(261 days)
The Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing are used in conjunction with the CrossFlow Integrated Arthroscopy Pump (K123441) to provide fluid distension and irrigation of the knee, shoulder, hip, elbow, ankle, and wrist joint cavities during diagnostic and operative arthroscopic procedures.
The Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing (hereafter referred to as "proposed device") is an accessory to the FDA-cleared CrossFlow Integrated Arthroscopy Pump, cleared under K123441, which provides fluid to the surgical site. CrossFlow Day-Use Inflow Cassette Tubing includes an inflow cassette with spikes for irrigation fluid units and connector. The Patient-Use Tubing includes a connector that attaches the Patient-Use Tubing to Day-Use Tubing, and luer-lock connector that attaches the Patient-Use Tubing to an inflow cannula. The Stryker CrossFlow Day-Use Inflow Cassette Tubing is a 24-hour multi-patient device that remains in the arthroscopy pump throughout a day of cases (no more than 10 cases, 8 hours of active use, or 24 hours from the point of first use). Day-Use Inflow Cassette Tubing provides an alternate to complete replacement of the irrigation tubing after each patient. The backflow check valve of the Patient-Use Tubing prevents contaminated fluid from reaching the Day-Use Tubing and permits its use for a day of cases. Patient-Use Tubing is a single-use device that is discarded after every case.
The provided text describes the regulatory clearance for the Stryker® CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing. This device is an irrigation tubing set used during arthroscopic procedures. The documentation focuses on demonstrating substantial equivalence to a predicate device, as required for a 510(k) submission.
The "acceptance criteria" and "device performance" in this context refer to the successful completion of various bench performance tests designed to ensure the safety and effectiveness of the device, particularly regarding its multi-patient use claim and prevention of cross-contamination.
Here's the breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for each test are implicitly that the device "Pass"ed the test, indicating it met the design specifications and applicable standards. The specific quantitative acceptance criteria are not detailed in the provided summary but are generally implied by the nature of the tests.
| Test | Description | Reported Device Performance |
|---|---|---|
| Microbial Ingress | Confirms ability to maintain sterility of inner tubing during lifetime of device | Pass |
| Day-Use Inflow Cassette Tubing Disinfection | Confirms safety of the device over multiple uses by achieving disinfection of the Day-Use Tubing to Patient-Use Tubing connection | Pass |
| Backflow Valve | Confirms no cross contamination occurs between cases by analyzing if microorganisms can breach the backflow valve | Pass |
| Backflow Valve Limit of Quantification | Measures limit of quantification for backflow valve test method | Pass |
| Backflow Valve Time to Close | Ensures adequate closure of backflow valve to prevent backwards movement of fluid | Pass |
| Backflow Valve Failure | Confirms complete backflow valve failure is not a risk to the patient | Pass |
| Regurgitation | Ensures fluid moving in the opposite direction to the normal direction does not pose risk of cross contamination | Pass |
| General requirements and performance | Verified all components against their design specifications | Pass |
| Viral Pathogens | Confirms no cross contamination occurs between cases by analyzing if viral pathogens can breach the backflow valve based on model system | Pass |
| Biocompatibility | Verified the biocompatibility of all patient contacting materials in accordance with ISO 10993-1:2009 | Pass |
| Sterilization | Ensured the EO sterilization process for all single-use, sterile components in accordance with ISO 11135-1:2014 to a sterility assurance level (SAL) of 10-6 and verified that the EO and ECH residuals are within the limits defined in ISO 10993-7:2008 | Pass |
| Real Time Aging | Verified functionality of all single-use, sterile components and integrity of all packaging after claimed shelf life | Pass |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes (number of units tested) for each of the performance tests listed. It states that "Bench performance testing was conducted to ensure that the device functioned as intended and met design specifications and acceptance criteria based off of predicate data." It further indicates that "Full test summaries and results can be found in Section 18 - Performance Testing -Bench," but this section is not included in the provided text.
