(100 days)
Turbo-Power™ Laser Atherectomy Catheted for laser atherectomy of de novo or restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).
The Turbo-Power System (Laser Atherectomy Catheter) is a laser atherectomy device designed for use with the CVX-300™ Excimer Laser System. The Turbo-Power Laser Atherectomy Catheter is a sterile, single use, prescription only device used for peripheral atherectomy.
Turbo-Power is used to directionally ablate infrainguinal concentric and eccentric lesions in vessels that are 3.5mm or greater in diameter. Turbo-Power Laser Atherectomy Catheter is comprised of 2 subassemblies:
-
- Catheter Subassembly
-
- Motor Drive Unit (MDU) Subassembly
The working length of the Turbo Power Laser Atherectomy Catheter is constructed of multiple optical fibers arranged eccentrically around a 0.018" (0.46 mm) quidewirecompatible lumen. The PTFE guidewire lumen tip is attached to a stainless steel torque wire which is connected to the MDU at the proximal end of the working length. The multifiber laser catheter transmits ultraviolet energy from the Spectranetics CVX-300 Excimer Laser System through the tip of the laser to an obstruction in the patient's artery. The outer surface of the laser catheter working length is hydrophilic-coated, and the distal tip of the catheter contains a radiopaque marker band for in situ visibility. The ultraviolet energy transmitted from the CVX-300 laser system is used to photoablate multiple morphology lesions which mav be comprised of atheroma, fibrosis, calcium, and thrombus, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.
This document, a 510(k) summary for the Spectranetics Turbo-Power Laser Atherectomy Catheter (K152181), does not contain information about the device's acceptance criteria or a study directly proving the device meets specific acceptance criteria in the format requested.
Instead, it focuses on demonstrating substantial equivalence to predicate devices (Turbo-Tandem and Turbo-Elite) based on various performance data and existing clinical studies of the predicate devices.
Here's an breakdown of the information that is available and a note on what is missing based on your request:
Missing Information:
- A table of acceptance criteria and the reported device performance: This document does not provide a table specifying predefined quantifiable acceptance criteria (e.g., specific thresholds for success, failure rates, accuracy) for the Turbo-Power device itself, nor does it present its performance against such criteria. The "Performance Data" section lists types of tests conducted but not the results or the criteria for those results.
- Sample size used for the test set and the data provenance: While a "pre-clinical GLP study" is mentioned, details about its sample size, specific test set, and data provenance are not provided. The clinical data leveraged are from studies of the predicate devices.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no direct acceptance criteria study with expert ground truth is described for the Turbo-Power device.
- Adjudication method for the test set: Not applicable for the same reason as above.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a laser atherectomy catheter, not an AI diagnostic tool.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is not an AI algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the predicate device studies (EXCITE, PATENT, CELLO), outcomes data (MAE, TLR, restenosis reduction, acute procedural success) were used as endpoints, which can be considered a form of ground truth for assessing clinical effectiveness. However, no ground truth is explicitly described for the Turbo-Power device's direct performance against acceptance criteria.
- The sample size for the training set: Not applicable, as this is not an AI device.
- How the ground truth for the training set was established: Not applicable.
Information that is available (or can be inferred) from the document, relevant to the spirit of your request, but focused on substantial equivalence rather than direct acceptance criteria fulfillment:
The document states that the following testing was conducted to validate and verify that the subject device met all specifications:
- Simulated Use Testing
- Functional Testing
- Physical Testing
- Laser Testing
- Software Testing
- Sterility/Biocompatibility/Physiochemical Testing
Pre-clinical Data:
- Study Type: A pre-clinical GLP study was conducted to compare the usability and procedural safety of the Turbo-Power and Turbo-Tandem laser catheters.
- Purpose: To support the premarket notification and demonstrate similarity to the predicate device.
- Sample Size/Data Provenance: Not specified in this document.
