(276 days)
3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Custom metal bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodonic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned the amment objectives.
The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
3Shape's Ortho System™ is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analyzinq, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes only Export Model (also called dental casts), Custom Metal Bands (also called metal bands), and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). All devices are to be fabricated from FDA cleared materials.
The device has no patient contact.
The provided FDA 510(k) summary for the 3Shape Ortho System™ does not contain acceptance criteria or a study proving the device meets said criteria in the traditional sense of a performance study with defined metrics (sensitivity, specificity, accuracy, etc.) against a specific ground truth.
Instead, this submission focuses on demonstrating substantial equivalence to predicate devices through a comparison of intended use, technical characteristics, and nonclinical testing that confirms the software acts as intended.
Here's a breakdown of the requested information based on the provided text, highlighting what is implicitly or explicitly stated and what is absent:
1. Table of Acceptance Criteria and Reported Device Performance
As noted, the document does not present a table of specific quantitative acceptance criteria (e.g., "accuracy > 95%") nor does it report performance against such criteria. The "performance" assessment is primarily qualitative, focusing on whether the device's features match or are equivalent to the predicate devices.
| Acceptance Criteria Category | Specific Acceptance Criteria (from document) | Reported Device Performance (from document) |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional Equivalence | The device must demonstrate the same intended uses and technical characteristics as predicate devices. | The 3Shape Ortho System™ Software is found to have the same intended use as the primary predicate OrthoCAD iQ (K082207) and similar technical characteristics to both OrthoCAD iQ and reference predicate Dolphin Imaging (K110430) across supported anatomic areas, patient/case management, study material collection/alignment/measurement/analysis, treatment simulation, and virtual appliance design. A detailed feature comparison table is provided (Page 5). |
| Safety and Effectiveness | The device must be as safe and effective as the predicate devices. | Based on nonclinical testing (software, hardware, and integration V&V) and comparison to predicates, the Ortho System™ is found to be "as safe and effective as the predicate devices." (Page 7). Risk management procedures related to device hazards were also validated. |
| Software Verification and Validation (V&V) | All verification and validation activities should be performed according to FDA guidance specific to software in medical devices. | Software, hardware, and integration V&V testing was performed according to the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Issued on May 11, 2005). All test results were reviewed and approved. |
2. Sample Size Used for the Test Set and the Data Provenance
- Sample Size: Not explicitly stated as a separate "test set" in the context of clinical performance evaluation. The "nonclinical testing" involved software, hardware, and integration V&V, which typically uses various test cases and scenarios rather than a dataset of patient cases for clinical performance evaluation.
- Data Provenance: Not mentioned. Since it's a nonclinical V&V, synthetic data or internal test data would likely be used, but this is not specified. It's not a retrospective or prospective study on patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- This information is not provided. As "clinical testing is not a requirement and has not been performed" (Page 7), there was no need for expert ground truth establishment for a patient-level test set.
4. Adjudication Method for the Test Set
- This information is not provided as there was no patient-level test set requiring ground truth adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- A MRMC comparative effectiveness study was not performed. The document explicitly states: "Clinical testing is not a requirement and has not been performed." (Page 7). The device is a "medical front-end device providing tools" rather than an AI-driven diagnostic aid for human readers.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
- A standalone performance study in the sense of accuracy metrics (sensitivity, specificity, etc.) was not performed. The V&V testing ensures the software functions as designed (e.g., measurements are calculated correctly, models display properly), but not against a "ground truth" to measure diagnostic or predictive performance. The device is a tool to be used by a trained orthodontist.
7. The Type of Ground Truth Used
- For the nonclinical V&V, the "ground truth" would be the expected output or behavior of the software based on its design specifications and requirements. This is not explicitly detailed but is intrinsic to software testing. There's no use of expert consensus, pathology, or outcomes data, as no clinical study was conducted.
8. The Sample Size for the Training Set
- This device is described as a "software system" programmed in Delphi that applies "digital imaging tools." While it mentions "encrypted libraries of the bracket geometry provided by the manufacturers," this refers to pre-programmed data, not a "training set" for a machine learning model. There is no mention of a training set in the context of machine learning, as this does not appear to be an AI/ML device in the modern sense (e.g., deep learning classification).