The data provenance is from bench testing conducted by the manufacturer, Stryker Corporation. It is retrospective in the sense that it's part of a submission for an already developed device. The country of origin for the data is implicitly the USA, where Stryker is headquartered and the submission was made.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This type of information is not applicable to this submission. The "ground truth" for this device is established by engineering specifications, recognized standards (like ISO 10993-1, ISO 11135, ISO 10993-7), and the ability of the device to perform its intended function without failure, particularly regarding preventing contamination. There's no clinical trial or diagnostic outcome being assessed that would require expert consensus on clinical findings.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods like "2+1" or "3+1" are used in clinical trials, especially for imaging or diagnostic devices, where multiple readers interpret data and a final "ground truth" decision is made through a consensus process. For bench testing of a medical device like an irrigation tubing set, the results are typically objectively measured against predefined engineering specifications (e.g., presence/absence of microbial growth, flow rate, pressure, integrity). There's no subjective interpretation requiring an adjudication process.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices, particularly those involving image interpretation where AI assistance could augment human reader performance. This device is a tubing set for arthroscopy, not an AI-powered diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable. The device described is a physical medical device (tubing), not an algorithm or AI system. Therefore, standalone algorithm performance is not a concept related to this submission.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The "ground truth" in this context is based on predefined engineering specifications, recognized international standards (e.g., ISO for biocompatibility and sterilization), and objective physical/microbiological measurements. For instance:
- Microbial Ingress/Backflow Valve tests: Ground truth is the absence of microbial or viral breach.
- Biocompatibility: Ground truth is compliance with ISO 10993-1.
- Sterilization: Ground truth is achieving a sterility assurance level (SAL) of 10-6 and meeting ISO 10993-7 residuals limits.
- Functional tests (Regurgitation, Time to Close, General requirements): Ground truth is successful operation within design parameters.
8. The Sample Size for the Training Set
This is not applicable. This device is a physical product and not an AI/algorithm-based device that would require a "training set" for machine learning.
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the same reason as above (not an AI/algorithm device).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping heads suggesting community and connection. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2016
Stryker Corporation Somi Ekwealor Regulatory Affairs Analyst 5900 Optical Ct San Jose, California 95138
Re: K151932
Trade/Device Name: Stryker Crossflow Day-Use Inflow Cassette Tubing and Patient-Use Tubing Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: February 25, 2015 Received: February 29, 2016
Dear Somi Ekwealor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
к151932
Device Name: Stryker® CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing
510(k) Number (if known): K151932
Indications for Use:
The Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing are used in conjunction with the CrossFlow Integrated Arthroscopy Pump (K123441) to provide fluid distension and irrigation of the knee, shoulder, hip, elbow, ankle, and wrist joint cavities during diagnostic and operative arthroscopic procedures.
| Prescription Use X | AND/OR | Over-The-Counter Use |
|---|---|---|
| (Part 21 CFR 801 Subpart D) | (Part 21CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
1. General Information
| 510(k) Sponsor | Stryker Endoscopy |
|---|---|
| Address | 5900 Optical CourtSan Jose, CA 95138 |
| FDA Registration Number | 2936485 |
| Correspondence Person | Mr. Somi Ekwealor, MSRS, RACRegulatory Affairs AnalystStryker Endoscopy |
| Contact Information | Email: somi.ekwealor@stryker.comPhone: (408) 754-2356Fax: (408) 754-2507 |
2. Device Identification
Proposed Device:
| Proprietary Name | Stryker® CrossFlow Day-Use Inflow CassetteTubing and Patient-Use Tubing |
|---|---|
| Common Name | Irrigation Tubing Sets |
| Classification Name | Arthroscope |
| Regulation Number | 21 CFR 888.1100 |
| Product Code | HRX |
| Regulatory Class | II |
Predicate Devices:
| Proprietary Name | Arthrex Continuous Wave III Arthroscopy Pump(Tubing Only) |
|---|---|
| Common Name | Pump |
| Premarket Notification | K024291 |
| Classification Name | Arthroscope |
| Regulation Number | 21 CFR 888.1100 |
| Product Code | HRX |
| Regulatory Class | II |
Reference Device:
| Proprietary Name | Stryker CrossFlow Arthroscopy Pump |
|---|---|
| Common Name | Pump |
| Premarket Notification | K123441 |
| Classification Name | Arthroscope |
{4}------------------------------------------------
| Regulation Number | 21 CFR 888.1100 |
|---|---|
| Product Code | HRX |
| Regulatory Class |
A reference device is used in this premarket notification to support scientific methodology because the reference device and proposed device share materials and test methods. The proposed device is also an accessory to, and indicated for use with the reference device. The reference device is discussed throughout this premarket notification when discussing the device description, indications for use, and performance testing.
