(212 days)
The IV-eye is a hand-held, non-invasive imaging device that assists medical personnel, trained in vascular access procedure, to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture.
The IV-eye should only be used in conjunction with standard techniques of visualization and palpation in assessing and locating veins.
The IV-eye is intended only for skin contact via a disposable single use cover.
The IV-eye is a hand-held, battery powered electronic non-invasive imaging device that assists medical personnel, trained in vascular access procedure, to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture. The IV-eye transmits near infrared light into a patient's tissue at a wavelength of 850nm. As the light hits a vascular structure it is absorbed by the hemoglobin in the blood, whereas it passes through other tissue.
The camera in the device captures the light that has passed through the patient and, in identifying the blocked light and applying a number of algorithms, the IV-eye is able to produce an image on its LCD display of the patient's vascular structure directly underneath the device. This appears as a darker color to the contrasting tissue. The picture is updated in real-time and is close to actual size. Trained medical personnel can use the image of the patient's vascular structure to assist them in choosing a suitable vein for cannulation and venipuncture.
The IV-eye is intended to be used only by trained medical personnel to assist them in locating suitable veins for venipuncture and cannulation. It does not differentiate between arteries and veins and should therefore only be used in conjunction with standard techniques of locating veins. Other than regular cleaning and replacement of batteries, the IV-eye requires no routine or preventative maintenance.
The IV-eye includes a single-use cover which prevents the lower casing and wings of the device having direct contact with the patient, and is intended to reduce cross-contamination risk in using the device. The IV-eye is intended only for skin contact via this disposable single use cover.
The provided text describes a 510(k) premarket notification for the "IV-eye" device. As such, it focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed clinical study report with specific performance metrics against pre-defined acceptance criteria as would be found in a full PMA submission or a comprehensive clinical trial publication.
However, based on the information provided, I can infer and extract some details regarding acceptance criteria and the study conducted.
1. Table of Acceptance Criteria and Reported Device Performance
The submission doesn't explicitly state quantitative acceptance criteria for device performance. Instead, the clinical study aims to "Demonstrating the ability of the IV-eye in identifying peripheral veins suitable for venipuncture and cannulation" and "Quantify the measurement depth capability of the IV-Eye in relation to the depth capability of the predicate VeinSite device from VueTek in order to establish equivalence."
The reported outcome is a qualitative statement of success: "Results of the clinical investigation support the indications for use of the IV-eye to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture and confirmed that the device can be considered equivalent to the Veinsite device from VuTek in terms of performance."
Without specific quantitative acceptance criteria, a table cannot be fully populated. However, if we interpret "equivalence in terms of performance" as the acceptance criterion, the device met this.
| Acceptance Criteria (Inferred from Study Goal) | Reported Device Performance |
|---|---|
| Ability to identify peripheral veins suitable for venipuncture and cannulation | Demonstrated ability to identify peripheral veins suitable for venipuncture and cannulation. |
| Equivalence to the VeinSite device's measurement depth capability | Confirmed to be equivalent to the VeinSite device from VuTek in terms of performance. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 30 healthy volunteers.
- Data Provenance: The document does not explicitly state the country of origin for the clinical study. It is a prospective clinical investigation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not specify the number of experts or their qualifications used to establish ground truth for the clinical study. It only mentions that the device "assists medical personnel, trained in vascular access procedure." It is implied that these "trained medical personnel" would assess the suitability of veins identified.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the clinical performance assessment. The results statement is general, implying an overall assessment rather than a detailed adjudicated reading protocol.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not explicitly described. The study focused on demonstrating the IV-eye's ability to identify veins and its equivalence to a predicate device, not on quantifying improvement in human reader performance with the device.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
The IV-eye device is described as "assists medical personnel" and "should only be used in conjunction with standard techniques of visualization and palpation." This indicates it is a human-in-the-loop device, not a standalone algorithm. Therefore, a standalone (algorithm only) performance study would not be applicable and was not described.
7. The Type of Ground Truth Used
The ground truth appears to be based on the assessment by "trained medical personnel" regarding the suitability of "peripheral veins for the purposes of cannulation and venipuncture." This suggests a form of expert assessment/clinical judgment as the ground truth. It is not explicitly stated if this was confirmed by subsequent successful cannulation/venipuncture or other objective measures (like pathology or imaging beyond the device itself).
8. The Sample Size for the Training Set
The document does not provide any information about a training set for the IV-eye. This is because the IV-eye is not described as an AI/ML device that requires a training set in the contemporary sense. It's an imaging device that uses "a number of algorithms" to produce an image, but it's not a learning algorithm that would have a separate training phase with a distinct training dataset.