- Ground Truth: Usability and procedural safety endpoints, likely determined by direct observation and measurement in the GLP study, but the specific metrics are not detailed.
Clinical Data (Leveraged from Predicate Devices to Support Indications for Use):
The document explicitly states: "New clinical data was not required to demonstrate substantial equivalence." Instead, the following clinical studies of the predicate devices were leveraged:
-
The EXCITE Trial:
- Device: Spectranetics Turbo Elite Laser Ablation Catheter and Spectranetics Turbo-Tandem Laser Guide Catheter with Laser Atherectomy Catheter.
- Focus: Safety and efficacy of Excimer Laser Atherectomy (ELA) with adjunctive PTA vs. PTA alone for femoropopliteal bare nitinol in-stent restenosis (ISR) in vessels ≥5mm.
- Study Design: Prospective randomized controlled trial (2:1 randomization).
- Endpoints/Ground Truth:
- Primary Safety: Freedom from Major Adverse Event (MAE) through 30 days (all-cause death, major amputation in target limb, Target Lesion Revascularization (TLR)).
- Primary Efficacy: Freedom from TLR through 6 months.
- Results: Both primary safety and efficacy hypotheses were met. No statistical difference in major amputation rates, mortality, serious adverse events, or adverse events between groups.
- Sample Size/Provenance: Not specified in this document, but implied to be a multi-center trial.
-
The PATENT Study:
- Device: Spectranetics peripheral atherectomy laser catheters (Turbo Elite) used with Turbo-Booster catheters.
- Focus: Safety and performance for in-stent restenosis of nitinol stents in femoropopliteal arteries.
- Study Design: Multicenter prospective registry.
- Endpoints/Ground Truth: Safety and performance (specific metrics not detailed in this summary).
- Sample Size/Provenance: Involved 5 EU centers (specific sample size not detailed).
-
The CELLO (CLiRpath Excimer Laser System to Enlarge Lumen Openings) Study:
- Device: Turbo-Elite used with Turbo-Booster catheters.
- Focus: Safety and performance for treatment of Peripheral Artery Disease (PAD).
- Endpoints/Ground Truth:
- Primary Effectiveness: ≥ 20 percent reduction in percent diameter stenosis (assessed by angiographic core lab).
- Secondary Effectiveness: Acute procedural success (visual assessment of final residual stenosis).
- Primary Safety: Occurrence of major adverse events (clinical perforation, major dissection requiring surgery, major amputation, CVA, MI, death) at procedure time and through six months.
- Results: Demonstrated safety for the treatment of patients with stenoses and occlusions crossable by a guidewire in SFA and popliteal artery, with no occurrence of major adverse events through six-month follow-up.
- Sample Size/Provenance: Not specified in this document.
Conclusion for Substantial Equivalence:
The document concludes that the Turbo-Power System is substantially equivalent to the Turbo-Tandem and Turbo-Elite based on:
- Similarities in design.
- Performance data (referring to the listed testing).
- Pre-clinical data (GLP study).
- Leveraged clinical data from the predicate devices.
The implication is that the Turbo-Power device, having successfully completed its internal specifications and pre-clinical testing, is expected to perform similarly to the predicate devices whose clinical safety and effectiveness have already been established by the EXCITE, PATENT, and CELLO trials against their respective clinical endpoints.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 12, 2015
Spectranetics, Inc. Ms. Priscila Tapia Regulatory Specialist 9965 Federal Drive Colorado Springs, Colorado 80921
Re: K152181
Trade/Device Name: Turbo-Power Laser Atherectomy Catheter Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: October13, 2015 Received: October 14, 2015
Dear Ms. Tapia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152181
Device Name
Turbo-Power™ Laser Atherectomy Catheter
Indications for Use (Describe)
Turbo-Power™ Laser Atherectomy Catheted for laser atherectomy of de novo or restenotic lesions in native infrainguinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the Spectranetics logo with the tagline "Always Reaching Farther" underneath. The logo is in blue and features a stylized design. Above the logo, the text "K152181" is present.
