K Number
K152257
Date Cleared
2017-07-21

(711 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fujifilm Duodenoscope Model ED-530XT is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Device Description

The Fujifilm Duodenoscope Model ED-530XT is comprised of three main sections: an operation section, an insertion portion, and an umbilicus. The operation section controls the angulation (up/down/left/right) of the distal end of the endoscope. The insertion portion contains glass fiber bundles, several channels and a CCD image sensor. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD sensor to capture an image and display it on the monitor. The endoscope also contains several channels to deliver air/water and provide suction, as well as a forceps channel. The forceps channel is used to introduce endoscope accessories such as biopsy forceps during the procedure. The umbilicus section consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called the Fujifilm Duodenoscope Model ED-530XT. It primarily focuses on demonstrating substantial equivalence to a predicate device (Fujinon G5 Duodenoscope ED-450XT5 / ED-250XT5) rather than detailing specific acceptance criteria and a study proving device performance against those criteria in a typical clinical or AI context.

Therefore, many of the requested categories (like sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance with AI, and specific ground truth types) are not applicable or not explicitly detailed in this document, as the submission relies on bench testing and comparison to an existing predicate device rather than a clinical trial demonstrating new performance metrics against specific acceptance thresholds.

Here's an analysis based on the available information:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not provide a formal table of specific acceptance criteria with quantifiable metrics (e.g., sensitivity, specificity, accuracy) that are typically seen for diagnostic devices or AI algorithms. Instead, the performance is demonstrated through various engineering and safety tests, with the implied "acceptance criterion" being that the device meets the standards and performs comparably to the predicate.

Test CategoryImplied Acceptance Criterion / Standard MetReported Device Performance
Safety & Regulatory Compliance
EMC & Electrical SafetyIEC 60601-1-2:2007, ANSI/AAMI ES60601-1:2005, IEC 60601-2-18:2009Device met performance specifications.
BiocompatibilityISO 10993Device was evaluated in accordance with ISO 10993.
Cleaning, Disinfection, SterilizationAAMI TIR12:2010, AAMI TIR30:2011, FDA guidance "Reprocessing Medical Devices..." (March 2015)Validation performed, indicating instructions are effective.
Endoscope Specific TestingISO8600-1:2013Device met performance specifications.
Device Functionality
Field of ViewPerformance specificationsMet performance specifications.
Bending CapabilityPerformance specificationsMet performance specifications.
Rate of Air SupplyPerformance specificationsMet performance specifications.
Rate of Water SupplyPerformance specificationsMet performance specifications.
Rate of SuctionPerformance specificationsMet performance specifications.
ResolutionPerformance specificationsMet performance specifications.
Light Guide (LG) OutputPerformance specificationsMet performance specifications.
Elevator Wire Channel SealSafety and effectiveness analysisDemonstrated safety and effectiveness.
Distal End Cap Seal Strength & IntegritySafety and effectiveness analysisDemonstrated safety and effectiveness.
Overall EquivalenceSubstantial equivalence to predicate device (ED-450XT5)"As safe and effective, and performs as well or better, than the predicate device."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified. The performance data described are primarily from bench testing of the device itself and its components, rather than a "test set" of patient data or clinical cases.
  • Data Provenance: Not applicable in the context of clinical data. The tests are lab-based, engineering performance tests.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not applicable. The ground truth for these engineering and safety tests is established by adherence to industry standards, measurement protocols, and physical testing, not by expert medical interpretation of images or clinical outcomes.

4. Adjudication Method for the Test Set

  • Not applicable. This concept is relevant for studies involving human interpretation or clinical endpoints, not for bench testing of physical device performance.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. This type of study is typically done for diagnostic aids or AI algorithms where human reader performance is a key outcome. This 510(k) submission focuses on the safety and functional equivalence of an endoscope.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Not applicable. This device is an endoscope, which is a physical medical instrument, not an AI algorithm. Therefore, "standalone" performance in the AI sense is not relevant.

7. The Type of Ground Truth Used

  • The "ground truth" for the tests described is based on engineering specifications, established consensus standards (e.g., ISO, AAMI, IEC), and documented test methods for device performance, safety, and biocompatibility.

8. The Sample Size for the Training Set

  • Not applicable. This device is not an AI algorithm that undergoes "training."

