K Number
K152452
Date Cleared
2016-05-19

(265 days)

Product Code
Regulation Number
878.4635
Panel
SU
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Medtek Lighting, LLC's Peacock Tanning, Express Tan Models 1000HT, 2000CT, 2000HT, 3200 are intended to provide ultraviolet light to tan the skin, in male and female users, not under the age of 18 years.

Device Description

The Express Tan 1000HT device is a skin tanning device, wall mounted or used on an accessory stand with a single lighting panel with 10 Cosmedico (T12/F71"/100w) lamps, driven by standard electrical lighting systems at 120 Volts, producing primary UV-A emissions within an enclosed metal (structure) and acrylic (exposure) housing and utilized only with included protective eyewear, as outlined in 21 CFR 1040.20 (c). A digital timer is used to program exposure times. Cabinet dimensions: 85.5" high, 5" deep, 22" wide - when door panels are closed. A digital control timer module is used to program exposure times.

The Express Tan models1600HT/2000CT/2000HT/3200 are skin tanning devices, wall mounted with 4 lighting panels containing an array of a total of 16, 20, or 32 Cosmedico (T12/F71"/100w) lamps, driven by standard electrical lighting systems at 120 Volts, producing UV-A emissions within an enclosed metal (structure) and acrylic (exposure) housing, and that is utilized only with included protective eyewear, as outlined in 21 CFR 1040.20 (c). A digital timer is used to program exposure times. Cabinet dimensions: 92" high, 8" deep, 50" wide - when panels are closed.

AI/ML Overview

This document is a 510(k) premarket notification for a tanning device, not an AI/ML medical device. Therefore, the requested information about acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, and MRMC studies is not applicable in the context of an AI/ML device.

The document discusses the substantial equivalence of the "Peacock Tanning, Express Tan Models" to predicate devices based on intended use, technological characteristics, and conformance to special controls and performance standards.

Here's the relevant information that can be extracted from the provided text, adapted to the closest interpretations:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Intended Use EquivalenceThe device must have substantially equivalent intended use as predicate devices.The devices are "intended to provide ultraviolet light to tan the skin, in male and female users, not under the age of 18 years," which is the same as the predicate devices.
Technological Characteristics EquivalenceThe device must have substantially equivalent technological characteristics as predicate devices.The proposed devices (Express Tan models 1000HT/1600HT/2000CT/2000HT/3200) are described as having the "same technological characteristics as the predicate devices" and conforming to special controls. They are wall-mounted or freestanding units with T12/F71"/100w Cosmedico lamps, standard electrical systems (120 Volts), UV-A emissions, enclosed metal/acrylic housing, and digital timers. The primary predicate devices (8920525) are stated as "identical to the proposed Express Tan devices and no differences exist." The secondary predicates (K944015, K871237) have "additional safety features and newer components" but maintain the same technological characteristics.
Performance Standards ComplianceThe device must conform to applicable international consensus standards for electrical safety and electromagnetic compatibility (EMC), and performance standards under Section 1040.20 of the Federal Food, Drug, and Cosmetics Act, including radiation limits and protective eyewear transmittance.The devices "have been tested and conform to international consensus standards":
  • Electrical Safety: IEC/EN 60601-1:2005 Edition 3 / A1:2012 (General II (ES/EMC))
  • EMC: IEC 60601-1-2 Edition 4: 2014 (General II (ES/EMC))
    They are also "in compliance with performance standards that have been established for such devices under Section 1040.20 of the Federal Food, Drug, and Cosmetics Act."
  • Specific performance testing (spectral analysis) was done to measure irradiance to ensure compliance with radiation limits set out in 21 CFR 1040.20.
  • Specific performance testing was done on the included protective eyewear (goggles) to ensure spectral transmittance did not exceed value limits set out in 21 CFR 1040.20 while allowing visibility for timer reset. |
    | Safety Features | Devices for home use (Models 1000HT/1600HT/2000HT) must include specific safety training, warnings, contraindications, precautions, emergency shut-off, photosensitizing agents list, an exposure log, and a risk acknowledgment certification for consumers. | For home environment models, the sponsor "provides training to the consumer before the product is shipped out." This training includes a User Manual overview, help in determining skin type/exposure schedules, warnings, contraindications, precautions, safety features (e.g., emergency shut-off), a list of photosensitizing agents, a "Tanning Exposure Log," and a "Risk Acknowledgement Certification" form, which consumers must sign. |
    | Age Restriction | The device should be used by individuals not under the age of 18 years. | The devices are "intended to provide ultraviolet light to tan the skin, in male and female users, not under the age of 18 years." The consumer training emphasizes "Tanning bed is not to be used by any person under the age of 18 years." |

Since this is not an AI/ML device, the following points are not applicable (N/A):

  1. Sample sized used for the test set and the data provenance: N/A (no AI/ML algorithm or test set in this context)
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (no ground truth in the AI/ML sense)
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: N/A
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: N/A
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): N/A
  7. The sample size for the training set: N/A
  8. How the ground truth for the training set was established: N/A

§ 878.4635 Sunlamp products and ultraviolet lamps intended for use in sunlamp products.

(a)
Identification. A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.(b)
Classification. Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
Labeling —(i)Sunlamp products. (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
Sunlamp products and UV lamps intended for use in sunlamp products. Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
Performance standard. Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.