(158 days)
The TRI® Dental Implant System is intended for placement in the bone of the maxillary or mandibular arch for the rehabilitation of edentulous and partially edentulous patients. TRI® Dental Implant System allows for one and two-stage surgical procedures. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading.
TRI Dental Implant System 6.5 mm implants are intended for delayed loading only.
The TRI-Narrow and TRI-Vent implants are root form endosseous dental implants. They are available with two collar options, blasted or machined. TRI-Narrow is available in one implant diameter (3.3 mm) and three lengths (11.5, 13, and 16 mm). TRI-Narrow abutments include all abutment options with the exception of the Gingiva Former and the angled Screw Retained Abutment.
TRI-Vent is available in three implant diameters (3.75, 4.1 and 4.7 mm) and six lengths (6.5, 8, 10, 11.5, 13, and 16 mm). Only the 4.1 mm and 4.7 mm diameter implant bodies are available in the 6.5 mm length. All three TRI-Vent implant diameters have the same platform diameter (3.5 mm). TRI-Vent abutments include Healing Collar, Straight Temporary PEEK Abutment, Straight and 20° Angled Abutments, Straight and 30° Angled Screw Retained Abutment, PEEK Gingiva Former, Straight and 15° Angled Contoured Abutment, Gold-castable Abutment, Screw- retained Abutment, Ball Abutment, and applicable abutment fixation screws.
The TRI-Octa implants are root form endosseous dental implants. They have a pink, transgingival collar. TRI-Octa is available in three implant diameters (3.75, 4.1 and 4.7 mm) and six lengths (6.5, 8, 10, 11.5, 13, and 16 mm). Only the 4.1 mm and 4.7 mm diameter implant bodies are available in the 6.5 mm length. All three TRI-Octa implant diameters have the same platform diameter (4.8 mm). TRI-Octa abutments include Healing Collar, Straight PEEK Temporary Abutment, Straight and 20° Angled Abutments, Straight Screw Retained Abutment, Ball Abutment, and applicable abutment fixation screws.
Here's a breakdown of the acceptance criteria and study information for the TRI® Dental Implant System, based on the provided text:
Preamble: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to existing legally marketed devices. It typically describes non-clinical testing performed to show that the new device performs as safely and effectively as predicates, rather than extensive clinical efficacy studies with specific acceptance criteria in the clinical sense. Therefore, the "acceptance criteria" here refer more to the performance standards the device was tested against to prove equivalence, particularly in mechanical and biological compatibility. Clinical outcomes data (e.g., success rates in patients) often come from post-market surveillance or larger clinical trials not detailed in a 510(k) submission.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria/Standard | Reported Device Performance |
|---|---|---|
| Sterilization | ISO 11137-1, ISO 11137-2, ISO 17665-1, ISO 17665-2 | Sterilization validation performed (demonstrates substantial equivalence to predicate processes). |
| Biocompatibility | ISO 10993-1, ISO 10993-5 | Biocompatibility evaluation and testing performed (demonstrates substantial equivalence to predicate materials). |
| Mechanical Performance | ISO 14801: Dentistry – Implants – Dynamic fatigue test for endosseous dental implants | Fatigue testing demonstrates increased abutment angulation is substantially equivalent to identified predicates with respect to performance. |
| Dimensional Analysis | N/A (Engineering dimensional analysis) | Engineering dimensional analysis performed (implied to match or be acceptably similar to predicates). |
| Material Composition | Titanium Alloy, CPTi, PEEK, Gold alloy | "Slight differences in material composition... are demonstrated substantially equivalent by biocompatibility testing." The device utilizes Titanium Alloy and other materials consistent with predicates. |
| Surface Area (Osseointegration) | N/A (Surface area analysis) | Surface area analysis demonstrates smallest diameter/shortest length combinations are substantially equivalent to reference predicate K092035. |
| Overall Equivalence | Same intended use, operating principle, basic design, similar/same materials, similar packaging/sterilization process as predicates. | Concluded that the data demonstrate substantial equivalence to predicate devices. |
Detailed Study Information:
-
Sample size used for the test set and the data provenance:
- The document does not explicitly state sample sizes for each non-clinical test (e.g., number of implants tested for fatigue). It only mentions that "Non-clinical testing data submitted, referenced, or relied upon" was provided.
