(186 days)
The TAURUS C1 Dental System and accessories are intended for use in general dental applications by providing the dental practitioner a user interface to control operation of the dental dental devices. The system delivers air, water, vacuum and electricity to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
TAURAUS C1 is a dental chair and units to be used to take for diagnosis and therapy of patients. It consists of the main unit and accessories.
The dental chair is used for patient to sit during the dental diagnosis, treatment and /or operation. The dental unit consists of patient chair, dentist table with instrument holders, assistant table, water unit, arm system, cuspidor (spittoon) and dental light. The dental unit system is used by trained dentists and /or dental assistants to supply air, water, vacuum and electrical power to dental devices and accessories to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
The provided text is a 510(k) summary for the TAURUS C1 Dental System. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving stand-alone clinical accuracy or comparative effectiveness with human readers through typical acceptance criteria studies.
Therefore, the document does not contain the information requested regarding acceptance criteria related to device performance in terms of diagnostic accuracy, sensitivity, specificity, or other performance metrics commonly associated with AI/algorithm-driven devices. It also does not include information on sample sizes for test sets, data provenance, expert ground truth adjudication, MRMC studies, or training set details.
The "performance data" section (Section VII) focuses on non-clinical testing to ensure safety and compliance with relevant standards, rather than clinical performance metrics.
Here's a breakdown of why the requested information is not present based on the provided text:
-
A table of acceptance criteria and the reported device performance: Not applicable in this context. The "acceptance criteria" here are compliance with international safety and performance standards (e.g., IEC, ISO). The "reported device performance" refers to the successful completion of these non-clinical tests.
Acceptance Criterion (Standard Compliance) Reported Device Performance IEC 60601-1 (Basic safety & essential performance) Fulfilled IEC 60601-1-2 (EMC) Fulfilled ISO 7494-1 (Dental units - General req.) Pass ISO 7494-2 (Dental units - Water & air supply) Pass ISO 6875 (Dental patient chair req.) Pass Biocompatibility Materials already cleared for patient contact, similar to predicate device -
Sample size used for the test set and the data provenance: Not applicable. There is no "test set" in the sense of clinical data or images for performance evaluation. The testing involved physical device compliance with standards.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. There is no ground truth established by experts for performance evaluation. The "ground truth" relates to the technical specifications and safety requirements of the device.
-
Adjudication method for the test set: Not applicable.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a dental operative unit, not an AI-driven diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an algorithm-only device.
-
The type of ground truth used: For the purposes of this submission, the "ground truth" is compliance with the technical specifications and safety standards (e.g., a chair must articulate in a certain way, the electrical system must meet specific safety ratings).
-
The sample size for the training set: Not applicable. There is no "training set" as this is not an AI/machine learning device.
-
How the ground truth for the training set was established: Not applicable.
In summary, the provided document details a 510(k) submission for a physical dental operative unit. The "performance data" section refers to engineering and safety compliance testing against recognized international standards, not clinical performance or artificial intelligence algorithm evaluation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 22, 2016
Shinhung Company, Ltd. c/o Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway Houston, Texas 77025
Re: K151996
Trade/Device Name: Taurus C1 Regulation Number: 21 CFR 872.6640 Regulation Name: Operative Dental Unit and Accessories Regulatory Class: I Product Code: EIA Dated: December 9, 2015 Received: December 15, 2015