3. Device Description
The Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing (hereafter referred to as "proposed device") is an accessory to the FDA-cleared CrossFlow Integrated Arthroscopy Pump, cleared under K123441, which provides fluid to the surgical site. CrossFlow Day-Use Inflow Cassette Tubing includes an inflow cassette with spikes for irrigation fluid units and connector. The Patient-Use Tubing includes a connector that attaches the Patient-Use Tubing to Day-Use Tubing, and luer-lock connector that attaches the Patient-Use Tubing to an inflow cannula. The Stryker CrossFlow Day-Use Inflow Cassette Tubing is a 24-hour multi-patient device that remains in the arthroscopy pump throughout a day of cases (no more than 10 cases, 8 hours of active use, or 24 hours from the point of first use). Day-Use Inflow Cassette Tubing provides an alternate to complete replacement of the irrigation tubing after each patient. The backflow check valve of the Patient-Use Tubing prevents contaminated fluid from reaching the Day-Use Tubing and permits its use for a day of cases. Patient-Use Tubing is a single-use device that is discarded after every case.
4. Indication for Use
The Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing are used in conjunction with the CrossFlow Integrated Arthroscopy Pump (K123441) to provide fluid distension and irrigation of the knee, shoulder, hip, elbow, ankle, and wrist joint cavities during diagnostic and operative arthroscopic procedures.
5. Comparison of Technological Characteristics with the Predicate Device
The Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing and predicate devices tubing are similar in technological characteristics and design.
Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing are substantially equivalent to the predicate devices. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the proposed device compares to the predicate device is
{5}------------------------------------------------
provided below. A complete comparison of technological characteristics between the proposed and predicate devices is provided in Section 12 – Substantial Equivalence Discussion.
| Characteristics | Proposed Devices | Predicate Devices (K024291) |
|---|---|---|
| Inner Diameter ofTubing (mm) | Day-Use Tubing: 6.20±0.05Patient-Use Tubing: 6.20±0.05 | Arthrex Reduce Tubing, PumpTubing (AR-6411): 4.83Reduce Tubing, Patient Tubing(AR-6241): 6.35 |
| Tubing Length (mm) | Day-Use Tubing: 1200Patient-Use Tubing: 2800 | Arthrex Reduce Tubing, PumpTubing (AR-6411): 2310Reduce Tubing, Patient Tubing(AR-6241): 2347 |
| Pressure SensingMechanism | Internal pressure dampeningmembrane | External pressure chamber |
| Tubing CassetteLoaded | Yes | No |
| Cassette Material | ABS | N/A |
| RFID | Yes | No |
| Disinfection Method | Included and in accordance withGuidance for Industry and FDAStaff - Reprocessing MedicalDevices in Health Care Settings:Validation Methods and Labeling –March 17, 2015 | None |
| Device Lifetime Claim | Single day's cases (no more than10 cases, 8 hours of active use, or24 hours from point of first use) | 1 surgical day |
Differences between the Proposed and Predicate Devices
Tubing Specifications:
The difference in inner diameter is negligible as verification testing provided in this submission demonstrates that this difference has no effect on the safety or effectiveness of the device.
The difference in length of the tubing is also negligible as verification testing provided in this submission demonstrates that the difference has no effect on the safety or effectiveness of the device.
{6}------------------------------------------------
These minor differences in the proposed device raise no new questions of safety or effectiveness, therefore the proposed device is substantially equivalent to the predicate device.
Pressure Sensing Mechanism:
The proposed device aids the CrossFlow Integrated Arthroscopy pump in measuring pressures in the joint space via a pressure dampening chamber housed in the cassette of Day-Use Inflow Cassette Tubing. The predicate device contains an external pressure chamber that acts similarly to the pressure dampening chamber in sensing pressure fluctuations.
This minor difference in how pressure is measured in the proposed device raises no new questions of safety or effectiveness, therefore the proposed device is substantially equivalent to the predicate device.
Tubing Cassette
Both devices are accessories to irrigation pumps that operate on the peristaltic principle where roller wheels are spun against soft tubing to create the flow of fluid within the tubing. The devices are different in how the tubing interfaces with the roller wheels of the pump. The predicate device requires the user to manually place the tubing inside the console and around the roller wheels while the proposed device simply requires the user to insert a cassette containing the peristaltic tubing into the cassette.
The cassette includes a Radio-frequency Identification (RFID) component that allows the CrossFlow Arthroscopy Pump to recognize tubing cassettes as either new or expired. The predicate device does not include any technology that allows the pump to differentiate between new tubing and tubing that has been used after the intended product lifetime (one surgical day).
This minor difference in the proposed device raises no new questions of safety or effectiveness, therefore the proposed device is substantially equivalent to the predicate device.
Disinfection
A disinfection method was developed for the proposed device to prevent cross-contamination of the connector between cases. Inadvertent and undetected contamination of the device is prevented by disinfecting the connector after each case. Complete disinfection are provided in Section 13 - Labeling. The predicate device labeling does not contain disinfection instructions to ensure safe reprocessing of the device.
