(158 days)
Not Found
No
The summary describes a mechanical and electrosurgical biopsy device with a controller. There is no mention of any software features that would suggest the use of AI or ML for image analysis, decision support, or any other function. The performance studies focus on safety, electrical, software verification (likely basic control software), usability, and bench performance, none of which indicate AI/ML components.
No
The device is described as a biopsy tool intended for tissue sampling and removal of breast abnormalities for histologic examination, which is a diagnostic purpose, not therapeutic.
No
The device is described as a biopsy tool intended to provide breast tissue for histologic examination, which is a method for obtaining tissue for diagnosis, but the device itself does not perform the diagnosis.
No
The device description clearly lists multiple hardware components including a probe/wand, handle, controller, table mount, and cart.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to obtain breast tissue samples for histologic examination. It is a tool for tissue sampling and removal.
- Device Description: The device is described as a "biopsy tool" that uses electrosurgical cutting and vacuum assistance to remove tissue.
- Lack of Diagnostic Function: The device itself does not perform any diagnostic testing on the tissue sample. It is solely focused on the process of acquiring the sample. The diagnostic examination (histologic examination) is performed after the tissue is obtained by this device, likely in a laboratory setting using other methods.
IVD devices are typically used to examine specimens (like tissue, blood, urine, etc.) in vitro (outside the body) to provide information for diagnosis, monitoring, or screening. This device's function is to collect the specimen, not to analyze it for diagnostic purposes.
N/A
Intended Use / Indications for Use
- · The Intact® Gen2 System is indicated to provide tissue sampling of breast abnormalities.
- · The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality.
- · The Intact · Gen2 System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
- The Intact Gen2 System is intended to preserve lesion architecture in samples with a diameter of 12–30 mm. .
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact Gen2 System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue marqins be examined for completeness of removal using standard surgical procedures.
Product codes (comma separated list FDA assigned to the subject device)
KNW
Device Description
The intent of the device is to be a biopsy tool to capture a sample of breast tissue in one intact sample to preserve the tissue architecture for histologic examination. The device is a highfrequency, vacuum-assisted electrosurgical device used to remove tissue by electrosurgical cutting and simultaneous capture of an incised tissue volume. The device includes the following components:
- Probe/Wand
Intact® Gen2 Probe/wand (Sizes: 12mm, 15mm, 20mm and 30mm) - Handle
- Gen2 Controller with Power Cord and Foot Pedal
- Table Mount
- Cart
Probe/Wand:
The probes/wands are sterile and intended for single use. They utilize a sharp blade to access the target lesion.
The Cut/Capture Electrode is located at the distal end of the wand. It consists of five (5) small diameter wire electrodes that cut tissue with a mono-polar electrosurgical cutting current and purse down to close the distal end of the cutting/capture element to make a circumscribing incision and capture of the target tissue.
The wand contains a resistor of unique value for each wand size (12mm, 15mm, 20mm, or 30mm) that is used by the Controller to identify the specific wand size installed in the handle.
Handle:
The handles consist of a hand piece at the distal end of a cable that mates with the Probe/Wand, and a connector at the proximal end of a cable that mates with the Controller. The handle is either mated to a stereotactic table using a mount or handheld for ultrasound-guided procedures.
Controller:
The Intact Gen2 Controller consists of a radiofrequency generator, which operates at a single, preset frequency. The Intact Gen2 Controller is designed to apply a high-frequency voltage during the biopsy specimen capture phase (involving Capture Electrodes). The power applied for each wand size is optimized to (1) be sufficiently high to form the electrical arc essential to effect the highly localized vaporization of cellular fluid which, thereby, enables electrosurgical cutting and (2) be sufficiently low to minimize the depth of necrosis induced in the biopsy specimen and the surrounding tissue. The controller incorporates a microprocessor and software.
Power Cord:
The Intact® Controller Power Cord is a standard detachable power cord that connects the Controller to the line voltage output receptacle.
Foot Pedal:
The foot pedal assembly is a floor-mounted, two-pedal device, which attaches to the rear of the Intact" Gen2 Controller. The two pedals provide the Arm and Capture functions, analogous to those buttons on the handle. Either the footswitch or the handle controls can be used interchangeably.
Cradle:
Multiple versions of the cradle are available to fit commercially available stereo-tactic tables. .
Cart:
This is an off-the-shelf moveable cart that houses the components of the system.