9. How the Ground Truth for the Training Set Was Established
As no training set is mentioned (see point 8), there is no information on how its ground truth would have been established.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be part of a single, flowing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2016
Novarix Limited c/o Ms. Deirdre Barrow Senior Regulatory Affairs Consultant Emergo Group 816 Congress Avenue, Suite 1400 Austin, Texas 78701
Re: K152088
Trade/Device Name: Iv-eye Regulation Number: 21 CFR 880.6970 Regulation Name: Liquid crystal vein locator Regulatory Class: I Product Code: KZA Dated: December 10, 2015 Received: December 10, 2015
Dear Ms. Barrow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang 5
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name IV-eye
Indications for Use (Describe)
The IV-eye is a hand-held, non-invasive imaging device that assists medical personnel, trained in vascular access procedure, to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture.
The IV-eye should only be used in conjunction with standard techniques of visualization in assessing and locating veins.
The IV-eye is intended only for skin contact via a disposable single use cover.
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K152088
510(k) Summary
for
IV-eye
1. Submission Sponsor
Novarix Limited The Core Business Centre Milton Hill Abingdon Oxfordshire, OX13 6AB United Kingdom Phone: +44 (0)1235 828 292 Fax: +44 (0)1235 861 041 John Scott, Chief Executive Officer
2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: 00 353 86 8733815 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Deirdre Barrow, Senior Consultant, RA Email: project.management@emergogroup.com
3. Date Prepared
January 5th, 2016
4. Device Identification
| Trade/Proprietary Name: | IV-eye |
|---|---|
| Common/Usual Name: | device, vein location, liquid crystal |
| Classification Name: | Liquid crystal vein locator |
| Classification Regulation: | 21 CFR 880.6970 |
| Product Code: | KZA |
| Device Class: | Class I |
| Classification Panel: | General Hospital |
{4}------------------------------------------------
5. Legally Marketed Predicate Device(s)
| Device | 510(k) | Classification | ||
|---|---|---|---|---|
| Name | No. | Product Code | Regulation | Sponsor |
| IR Viewer | K042679 | KZA | 21 CFR 880.6970 | Infrared ImagingSystem |
| VTS1000(Veinsite) | K101838 | KZA | 21 CFR 880.6970 | VueTek Scientific |
Table 5A: Legally Marketed Predicates
6. Device Description
The IV-eye is a hand-held, battery powered electronic non-invasive imaging device that assists medical personnel, trained in vascular access procedure, to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture. The IV-eye transmits near infrared light into a patient's tissue at a wavelength of 850nm. As the light hits a vascular structure it is absorbed by the hemoglobin in the blood, whereas it passes through other tissue.
The camera in the device captures the light that has passed through the patient and, in identifying the blocked light and applying a number of algorithms, the IV-eye is able to produce an image on its LCD display of the patient's vascular structure directly underneath the device. This appears as a darker color to the contrasting tissue. The picture is updated in real-time and is close to actual size. Trained medical personnel can use the image of the patient's vascular structure to assist them in choosing a suitable vein for cannulation and venipuncture.
The IV-eye is intended to be used only by trained medical personnel to assist them in locating suitable veins for venipuncture and cannulation. It does not differentiate between arteries and veins and should therefore only be used in conjunction with standard techniques of locating veins. Other than regular cleaning and replacement of batteries, the IV-eye requires no routine or preventative maintenance.
The IV-eye includes a single-use cover which prevents the lower casing and wings of the device having direct contact with the patient, and is intended to reduce cross-contamination risk in using the device. The IV-eye is intended only for skin contact via this disposable single use cover.
{5}------------------------------------------------
7. Indication for Use Statement
The IV-eye is a hand-held, non-invasive imaging device that assists medical personnel, trained in vascular access procedure, to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture.
The IV-eye should only be used in conjunction with standard techniques of visualization and palpation in assessing and locating veins.
The IV-eye is intended only for skin contact via a disposable single use cover.
8. Substantial Equivalence Discussion
Table 5A compares the IV-eye to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
One key difference between the proposed device and the predicate device, IR viewer, is that the proposed device and the other predicate, VTS1000 (Veinsite), are handheld and the IR viewer is not hand held. The IR Viewer is a much larger piece of equipment which is on a roll stand comprised of a nIR camera, computer and display screen with touchscreen controls. This technological characteristic difference is not considered to negatively impact the substantial equivalence argument as it does not raise new questions with regard to safety and effectiveness. It has been shown in this 510(k) that the IV-eye device is substantially equivalent in its function as a non-invasive medical device that assists medical personnel, trained in vascular access procedure, to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture.
