K Number
K152483

Validate with FDA (Live)

Date Cleared
2015-09-29

(29 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure on an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm.

Device Description

Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel. The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading. The key components of device are: a pressure sensor, an electric valve and an electronic control module together with an electric pump.

AI/ML Overview

The information provided describes the Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series. Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for blood pressure measurement devices are typically defined by international standards like ISO 81060-2. The document explicitly references compliance with IEC 81060-2:2009 (and
the new device uses IEC 81060-2:2013). While the document does not list the specific numerical acceptance criteria from this standard, it states that the device was tested in accordance with it and shows the reported performance for key metrics.

Acceptance Criteria (Based on ISO 81060-2 standards)*Reported Device Performance (Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series)
Blood Pressure Measurement Accuracy:± 3 mmHg (Comparable to predicate device)
- Mean difference (systolic) should be within ±5 mmHgImplicitly met by ±3 mmHg accuracy claim
- Standard deviation (systolic) should be within 8 mmHgImplicitly met by ±3 mmHg accuracy claim
- Mean difference (diastolic) should be within ±5 mmHgImplicitly met by ±3 mmHg accuracy claim
- Standard deviation (diastolic) should be within 8 mmHgImplicitly met by ±3 mmHg accuracy claim
Pulse Rate Measurement Accuracy:± 5 % of the reading

Note: The document states compliance with ISO 81060-2. The numerical acceptance criteria for mean difference and standard deviation are standard requirements within ISO 81060-2 for evaluating automated sphygmomanometers. While not explicitly listed with these specifics in the document, the stated "± 3 mmHg" for blood pressure accuracy implies compliance with these underlying statistical criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: One hundred patients (49 males and 51 females).
  • Data Provenance: The document does not specify the country of origin for the data. The study appears to be prospective as patients were "invited for the study" and measurements were taken during the study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: The document does not explicitly state the number of experts. However, it indicates that "Standard auscultation method was used as the reference blood pressure monitor," which implies measurement by trained human observers (likely medical professionals) using a stethoscope and sphygmomanometer.
  • Qualifications of Experts: Not specified. However, for a clinical accuracy study using auscultation as the reference, the operators are typically trained medical personnel (e.g., physicians, nurses, or trained technicians) who are proficient in the auscultatory method.

4. Adjudication Method for the Test Set

  • The document implies that the auscultation method serves as the ground truth comparator. It states, "Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2." This suggests a direct comparison method rather than an adjudication process between multiple automated device readings or between the device and multiple human observers. There is no mention of 2+1, 3+1, or other adjudication methods.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study focuses on the accuracy of the device against a gold standard (auscultation), not on how human readers might improve with or without AI assistance, as AI is not a component of this blood pressure monitor. The device is a standalone measurement system.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, a standalone performance study was done. The clinical test evaluated the Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series as a standalone device, measuring its accuracy against the reference auscultation method without human intervention in the device's measurement process.

7. Type of Ground Truth Used

  • The ground truth used was established by the standard auscultation method. This is considered the clinical gold standard for non-invasive blood pressure measurement.

8. Sample Size for the Training Set

  • There is no mention of a training set in the provided document. This device is a traditional medical device (blood pressure monitor) that operates based on a pre-programmed algorithm (oscillometric methodology) rather than a machine learning or AI algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

  • As there is no mention of a training set, the question is not applicable.

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human figures, one behind the other, with their arms raised.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 29, 2015

Grandway Technology (shenzhen) Limited Mr. Patrick Chow General Manager Block 7, Zhu Keng Industrial Zone, Ping Shan District, Shenzhen, 518118 CN

Re: K152483

Trade/Device Name: Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 31, 2015 Received: August 31, 2015

Dear Mr. Patrick Chow,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152483

Device Name

Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series

Indications for Use (Describe)

This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure on an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K152483

Page 1 of 4

1. Submitter identification

510(k) SubmitterGRANDWAY TECHNOLOGY (SHENZHEN) LIMITED
AddressBlock 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang District, 518118,Shenzhen, People's Republic of China
Phone Number(00852)-2851-6789
Fax Number(00852)-2851-6278
Contact PersonMr. Patrick Chow
Date of Submission28-Aug-2015

Device identification 2.

