(210 days)
The Ventix Knotless Anchor is indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/ foot, elbow, and knee. Specific indications as follows:
Shoulder - Bankart repair, SLAP repair, acromio-clavicular separation, rotator cuff repair or capsulolabral reconstruction, biceps tenodesis, and deltoid repair.
Wrist/Hand - Scapholunate ligament reconstruction and ulnar/radial collateral ligament reconstruction.
Ankle/Foot - Lateral stabilization, medial stabilization, Achilles tendon repair/reconstruction, hallux valgus reconstruction, and mid- and forefoot reconstruction.
Elbow - Ulnar or radial collateral ligament reconstruction, biceps tendon reconstruction, and lateral epicondylitis repair
Knee - Medial collateral ligament repair, lateral ligament repair, posterior oblique ligament repair, joint capsule closure, iliotibial band tenodesis, patellar ligament/tendon repair, patellar realignment and repair, and Vastus Medialis Obliquus (VMO) Muscle Advancement.
The Ventix Knotless anchors are also indicated for supplementary fixation when used in conjunction with a primary fixation device in ACL repair and reconstruction surgical procedures requiring graft fixation.
The Ventix anchor is a two-piece design featuring a winged tip that fits into the anchor body. The anchor body is vented. The Ventix anchors will be offered in two sizes: 4.75mm and 5.5mm and is manufactured from polyetheretherketone (PEEK). The Ventix anchor will be loaded onto a disposable inserter.
The provided text is a 510(k) premarket notification for the Ventix Knotless Anchors. It does not contain information about acceptance criteria, device performance tables, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone algorithm performance, or training set details that would be relevant to an AI/ML medical device.
The document describes a mechanical medical device (surgical anchors) and its substantial equivalence to a predicate device. The performance data discussed is related to the pull-out strength of the anchor, which is a biomechanical characteristic, not an AI/ML performance metric.
Therefore, I cannot fulfill your request for information regarding acceptance criteria and studies related to an AI/ML device based on the provided text. The information you are asking for is typically found in FDA submissions for AI/ML-powered software as a medical device (SaMD) or AI-augmented medical devices.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of three faces in profile, stacked on top of each other.
March 8, 2016
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Biomet Manufacturing Corporation Jared Cooper, Ph.D. Regulatory Affairs Specialist, Sports Medicine 56 East Bell Drive Warsaw, Indiana 46581
Re: K152262
Trade/Device Name: Ventix Knotless Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: February 5, 2016 Received: February 8, 2016
Dear Dr. Cooper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152262
Device Name Ventix Knotless Anchor
Indications for Use (Describe)
The Ventix Knotless Anchor is indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/ foot, elbow, and knee. Specific indications as follows:
Shoulder - Bankart repair, SLAP repair, acromio-clavicular separation, rotator cuff repair or capsulolabral reconstruction, biceps tenodesis, and deltoid repair.
Wrist/Hand - Scapholunate ligament reconstruction and ulnar/radial collateral ligament reconstruction.
Ankle/Foot - Lateral stabilization, medial stabilization, Achilles tendon repair/reconstruction, hallux valgus reconstruction, and mid- and forefoot reconstruction.
Elbow - Ulnar or radial collateral ligament reconstruction, biceps tendon reconstruction, and lateral epicondylitis repair
Knee - Medial collateral ligament repair, lateral ligament repair, posterior oblique ligament repair, joint capsule closure, iliotibial band tenodesis, patellar ligament/tendon repair, patellar realignment and repair, and Vastus Medialis Obliquus (VMO) Muscle Advancement.