- Data provenance: Not specified (e.g., country of origin, retrospective/prospective). Given it's non-clinical lab testing against ISO standards, it would be prospective data generated for the submission.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in this context. The "ground truth" for non-clinical performance data is defined by the technical specifications outlined in the ISO standards themselves (e.g., a specific fatigue load, or a cytotoxicity response in a cell culture). There isn't typically expert consensus to establish a "ground truth" for these types of engineering and biocompatibility tests.
-
Adjudication method for the test set:
- Not applicable as this is non-clinical testing against ISO standards, not a study involving human readers or clinical adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This document describes a dental implant system, not an AI-powered diagnostic device. No MRMC studies were conducted or reported.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device (dental implant), not an algorithm.
-
The type of ground truth used:
- For non-clinical testing: The "ground truth" is defined by the physical, chemical, and biological properties and performance metrics specified in the referenced ISO standards (e.g., mechanical strength requirements, cytotoxicity levels, sterility assurance levels). It's based on established scientific and engineering criteria.
- For demonstrating equivalence: The ultimate "ground truth" for the 510(k) process is the safety and effectiveness of the legally marketed predicate devices.
-
The sample size for the training set:
- Not applicable. This is a physical medical device, not a machine learning algorithm requiring a "training set."
-
How the ground truth for the training set was established:
- Not applicable, for the same reason as above.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2015
Tri Dental Implants Int. AG c/o Ms. Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K151916
Trade/Device Name: TRI® Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: November 13, 2015 Received: November 16, 2015
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number
Device Name
TRI® Dental Implant System
Indications for Use (Describe)
The TRI® Dental Implant System is intended for placement in the bone of the maxillary or mandibular arch for the rehabilitation of edentulous and partially edentulous patients. TRI® Dental Implant System allows for one and twostage surgical procedures. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading.
TRI Dental Implant System 6.5 mm implants are intended for delayed loading only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
TRI Dental Implants Int. AG
TRI® Dental Implant System
December 18, 2015
ADMINISTRATIVE INFORMATION
| Manufacturer Name | TRI Dental Implants Int. AG |
|---|---|
| Lindenstrasse 14 | |
| Baar CH-6340 Switzerland | |
| Telephone: +41 32 510 1606 | |
| Fax: +41 32 510 1601 | |
| Official Contact | Sandro Venanzoni, Chief Technology Officer |
| Representative/Consultant | Linda K Schulz, BSDH, RDH |
| Floyd G. Larson, MS, MBA PaxMed International, LLC | |
| 12264 El Camino Real, Suite 400 | |
| San Diego, CA 92130 | |
| Telephone: +1 (858) 792-1235 | |
| Fax: +1 (858) 792-1236 | |
| Email: lschulz@paxmed.comflarson@paxmed.com |
DEVICE NAME AND CLASSIFICATION
Reviewing Branch
| Trade/Proprietary Name | TRI® Dental Implant System |
|---|---|
| Common Name | Endosseous dental implant |
| Classification Name | Endosseous Dental Implant |
| Classification Regulations | 21 CFR 872.3640, Class II |
| Product Code | DZE, NHA |
| Classification Panel | Dental Products Panel |
The primary predicate for the TRI®-Narrow and TRI®-Vent implants is: K101880, Tapered Screw-Vent® T Implant and Tapered Screw-Vent® P Implant, Zimmer Dental, Inc.
The primary predicate for the TRI®-Octa implants is: K012757, ITT® Dental Implant System, Straumann USA.
The primary predicate for TRI®-Dental Implant System abutments is: K083496, CAMLOG Implant System Modified Implants and Abutments, Altatec GmbH.
Dental Devices Branch
{4}------------------------------------------------
Reference Predicates:
K101545, Genesis Implant System, Keystone Dental; K113779, CONELOG Implant System, Altatec GmbH; K072570, NobelActive Multi-Unit Abutment, Nobel Biocare AB; and K092035, Bicon Implants with a 2.5 mm Internal Connection, Bicon, LLC.
INTENDED USE
The TRI® Dental Implant System is intended for placement in the bone of the maxillary or mandibular arch for the rehabilitation of edentulous and partially edentulous patients. TRI® Dental Implant System allows for one and two-stage surgical procedures. When a one-stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and with appropriate occlusal loading.
TRI Dental Implant System 6.5 mm implants are intended for delayed loading only.
DEVICE DESCRIPTION
The TRI-Narrow and TRI-Vent implants are root form endosseous dental implants. They are available with two collar options, blasted or machined. TRI-Narrow is available in one implant diameter (3.3 mm) and three lengths (11.5, 13, and 16 mm). TRI-Narrow abutments include all abutment options with the exception of the Gingiva Former and the angled Screw Retained Abutment.
