K Number
K153415

Validate with FDA (Live)

Date Cleared
2016-01-13

(49 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reprocessed Stryker External Fixation Devices are indicated for use in patients requiring external fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

Device Description

External fixation devices are specially designed frames, clamps, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, specifically Reprocessed Stryker External Fixation Devices. It does not describe an AI/ML powered device, but rather a reprocessed conventional medical device. Therefore, many of the requested categories related to AI/ML device studies will not be applicable.

Here's an analysis of the provided text based on the request:

Device Name: Reprocessed Stryker External Fixation Devices

Device Description: External fixation devices are specially designed frames, clamps, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.


Acceptance Criteria and Study to Prove Device Meets Criteria:

1. Table of acceptance criteria and the reported device performance:

Acceptance CriteriaReported Device Performance
Device functionalityDemonstrated to perform as originally intended.
Post-reprocessing inspection of componentsMet pre-determined acceptance criteria.
SafetyConcluded to be safe.
EffectivenessConcluded to be effective.
Substantial Equivalence to predicate devicesConcluded to be substantially equivalent.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

  • Sample Size for Test Set: Not specified. The document states "Verification analyses consisting of device functionality testing and post-reprocessing inspection of components were performed to evaluate the effects of the reprocessing procedures on the subject device." It does not detail the number of devices tested.
  • Data Provenance: Not specified, but implied to be from testing conducted by Stryker Sustainability Solutions. The study is a prospective verification and validation study of the reprocessed devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This device is a reprocessed physical medical device, not an AI/ML algorithm requiring expert ground truth for classification or diagnosis. The "ground truth" here is the original performance specifications of the predicate devices and the physical integrity of the reprocessed components.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. This is not a study requiring human adjudication for diagnostic or classification tasks. The acceptance criteria were based on functionality and inspection.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI-powered device.

6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done:

  • The performance testing described ("device functionality testing and post-reprocessing inspection of components") is inherently "standalone" in the sense that it evaluates the device itself against its specifications, without human intervention in its function. However, this is not an algorithm's standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" implicitly used for the Reprocessed Stryker External Fixation Devices is the original performance specifications and safety profiles of the new, predicate versions of the external fixation devices. The goal of the reprocessing verification was to demonstrate that the reprocessed devices perform as originally intended and meet the same standards as the new devices.

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI/ML device that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other, representing the human element of the department's work.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 13, 2016

Stryker Sustainability Solutions Ms. Kelli Drver Senior Regulatory Affairs Specialist 1810 West Drake Drive Tempe, Arizona 85283

Re: K153415 Trade/Device Name: Reprocessed Stryker External Fixation Devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT, JEC, KTW Dated: December 9, 2015 Received: December 14, 2015

Dear Ms. Dryer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153415

Device Name

Reprocessed Stryker External Fixation Device

Indications for Use (Describe)

Reprocessed Stryker External Fixation Devices are indicated for use in patients requiring external fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

Type of Use (Select one or both, as applicable)
✘ Prescription Use (Part 21 CFR 801 Subpart D)
❏ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 4: 510(k) Summary

SUBMITTER:

Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, Arizona 85283

CONTACT:

Kelli Dryer Senior Regulatory Affairs Specialist 480-763-5324 (direct) 480-763-5310 (fax)

November 24, 2015 Date of Preparation:

NAME OF DEVICE:

Trade/Proprietary Name:Reprocessed Stryker External Fixation Devices
Common Name:External Fixation Devices, Fixation Appliance, Single/MultipleComponent
Classification Name:Single/Multiple Component Metallic Bone Fixation Appliances andAccessories and Smooth or Threaded Metallic Bone Fixation Fastener

DEVICE CLASS

Class:II
Review Panel:Orthopedic
Product Code:KTT, JEC, KTW
Regulation Number Citation:888.3030, 888.3040, 888.3030

Table 1. Device Class Information

{4}------------------------------------------------

PREDICATE DEVICES:

510(k) Number510(k) TitleOriginal Manufacturer
K052062 -PrimaryModification to Reprocessed HowmedicaExternal Fixation Devices [Non-Sterile]Alliance Medical Corporation
K012648 -SecondaryReprocessed Howmedica External FixationDevices [Sterile]Alliance Medical Corporation

Table 2. Predicate Devices

DEVICE DESCRIPTION:

External fixation devices are specially designed frames, clamps, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

INTENDED USE:

The Reprocessed Stryker External Fixation Devices intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

INDICATIONS STATEMENT:

Reprocessed Stryker External Fixation Devices are indicated for use in patients requiring external fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.

{5}------------------------------------------------

SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:

The design, materials, and intended use of the Reprocessed Stryker External Fixation Devices are the same as the predicate devices. The mechanism of the Reprocessed Stryker External Fixation Device is identical to the predicate devices in the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specification, or method of operation. The line extension is to add new model numbers to the predicate device line. Although the predicate devices will continue to be marketed, the Reprocessed Stryker External Fixation Devices will offer additional components and include minor design and material modifications.

PERFORMANCE TESTING:

Verification analyses consisting of device functionality testing and post-reprocessing inspection of components were performed to evaluate the effects of the reprocessing procedures on the subject device. Performance testing demonstrates that the Reprocessed Stryker External Fixation Devices perform as originally intended.

CONCLUSION:

The information summarized in the Design Control Activities Summary demonstrates that the Reprocessed Stryker External Fixator Device met the pre-determined acceptance criteria for the verification activities identified.

There is no change to the fundamental scientific technology, indicated use or material types or the proposed line extension devices.

Stryker Sustainability Solutions concludes that the line extension device (the Reprocessed Stryker External Fixation Device) is safe, effective and substantially equivalent to the predicate devices, as described herein.

N/A