(267 days)
A surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
The proposed device is a sterile latex powder-free surgical glove that is formulated using natural rubber latex and is brown in color. The glove is coated with hydrogel polymer coating. The glove is manufactured using molds that feature an independent thumb and mechanically locking cuffs to help prevent cuff roll down. They are offered powder-free and sterile. This glove is for single use only. This glove is suitable for use as a specialty surgical glove intended for orthopedic procedures or other procedures where a thicker barrier may be desired such as trauma and reconstructive surgery.
The provided document describes the predicate device and the performance data that supports substantial equivalence for the "Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating." This is a medical device approval summary, and as such, it focuses on non-clinical tests to demonstrate performance in line with existing standards for gloves, rather than a study for an AI-powered diagnostic device.
Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone study, type of ground truth, training set size, ground truth for training set) are not applicable to this type of device and submission.
Here's an analysis of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Standard/Test/FDA Guidance) | Reported Device Performance (New Device) |
|---|---|---|
| Biocompatibility (Overall) | ISO 10993-1 | Meets requirements |
| Primary Skin Irritation | ISO 10993-10 | Pass (Under conditions of the study, gloves are non-irritating.) |
| Guinea Pig Maximization | ISO 10993-10 | Pass (Under the conditions of the study, gloves do not display any potential for sensitization.) |
| Dimensions | ASTM D3577 | Meets requirements |
| Physical Properties | ASTM D3577 | Meets requirements for rubber surgical gloves |
| Freedom from Holes | 21 CFR 800.20 & ASTM D3577 | Meets requirements (Tested in accordance with ASTM D5151 with acceptable results) |
| Powder Residual | ASTM D3577 | Meets requirements of ≤ 2.0 mg/glove for "Powder-Free" designation (Results generated values < 2mg of residual powder per glove.) |
| Protein Content | FDA Medical Glove Guidance Manual, ASTM D5712 | Pass (Gloves yielded the results of less than 50 µg/dm² of total water extractable protein per glove.) |
| Puncture Resistance | AS/NZ 4179 | Pass (Tested in accordance with AS/NZ with acceptable results) |
2. Sample size used for the test set and the data provenance
- Sample Size: Not explicitly stated. For medical gloves, testing is typically done on a representative sample from production lots according to the specified ASTM standards. The standards themselves define the sampling plans.
- Data Provenance: The tests are non-clinical (laboratory/material tests). As such, there is no "country of origin of the data" in the sense of patient data. The tests were performed by the manufacturer (Cardinal Health 200, LLC) in a laboratory setting to demonstrate compliance with international and national standards. These are retrospective analyses of material properties.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. These are objective physical and chemical property tests, not expert-adjudicated diagnostic interpretations. The "ground truth" is defined by the technical specifications and methodologies of the referenced ASTM, ISO, and FDA standards.
4. Adjudication method for the test set
- Not applicable. The tests involve objective measurements against defined criteria, not subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI-powered diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an AI-powered diagnostic device.
7. The type of ground truth used
- Ground Truth: The "ground truth" is defined by the Pass/Fail criteria and quantitative limits specified in the referenced international standards (ASTM D3577, ISO 10993-10, AS/NZ 4179) and FDA guidance for medical gloves (e.g., 21 CFR 800.20, FDA Medical Glove Guidance Manual, ASTM D5712). These standards establish the acceptable physical, chemical, and biological properties for safe and effective surgical gloves.
8. The sample size for the training set
- Not applicable. This device does not involve a training set as it is not a machine learning or AI algorithm.
9. How the ground truth for the training set was established
- Not applicable. This device does not involve a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2016
Cardinal Health 200, LLC. Ms. Caroline Miceli Manager, Regulatory Affairs 1500 Waukegan Road Waukegan, Illinois 60085
Re: K151778
Trade/Device Name: Cardinal Health 111 Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: February 18, 2016 Received: February 22, 2016
Dear Ms. Caroline Miceli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151778
Device Name
Cardinal Health TM Sterile Latex Powder-Free Orthopedic Surgical Gloves With Hydrogel Coating
Indications for Use (Describe)
A surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized red bird-like shape above the words "CardinalHealth" in a bold, sans-serif font. The bird shape is made up of several curved lines that suggest movement or flight.
Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating K151778
| Manufacturer: | Cardinal Health 200, LLC1500 Waukegan RoadWaukegan, IL 60085 |
|---|---|
| Regulatory Affairs Contact: | Caroline Miceli1500 Waukegan RoadWaukegan, IL 60085 |
| Telephone Number: | (847) 887-6864 |
| Fax Number: | (847) 887-2461 |
| Date Summary Prepared: | March 23, 2016 |
| Product Trade Name: | Cardinal Health™ Sterile Latex Powder-Free Orthopedic SurgicalGloves with Hydrogel Coating |
| Common Name: | Orthopedic Surgeon's Gloves |
| Classification Name: | Surgeon's Gloves |
| Classification Panel: | General and Plastic Surgery |
| Regulation: | 21 CFR 878.4460 |
| Product Code: | KGO |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Cardinal Health. The logo consists of a stylized red bird-like design above the company name. The text "CardinalHealth" is written in a bold, sans-serif font, with the first letter capitalized and the rest in lowercase.
Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating K151778
| Predicate Devices: | K111015 – ProtexisTM Sterile Latex Powder-Free Surgical Gloveswith Hydrogel Coating |
|---|---|
| K903987- TriflexTM Orthopedic Sterile Surgeon's Glove | |
| Reason for 510(k)Submission: | New device |
| Device Description: | The proposed device is a sterile latex powder-free surgical glovethat is formulated using natural rubber latex and is brown in color.The glove is coated with hydrogel polymer coating. |
| The glove is manufactured using molds that feature an independentthumb and mechanically locking cuffs to help prevent cuff rolldown. They are offered powder-free and sterile. This glove is forsingle use only. | |
| This glove is suitable for use as a specialty surgical glove intendedfor orthopedic procedures or other procedures where a thickerbarrier may be desired such as trauma and reconstructive surgery. | |
| Intended Use: | A surgeon's glove is a device made of natural rubber intended to beworn by operating room personnel to protect a surgical woundfrom contamination. |
| Substantial Equivalence: | The proposed device is substantially equivalent to the predicatedevices identified in this 510(k) summary. Substantial equivalencecan be established in regard to intended use, physical propertiesand characteristics, design and product features. All gloves aremade of natural rubber latex using similar manufacturingprocesses. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Cardinal Health. The logo features a stylized red bird-like design above the company name. The text "CardinalHealth" is written in a bold, sans-serif font.
Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating K151778
| New DeviceCardinal Health™Sterile Latex Powder-Free OrthopedicSurgical Gloves withHydrogel Coating(K151778) | Predicate Device 1Protexis™ Sterile LatexPowder-Free SurgicalGloves with HydrogelCoating (K111015) | Predicate Device 2Triflex™ OrthopedicSterile Surgeon's Glove(K903987) | |
|---|---|---|---|
| Characteristic | Sterile Latex Powder-Free OrthopedicSurgical Gloves withHydrogel Coating | Sterile Latex Powder-Free Surgical Gloveswith Hydrogel Coating | Orthopedic SterileLatex PowderedSurgeon's Glove |
| Material Composition | Natural Rubber Latex | Natural Rubber Latex | Natural Rubber Latex |
| Design | Single UseSterilePowder-freeHand SpecificIndependent ThumbBeaded CuffLubricatedHydrogel PolymerCoating | Single UseSterilePowder-freeHand SpecificIndependent ThumbBeaded CuffLubricatedHydrogel PolymerCoating | Single UseSterilePowderedHand SpecificIndependent ThumbBeaded Cuff |
| Label Claim | • Latex• Single Use• Sterile• Powder-free• Hydrogel Coating• Contains less than50 µg/dm² of totalwater extractableprotein per glove• PunctureResistance $≥$ 5N | • Latex• Single Use• Sterile• Powder-free• Hydrogel Coating• Contains less than50 µg/dm² of totalwater extractableprotein per glove | • Latex• Single Use• Sterile• Powdered |
| New DeviceCardinal Health™Sterile Latex Powder-Free OrthopedicSurgical Gloves withHydrogel Coating(K151778) | Predicate Device 1Protexis™ Sterile LatexPowder-Free SurgicalGloves with HydrogelCoating (K111015) | Predicate Device 2Triflex™ OrthopedicSterile Surgeon's Glove(K903987) | |
| Intended Use/Indications for Use | Orthopedic Powder-Free Surgeon's Glove | Powder-Free Surgeon'sGlove | Orthopedic PowderedSurgeon's Glove |
| Dimensions &Physical Properties | Meets ASTM D3577 | Meets ASTM D3577 | Meets ASTM D3577 |
| Freedom from Holes | AQL meets 21CFR800.20 & ASTMD3577 requirements | AQL meets 21CFR800.20 & ASTMD3577 requirements | AQL meets 21CFR800.20 & ASTMD3577 requirements |
| Powder Residual | Meets requirements of≤ 2.0 mg/glove forPowder-Freedesignation per ASTMD3577 | Meets requirements of≤ 2.0 mg/glove forPowder-Freedesignation per ASTMD3577 | N/A |
| Protein Contents | Contains less than 50µg/dm² of total waterextractable protein perglove as tested perASTM D5712 | Contains less than 50µg/dm² of total waterextractable protein perglove as tested perASTM D5712 | N/A |
| Puncture Resistance | PunctureResistance ≥ 5Nas tested per AS/NZ4179 | N/A | N/A |
Performance Data
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized red bird-like design above the words "CardinalHealth" in black font. The bird design is made up of several curved lines that resemble feathers or wings.
Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating K151778
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized red bird-like design above the words "CardinalHealth" in black font. The bird design is made up of several curved lines that resemble feathers or wings.
Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating K151778
Summary of Non-Clinical Tests Conducted for Determination of Substantial Equivalence Dorfo - Now Davia
| Performance Test Summarv-New Device | |||
|---|---|---|---|
| Company of the country of the finance of the finance of the finance and the first and the first and the first and |
| Characteristics | Standard/Test/FDAGuidance | Results Summary |
|---|---|---|
| Biocompatibility: | ||
| Primary Skin Irritation | ISO 10993-10 | Under conditions of the study,gloves are non-irritating. |
| Guinea PigMaximization | ISO 10993-10 | Under the conditions of thestudy, gloves do not displayany potential for sensitization. |
| Physical Characteristics: | ||
| Dimensions | ASTM D3577 | Meet requirements |
| Physical Properties | ASTM D3577 | Meet requirements for rubbersurgical gloves |
| Freedom from Holes | 21 CFR 800.20 & ASTMD3577 | Tested in accordance withASTM D5151 with acceptableresults |
| Powder Residual | ASTM D3577 tested usingASTM standard D6124 | Gloves meet powder levelrequirements for "Powder-Free" designation. Resultsgenerated values < 2mg ofresidual powder per glove. |
| Protein Content | FDA Medical Glove GuidanceManual, ASTM D5712 | Gloves yielded the results ofless than 50 µg/dm² of totalwater extractable protein perglove. |
| Puncture Resistance | AS/NZ 4179 | Tested in accordance withAS/NZ with acceptable results |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image features the Cardinal Health logo. The logo consists of a stylized red bird-like design above the company name. The company name, "CardinalHealth", is written in a bold, sans-serif font, with the first letter of each word capitalized.
Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating K151778
| Characteristic | Requirement | New Device | PredicateDevice 1 | PredicateDevice 2 |
|---|---|---|---|---|
| Cardinal HealthTMSterile LatexPowder-FreeOrthopedicSurgicalGloves withHydrogelCoating(K151778) | ProtexisTMSterile LatexPowder-FreeSurgicalGloves withHydrogelCoating(K111015) | TriflexTMOrthopedicSterileSurgeon'sGlove(K903987) | ||
| Biocompatibility | ISO 10993-1 | Meetsrequirements | Meetsrequirements | Meetsrequirements |
| Primary SkinIrritation | ISO 10993-10 | Pass | Pass | Pass |
| Guinea PigMaximization | ISO 10993-10 | Pass | Pass | Pass |
| Dimensions | ASTM D3577 | Meetsrequirements | Meetsrequirements | Meetsrequirements |
| Physical Properties | ASTM D3577 | Meetsrequirements | Meetsrequirements | Meetsrequirements |
| Freedom from Holes | 21CFR800.20,ASTM D3577 | Meetsrequirements | Meetsrequirements | Meetsrequirements |
| Powder Residual | ASTM D3577 | Meetsrequirements | Meetsrequirements | N/A |
| Protein Content | ASTM D5712 | Pass | Pass | N/A |
| Puncture Resistance | AZ/NZ 4179 | Pass | N/A | N/A |
Comparative Performance Information Summary
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Cardinal Health. The logo consists of a stylized red bird-like design above the company name. The text "CardinalHealth" is written in a bold, sans-serif font, with the first letter capitalized and the rest in lowercase.
Cardinal Health™Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating K151778
Summary of Clinical Tests Conducted for Determination of Substantial Equivalence and/or Clinical Information:
Clinical Data is not required.
Conclusions Drawn from Non-Clinical Data:
Non-clinical data demonstrates that the Cardinal Health™ Sterile Latex Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating meet the technological characteristics of the ASTM D3577 standard. These gloves are as safe and as effective and performs as well as the legally marketed devices identified in this submission.
§ 878.4460 Non-powdered surgeon's glove.
(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).