This difference in the proposed device raises no new questions of safety or effectiveness, therefore the proposed device is substantially equivalent to the predicate device.
{7}------------------------------------------------
Device Lifetime
The proposed device can be used for a single day's cases - no more than 10 cases, 8 hours of active use, or 24 hours from point of first use, which is based on data from a high case-volume day at a high volume arthroscopy surgery center. This claim is supported by the performance testing. Full test summaries and results can be found in Section 18 - Performance Testing -Bench. The predicate device is also intended to be used for a single day's cases or one surgical day however they provide no definition of one surgical day. Based on the data obtained, one surgical day will not exceed 8 hours of active use or 10 cases performed within a 24 hour period.
This difference in the proposed device labeling raises no new questions of safety or effectiveness, therefore the proposed device is substantially equivalent to the predicate device.
Any technological differences between the Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing and the predicate devices do not raise new questions of safety or effectiveness.
6. Performance Testing
The Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing were evaluated in accordance with internal design specification and with the applicable performance standards as required by the risk analysis. The assessment of non-clinical performance data supports the substantial equivalence determination. The following non-clinical tests were conducted and are summarized in this submission:
Biocompatibility was assessed in accordance with ISO 10993-1:2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" and related collateral standards for direct and indirect patient contacting materials (see Section 15 - Biocompatibility).
Sterilization was assessed in accordance with ISO 11135:2014 - "Sterilization of health-care products - Ethylene Oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices". Sterilization residuals were assessed in accordance with ISO 10993-7:2008 - "Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals" (see Section 14 - Sterilization and Shelf Life).
Bench performance testing was conducted to ensure that the device functioned as intended and met design specifications and acceptance criteria based off of predicate data. The following performance tests were executed in support of the substantial equivalence determination:
{8}------------------------------------------------
| Test | Description | Results |
|---|---|---|
| MicrobialIngress | Confirms ability to maintain sterility of inner tubing duringlifetime of device | Pass |
| Day-Use InflowCassette TubingDisinfection | Confirms safety of the device over multiple uses byachieving disinfection of the Day-Use Tubing to Patient-Use Tubing connection | Pass |
| Backflow Valve | Confirms no cross contamination occurs between cases byanalyzing if microorganisms can breach the backflow valve | Pass |
| Backflow ValveLimit ofQuantification | Measures limit of quantification for backflow valve testmethod | Pass |
| Backflow ValveTime to Close | Ensures adequate closure of backflow valve to preventbackwards movement of fluid | Pass |
| Backflow ValveFailure | Confirms complete backflow valve failure is not a risk tothe patient | Pass |
| Regurgitation | Ensures fluid moving in the opposite direction to thenormal direction does not pose risk of cross contamination | Pass |
| Generalrequirementsand performance | Verified all components against their design specifications | Pass |
| ViralPathogens | Confirms no cross contamination occurs between cases byanalyzing if viral pathogens can breach the backflow valvebased on model system | Pass |
| Biocompatibility | Verified the biocompatibility of all patient contactingmaterials in accordance with ISO 10993-1:2009 | Pass |
| Sterilization | Ensured the EO sterilization process for all single-use,sterile components in accordance with ISO 11135-1:2014 toa sterility assurance level (SAL) of 10-6 and verified that theEO and ECH residuals are within the limits defined in ISO10993-7:2008 | Pass |
| Real TimeAging | Verified functionality of all single-use, sterile componentsand integrity of all packaging after claimed shelf life | Pass |
Test results obtained verified the safety and effectiveness of the device per design specifications and applicable standards when compared to bench testing. The test results support the substantial equivalence of the Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing compared to the predicate devices. Full test summaries and results can be found in Section 18 – Performance Testing – Bench.
{9}------------------------------------------------
7. Nonclinical Testing
When the proposed device is in use certain components will indirectly contact the patient. The materials used were analyzed in accordance with ISO 10993-1:2009. Test results show that the device meets all applicable requirements.
The nonclinical tests conducted and referenced in this premarket notification show that the proposed device meets or exceeds the performance requirements, and demonstrate the device is as safe, as effective, and performs as well as or better than the predicate device based on competitor test data.
8. Conclusion
The proposed and predicate devices are similar in design and performance specifications. Clinical testing was not required to demonstrate substantial equivalence. Based on the evaluation of the indication for use, technological characteristics, performance testing, nonclinical testing, and comparison to the predicate devices, the Stryker CrossFlow Day-Use Inflow Cassette Tubing and Patient-Use Tubing raises no new questions of safety and effectiveness when compared to the predicate device and is thus considered substantially equivalent.
§ 888.1100 Arthroscope.
(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.