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with a stylized wing-like shape extending from the top profile.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 21, 2017

FUJIFILM Medical Systems U.S.A., Inc. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, DC 20004

Re: K152257

Trade/Device Name: Fujifilm Duodenoscope Model ED-530XT Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDT Dated: May 30, 2017 Received: May 30, 2017

Dear John J. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

Page 2 - John Smith, M.D., J.D.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, sans-serif font. The text is black and appears to be centered. The background is a light color, possibly white, with a faint watermark or design element visible behind the text.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K152257

Device Name

Fujifilm Duodenoscope Model ED-530XT

Indications for Use (Describe)

The Fujifilm Duodenoscope Model ED-530XT is intended for the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Type of Use (Select one or both , as applicable)
----------------------------------------------------------------

区 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

FUJIFILM Medical Systems U.S.A., Inc.'s Fujifilm Duodenoscope Model ED-530XT

Submitter's Information

FUJIFILM Medical Systems U.S.A., Inc. Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293

Contact Person:

Shraddha More Specialist, Regulatory Affairs and Quality Assurance Telephone: (973) 686-2627 Facsimile: (973) 633-8818 E-Mail: smore@fujifilm.com

Larry Picciano Senior Director, Quality and Regulatory Affairs Telephone: (973) 686-2479 Facsimile: (973) 633-8818 E-Mail: larry.picciano@fujifilm.com

Date Prepared: May 30, 2017

ldentification of the Subject Device:

Proprietary/Trade Name:Fujifilm Duodenoscope Model ED-530XT
Common Name:Video Endoscope
Device Class:Class II
Review Panel:Gastroenterology/Urology

Classification Information:

Endoscope and Accessories (Flexible/Rigid), 21 C.F.R. § 876.1500 Product Code: FDT

Primary Predicate

Fujinon G5 Duodenoscope ED-450XT5 / ED-250XT5 (K042076)

{4}------------------------------------------------

Intended Use / Indications for Use

The Fujifilm Duodenoscope Model ED-530XT is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Technological Characteristics

The Fujifilm Duodenoscope Model ED-530XT is comprised of three main sections: an operation section, an insertion portion, and an umbilicus. The operation section controls the angulation (up/down/left/right) of the distal end of the endoscope. The insertion portion contains glass fiber bundles, several channels and a CCD image sensor. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD sensor to capture an image and display it on the monitor. The endoscope also contains several channels to deliver air/water and provide suction, as well as a forceps channel. The forceps channel is used to introduce endoscope accessories such as biopsy forceps during the procedure. The umbilicus section consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source.

The subject device is used in combination with Fujifilm's EPX-4400 (K944620), EPX-4400HD (K944620), and EPX-4440HD (K102466, K140149) video processors and light sources, as well as peripheral devices such as water tank, endoscope accessories, monitor, printer, electrosurgical instruments, foot switch, and cart. All of the peripheral accessories are substantially similar to those cleared in the predicate submission, with the exception of a new single use Elevator Cleaning Brush WB1318DE.

The differences between the subject and predicate device are as follows:

  • Distal end design differences (e.g., changes in materials and dimensions);
  • . Reduction in the diameter of the flexible portion and insertion portion;
  • Optical differences (e.g., modified viewing direction and CCD sensor); and ●
  • Modifications to the device's labeled reprocessing procedure.

Performance Data

EMC and electrical safety of the subject devices were evaluated using the following consensus standards: IEC 60601-1-2:2007; ANSI/AAMI ES60601-1:2005; and IEC 60601-2-18:2009.

Biocompatibility of the subject device was evaluated in accordance with ISO 10993.

Cleaning, disinfection, and sterilization were evaluated according to the following consensus standards: AAMI TIR12:2010, AAMI TIR30:2011. Validation of the cleaning, disinfection, and sterilization instructions was performed in accordance with FDA's quidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17, 2015.

Endoscope specific testing was conducted using the consensus standard ISO8600-1:2013.

{5}------------------------------------------------

The subject device met performance specifications in the following additional testing:

  • Field of View ●
  • Bending capability
  • Rate of air supply ●
  • Rate of water supply ●
  • Rate of suction
  • Resolution ●
  • Light Guide (LG) Output ●

Performance testing and design analysis was conducted to demonstrate the safety and effectiveness of the elevator wire channel seal, as well as distal end cap seal strength and integrity.

Substantial Equivalence

The ED-530XT is substantially equivalent the ED-450XT5. The ED-530XT has the same intended use, indications, and principles of operation, and similar technological characteristics, as its predicate device. The minor technological differences between the ED-530XT and its predicate devices were made for the purpose of overall product enhancement and general technological advancement, and do not raise different questions of safety or effectiveness. Performance data demonstrate that the ED-530XT is substantially equivalent to the ED-450XT5. Thus, the ED-530XT is substantially equivalent to the predicate device.

Conclusions

Based on device description, comparison to predicate device, and performance bench data described above, Fujifilm believes that the ED-530XT is substantially equivalent to the predicate device ED-450XT5 based on intended use/indications for use and technological characteristics. The minor technological differences between the subject endoscope and its predicate device raise no new issues of safety or effectiveness. Bench testing data demonstrated that the subject endoscope is as safe and effective, and performs as well or better, than the predicate device. Thus, the device is substantially equivalent to the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.