(190 days)
The Liofilchem MIC Test Strip (MTS) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of non-fastidious Gram negative and Gram positive aerobic bacteria (for example, Enterobacteriaceae, Pseudomonas, Enterococcus and Staphylococus species) and fastidious bacteria (for example, anaerobes, Haemophilus and Streptococcus species and N. gonorthoeae). MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in up/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures.
The purpose of the 510(k) is specifically for the dalbavancin MTS at concentrations of 0.002-32 ug/mL interpreted after 16-20 hours incubation.
Dalbavancin has been shown to be active against the following bacteria, both clinically and in vitro according to the FDA label:
Staphylococcus aureus (including methicillin resistant strains) Enterococcus faecalis (vancomycin susceptible strains)
MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in up/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures.
The provided text describes the 510(k) premarket notification for the Liofilchem MIC Test Strip (MTS) - Dalbavancin at 0.002 - 32 µg/ml. While it outlines the device's indications for use and FDA's determination of substantial equivalence, it does not contain the detailed acceptance criteria and study data requested in the prompt.
Specifically, the document does not provide:
- A table of acceptance criteria and reported device performance.
- Sample sizes or data provenance for a test set.
- Information on experts used to establish ground truth or their qualifications.
- Details on an adjudication method.
- Results of a multi-reader multi-case (MRMC) comparative effectiveness study, including effect size.
- Information about a standalone algorithm performance study.
- The type of ground truth used (e.g., pathology, outcomes data).
- The sample size for the training set.
- How ground truth for the training set was established.
The document is an FDA clearance letter, which confirms that the device has been found substantially equivalent to a predicate device. It briefly mentions that Dalbavancin has been shown to be active against certain bacteria "both clinically and in vitro according to the FDA label," but this statement refers to the drug's efficacy, not the performance of the MIC Test Strip device itself or details of a study assessing its accuracy.
Therefore,Based on the provided text, I cannot describe the acceptance criteria and the study that proves the device meets those criteria because the necessary information is not present.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2016
LIOFILCHEM s.r.l. C/O LAURA KOETH LABORATORY SPECIALISTS, INC. 1651-A CROSSINGS PARKWAY WESTLAKE OH 44145
Re: K153569
Trade/Device Name: Liofilchem MIC Test Strip (MTS) - Dalbavancin at 0.002 - 32 µg/ml Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: JWY Dated: April 15, 2016 Received: April 20, 2016
Dear Ms. Koeth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ribhi Shawar -S
For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K153569
Device Name
Liofilchem MIC Test Strip (MTS) Dalbavancin 0.002-32 ug/mL
Indications for Use (Describe)
The Liofilchem MIC Test Strip (MTS) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of non-fastidious Gram negative and Gram positive aerobic bacteria (for example, Enterobacteriaceae, Pseudomonas, Enterococcus and Staphylococus species) and fastidious bacteria (for example, anaerobes, Haemophilus and Streptococcus species and N. gonorthoeae). MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in up/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures.
The purpose of the 510(k) is specifically for the dalbavancin MTS at concentrations of 0.002-32 ug/mL interpreted after 16-20 hours incubation.
Dalbavancin has been shown to be active against the following bacteria, both clinically and in vitro according to the FDA label:
Staphylococcus aureus (including methicillin resistant strains) Enterococcus faecalis (vancomycin susceptible strains)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
§ 866.1640 Antimicrobial susceptibility test powder.
(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).