The proposed device shares the same intended use as the predicate devices but has a more limited indication for use in that the IV-eye is indicated for veins only whereas the predicate devices are indicated for peripheral vessels. The proposed device clearly falls within the intended use of the predicate devices therefore this difference in indications for use does not impact the substantial equivalence argument.
A potential difference exists between the devices in terms of patient contacting materials, however, it has been established in this 510(k) through appropriate biocompatibility testing that the device does not introduce any new concerns regarding safety of the patient contacting materials. Therefore it can be assumed that any potential differences will not impact the substantial equivalence argument.
The IR viewer predicate product allows the option for a sterile single use disposable light source whereas the proposed IV-eye product and the VTS1000 do not. The proposed device contraindicates against the use of the device on broken or damaged skin and the skin contacting element of the IV-eye device is for single use only. It is noted that the IR viewer also includes the option of a non-sterile single use disposable patient contacting element. Therefore the use of single-use non-sterile patient contacting material in the proposed device does not raise new questions with regard to safety and effectiveness and the device is deemed substantially equivalent to the predicate devices.
{6}------------------------------------------------
The proposed device and both predicates are digital units however the proposed IV-eye and VTS1000 device are battery operated units and the predicate device, IR viewer is AC powered. This difference does not impact the device operation and the overall technical and functional capabilities as demonstrated by appropriate bench testing and compliance with appropriate international standards.
In summary, the IV-eye device shares the same intended use, similar indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent to the predicate device. The IV-eye is similar in function to the predicate devices for the modes of operation and use as each device is intended to be used in conjunction with standard techniques of visualization and palpation in assessing and locating veins. The design of the predicates differ somewhat from each other and from that of the proposed device, however each device is non-invasive and utilizes infrared red light of the same wavelength, the proposed device and that of the IR-viewer are patient contacting but both devices use a disposable unit for the patient contacting element.
The differences in design do not affect performance, or raise different questions of safety and effectiveness as demonstrated through bench and clinical testing provided in support of this 510(k).
{7}------------------------------------------------
| Manufacturer | Novarix Limited | Infrared Imaging System | VueTek Scientific | Significant Differences |
|---|---|---|---|---|
| Trade Name | IV-eye | Originally IR viewer now | VTS1000 (Veinsite) | |
| known as Vascular ViewerTM | ||||
| 510(k) Number | n/a | K042670 | K101838 | N/A |
| Product Code | KZA | KZA | KZA | Same |
| Regulation Number | 880.6970 | 880.6970 | 880.6970 | Same |
| Regulation Name | Liquid crystal vein locator. | Liquid crystal vein locator. | Liquid crystal vein locator. | Same |
| Intended Use | Intended Use is to indicate | Intended Use is to indicate | Intended Use is to indicate | Same |
| the location of blood vessels | the location of blood vessels | the location of blood vessels | ||
| Indications for Use | The IV-eye is a hand-held,non-invasive imaging devicethat assists medicalpersonnel, trained in vascularaccess procedure, to identifyand locate suitable peripheralveins for the purposes ofcannulation andvenipuncture.The IV-eye should only beused in conjunction withstandard techniques ofvisualization and palpation inassessing and locating veins.The IV-eye is intended onlyfor skin contact via adisposable single use cover. | The IR Viewer is a non-invasive, electronic devicefor visualization of patientvasculature to supplementnormal, line-of-sight viewingof vascular structures.It is indicated for use inprocedures for inserting aneedle or catheter insuperficial, peripheralvessels. | The VTS 1000 is a non-invasive electronic device toaid in the visualization ofsuperficial vasculature.It is indicated for use duringprocedures requiringvascular or peripheral vesselaccess. | No Significant Difference; the differences between them are:• The IR viewer is not hand held; it is a Digital (solid state) device on a roll stand comprised of a nIR camera, computer and display screen with touchscreen controls.• Both predicates are indicated for peripheral vessels whereas the IV-eye is a subset of this indication as it is indicated for Veins only.• General differences in the choice of |
| Manufacturer | Novarix Limited | Infrared Imaging System | VueTek Scientific | Significant Differences |
| Trade Name | IV-eye | Originally IR viewer nowknown as Vascular Viewer™ | VTS1000 (Veinsite) | |
| Patient ContactingMaterial | Disposable cover: Non-Woven fabric ,Clear LDPEFilm and Hypoallergeniccontact adhesiveDevice cover: Plastic | Foil backing that seals to theskin and hydrogel adhesive | Unknown | No known significantdifference |