510(k) Summary
This 510(k) summary was prepared in accordance with 21 CFR 807.92(c) Prepared on 07 October 2015
| 510(k) Submitter / Holder: | Spectranetics9965 Federal DriveColorado Springs, CO 80921.3617 |
|---|---|
| Establishment Registration No: 3007284006 | |
| Contact: | Ms. Priscila TapiaRegulatory Affairs SpecialistOffice: 719.447.2587Mobile: 719.393.5354Fax: 719.447.2070Email: Priscila.Tapia@spnc.com |
Subject Device
| 510(k) Number: | K152181 |
|---|---|
| Device Trade Name: | Turbo-Power™ Laser Atherectomy Catheters |
| Device Common Name: | Laser Atherectomy Catheter |
| Device Class: | II |
| Classification Regulation: | 21 CFR 870.4875, Intraluminal Artery Stripper |
| Regulation Description: | Cardiovascular |
| Product Code: | MCW |
| 510(k) Type: | Traditional |
| Model Numbers: | 423-050 |
Predicate Devices
The Turbo-Power Laser Atherectomy Catheter is being compared to the following legally marketed predicate devices:
| Turbo-Tandem | |
|---|---|
| 510(k) Number: | K140775 |
| Manufacturer: | The Spectranetics Corporation |
| Trade Name: | Turbo-Tandem Laser Guide Catheter withLaser Atherectomy Catheter |
| Device Common Name: | Percutaneous Laser Ablation Catheter |
| Turbo-Elite | |
| 510(k) Number: | K140775 |
| Manufacturer: | The Spectranetics Corporation |
| Trade Name: | Turbo-Elite Laser Atherectomy Catheter |
| Device Common Name: | Laser Atherectomy Catheter |
{4}------------------------------------------------
Intended and Indications for Use
Turbo-Power™ Laser Atherectomy Catheter is indicated for laser atherectomy of de novo or restenotic lesions in native infrainquinal arteries and for the treatment of femoropopliteal artery in-stent restenosis (ISR) in bare nitinol stents, with adjunctive Percutaneous Transluminal Angioplasty (PTA).
Device Description
The Turbo-Power System (Laser Atherectomy Catheter) is a laser atherectomy device designed for use with the CVX-300™ Excimer Laser System. The Turbo-Power Laser Atherectomy Catheter is a sterile, single use, prescription only device used for peripheral atherectomy.
Turbo-Power is used to directionally ablate infrainguinal concentric and eccentric lesions in vessels that are 3.5mm or greater in diameter. Turbo-Power Laser Atherectomy Catheter is comprised of 2 subassemblies:
-
- Catheter Subassembly
-
- Motor Drive Unit (MDU) Subassembly
The working length of the Turbo Power Laser Atherectomy Catheter is constructed of multiple optical fibers arranged eccentrically around a 0.018" (0.46 mm) quidewirecompatible lumen. The PTFE guidewire lumen tip is attached to a stainless steel torque wire which is connected to the MDU at the proximal end of the working length. The multifiber laser catheter transmits ultraviolet energy from the Spectranetics CVX-300 Excimer Laser System through the tip of the laser to an obstruction in the patient's artery. The outer surface of the laser catheter working length is hydrophilic-coated, and the distal tip of the catheter contains a radiopaque marker band for in situ visibility. The ultraviolet energy transmitted from the CVX-300 laser system is used to photoablate multiple morphology lesions which mav be comprised of atheroma, fibrosis, calcium, and thrombus, thus recanalizing diseased vessels. Photoablation is the process by which energy photons cause molecular bond disruption at the cellular level without thermal damage to surrounding tissue.