9. How the Ground Truth for the Training Set Was Established
- Since there's no explicit mention of a training set or machine learning, this information is not applicable or provided.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of a human figure, represented by three stylized profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2016
3Shape A/S Ms. Hanne Nielsen Regulatory Affairs Manager Holmens Kanal 7 DK-1060 Copenhagen DENMARK
Re: K152086
Trade/Device Name: 3Shape Ortho System TM Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: PNN, LLZ Dated: March 15, 2016 Received: March 18, 2016
Dear Ms. Nielsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K152086
Device Name 3Shape Ortho System™
Indications for Use (Describe)
3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Custom metal bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodonic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned the amment objectives.
The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
351
510(K) SUMMARY - Traditional 510(K)
Submitter Information
| A Company Name: | 3Shape A/S | |
|---|---|---|
| B Company Address: | Holmens Kanal 7DK-1060 Copenhagen K | |
| C Company Phone:Company Fax: | +45 7027 2620+45 7027 2621 | |
| D Contact Person: | Hanne NielsenRegulatory Affairs Manager | |
| E Date Summary Prepared: | April 25, 2016 | |
| Device Identification | ||
| A Trade/proprietary Name: | 3Shape Ortho System™ | |
| B Common Name: | Orthodontic Plastic Brackets | |
| C Device Classification Name: | Orthodontic Treatment Planning andDiagnosis Software | |
| C Regulation Number: | 872.5470 | |
| C Classification: | Class II | |
| D Product Code: | PNN (Orthodontic Software): LLZ |
Predicate Device
The 3Shape Ortho System™ Software has the same intended uses and technical characteristics as primary predicate OrthoCAD iQ (K082207) from Cadent Inc. and reference predicate Dolphin Imaging from Patterson Dental Supply (K110430) as listed in "Volume 005, Comparison to Predicate Devices, Table 1: Predicates".
Based on the information and supporting documentation provided, the 3Shape Ortho System™ Software and the primary predicate (OrthoCAD) have same intended use. Both software devices are used by Dental Professionals in orthodontic treatment planning (before, during, after treatment) covering management of patients and models, inspection, 2D and 3D measurement and orthodontic analysis of models, 2D & 3D treatment simulation, as well as virtual appliance preparation, handling and export, and they are both providing device output. Additionally, the indirect bonding functionality of both systems is intended for use with commercially available brackets and wires. Therefore, the 3Shape Ortho System™ Software and the primary predicate (OrthoCAD) are found to be similar in their intended use, supported anatomic areas and the majority of the available features and functionalities.
{4}------------------------------------------------
Indications for Use
3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
Device Description
3Shape's Ortho System™ is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analyzinq, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes only Export Model (also called dental casts), Custom Metal Bands (also called metal bands), and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). All devices are to be fabricated from FDA cleared materials.
The device has no patient contact.
Scientific Concept
The underlying scientific concept of the Ortho System™ is to apply digital imaging tools for in orthodontic case archiving, diagnosis, treatment planning and CAD design of customized appliances.
Virtual positioning of brackets is possible with the use of encrypted libraries of the bracket geometry provided by the manufacturers and available through a dedicated download center in the software.
The system supports the following types of digital data: DICOM, STL, JPG, BMP, PNG.
