K Number
K153234
Manufacturer
Date Cleared
2016-07-07

(241 days)

Product Code
Regulation Number
868.5270
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AirLife Adult Heated Wire Circuit is intended to warm breathing gases before they enter a patient's airway. The AirLife Adult Heated Wire Circuit is used with the adult patient population that requires mechanical ventilation, positive pressure breathing or general medical gases. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Adult Heated Wire Circuit is compatible to the Fisher & Paykel MR850 humidifier.

Device Description

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. It is provided non sterile for single patient use. The heated breathing circuit is used with Fisher and Paykel heated humidifier.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called the "AirLife Adult Heated Wire Circuit." This type of document is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a new device's performance against clinical acceptance criteria in the same way a novel device might.

The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and performance tests against recognized standards for breathing circuits and humidifiers, not clinical performance measures like sensitivity, specificity, or reader agreement for an AI diagnostic device.

Here's the breakdown of the information based on the document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance statistics in a pass/fail format typical for a clinical study of an AI device. Instead, it lists the performance characteristics tested and the standards they comply with, implying that meeting these standards constitutes the acceptance criteria. The "reported device performance" is implicitly that the device did meet these standards, as stated in the conclusion.

Performance CharacteristicStandard (Implicit Acceptance Criteria)Reported Device Performance
LengthBS EN ISO 5367: 2014Compliant
Resistance to FlowBS EN ISO 5367: 2014Compliant
Resistance to Flow with BendingBS EN ISO 5367: 2014Compliant
Conical ConnectorsBS EN ISO 5367: 2014Compliant
Conical ConnectorsBS EN ISO 8185: 2009Compliant
Security of AttachmentISO 5356-1Compliant
Security of AttachmentBS EN ISO 5367: 2014Compliant
LeakageBS EN ISO 5367: 2014Compliant
ComplianceBS EN ISO 5367: 2014Compliant
Resistance to MeltBS EN ISO 8185: 2009Compliant
Security of Engagement Temperature SensorBS EN ISO 8185: 2009Compliant
Leakage from Sensing PortBS EN ISO 8185: 2009Compliant
Specific EnthalpyBS EN ISO 8185: 2009Compliant
Surface TemperatureBS EN ISO 8185: 2009Compliant
Humidity Output InvasiveBS EN ISO 8185: 2009Compliant
Humidity Output Non-InvasiveBS EN ISO 8185: 2009Compliant
Electrical SafetyIEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + A1:2012Compliant
Electromagnetic Compatibility60601-1-2 Edition 3: 2007-03Compliant
Biocompatibility (Cytotoxicity, Sensitization, Irritation, Muscle Implantation, Genotoxicity and Extractables/Leachables)ISO 10993-1, -3, -5, -6, -10, -17, -18Compliant

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This document does not specify a "test set" in the context of clinical data for AI benchmarking. The tests performed are engineering and performance evaluations of the device itself (e.g., length, resistance, leakage). The document does not provide details on the sample size (number of devices tested), the provenance of the data (country of origin), or whether the testing was retrospective or prospective. These are typically internal validation tests by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This device is not an AI diagnostic tool that requires expert human interpretation to establish ground truth for a test set. The "ground truth" here is defined by the technical specifications outlined in the referenced ISO and IEC standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. There is no human adjudication process involved as this is a technical device performance evaluation against standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device, and therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device (heated wire circuit), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is established by the technical specifications and test methodologies outlined in the referenced international standards (e.g., ISO 5367, ISO 8185, IEC 60601-1, ISO 10993 series). The device's performance is measured against these quantitative and qualitative criteria defined by the standards.