Technological Characteristics
The debulking performance of the Turbo-Power was developed to be similar to the Turbo-Tandem system, while maintaining the ability to cross difficult lesion morphologies, un-crossable with a quidewire, similar to the Turbo-Elite laser catheter, The Turbo-Power allows physicians to use a single tool for the full atherectomy procedure. To improve ease of use, the Turbo-Power device has mechanized the rotation process of the Turbo-Tandem system. In addition, the distal tip design of the subject device has been simplified by removing the tapered biasing tip of the Turbo-Tandem from the Turbo-Power design.
{5}------------------------------------------------
Performance Data
The following testing was conducted to validate and verify that the subject device met all specifications:
- Simulated Use Testing o
- Functional Testing ●
- o Physical Testing
- o Laser Testing
- Software Testing ●
- Sterility/Biocompatibility/Physiochemical Testing o
Pre-clinical and Clinical Data:
In addition to the design verification and validation tests, a pre-clinical GLP study was conducted to compare the usability and procedural safety of the Turbo-Power and Turbo-Tandem laser catheters, and support this premarket notification. New clinical data was not required to demonstrate substantial equivalence. The following clinical studies were leveraged to support the indications for use for Turbo-Power:
- Clinical Data: ●
- o The EXCITE trial evaluated the safety and efficacy of Excimer Laser Atherectomy (ELA) using the Spectranetics Turbo Elite Laser Ablation Catheter and the Spectranetics Turbo-Tandem Laser Guide Catheter with Laser Atherectomy Catheter with adjunctive percutaneous transluminal angioplasty (PTA) in comparison with PTA alone in the treatment of femoropopliteal bare nitinol in-stent restenosis in vessels ≥5mm. This trial was a prospective randomized controlled trial performed respectively in a 2:1 randomization scheme. The primary safety hypothesis was that freedom from a major adverse event (MAE) through 30 days with ELA+PTA, which included all-cause death, maior amputation in the target limb, or target lesion revascularization (TLR), would be non-inferior to PTA. The primary efficacy hypothesis was that freedom from TLR through 6 months with ELA+PTA would be superior to PTA. Both primary safety and efficacy hypothesis were met. There was no statistical difference in major amputation rates, mortality, serious adverse events, or adverse events between groups.
- The PATENT study was a multicenter prospective registry involving 5 EU o centers. The PATENT study was intended to evaluate the safety and performance of the Spectranetics peripheral atherectomy laser catheters (Turbo Elite) used in conjunction with Turbo-Booster catheters for the treatment of in-stent restenosis of nitinol stents implanted in the femoropopliteal arteries.
- The CELLO (CLiRpath Excimer Laser System to Enlarge Lumen o Openings) Study, was intended to study the safety and performance of the Turbo-Elite used in conjunction with Turbo-Booster catheters for the treatment of Peripheral Artery Disease (PAD). The primary effectiveness
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Spectranetics logo with the text "Always Reaching Farther" underneath. The logo is in blue and features a stylized diamond shape to the left of the word "Spectranetics". Above the logo is the text "K152181".
endpoint was ≥ 20 percent reduction in percent diameter stenosis, on average, as assessed by an angiographic core lab. The secondary effectiveness endpoint was acute procedural success (visual assessment of final residual stenosis). The primary safety endpoint measured was the occurrence of major adverse events, defined as clinical perforation, major dissection requiring surgery, major amputation, cerebrovascular accidents (CVA), myocardial infarction, and death at the time of procedure, through six (6) months. The study demonstrated that the Turbo-Booster is safe for the treatment of patients with stenoses and occlusions crossable by a guidewire in the superficial femoral artery and popliteal artery as evident by no occurrence of major adverse events through the six-month followup.
Substantial Equivalence
Based on the similarities in design between the subject and predicate devices currently in use, and the performance and pre-clinical data, the use of the Turbo-Power System for the proposed indication does not raise new questions related to safety and effectiveness compared with the predicates. Therefore Turbo-Power is substantially equivalent to Turbo-Tandem and Turbo-Elite.
§ 870.4875 Intraluminal artery stripper.
(a)
Identification. An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)(b)
Classification. Class II (performance standards).