Summary of the technological characteristics
Ortho System™ is a software only device programmed in Delphi and has the following PC/laptop hardware requirements:
| Item | Minimum Requirements |
|---|---|
| OS: | Windows 7 or 8 64-bit |
| RAM: | 8 GB |
| Monitor Resolution: | 1280x800 or similar |
| Video Card Memory: | 1 GB GeForce |
{5}------------------------------------------------
ssha
| Available HDD Space: | 250 GB |
|---|---|
| CPU: | IntelCore i5 or equivalent |
| Network: | Network Internet connection |
| Mouse: | With the wheel button |
The Ortho System™ Software has the same intended uses and technical characteristics as OrthoCAD iQ (K082207) from Cadent Inc. and Dolphin Imaging from Patterson Dental Supply (K110430):
| Feature name | 3ShapeOrthoSystem™ | PrimarypredicateCadent Inc.OrthoCAD iQ(K082207) | ReferencepredicatePattersonDolphinImaging(K110430) |
|---|---|---|---|
| Supported anatomic areas | Maxilla | Maxilla | Maxilla |
| Mandible | Mandible | Mandible | |
| Intended use | |||
| Managing patient and case base data | Yes | Yes | Yes |
| Collection of study material | Yes | Yes | Yes |
| Alignment of study material | Yes | Yes | Yes |
| Measuring study material | Yes | Yes | Yes |
| Analyzing study material | Yes | Yes | Yes |
| Treatment simulation | Yes | Yes | Yes |
| Virtual appliance design | Yes | Yes | Yes |
| Supported PC formats | Windows | Windows | Windows |
| Managing patient and case base data | |||
| Creating, editing, deleting and copying patient data | Yes | Yes | Yes |
| Creating, editing, deleting and copying case data | Yes | Yes | Yes |
| Feature name | 3ShapeOrthoSystem™ | PrimarypredicateCadent Inc.OrthoCAD iQ(K082207) | ReferencepredicatePattersonDolphinImaging(K110430) |
| Collection of study material | |||
| Surface scan for intra-oral scanner | Yes | Yes | Yes |
| Surface scan from STL file | Yes | Yes | Yes |
| CT image data | DICOM | No | DICOM |
| 2D overlay | PNG, JPG,BMP | JPEG | PNG, JPG,BMP |
| Alignment of study material | |||
| Aligning surface scan and CT image | Yes | No | Yes |
| Aligning cephalometric images | Yes | No | Yes |
| Alignment of 2D overlays (e.g. ideal arch) | Yes | No | Yes |
| Ability to check/adjust DICOM visibility | Yes | No | Yes |
| DICOM scan segmentation | No | No | Yes |
| Measuring study material | |||
| 2D measurement toolbox | Yes | Yes | Yes |
| 3D measurement toolbox | Yes | Yes | Yes |
| Analyzing study material | |||
| Arch shape | Yes | Yes | Yes |
| Wire length | Yes | Yes | Yes |
| Tooth width | Yes | Yes | Yes |
| Bolton | Yes | Yes | Yes |
| Space analysis | Yes | Yes | Yes |
| Overjet/overbite | Yes | Yes | Yes |
| Occlusion map | Yes | Yes | Yes |
| Treatment simulation | |||
| 2D & 3D simulation | Yes | Yes | Yes |
| Feature name | 3ShapeOrthoSystem™ | PrimarypredicateCadent Inc.OrthoCAD iQ(K082207) | ReferencepredicatePattersonDolphinImaging(K110430) |
| Virtual appliance design | |||
| Orthodontic appliance search | Yes | Yes | Yes |
| Orthodontic appliance virtual preparation | Yes | Yes | Yes |
| Orthodontic appliance design | Yes | No | Yes |
| Orthodontic appliance export | Yes | Yes | Yes |
{6}------------------------------------------------
3SHAPE ORTHO SYSTEM™ SOFTWARE 510(K) SUBMISSION
ssha
{7}------------------------------------------------
3sh
Nonclinical Testing
Software, hardware, and integration verification and validation testing was performed in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Issued on May 11, 2005).
The validation suite includes validation of implemented mitigations related to device hazards identified in the risk management procedures.
All test results have been reviewed and approved, showing the Ortho System™ to be substantially equivalent to the predicates.
Clinical Testing
Clinical testing is not a requirement and has not been performed.
Conclusion
Based on a comparison of intended use, indications, principle of operations, features and technical data, and the test results, the Ortho System™ is found to be as safe and effective as the predicate devices. Intended use and performance is found to be substantially equivalent to the Predicate Devices.
§ 872.5470 Orthodontic plastic bracket.
(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.