biocompatibility and proved to be suitable.
After these tests, the Spirit Hair Removal laser Family underwent the testing of the safety and EMC properties so it could be compared with the predicated device.
In order to verify its safety and EMC properties, the Spirit Hair Removal laser Family was tested by SII according to IEC 60601-1 and IEC 60601-1-2. The results showed us that the Spirit Hair Removal laser Family is a safe and electrically compatible device just like the predicated device.
The combined result of all the tests led us to the opinion that the Spirit Hair Removal laser Family is substantially equivalent to the predicated device.
| No. | Test | Guidance Document |
|---|---|---|
| 1. | Risk Analysis | ISO 14971:2007, 2012 |
| 2. | Biocompatibility | ISO 10993-1:2003 |
| 3. | Safety | IEC60601-1:2007 |
| 4. | EMC | IEC 60601-1-2:2007 |
| 5. | Safety & performance | IEC 60601-2-22:2007, IEC 60825-1:2007 |
| 6. | Substantial Equivalence | The 510(k) Program: Evaluating SubstantialEquivalence in Premarket Notifications[510(k)]” Dated: July 28, 2014 |
Guidance documents referenced\ used for the testing
Table 2 - Tests and guidance documents
FDA recognized consensus standards used upon testing
| No. | recognized consensus standard | Connotation |
|---|---|---|
| 19-4 | IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007): Part 1 - Medical Electrical Equipment Part1:General requirements for basic safety and essentialperformance US National standard ANSIIAAMIES60601-1: 2005 / A2:2010 | Safety testcomparison |
| 037 | IEC 60601-1-2: Third Edition (2007): Medicalelectrical equipment Part 1-2: Collateral Standard:Electromagnetic compatibility-Requirements and tests. | EMC testcomparison |
| 12-273 | 60825-1: 2007 (2nd Edition): Safety of laser productsPart 1: Equipment classification and requirements | Safety &performancetestcomparison |
| 12-208 | IEC 60601-2-22: 2007 (Third Edition): Medicalelectrical equipment, Part 2: Particular requirements forthe safety of diagnostic and therapeutic laser equipment | Safety &performancetestcomparison |
Table 3 - recognized consensus standards used
{7}------------------------------------------------
Conclusion - Substantial Equivalency
Since the predicate device and the Spirit Hair Removal laser Family share the same indications for use, substantially equivalent technological characteristics, and based on the bench tests and the assessment of clinical data,it is Active Optical Systems'opinion that the Spirit Hair Removal laser Family is substantially equivalent in terms of safety and effectiveness to the predicate device.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.