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 7, 2016

CareFusion, Inc. Erika Fernandez Regulatory Affairs Manager 75 N Fairway Drive Vernon Hills, Illinois 60061

Re: K153234

Trade/Device Name: AirLife Adult Heated Wire Circuit Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing System Heater Regulatory Class: Class II Product Code: BZE Dated: May 27, 2016 Received: May 31, 2016

Dear Erika Fernandez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved OMB No 0910-0120 Expiration Date January 31, 2017 See PRA Statement below

510(k) Number (if known)

K153234

Device Name AirLife Adult Heated Wire Circuit

Indications for Use (Describe)

The AirLife Adult Heated Wire Circuit is intended to warm breathing gases before they enter a patient's airway. The AirLife Adult Heated Wire Circuit is used with the adult patient population that requires mechanical ventilation, positive pressure breathing or general medical gases. The product is single use device, non-sterile and used in professional healthcare environments and intra-hosport environments under a doctor's supervision and by skilled clinicians. The AirLife Adult Heated Wire Circuit is compatible to the Fisher & Paykel MR850 humidifier.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda hhs gov

"An agency may not conduct or sponsor, and a person is not requred to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in black. Below the text, there is a smaller text that says "has joined BD".

Summary of Safety and Effectiveness 510k summary complying with 21 CFR 807.92.

1. SUBMITTER

CareFusion 2200, Inc. 75 N Fairway Drive, Vernon Hill, IL 60061 847-362-8097 Phone: Fax: 312 949-0731

Contact Person: Erika Fernandez

Date Prepared: July 6, 2016

2. Device

Product Name:AirLife Adult Heated Wire Circuit
Device Name:Heated Breathing Circuit
Common Name:Heater, breathing system w/wo controller (not humidifier ornebulizer)
Classification Name:Breathing system heater (21 CFR 868.5270)
Regulatory Class:II
Product Code:BZE

3. Predicate Device

This submission demonstrates substantial equivalence to the AirLife Heated Ventilator and Anesthesia Breathing Circuits, K000697 was cleared on March 30, 2000. This predicate device has not been subject to a design-related recall.

4. Device Description

The heated breathing circuit is intended to deliver and warm breathing gases before they enter the patient's airway. It is provided non sterile for single patient use. The heated breathing circuit is used with Fisher and Paykel heated humidifier.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CareFusion. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, in smaller gray font, it says "has joined BD".

5. Indication for use

The AirLife Adult Heated Wire Circuit is intended to warm breathing gases before they enter a patient's airway. The AirLife Adult Heated Wire Circuit is used with the adult patient population that requires mechanical ventilation, positive pressure breathing or general medical gases. The product is single use device, non-sterile and used in professional healthcare environments and intra-hospital transport environments under a doctor's supervision and by skilled clinicians. The AirLife Adult Heated Wire Circuit is compatible to the Fisher & Paykel MR850 humidifier.

Comparison of technological characteristics with the predicate device 6.

The fundamental scientific technology is the same for both proposed and predicate device. It is based on acting as an airway conduit between a breathing machine and the patient (typically attached to an endotracheal or tracheal tube previously insert into the patient's airway). The AirLife Adult Heated Wire Circuit is substantially equivalent to the predicate device AirLife Heated Wire Breathing Circuits regarding safety, effectiveness, design (technology), materials and intended use. The proposed AirLife Adult Heated Wire Circuit is designed to operate at the same minimum flow rate of 3 LPM as the predicate device. Successful test results (specific enthalpy, humidity output and surface temperature) ensured that the flow rate of the proposed device does not raise any different question of safety and effectiveness

Element of comparisonProposed DevicePredicate Device
Intended UseIntended to warm breathinggases before they enter apatient's airwayIntended to warm breathinggases before they enter apatient's airway
Principal of OperationResistance wires within thetubing generate heat tomaintain temperatures andhumidityResistance wires within thetubing generate heat tomaintain temperatures andhumidity
Circuit Characteristics
Limb Length (Inspiratoryand Expiratory)5 ft5 ft
Dryline Length1.7 ft1.7 ft
Intended Patient UseAdultAdult
UsageDisposableDisposable
DesignDual and Single LimbDual and Single Limb
Tube Specifications
Nominal Inside Diameter21 mm21 mm
DesignCorrugated Breathing Tube with22mm AdaptersCorrugated Breathing Tubewith 22mm Adapters

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text. Below the word "CareFusion" is the phrase "has joined BD" in a smaller, lighter gray font.