Trade NameDigital Automatic Blood Pressure Monitor BPM30 and BPM31 Series
Common NameNon-invasive Blood Pressure Measurement System
Classification NameNon-invasive Blood Pressure Measurement System(CFR 870.1130, Class II, Product Code DXN)

3. Predicate device

Predicate DeviceDigital Automatic Blood Pressure Monitor BPM18 Series
ManufacturerGRANDWAY TECHNOLOGY (SHENZHEN) LIMITED
510(k) NumberK133619

Device Description 4.

Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel.

The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading.

The table below illustrate the feature presence in Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series.

ModelBPMeasurementPulse RateMeasurementWHOClassificationIHBLCD TypeBacklightDCJackV.A Size
MD3000Positive reflectiveXX60 x 70 mm
MD3010Positive reflectiveX60 x 70 mm

{4}------------------------------------------------

ModelBPMeasurementPulse RateMeasurementWHOClassificationIHBLCD TypeBacklightDCJackV.A Size
MD3020Negative transmissiveX60 x 70 mm
MD3030Negative transmissive60 x 70 mm
MD3040Positive transmissiveX60 x 70 mm
MD3050Positive transmissive60 x 70 mm
MD3001Positive reflectiveXX50 x 60 mm
MD3011Positive reflectiveX50 x 60 mm
MD3021Negative transmissiveX50 x 60 mm
MD3031Negative transmissive50 × 60 mm
MD3041Positive transmissiveX50 x 60 mm
MD3051Positive transmissive50 x 60 mm
MD3100Positive reflectiveXX60 x 70 mm
MD3110Positive reflectiveX60 × 70 mm
MD3120Negative transmissiveX60 x 70 mm
MD3130Negative transmissive60 x 70 mm
MD3140Positive transmissiveX60 x 70 mm
MD3150Positive transmissive60 x 70 mm
MD3101Positive reflectiveXX50 x 60 mm
MD3111Positive reflectiveX50 x 60 mm
MD3121Negative transmissiveX50 x 60 mm
MD3131Negative transmissive50 x 60 mm
MD3141Positive transmissiveX50 x 60 mm
MD3151Positive transmissive50 x 60 mm

5. Indication for use

This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure on an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm.

Comparison of technological characteristics between new Device and 6. predicate Devices

Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series is compared to the predicate device, Digital Automatic Blood Pressure Monitor BPM18 Series (K133619) in the device comparison table below.