The Ventix Knotless anchors are also indicated for supplementary fixation when used in conjunction with a primary fixation device in ACL repair and reconstruction surgical procedures requiring graft fixation.
| Type of Use (Select one or both, as applicable) |
|---|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Ventix Knotless Anchor 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet Inc.56 East Bell DrivePO Box 587Warsaw, IN 46581Establishment Registration Number: 1825034Phone: (574) 267-6639Fax: (574) 267-8137 |
|---|---|
| Contact: | Jared Cooper, Ph.DRegulatory Affairs ManagerBiomet Sports Medicine |
| Date: | February 5, 2016 |
| Subject Device: | Trade Name: Ventix Knotless AnchorCommon Name: Soft Tissue Anchor |
| Classification Name: | • MBI– Fastener, Fixation, Nondegradable, Soft tissue (2888.3040) |
Legally marketed devices to which substantial equivalence is claimed:
- . K070389 - Biomet PEEK Knotless Anchors
Device Description
The Ventix anchor is a two-piece design featuring a winged tip that fits into the anchor body. The anchor body is vented. The Ventix anchors will be offered in two sizes: 4.75mm and 5.5mm and is manufactured from polyetheretherketone (PEEK). The Ventix anchor will be loaded onto a disposable inserter.
Intended Use and Indications for Use
The Ventix Knotless Anchors are indicated for use in soft tissue reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, and knee. Specific indications as follows:
Shoulder - Bankart repair, SLAP repair, acromio-clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, and deltoid repair.
Wrist/Hand - Scapholunate ligament reconstruction and ulnar/radial collateral ligament reconstruction.
CFR
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "BIOMET" in a sans-serif font. The letters are all capitalized and are a dark gray color. The letters are evenly spaced and the word is centered in the image. The background is white.
Ankle/Foot - Lateral stabilization, medial stabilization, Achilles tendon repair/reconstruction, hallux valgus reconstruction, and mid- and forefoot reconstruction.
Elbow - Ulnar or radial collateral ligament reconstruction, biceps tendon reconstruction, and lateral epicondylitis repair
Knee - Medial collateral ligament repair, lateral ligament repair, posterior oblique ligament repair, joint capsule closure, iliotibial band tenodesis, patellar ligament/tendon repair, extra-capsular repair, patellar realignment and repair, and Vastus Medialis Obliquus (VMO) Muscle Advancement.
The Ventix Knotless anchors are also indicated for supplementary fixation when used in conjunction with a primary fixation device in ACL repair and reconstruction surgical procedures requiring graft fixation.
Summary of Technological Characteristics
The technological characteristics (materials, design, sizing, and indications) of the Ventix Knotless Anchors are similar or identical to the predicate devices or other previously cleared devices.
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: The proposed Ventix Knotless Anchors are intended for use in soft tissue . reattachment procedures in the shoulder, wrist/hand, ankle/foot, elbow, and knee which are identical to the predicate device.
- Indications for Use: Indications are identical to the predicate device with exception of the expanded indications which are supported in this submission.
- Materials: The proposed Ventix Knotless Anchor uses the identical implant material as the predicate device.
- Design Features: The proposed Ventix Knotless Anchor incorporates similar design features as the predicate device, in that it is a cannulated fully threaded two-piece design with similar dimensions. The differences are in the symmetrical winged anchor tip compared to the asymmetrical anchor tip of the predicate. Further, the predicate two-piece anchor was loaded on the inserter where the tip and body were not connected. The two-piece Ventix anchor is loaded on the inserter as a single connected anchor.
- Sterilization: The proposed Ventix Knotless Anchor is provided sterile via gamma irradiation, the same sterilization method utilized for the predicate device.
Summary of Performance Data (Nonclinical and/or Clinical)
- . Non-Clinical Tests
- o Non-clinical laboratory testing was performed to verify the pull-out strength of the Ventix Knotless Anchor by mechanical testing as compared to the Rattler PEEK Knotless Anchor. The test results indicate that the Ventix Knotless Anchor provides greater fixation strength to the predicate device and would be functional within the intended use.
- Clinical Tests
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "BIOMET" in all caps. The letters are a dark gray color and are set against a white background. The font is sans-serif and appears to be bold.
- None provided as a basis for substantial equivalence. O
Substantial Equivalence Conclusion
The proposed Ventix Knotless Anchor has similar intended use, technological characteristics, and mechanical performance as the predicate devices. The performance testing data identified no new risks and substantial equivalence to the legally marketed predicate devices.
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.