| Sterile | No, patient contact areacovered by single usedisposable covers | The single use disposablelight source (DLS) isprovided in both a sterileand non-sterile format | No - device is not a patientcontacting device | No Significant differences;the differences betweenthem are the fact that:• The IR viewerpredicate productallows the option fora sterile single usedisposable lightsource whereas theproposed IV-eyeproduct and theVTS1000 do not.• VTS1000 is not apatient contactingdevice while the IV-eye and the IR viewerinclude a non-sterilesingle use disposablepatient contactingelement. |
| Single-Use | No - disposable cover is asingle use but device is amultiple use device | No - disposable light sourceis a single use but device is amultiple use device | No | No significant difference |
| Manufacturer | Novarix Limited | Infrared Imaging System | VueTek Scientific | Significant Differences |
| Trade Name | IV-eye | Originally IR viewer nowknown as Vascular ViewerTM | VTS1000 (Veinsite) | |
| Shelf Life | 2 Years | Unknown | Unknown | Unknown; predicatedevice shelf life is notavailable |
| Light Source | Infrared light via a LED ofwavelength 850nm | Infrared light via a LED ofwavelength 850nm | Infrared light via a LED ofwavelength 850nm | No difference |
| Battery Operated | Yes - 2 x 1.5v AA alkalinebatteries that are notsuitable for rechargingwhereas the IV-viewer isplugged into the mains | No | Yes sealed 7.4V lithium ionrechargeable battery | All units are digital unitshowever the proposed IV-eye and VTS1000 deviceare battery operatedunits and the predicatedevice is AC powered |
| AC Powered | No | Yes | No | All units are digital unitshowever the proposed IV-eye and VTS1000 deviceare battery operatedunits and the predicatedevice is AC powered |
| Complies with ISO10993-1 | Yes | Unknown | Unknown | No known difference |
| ElectricalPerformanceTesting Passed | Yes | Yes | Yes | No difference |
Table 5A – Device Comparison Chart: Similarities and Differences to predicate device
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
9. Non-Clinical Performance Data
The following testing has been performed successfully to support substantial equivalence:
- . Biocompatibility;
- o ISO 10993-1: 2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing
- ISO 10993-5: 2009, Biological Evaluation of Medical Devices Part 5: Test for In O Vitro Cytotoxicity
- ISO 10993-10: 2010, Biological Evaluation of Medical Devices Part 10: Tests for o Irritation and Delayed-Type Hypersensitivity
- Laser Performance;
- o EN 60825-1: 2007, Safety of Laser Products Part 1: Equipment Classification and Requirement
- Electrical Performance; ●
- IEC 60601-1: 2005, Medical Electrical Equipment Part 1 General Requirements for Basic Safety and Essential Performance
- EN 60601-1-2: 2007, Medical Electrical Equipment Part 1-2: General o Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- Environmental & Transportation - ISTA 1A
- Cleaning
- Mechanical Testing
As part of demonstrating the acceptable performance of the IV-eye and in showing substantial equivalence to the predicate device, Novarix completed a number of tests. The IV-eye meets all the requirements for overall design, biocompatibility, laser and electrical performance and Novarix has confirmed that the output meets the design inputs and specifications. The IV-eye passed all testing stated above as shown by the acceptable results obtained.
The device passed all the testing in accordance with national and international standards including those referenced above and IEC 62366-2007 and IEC 62304:2006. International standards associated with risk management; ISO 14971-2012 and labelling symbols; ISO 15223-1:2012 were also employed to ensure the IV-eye device is safe and effective.
10. Clinical Performance Data
A clinical study was performed on 30 healthy volunteers with different BMI, gender and skin color and age to examine the performance characteristics of the IV-eye with regards to:
-
- Demonstrating the ability of the IV-eye in identifying peripheral veins suitable for venipuncture and cannulation
-
- Quantify the measurement depth capability of the IV-Eye in relation to the depth capability of the predicate VeinSite device from VueTek in order to establish equivalence.
Results of the clinical investigation support the indications for use of the IV-eye to identify and locate suitable peripheral veins for the purposes of cannulation and venipuncture and
{11}------------------------------------------------
confirmed that the device can be considered equivalent to the Veinsite device from VuTek in terms of performance.
11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics, but can be demonstrated to be substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the IV-eye and the predicate devices do not raise any new questions regarding its safety and effectiveness. The IV-eye, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
§ 880.6970 Liquid crystal vein locator.
(a)
Identification. A liquid crystal vein locator is a device used to indicate the location of a vein by revealing variations in the surface temperature of the skin by displaying the color changes of heat sensitive liquid crystals (cholesteric esters).(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9.