(85 days)
The Caspian OCT/MESA Mini/ DENALI Mini Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (Cl to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudoarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Caspian OCT/MESA Mini/DENALI Mini Spinal System is also intended to restore the integrity of the spinal column eyen in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Caspian OCT/MESA Mini/DENALI Mini Spinal System may be connected to Range/MESA/DENALI Spinal System and Everest Spinal System components via the rod to rod connectors or transition rods.
The Caspian OCT/MESA Mini/ DENALI Mini Spinal System is a top-loading, multiple component, posterior (occipital-cervical-thoracic) spinal fixation system. The purpose of the current submission is to add modified plates and rods and to revise the Indications for Use.
The provided text is an FDA 510(k) Premarket Notification for a spinal system, not a study evaluating an AI/CADe medical device. Therefore, it does not contain the information required to answer your prompt regarding acceptance criteria and a study that proves the device meets those criteria, specifically concerning AI/CADe performance, ground truth, sample sizes, or expert adjudication.
The document describes a spinal fixation system, its indications for use, and a comparison to predicate devices, focusing on mechanical testing and material equivalence rather than diagnostic performance metrics.
To answer your prompt, I would need a different type of document, such as a clinical study report or a 510(k) submission for an AI/CADe device that details its performance evaluation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol depicts a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 16, 2016
K2M, Incorporated Ms. Nancv Giezen Manager, Regulatory Affairs 751 Miller Drive Southeast Leesburg, Virginia 20175
Re: K153370
Trade/Device Name: Caspian OCT/MESA Mini/DENALI Mini Spinal System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: November 20, 2015 Received: November 23, 2015
Dear Ms. Giezen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K153370
Device Name
Caspain OCT/MESA Mini/DENALI Mini Spinal System
Indications for Use (Describe)
The Caspian OCT/MESA Mini Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fe.g., pseudoarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Casplan OCT/MESA Mini/ DENALI Mini Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Caspian OCT/ MESA Mini/DENALI Mini Spinal System may be connected to Range/MESA/DENAL1 Spinal System and Everest Spinal System components via the rod to rod connectors or transition rods.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
STATUAL FOR FOR FOR FOR USE ONLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
510(k) SUMMARY Caspian OCT/MESA Mini/DENALI Mini Spinal System
Submitter
K2M, Inc. 751 Miller Drive SE Leesburg, VA 20175
Contact Person: Nancy Giezen Telephone: 703-777-3155 Date Prepared: 2/12/2016
Classification
Trade Name: Caspian OCT/MESA Mini/ DENALI Mini Spinal System Common Name: Spinal Fixation System Regulatory Class: Unclassified (NKG), Class II (K WP)
Classification Name(s):
Orthosis, cervical pedicle screw spinal fixation (Unclassified, Product Code NKG) Appliance, fixation, spinal interlaminal (21 CFR 888.3050, Product Code KWP)
Predicate Device(s)
Primary Predicate: Nuvasive VuePoint II OCT System (K150474) Additional Predicates: K2M Caspian OCT/MESA Mini/DENALI Mini Spinal System (K081107) K2M Caspian OCT/MESA Mini/DENALI Mini Spinal System (K133288) Medtronic Vertex Reconstruction System (K152338) Stryker Oasys System (K151755) DePuy Summit SI OCT Spinal Fixaiton System, Mountaineer OCT Spinal System (K151885) Globus Ellipse and Protex CT Occipitio-Cervico-Thoracic Spinal Systems (K150552)
Device Description
The Caspian OCT/MESA Mini/ DENALI Mini Spinal System is a top-loading, multiple component, posterior (occipital-cervical-thoracic) spinal fixation system. The purpose of the current submission is to add modified plates and rods and to revise the Indications for Use.
Function: The system functions as an adjunct to fusion to provide immobilization of stabilization of the posterior cervical and thoracic spine.
Indications for Use
The Caspian OCT/MESA Mini/DENALI Mini Spinal System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (Cl to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudoarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.
{4}------------------------------------------------
The Caspian OCT/MESA Mini/DENALI Mini Spinal System is also intended to restore the integrity of the spinal column eyen in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the Caspian OCT/MESA Mini/DENALI Mini Spinal System may be connected to Range/MESA/DENALI Spinal System and Everest Spinal System components via the rod to rod connectors or transition rods.
Technological Comparison to Predicate(s)
The proposed modifications were compared to predicate systems and the design features, materials and sizes were found to be substantially the same as these systems.
Non-clinical Performance Evaluation
Published literature and mechanical testing per ASTM F1717 and F2706 (static torsion, static compression, dynamic torsion, dynamic compression), and biomechanical (cadaver) testing demonstrates the Caspian OCT/MESA Mini/DENALI Mini Spinal System is substantially equivalent to the predicate devices.
Conclusion
There are no significant differences between the Caspian OCT/MESA Mini/ DENALI Mini Spinal System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.
§ 888.3075 Posterior cervical screw system.
(a)
Identification. Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (
e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.(b)
Classification. Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
e.g., CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.