Required for the procedure but not included in the system are: An off-the-shelf vacuum unit and a patient return electrode.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
breast
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Verification Testing Summary Table:
- Test Type: Safety and Essential Performance Tests; Proposed Intact® Gen2 System: Pass; Predicate Intact® BLES System Intact BLES (K142477): Pass; Comments: Equivalent Both systems meet requirements of IEC 60601-1, 3rd Edition and IEC 60601-2-2
- Test Type: EMC Tests; Proposed Intact® Gen2 System: Pass; Predicate Intact® BLES System Intact BLES (K142477): Pass; Comments: Equivalent Both systems meet requirements of IEC 60601-1-2
- Test Type: Software Verification Tests; Proposed Intact® Gen2 System: Pass; Predicate Intact® BLES System Intact BLES (K142477): Not applicable; Comments: Proposed software meets requirements of IEC 62304. Predicate contains no software, utilized hardware for all functions.
- Test Type: Usability Tests; Proposed Intact® Gen2 System: Pass; Predicate Intact® BLES System Intact BLES (K142477): Pass; Comments: Equivalent Both systems meet the requirements of EN 60601-1-6.
- Test Type: Bench Performance Verification Test; Proposed Intact® Gen2 System: Pass; Predicate Intact® BLES System Intact BLES (K142477): Pass; Comments: Equivalent Demonstrated that the proposed system does not deliver more total energy to the biopsy tissue sample than the predicate.
- Test Type: Thermal Artifact Verification Testing; Proposed Intact® Gen2 System: Pass; Predicate Intact® BLES System Intact BLES (K142477): Pass; Comments: Equivalent Demonstrated that the thermal artifacts in biopsy tissue samples are equivalent
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 876.1075 Gastroenterology-urology biopsy instrument.
(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 15, 2016
Intact Medical Corporation % Dr. John Smith, M.D., J.D. Hogan Lovells US LLP 555 Thirteenth St. NW Washington, District of Columbia 22204
Re: K152596
Trade/Device Name: Intact® Gen2 System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: January 19, 2016 Received: January 19, 2016
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
510(k) Number (if known)
K152596
Intact® Gen2 System Device Name:
Indications for Use (Describe)
- · The Intact® Gen2 System is indicated to provide tissue sampling of breast abnormalities.
- · The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality.
- · The Intact · Gen2 System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
- The Intact Gen2 System is intended to preserve lesion architecture in samples with a diameter of 12–30 mm. .
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact Gen2 System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue marqins be examined for completeness of removal using standard surgical procedures.
Type of Use (Select one or both, as applicable) | |
---|---|
X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human
Services Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA)
Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary
Intact® Gen2 System | |
---|---|
Applicant | Intact Medical, Inc. |
Establishment Registration Number | 1226766 |
Contact Person | John Vacha |
Intact Medical Corporation | |
550 Cochituate Road - Suite 25 | |
East Wing, Floor 4 | |
Framingham, MA 01701 | |
Phone: 508-655-7820 | |
Fax: 508-655-7822 | |
Summary Date | February 12, 2016 |
Proprietary Name | Intact® Gen2 System |
Classification | Class II |
Classification Name | Biopsy Instrument and Electrosurgical Cutting & |
Coagulation Device & Accessories | |
Regulation Number | 21 C.F.R. § 876.1075 |
Classification | Product Code KNW |
Predicate Device | Intact® BLES with Intact® Excision XL wand and |
handle (K142477) |
Device Description
The intent of the device is to be a biopsy tool to capture a sample of breast tissue in one intact sample to preserve the tissue architecture for histologic examination. The device is a highfrequency, vacuum-assisted electrosurgical device used to remove tissue by electrosurgical cutting and simultaneous capture of an incised tissue volume. The device includes the following components:
1. Probe/Wand
Intact® Gen2 Probe/wand (Sizes: 12mm, 15mm, 20mm and 30mm)
-
- Handle
-
- Gen2 Controller with Power Cord and Foot Pedal
-
- Table Mount
4
Available to fit stereotactic tables (Fischer and Lorad mounts).
5. Cart
Probe/Wand:
The probes/wands are sterile and intended for single use. They utilize a sharp blade to access the target lesion.
The Cut/Capture Electrode is located at the distal end of the wand. It consists of five (5) small diameter wire electrodes that cut tissue with a mono-polar electrosurgical cutting current and purse down to close the distal end of the cutting/capture element to make a circumscribing incision and capture of the target tissue.
The wand contains a resistor of unique value for each wand size (12mm, 15mm, 20mm, or 30mm) that is used by the Controller to identify the specific wand size installed in the handle.
