K Number
K153205
Date Cleared
2016-08-01

(271 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IDSS-ForceTriad Control Module is designed for integration in the IDSS(Integrated Digital Surgical Suite) and enables the Covidien ForceTriad Energy Platform to be controlled remotely.

Device Description

The IDSS ForceTriad Control Module is an additional function to the IDSS, which is an integrated operating room system controlling video displays, observation cameras, audio video equipment, teleconferencing and the routing of video and images from multiple sources to multiple destinations via a touch screen interface. With the IDSS ForceTriad Control Module, operation room staff is able to control the ForceTriad setup from the touch panel location rather than using the ForceTriad unit itself.

AI/ML Overview

The provided FDA 510(k) summary (K153205) describes the IDSS-ForceTriad Control Module. This device is an accessory that enables remote control of the Covidien ForceTriad Energy Platform within an Integrated Digital Surgical Suite (IDSS). The primary purpose of this submission is to demonstrate substantial equivalence to a predicate device, not to showcase the performance of an AI/ML algorithm that interprets medical images or data.

Therefore, many of the requested criteria regarding AI/ML performance studies, such as effect size of human readers with AI assistance, standalone algorithm performance, number of experts for ground truth, and training set information, are not applicable to this document.

Here's an analysis of the provided information, focusing on the context of a medical device accessory and its validation:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied)Reported Device Performance
Software Verification and Validation of interfaces, features, and non-functional reliability.Software Verification and Validation testing were performed on interfaces, feature functional requirements, and non-functional reliability.
Risk analysis in accordance with established in-house acceptance criteria based on ISO 14971:2007.The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007.
Device meets the needs of users and does not raise new safety and efficacy issues compared to the predicate device.Testing and evaluation indicate that the system meets the needs of the users of the device and does not raise any new safety and efficacy of the predicate device. The IDSS –ForceTriad Control Module is substantially equivalent to the predicate device since intended use, operational principle, basic technology and design are similar.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This device is a control module, not a diagnostic or AI image analysis system that would typically use a "test set" of medical images or patient data. The testing is focused on software functionality and risk management.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. As noted above, this is a control module, not an AI/ML system requiring expert-adjudicated ground truth for medical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not an AI/ML system requiring adjudication of medical cases.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This filing concerns a control module for an energy platform, not an AI-assisted diagnostic or imaging device. There is no mention of human readers or AI assistance in the context of clinical interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI algorithm. Its function is to remotely control an existing medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the software validation and risk analysis, the "ground truth" would be related to the correct and expected functionality of the software according to its design specifications, user requirements, and safety standards (e.g., ISO 14971:2007). This is established through internal design documents, functional specifications, and compliance with regulatory standards, rather than clinical outcomes or expert consensus on medical findings.

8. The sample size for the training set

Not applicable. This is not an AI device that involves a "training set" in the machine learning sense.

9. How the ground truth for the training set was established

Not applicable. This is not an AI device.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, arranged in a cascading manner. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the top and left side of the profile graphic.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 1, 2016

Black Diamond Video, Inc. Mr. Ragunath Muniandy Regulatory Affairs Specialist 503 Canal Blvd Richmond, California 94804

Re: K153205

Trade/Device Name: IDSS ForceTriad Control Module Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: ODA Dated: June 27, 2016 Received: June 30, 2016

Dear Mr. Muniandy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Christopher J. Ronk -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153205

Device Name IDSS –ForceTriad Control Module

Indications for Use (Describe)

The IDSS-ForceTriad Control Module is designed for integration in the IDSS(Integrated Digital Surgical Suite) and enables the Covidien ForceTriad Energy Platform to be controlled remotely.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Black Diamond Video, which includes a stylized diamond shape on the left and the letters "BDV" in bold, black font on the right. The diamond shape is made up of smaller diamonds, with a blue diamond in the center. Below the letters "BDV" is the text "BLACK DIAMOND VIDEO" in a smaller, light blue font. The logo has a modern and professional look.

510(K) Summary

Date Prepared: 10/15/2015

510K Owner:Black Diamond Video503 Canal BlvdRichmond, CA94804
Contact:Ragunath Muniandy, RACRegulatory Affairs Specialist
Trade Name:IDSS -ForceTriad Control Module
Common Name:Endoscope and accessories
Classification Name:Endoscopic central control unit
Classification Panel:Gastroenterelogy/Urology
CFR section:21 CFR 876.1500
Class:2
Product Code:ODA
Predicate Device :Karl Storz SCB/Covidien ForceTriad Interface Module(K111165)
Device Description:The IDSS ForceTriad Control Module is an additional functionto the IDSS, which is an integrated operating room systemcontrolling video displays, observation cameras, audio videoequipment, teleconferencing and the routing of video andimages from multiple sources to multiple destinations via atouch screen interface. With the IDSS ForceTriad ControlModule, operation room staff is able to control the ForceTriadsetup from the touch panel location rather than using theForceTriad unit itself.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Black Diamond Video (BDV). The logo features a stylized diamond shape on the left, with the letters "BDV" in bold, sans-serif font to the right. Below the letters, the words "BLACK DIAMOND VIDEO" are written in a smaller font. The diamond shape is made up of smaller diamonds, with a blue diamond in the center.

Indications for use: The IDSS-ForceTriad Control Module is designed for integration in the IDSS (Integrated Digital Surgical Suite) and enables the Covidien ForceTriad Energy Platform to be controlled remotely..

Substantial Equivalence

(SE) Rational:The IDSS –ForceTriad Control Module is substantiallyequivalent to the predicate device since intended use,operational principle, basic technology and design are similar.The minor differences between the IDSS –ForceTriad ControlModule and the predicate device raise no new issues of safetyand effectiveness, as these differences have no effect on theperformance, function or general intended of use of the device.Therefore based on the applicable testing and the equivalenceinformation presented in this submission, Black DiamondVideo believes that IDSS –ForceTriad Control Module doesnot raise any new safety or efficacy issues.
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Summary of Nonclinical

Testing:- Software Verification and Validation testing were performedon interfaces, feature functional requirements, and non-functional reliability. The software validation activities wereperformed in accordance with the FDA Guidance “Guidancefor the content of premarket Submissions for SoftwareContained in Medical Devices”. The risk analysis was carriedout in accordance with established in-house acceptance criteriabased on ISO 14971:2007.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
  • Conclusion: Testing and evaluation indicate that the system meets the needs of the users of the device and does not raise any new safety and efficacy of the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.