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Dave Kim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151996
Device Name
TAURUS C1
Indications for Use (Describe)
The TAURUS C1 Dental System and accessories are intended for use in general dental applications by providing the dental practitioner a user interface to control operation of the dental dental devices. The system delivers air, water, vacuum and electricity to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Summary
This 510(k) summary is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510k summary prepared: December 10, 2015
I. SUBMITTER
| Submitter's Name : | Shinhung Co., Ltd. |
|---|---|
| Submitter's HQ Address: | Shinhung Bldg, 450, Cheongpa-ro, Jung-gu,Seoul, 100-858, Korea |
| Factory Address: | 42-25, 27beon-gil, Dongsan-ro, Danwon-gu,Ansan-si, Gyeonggi-Do, 425-852 Korea |
| Submitter's Telephone: | +82(2)6366-2124 |
| Contact person: | Sky Shin( sky@shinhung.co.kr) / Manager |
| Official Correspondent:(U.S. Designated agent) | Dave Kim (davekim@mtech-inc.net) |
| Address: | 8310 Buffalo Speedway, Houston, TX 77025 |
| Telephone: | +713-467-2607 |
| Fax: | +713-583-8988 |
II. DEVICE
| Trade/proprietary name | : | TAURUS C1 |
|---|---|---|
| Common or Usual Name | : | Dental Chair and Units |
| Regulation Name | : | Dental Operative Unit and Accessories |
| Regulation Number | : | 21 CFR 872.6640 (Product Code: EIA) |
| Regulatory Class | : | I Class |
| Prescription Use. |
{4}------------------------------------------------
III. PREDICATE DEVICE
| Primary Manufacturer | : | A-DEC, INC |
|---|---|---|
| Device Name | : | A-DEC 200 DENTAL SYSTEM |
| 510(k) Number | : | K102234 (Decision Date – Jan. 13, 2012) |
| Regulation Name | : | Dental Operative Unit and Accessories |
| Regulation Number | : | 21 CFR 872.6640 (Product Code: EIA) |
| Regulatory Class | : | I Class |
Prescription Use
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
DEVICE DESCRIPTION IV.
TAURAUS C1 is a dental chair and units to be used to take for diagnosis and therapy of patients. It consists of the main unit and accessories.
The dental chair is used for patient to sit during the dental diagnosis, treatment and /or operation. The dental unit consists of patient chair, dentist table with instrument holders, assistant table, water unit, arm system, cuspidor (spittoon) and dental light. The dental unit system is used by trained dentists and /or dental assistants to supply air, water, vacuum and electrical power to dental devices and accessories to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
{5}------------------------------------------------
INDICATIONS FOR USE: V.
The TAURUS C1 Dental System and accessories are intended for use in general dental applications by providing the dental practitioner a user interface to control operation of the dental chair and attached devices. The system delivers air, water, vacuum and electricity to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE
| Model | A-dec 200 (K102234) | TAURUS C1 for 510K | ||
|---|---|---|---|---|
| Indications for Use | The A-dec 200 DentalSystem and accessories areintended for use in generaldental applications byproviding the dentalpractitioner a user interfaceto control operation of thedental chair and attacheddental devices. The systemdelivers air, water, vacuumand electricity to allow thedental practitioner anintuitive control center forall common and normalpatient treatment proceduresperformed in the dentaloperatory | The TAURUS C1 DentalSystem and accessories areintended for use in generaldental applications byproviding the dentalpractitioner a user interfaceto control operation of thedental chair and attacheddental devices. The systemdelivers air, water, vacuumand electricity to allow thedental practitioner anintuitive control center forall common and normalpatient treatment proceduresperformed in the dentaloperatory. | ||
| Construction | CHAIR / UNIT / DOCTORTABLE /ASSISTANTTABLE / OPERATINGLIGHT / FOOTCONTROLLER / STOOL | CHAIR / UNIT/ DOCTORTABLE /ASSISTANTTABLE / OPERATINGLIGHT / FOOTCONTROLLER / STOOL | ||
| Power Supply /Frequency | AC100V/110V/120V,AC210/220/230V , 50/60Hz | AC100V/110V/115V,AC210/220/230V, 50/60Hz | ||
| PowerShockProtection | Class 1 | Class 1 | ||
| Electric Protection | B Type | B Type | ||
| EN 60601-1-2 | EN 60601-1-2 | |||
| EMC Standard | IEC 61000-3-2 | IEC 61000-3-2 | ||
| IEC 61000-3-3 | IEC 61000-3-3 | |||
| DentalunitsStandard | ISO 7494-1ISO 7494-2 | ISO 7494-1ISO 7494-2 | ||