TRI-Vent is available in three implant diameters (3.75, 4.1 and 4.7 mm) and six lengths (6.5, 8, 10, 11.5, 13, and 16 mm). Only the 4.1 mm and 4.7 mm diameter implant bodies are available in the 6.5 mm length. All three TRI-Vent implant diameters have the same platform diameter (3.5 mm). TRI-Vent abutments include Healing Collar, Straight Temporary PEEK Abutment, Straight and 20° Angled Abutments, Straight and 30° Angled Screw Retained Abutment, PEEK Gingiva Former, Straight and 15° Angled Contoured Abutment, Gold-castable Abutment, Screw- retained Abutment, Ball Abutment, and applicable abutment fixation screws.
The TRI-Octa implants are root form endosseous dental implants. They have a pink, transgingival collar. TRI-Octa is available in three implant diameters (3.75, 4.1 and 4.7 mm) and six lengths (6.5, 8, 10, 11.5, 13, and 16 mm). Only the 4.1 mm and 4.7 mm diameter implant bodies are available in the 6.5 mm length. All three TRI-Octa implant diameters have the same platform diameter (4.8 mm). TRI-Octa abutments include Healing Collar, Straight PEEK Temporary Abutment, Straight and 20° Angled Abutments, Straight Screw Retained Abutment, Ball Abutment, and applicable abutment fixation screws..
PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to ISO 11137-1 Sterilization of health care products -Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, ISO 11137-2 Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose, ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices, ISO 17665-2 Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1; biocompatibility evaluation and testing according to ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and ISO 10993-5 Biological evaluation of medical devices – Part 5: Tests for in vitro cvtotoxicity; engineering dimensional analysis, and static and dynamic compression-bending testing according to ISO 14801 Dentistry – Implants – Dynamic fatigue test for endosseous dental implants.
{5}------------------------------------------------
EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles predicate devices shown above. Below are summary tables showing technical comparison between the subject device and the primary predicate devices.
| Comparison | Subject Device | Primary Predicate Devices | |
|---|---|---|---|
| TRI Dental Int. AGTRI® Dental ImplantSystemK151916 | Zimmer Dental Inc.Tapered Screw-Vent® TImplant and TaperedScrew-Vent® P ImplantK101880 | Straumann USAITI® Dental ImplantSystemK012757 | |
| Indications for Use | The TRI® Dental Implant Systemis intended for placement in thebone of the maxillary ormandibular arch for therehabilitation of edentulous andpartially edentulous patients.TRI® Dental Implant Systemallows for one and two-stagesurgical procedures. When a one-stage surgical procedure isapplied, the implant may beimmediately loaded when goodprimary stability is achieved andwith appropriate occlusalloading.TRI Dental Implant System6.5 mm implants are intended fordelayed loading only. | The Tapered Screw-Vent® T Implantsare designed for use in the maxilla ormandible for immediate loading orfor loading after a conventionalhealing period. Implants may be usedto replace one or more missing teeth.Immediate loading is indicated whenthere is good primary stability and anappropriate occlusal load.The Tapered Screw-Vent® P Implantsare designed for use in the maxilla ormandible for loading after aconventional healing period. Implantsmay be used to replace one or moremissing teeth. | ITI® tapered implants areintended for immediate,delayed, or conventionalplacement in the maxillaryand/or mandibular arches tosupport crowns, bridges, oroverdentures in edentulous orpartially edentulous patients. |
| Implant Design | |||
| TRI-Narrow | |||
| Implant Diameter (mm) | 3.3 | 3.3, 3.7, 4.1, 4.7, 6.0 | |
| Platform Diameter (mm) | 3.2 | 3.5, 4.5, 5.7 | |
| Length (mm) | 11.5, 13, 16 | 8, 10, 11.5, 13, 16 | |
| Connection | Internal | Internal | |
| TRI-Vent | |||
| Implant Diameter (mm) | 3.75, 4.1, 4.7 | 3.3, 3.7, 4.1, 4.7, 6.0 | |
| Platform Diameter (mm) | 3.5 | 3.5, 4.5, 5.7 | |
| Length (mm) | 6.5, 8, 10, 11.5, 13, 16 | 8, 10, 11.5, 13, 16 | |
| Connection | Internal | Internal | |
| TRI-Octa | |||
| Implant Diameter (mm) | 3.75, 4.1, 4.7 | 3.3, 4.1, 4.8 | |
| Platform Diameter (mm) | 4.8 | 3.5, 4.8, 6.5 | |
| Length (mm) | 6.5, 8, 10, 11.5, 13, 16 | 6, 8, 10, 12, 14 | |
| Connection | Internal | Internal | |
| Material | |||
| Implant | Titanium Alloy | Titanium Alloy | CPTi |
| Comparison | Subject Device | Primary Predicate Device | |