Handle:
The handles consist of a hand piece at the distal end of a cable that mates with the Probe/Wand, and a connector at the proximal end of a cable that mates with the Controller. The handle is either mated to a stereotactic table using a mount or handheld for ultrasound-guided procedures.
Controller:
The Intact Gen2 Controller consists of a radiofrequency generator, which operates at a single, preset frequency. The Intact Gen2 Controller is designed to apply a high-frequency voltage during the biopsy specimen capture phase (involving Capture Electrodes). The power applied for each wand size is optimized to (1) be sufficiently high to form the electrical arc essential to effect the highly localized vaporization of cellular fluid which, thereby, enables electrosurgical cutting and (2) be sufficiently low to minimize the depth of necrosis induced in the biopsy specimen and the surrounding tissue. The controller incorporates a microprocessor and software.
Power Cord:
The Intact® Controller Power Cord is a standard detachable power cord that connects the Controller to the line voltage output receptacle.
Foot Pedal:
The foot pedal assembly is a floor-mounted, two-pedal device, which attaches to the rear of the Intact" Gen2 Controller. The two pedals provide the Arm and Capture functions, analogous to those buttons on the handle. Either the footswitch or the handle controls can be used interchangeably.
Cradle:
Multiple versions of the cradle are available to fit commercially available stereo-tactic tables. .
5
Cart:
This is an off-the-shelf moveable cart that houses the components of the system.
Required for the procedure but not included in the system are: An off-the-shelf vacuum unit and a patient return electrode.
Intended Use/Indications for Use
- · The Intact · Gen2 System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities.
- · The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality.
- · The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
- · The Intact® Gen2 System is intended to preserve lesion architecture in samples with a diameter of 12-30mm.
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact" Gen2 System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Technical and Operational Characteristics
The Intact Gen2 System works on the same principle of operation as the predicate device. Both require setting up a high-frequency voltage between an electrosurgical electrode (i.e., active electrode) and the tissue to be incised. This voltage creates an arc between the electrode and the tissue. This arc, by virtue of the high temperature, induced at the point of impingement with tissue, vaporizes the tissue in its vicinity.
This 510(k) notice implements technological modifications to the wands (addition of a resistor) and controller (addition of a microprocessor and software) to individualize the amount of power required to capture a sample based on probe size. Thus, less power is required to capture a tissue sample
6
for each sized probe. The proposed device is capable of performing to its intended use similar to the predicate device.
7
Summary of Testing
Verification Testing Summary Table | |||
---|---|---|---|
Test Type | Proposed | ||
Intact® Gen2 | |||
System | Predicate | ||
Intact® BLES | |||
System Intact | |||
BLES | |||
(K142477) | Comments | ||
Safety and | |||
Essential | |||
Performance Tests | Pass | Pass | Equivalent |
Both systems meet requirements of IEC | |||
60601-1, 3rd Edition | |||
and IEC 60601-2-2 | |||
EMC Tests | Pass | Pass | Equivalent |
Both systems meet requirements of IEC | |||
60601-1-2 | |||
Software | |||
Verification Tests | Pass | Not applicable | Proposed software meets requirements of |
IEC 62304. | |||
Predicate contains no software, utilized | |||
hardware for all functions. | |||
Usability Tests | Pass | Pass | Equivalent |
Both systems meet the requirements of | |||
EN 60601-1-6. | |||
Bench Performance | |||
Verification Test | Pass | Pass | Equivalent |
Demonstrated that the proposed system | |||
does not deliver more total energy to the | |||
biopsy tissue sample than the predicate. | |||
Thermal Artifact | |||
Verification Testing | Pass | Pass | Equivalent |
Demonstrated that the thermal artifacts in | |||
biopsy tissue samples are equivalent |
Rationale for Substantial Equivalence
The proposed modifications reduce the power introduced to the patient to capture a similar intact sample (size, shape, weight, architectural integrity, thermal artifact) as that of the predicate device. The proposed device is capable of performing to its intended use similar to the predicate device.
The Intended Use/Indications for Use statement and technological characteristics of the Intact Gen2 System device and the predicate devices were compared. The Intended Use/Indications for Use statements of the devices are identical. The proposed changes to the technology are supported by extensive bench and animal testing described above and are similar to that of the predicate device with the same intended use.
Conclusion
Based on similar technology and operation characteristics as well as the bench and animal performance testing, the Intact Gen2 System device is substantially equivalent to the predicate device.