| Dental Patientchair Standard | ISO 6875 | ISO 6875 | ||
| Safety Standard | EN 60601-1 | EN 60601-1 | ||
| Chair Operatingsystem | Hydraulic system | Hydraulic system | ||
| Description(Comparativetestingconfirmedequivalenceto the notedpredicate) | UNIT/CHAIRform type | Over-arm Contour type | Over-arm Contour type | |
| Vacuum system | Central Vacuum system | Central Vacuum system orAir Vacuum system | ||
| 3-Way Syringe | Adjusting the Water / Air /Spray function is used in thesyringe button. | Adjusting the Water / Air /Spray function is used in thesyringe button. | ||
| Cuspidor(Spittoon) | Spittoon bowl / Rinsing pipe/ Tumbler filler | Spittoon bowl / Rinsing pipe/ Tumbler filler | ||
| Water System | City Water / Distilled Water | City Water / Distilled Water | ||
| Comparative testingconfirmed equivalence tothe noted predicate | ||||
| Light headstructure | 3-axis head adjustmentmethod | 3-axis head adjustmentmethod | ||
| OperatingLight | ON/OFF control | Auto and manual ON/OFF | Auto and manual ON/OFF | |
| Headrest | Double-articulatingheadrest | Double-articulatingheadrest | ||
| Safety switchfunction | Chair lift arm and auxiliaryarm, 2 safety switches | Chair lift arm and auxiliaryarm, 2 safety switches | ||
| The safety switch stops thechair operation. | The safety switch stops thechair operation. | |||
| chair | chair | |||
| Programmablechair position | Programmableposition:Total 4 positions | Programmableposition:Total 5 positions | ||
| Deliverysystem | Handpiece holder | 5 angle adjustable holders | 6 integrated rotation typeholder | |
| Handl e | Adjustable handle | Fixed handle | ||
| Touchpad | Membrane panel | Membrane Panel | ||
| Handpiece waterand air cooling | Manually adjustablehandpiece water / air cooling | Manually adjustablehandpiece water / air cooling | ||
| Flexarm tension | Control the tension to adjustthe shift of flexarm rotationflow. | Control the tention to adjustthe shift of TABLE ARMrotation. | ||
| Accessoriestools | Accessories holder | HVE, SE, 3-WAY syringeattached | HVE, SE, 3-WAYsyringeattached | |
| Holder structure | Rotation and individualholder | Fixed, one-piece holder | ||
| Supportcenter | Cuspidor | Self-contained 2 liter waterbottle system | Attach/detachable GLASS,disassemble for cleaning | |
| Sol i ds c ol lec tor | Equipped | Equipped | ||
| Distilled waterreservoir | Equipped | Equipped | ||
| Cup water system | Manual water systemPress the cupfill button for atimed operation. Press andhold the cupfill button formanual operation. | Automatic and manual watersystem.Place the cup at the specifiedlocation for a timedoperation. | ||
| Accessoriesattached tothe device: | handpiece / Low speed handpiece / Scaler / Curing light / Dental Chair / 3-waysyringe | |||
| OperationMethod | Control panel / Assistant control panel / Foot controller |
{6}------------------------------------------------
{7}------------------------------------------------
VII. PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination.
Biocompatibility testing:
TAURUS C1 dental operative unit and accessories use the materials already cleared for patient contact components such as patient chair seat, arm rest, 3 way syringe,
{8}------------------------------------------------
vacuum syringe, and saliva ejector which are similar to the predicate device.
Non Clinical testing:
IEC 60601-1 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance. The requirements of specified standards were fulfilled.
IEC 60601-1-2 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance (collateral standards: electromagnetic compatibility. The requirements of specified standards were fulfilled. ISO 7494-1 Test was performed for general requirements and test methods: Pass ISO 7494-2 Test was performed for general requirements for water and air supply: Pass ISo6875 test was performed for general, electrical, and mechanical requirements: Pass
VIII. CONCLUSIONS
There are no significant differences between the TAURUS C1 and the predicate device.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Shinhung Co, Ltd. concludes that TAURUS C1 is substantially equivalent in comparison with Adec-200, the predicate device as described herein.
§ 872.6640 Dental operative unit and accessories.
(a)
Identification. A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.(b)
Classification. Class I (general controls). Except for dental operative unit, accessories are exempt from premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.