K Number
K152149

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-01-22

(172 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

indicated for use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry.

Device Description

VersaOne™ Bladed Trocar is available in diameter 5mm, cannula fixation and smooth, and length standard and short. VersaOne™ Bladed Trocar consists of an obturator component with a bladed tip and a cannula assembly. The devices are supplied sterile single-use.

AI/ML Overview

This FDA 510(k) summary is for a medical device (VersaOne™ Bladed Trocar), which means its "acceptance criteria" and "performance" are related to mechanical and biological safety and efficacy, not AI algorithm performance metrics like sensitivity/specificity.

Therefore, the requested information elements related to AI algorithm studies (such as sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance with AI, and type of ground truth) are not applicable to this document.

The document describes material and performance testing against specifications, rather than a clinical study evaluating diagnostic or prognostic algorithmic performance.

Here's an interpretation of the document in the context of the requested table, focusing on what is available:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (What the device must achieve)Reported Device Performance (What the device achieved)
Biocompatibility: Complies with ISO 10993-1 standard.Acceptable results demonstrated VersaOne™ Bladed Trocar complies with ISO 10993-1 standard.
Performance: Performs to specifications.Results from performance testing demonstrated the proposed VersaOne™ Bladed Trocar perform to specifications.
Comparability: Comparable to predicate devices (K062326 and K112349).Results from performance testing demonstrated the proposed VersaOne™ Bladed Trocar is comparable to predicate devices K062326 and K112349.
Functionality: Bladed tip cuts tissues, shield covers blade upon entry, seal system prevents loss of Pneumoperitoneum, cannula bevel reduces penetration forces, stopcock for insufflation/desufflation.Implicitly met: The device description outlines these functional characteristics, and performance testing confirmed it performs to specifications, implying these functions were successful.

The study that proves the device meets the acceptance criteria is referred to as "performance testing" and "biocompatibility evaluation."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable: This document describes a medical device's mechanical and material performance testing, not a clinical study involving human data or AI algorithms. The "test set" would refer to the physical devices tested. The number of devices tested is not specified, nor is their "provenance" in the context of country of origin for clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable: "Ground truth" in this context would refer to established engineering and biocompatibility standards (e.g., ISO 10993-1). The "experts" would be the personnel conducting the testing and evaluating the results against these standards, but their number or specific qualifications are not detailed in this summary, as it is standard practice in medical device testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable: This refers to clinical data adjudication, which is not relevant to medical device performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable: This applies to AI-assisted diagnostic or screening tools. This device is a surgical trocar.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable: This applies to AI algorithms.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: The ground truth for this device's performance is based on established engineering specifications, material science standards (e.g., ISO 10993-1 for biocompatibility), and comparability to predicate devices as demonstrated through physical and mechanical testing.

8. The sample size for the training set

  • Not Applicable: This applies to AI algorithms.

9. How the ground truth for the training set was established

  • Not Applicable: This applies to AI algorithms.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, layered on top of each other. The profiles are connected by a flowing line that forms a wing-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 22, 2016

Covidien Trang Huynh Senior Specialist, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473

Re: K152149

Trade/Device Name: VersaOne™ V2 Bladed Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: January 15, 2016 Received: January 20, 2016

Dear Trang Huynh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a white design inside. The design appears to be three horizontal bars, with the middle bar being slightly shorter than the top and bottom bars.

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: January 31, 2017
Indications for UseSee PRA Statement below.
510(k) Number ( if known )K152149
Device NameVersaOne™ Bladed Trocar
Indications for Use ( Describe )indicated for use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry.
Type of Use ( Select one or both, as applicable )
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)Page 1 of 1
---------------------------------------

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Covidien. The logo consists of a blue square with a white design inside, followed by the word "COVIDIEN" in blue, sans-serif font. The blue square is on the left side of the logo, and the word "COVIDIEN" is on the right side.

510(k) Summary

This 510(k) summary information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92.

SUBMITTER:Covidien60 Middletown AvenueNorth Haven, CT 06473 USA
CONTACT PERSON:Trang HuynhSenior Specialist, Regulatory Affairs
Phone: (203) 492-7473
Fax: (203) 492-5029
Email: trang.huynh27@medtronic.com
DATE PREPARED:July 30, 2015
TRADE/PROPRIETARY NAME:VersaOne™ Bladed Trocar
COMMON/USUAL NAME:Surgical Trocar
CLASSIFICATION NAME:Endoscope and Accessories
PRODUCT CODE:GCJ
FDA PANEL NUMBER:78
CLASS CODE:Pursuant to 21 CFR 876.1500, Class II device
PREDICATE DEVICES:Autosuture™ Modified Versaport™ Trocar with Fixation Sleeve(K062326)
Versaport™ V2 Bladeless Optical (K112349)
PURPOSE OF SUBMISSION:Obtain market clearance for VersaOne™ Bladed Trocar 5mm
INTENDED USE:Indicated for use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry
DEVICE DESCRIPTION:VersaOne™ Bladed Trocar is available in diameter 5mm, cannula fixation and smooth, and length standard and short. VersaOne™ Bladed Trocar consists of an obturator component with a bladed tip and a cannula assembly. The devices are supplied sterile single-use.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a white cross in the center. The cross is made up of two intersecting bars, one horizontal and one vertical. The overall impression is that of a corporate logo.

TECHNOLOGICAL /DESIGNCHARACTERISTICS:VersaOne™ Bladed Trocar obturator has a dolphin nose (conical)shaped tip with a sharp linear blade and a spring-loaded lockingshield. The blade is used to cut the tissues as the trocar isinserted into the body cavity. The shield advances to cover theblade upon entry into a free space. The cannula assembly iscomposed of a seal system, transparent cannula sleeve, and astopcock. The seal system prevents loss of Pneumoperitoneumwhen instruments are inserted or withdrawn. The cannula sleevehas a bevel shaped at the distal end to reduce penetration forcesduring insertion and transparent for better visualization. Thestopcock valve is for insufflation and desufflation.
MATERIALS:The biocompatibility evaluation was conducted in accordance withISO 10993-1. Acceptable results demonstrated VersaOne™Bladed Trocar complies with ISO 10993-1 standard.
PERFORMANCE DATA:Results from performance testing demonstrated the proposedVersaOne™ Bladed Trocar perform to specifications andcomparable to predicate devices K062326 and K112349.
CONCLUSION:VersaOne™ Bladed Trocar is substantially equivalent topredicates devices in terms of indications for use,contraindications, technological characteristics, labeling,sterilization, shelf life, biocompatibility, and performance.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.