| TRI Dental Int. AGTRI® Dental Implant System | Altatec GmbHCAMLOG Implant System ModifiedImplants and Abutments | ||
| K151916 | K083496 | ||
| Indications for Use | The TRI® Dental Implant System is intendedfor placement in the bone of the maxillary ormandibular arch for the rehabilitation ofedentulous and partially edentulous patients.TRI® Dental Implant System allows for oneand two-stage surgical procedures. When aone-stage surgical procedure is applied, theimplant may be immediately loaded whengood primary stability is achieved and withappropriate occlusal loading.TRI Dental Implant System 6.5 mm implantsare intended for delayed loading only. | Camlog Implant System implants are intended forimmediate or delayed placement in the bone of themaxillary or mandibular arch. Camlog ImplantSystem Abutments are intended for use as supportfor crowns, bridges or overdentures. When a one-stage surgical approach is applied, the implant maybe immediately loaded when good primary stabilitis achieved and the functional load is appropriate. | |
| Designs | |||
| Prosthesis Attachment | Screw-retainedCement-retained | Screw-retainedCement-retained | |
| Restoration | Single-unitMulti-unit | Single-unitMulti-unit | |
| Implant/AbutmentConnection | Internal | Internal | |
| Abutment Diameter, mm | 3.3, 3.5, 4.8 | 3.3, 3.8, 4.3, 5.0, 6.0 | |
| Abutment Angle | Straight, up to 30° | Straight, up to 20° | |
| Materials | |||
| Temporary Abutment | Titanium Alloy, PEEK | PEEK | |
| Gold Abutment | Gold alloy | Gold alloy | |
| Abutment | Titanium Alloy | Titanium Alloy, Zirconia | |
| Healing Cap | Titanium Alloy | NA | |
| Abutment Screw | Titanium Alloy | Titanium Alloy |
Device Comparison – Implants
{6}------------------------------------------------
Device Comparison - Abutments
In addition to the identified primary predicates, the reference predicates all have a similar intended use as they are for placement in the maxilla or mandible for restoration of chewing function by support of a prosthetic component such as crowns and bridges.
TRI-Narrow and TRI-Vent implants have a similar bone level design, tapered internal connection, and range of sizes as K101880. TRI-Octa implants have a similar trans-gingival design, tapered internal connection and range of sizes as K012757 and K101545. Smaller length of the TRI-Vent as compared to primary predicate K101880 (6.5mm length) is supported by the reference predicate K092035 which contained a 4.0x5.0 implant body. The 16mm length is supported by the primary predicates and, for TRI-Octa by the reference predicates K101545 and K113779. All three implant designs have the same materials used for implant body in the primary predicate K101880 and for the abutments in primary predicates K083496 and reference predicate K072570. Subject device abutment designs are equivalent to the primary predicate K083496. The differences are supported by the reference predicates K113779 and K072570. The primary predicate K083496 does not include pink anodization for optimal esthetics. The pink anodization of the transmucosal collar of the submission device is included in the reference predicate K101545. The primary predicate K083496 does not include a multi-unit abutment, this is included in the
{7}------------------------------------------------
510(k) Summary Page 5 of 5
reference predicate K072570.
The performance data included in this submission demonstrate substantial equivalence to the predicate devices with respect to any differences identified. The fatigue testing provided under ISO 14801 demonstrates that the increased abutment angulation is substantially equivalent to the identified predicates with respect to performance. The surface area analysis demonstrates that the smallest diameter/shortest length combinations of the submission device are substantially equivalent to the identified reference predicate K092035 with respect to surface area available for osseointegration. Slight differences in material composition, surface treatment, or manufacturing exposures are demonstrated substantially equivalent by biocompatibility testing.
CONCLUSION
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Overall, the subject device has the following similarities to the predicate devices:
- . has the same intended use,
- . uses the same operating principle,
- . incorporates the same basic design,
- incorporates the same or very similar materials, and
- has similar packaging and is sterilized using the same materials and processes.
§ 872.3640 Endosseous dental implant.
(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.