Element of comparisonProposed DevicePredicate Device
Circuit Specifications
Circuit DesignDual or SingleDual or Single
Maximum Power60W60W
Min circuit resistance perLimb16.15 Ω16.24 - 17.75 Ω
Maximum Power/foot ofLimb6.06.0
ConductorCopper/Ni alloyCopper/Ni alloy
Compatible HumidifiersFisher and Paykel MR850Fisher and Paykel MR850

7. Performance Data

The proposed device was tested to ensure compliance to the following standards:

Biocompatibility

Tests for an externally communicating, tissue by way of gas path and direct mucosal contact with prolonged contact (greater than 24 hours but less than 30 days): Cytotoxicity, Sensitization, Irritation, Muscle Implantation, Genotoxicity and Extractables/Leachables

Standards

Performance CharacteristicStandard
Biological Evaluation of Medical Devices Part 1: Evaluationand TestingFDA Guidance: Use of International Standard ISO- 10993,"Biological Evaluation of Medical Devices Part 1: Evaluationand Testing"ISO 10993-1:2009
Biological Evaluation of Medical Devices Part 3: Tests forgenotoxicity, carcinogenicity and reproductive toxicityISO 10993-3:2014
Biological Evaluation of Medical Devices Part 5: Tests for invitro cytotoxicityISO 10993-5:2009
Biological Evaluation of Medical Devices Part 6: Tests forlocal effects after implantationISO 10993-6:2007
Biological Evaluation of Medical Devices Part 10: Tests forirritation and skin sensitizationISO 10993-10:2010
Biological Evaluation of Medical Devices Part 17:Establishment of allowable limits for leachable substancesISO 10993-17:2002

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the word "CareFusion" in gray text. Below the word "CareFusion" is the phrase "has joined BD" in a smaller font.

Performance CharacteristicStandard
Biological Evaluation of Medical Devices Part 18: Chemicalcharacterization of materialsISO 10993-18:2005

Performance

The following tests were performed for the proposed device to support the substantial equivalence decision.

Performance CharacteristicStandard
LengthBS EN ISO 5367: 2014
Resistance to FlowBS EN ISO 5367: 2014
Resistance to Flow with BendingBS EN ISO 5367: 2014
Conical ConnectorsBS EN ISO 5367: 2014
Conical ConnectorsBS EN ISO 8185: 2009
Security of AttachmentISO 5356-1
Security of AttachmentBS EN ISO 5367: 2014
LeakageBS EN ISO 5367: 2014
ComplianceBS EN ISO 5367: 2014
Resistance to MeltBS EN ISO 8185: 2009
Security of Engagement Temperature SensorBS EN ISO 8185: 2009
Leakage from Sensing PortBS EN ISO 8185: 2009
Specific EnthalpyBS EN ISO 8185: 2009
Surface TemperatureBS EN ISO 8185: 2009

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. Below the text, there is a smaller text that says "has joined BD".

Performance CharacteristicStandard
Humidity Output InvasiveBS EN ISO 8185: 2009
Humidity Output Non-InvasiveBS EN ISO 8185: 2009
Electrical SafetyIEC 60601-1:2005 (Third Edition) + CORR.1:2006 + CORR. 2:2007 + A1:2012 (or IEC60601-1: 2012 reprint)
Electromagnetic Compatibility60601-1-2 Edition 3: 2007-03

8. Conclusion

The test results demonstrate that the device is as safe and effective as the predicate and therefore substantially equivalent to the predicate device.

§ 868.5270 Breathing system heater.

(a)
Identification. A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.(b)
Classification. Class II (performance standards).