ItemPredicate DeviceNew deviceComment
Indication for UseDigital Automatic Blood PressureMonitor BPM18 Series is for use bymedical professional or home user.The BPM18 Series is intended tomeasure the systolic and diastolicblood pressure, and pulse rate of anadult individual by using a non-invasive technique, in which aninflatable cuff is wrapped around theupper arm.This device is for use by medicalprofessional or home users. It isintended to measure the systolic anddiastolic blood pressure on an adultindividual by using a non-invasivetechnique, in which an inflatable cuffis wrapped around the upper arm.Equivalent
MeasurementMethodNon-invasive, OscillometricNon-invasive, OscillometricIdentical
IHB DetectionYesYesIdentical
ItemPredicate DeviceNew deviceComment
PatientPopulationAdultAdultIdentical
BPMeasurementRangeCuff Pressure: 0 - 300 mmHgSystolic Pressure: 50 - 250 mmHgDiastolic Pressure: 30 - 200 mmHgCuff Pressure: 0 - 300 mmHgSystolic Pressure: 50 - 250 mmHgDiastolic Pressure: 30 - 200 mmHgIdentical
Number of User2 independent users4 independent usersMore user
Memory Space2 users × 120 memory space4 users × 120 or 240 memory spaceMore usermemory
Resolution ofMeasurementBlood Pressure: 1 mmHg or 0.1kPaPulse Rate: 1 beat/ minBlood Pressure: 1 mmHg or 0.1kPaPulse Rate: 1 beat/ minIdentical
Blood PressureMeasurementAccuracy± 3 mmHg or 2% of reading± 3 mmHgEquivalent
Pulse RateMeasurementRange40 - 180 beats/min40 - 180 beats/minIdentical
Pulse RateMeasurementAccuracy± 5 % of the reading± 5 % of the readingIdentical
Display TypeLCDLCDIdentical
Power Source4 × 1.5 V AAA-batteries and/or ACAdaptor4 × 1.5 V AAA-batteries and/or ACAdaptorIdentical
PressurizationModeAutomatic InflationAutomatic InflationIdentical
Deflation ModeAutomatic Exhaust/ DeflationAutomatic Exhaust/ DeflationIdentical
OperatingConditionTemperature: +5 to +40 °CHumidity: 15 to 93 % R.H. maxAtmospheric Pressure: 700-1060 hPaTemperature: +5 to +40 °CHumidity: 15 to 93 % R.H. maxAtmospheric Pressure: 700-1060 hPaIdentical
Storage andTransportationConditionTemperature: -25 to +70 °CHumidity: up to 93% R.H. maxAtmospheric Pressure: 700-1060 hPaTemperature: -25 to +70 °CHumidity: up to 93% R.H. maxAtmospheric Pressure: 700-1060 hPaIdentical
MaterialResistances, capacitance, transistors,amplifiers, pressure sensor, CPU,PCB, cuff ABS button, ABS cabinet,batteries and packagingResistances, capacitance, transistors,amplifiers, pressure sensor, CPU,PCB, cuff ABS button, ABS cabinet,batteries and packagingIdentical
CompatibilitywithEnvironmentand OtherDevicesNo influence with environment andother deviceNo influence with environment andother deviceIdentical
ApplicableStandard- EN 1060-1:1995+A2:2009- EN 1060-3:1997+A2:2009- IEC 60601-1:2012- EN 60601-1-2:2007- FCC Part 15 Subpart B- IEC 81060-2:2009- EN 1060-1:1995+A2:2009- EN 1060-3:1997+A2:2009- IEC 60601-1:2012- IEC 60601-1-2:2007- FCC Part 15 Subpart B- IEC 81060-2:2013Equivalent

{5}------------------------------------------------

Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series is a non-invasive measuring device and utilizes the oscillometric methodology to measure the blood pressure reading. The key components of device are: a pressure sensor, an electric valve and an electronic control module together with an electric pump.

{6}------------------------------------------------

K152483

Page 4 of 4

The electric pump inflate (and deflate) the inflatable cuff automatically according to our designed architecture. The predicate device adopts exactly same methodology and key components for measuring blood pressure.

Clinical and Non-clinical Tests 7.

Clinical Test Summary

Testing to insure clinical accuracy of the device in accordance with ISO 81060-2 as documented in Clinical Test report.

One hundred patients (49 males and 51 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2.

Non-Clinical Test Summary

Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series has performed several non-clinical tests to show that all requirement specifications and standard requirements are met. The tests includes the follows:

  • � EN 1060-1:1995+A2:2009
  • � EN 1060-3:1997+A2:2009
  • � IEC 60601-1:2012
  • � IEC 60601-1-2:2007
  • ゃ FCC Part 15 Subpart B

All of the clinical and non-clinical testing performed on Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series are same as the predicate device. Therefore, no bench test is conducted to show the performance of Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series is equivalent to the predicate device.

Conclusion 8.

Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series has the same intended use and same technological characteristics as the predicate device, Digital Automatic Blood Pressure Monitor BPM18 Series (K133619). Moreover both clinical and non-clinical testing has demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, Digital Automatic Blood Pressure Monitor BPM30 and BPM31 Series is substantially equivalent to the predicate device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).