(84 days)
The Valleylab FT10 is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles. and lymphatics. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use this function for these procedures.
The Valleylab™ FT10 Electrosurgical Platform (VLFT10) provides radio frequency (RF) energy for monopolar and bipolar surgical applications, and tissue-fusion and vessel-sealing applications (LigaSure function). It is a combination of a full-featured general-surgery electrosurgical unit and a LigaSure vessel sealing system. The monopolar and bipolar sections, including the advanced bipolar/LigaSure section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery. The LigaSure section of the system provides power for vessel sealing.
The VLFT10 is used in hospitals and other health care facilities where surgical procedures are carried out.
The VLFT10 can be used with a variety of legally marketed accessories including monopolar and bipolar instruments, footswitches, and return electrode pads. The VLFT10 connects to electrical mains and operates at an input line frequency of 47-63 Hz.
The provided document is a 510(k) summary for the Valleylab™ FT10 Electrosurgical Platform. It describes the device, its intended use, technological characteristics, and performance testing to demonstrate substantial equivalence to a predicate device (ForceTriad™ Energy Platform).
However, the document does not describe acceptance criteria or a specific study that proves the device meets those acceptance criteria in the format requested. Instead, it outlines a series of tests performed to establish substantial equivalence.
Here's an analysis of the provided text in relation to your request:
1. A table of acceptance criteria and the reported device performance:
The document lists several types of testing performed, but it does not explicitly state specific quantitative acceptance criteria for each test nor present the reported device performance in a comparative table against those criteria. It generally states that tests were "successfully completed" or showed "comparable performance."
For example, for ex vivo and in vivo testing, it states "showed comparable performance with regard to thermal effects, and vessel sealing capabilities. Some incremental improvements in procedural speed were seen with the VLFT10." This is a qualitative statement rather than a quantitative comparison against defined acceptance criteria.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size: The document mentions "ex vivo and in vivo testing using porcine tissue and a porcine model," but does not specify the sample size (e.g., number of tissue samples, number of animals) used for these tests.
- Data Provenance: The document does not specify the country of origin of the data. The testing appears to be prospective as it was conducted specifically for the device's validation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided in the document. The tests described (e.g., thermal effects, vessel sealing capabilities, electrical safety) are typically assessed through objective measurements and engineering analysis rather than expert review to establish "ground truth" in the way it might be for a diagnostic imaging device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable and therefore not provided, as the tests described do not involve subjective interpretation requiring adjudication among experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable as the device is an electrosurgical generator, not a diagnostic imaging device that would involve human readers or AI assistance in interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device is an electrosurgical generator. It has "tissue sensing algorithms" and a "LigaSure algorithm" that have been refined. This suggests some algorithmic control or processing. However, the document does not describe "standalone algorithm performance" in terms of clinical outcomes independent of the surgical procedure and human operator. The performance is assessed in the context of the device's function (cutting, coagulation, sealing) during surgical tasks, inherently involving human interaction.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For the "ex vivo and in vivo testing," the "ground truth" would likely involve direct measurements of physiological effects such as thermal spread, tissue integrity, burst pressure of sealed vessels, and potentially histological examination of tissue, rather than expert consensus on subjective assessments. The document doesn't explicitly detail the "ground truth" type but implies objective biophysical measurements.
8. The sample size for the training set:
This information is not applicable as the document describes a hardware device with embedded algorithms, not a machine learning model that undergoes a distinct "training" phase with a labeled dataset in the conventional sense. The "refinement" of algorithms (e.g., tissue sensing, LigaSure algorithm) would likely involve iterative development and testing, but not a formally defined "training set" with a specified size.
9. How the ground truth for the training set was established:
This information is not applicable for the same reasons as point 8.
In summary:
The document focuses on demonstrating substantial equivalence through compliance with electrical safety and EMC standards, ex vivo and in vivo testing (thermal effects, vessel sealing, procedural speed), system verification (functionality, specifications), software verification and validation, and usability testing. While these tests are designed to show the device performs as intended and is safe and effective compared to its predicate, it does not structure this information as a direct comparison against pre-defined, quantitative "acceptance criteria" for clinical performance.
For a regulatory submission of this nature, "acceptance criteria" are often embedded within the test protocols and engineering specifications, which are not fully detailed in this summary. The "reported device performance" is generally described as meeting these implicit criteria or being "comparable" to the predicate.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 10, 2015
Covidien Ms. Sharon McDermott Senior Regulatory Affairs Product Specialist 5920 Longbow Drive Boulder, Colorado 80301
Re: K151649
Trade/Device Name: Valleylab™ FT10 Electrosurgical Platform Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: June 17, 2015 Received: June 18, 2015
Dear Ms. McDermott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151649
Device Name Valleylab™ FT10 Electrosurgical Platform
Indications for Use (Describe)
The Valleylab FT10 is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles. and lymphatics. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use this function for these procedures.
| Type of Use (Select one or both, as applicable) |
|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
6 510(k) Summary
510(k) Summary
Date summary prepared: September 10, 2015
510(k) Submitter/Holder
Covidien 5920 Longbow Drive Boulder, CO 80301
Contact
Sharon McDermott Sr. Regulatory Affairs Product Specialist Telephone: 303-581-6789 Fax: 303-530-6313 Email: sharon.mcdermott@medtronic.com
Name of Device
| Trade Name: | Valleylab™ FT10 Energy Platform |
|---|---|
| Common Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories(21 CFR § 878.4400, class II, GEI) |
Predicate Device
| Trade Name: | ForceTriad™ Energy Platform |
|---|---|
| Common Name: | Electrosurgical Generator |
| Catalog Number: | FORCETRIAD |
| 510(k) Number: | K110268 |
| Manufacturer: | Covidien |
Device Description
The Valleylab™ FT10 Electrosurgical Platform (VLFT10) provides radio frequency (RF) energy for monopolar and bipolar surgical applications, and tissue-fusion and vessel-sealing applications (LigaSure function). It is a combination of a full-featured general-surgery electrosurgical unit and a LigaSure vessel sealing system. The monopolar and bipolar sections, including the advanced bipolar/LigaSure section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery. The LigaSure section of the system provides power for vessel sealing.
The VLFT10 is used in hospitals and other health care facilities where surgical procedures are carried out.
The VLFT10 can be used with a variety of legally marketed accessories including monopolar and bipolar instruments, footswitches, and return electrode pads. The VLFT10 connects to electrical mains and operates at an input line frequency of 47-63 Hz.
{4}------------------------------------------------
Principle of Operation
The Valleylab FT10 is a radio-frequency (RF) electrosurgical generator that delivers energy to compatible surgical instruments. The concentration of energy at the instrument in conjunction with tissue characteristics produces heat. The heating of tissue provides the desired surgical effect (cutting, coagulation, sealing). Variations in the waveform result in the different surgical effects achieved by different modes. In the case of LigaSure vessel sealing, the generator senses changes in tissue characteristics to determine when a seal cycle is complete.
Indications for Use
The Valleylab FT10 is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.
Technological Characteristics
The technological characteristics of the proposed VLFT10 are the same as the predicate, ForceTriad; both are a combination of a full-featured general-surgery electrosurgical unit and a LigaSure vessel sealing system. The VLFT10 provides several incremental improvements over the predicate while maintaining the same basic functionality and intended use. These are summarized below.
- . Tissue sensing algorithms have been upgraded to sense electrical impedance at a faster rate.
- Monopolar capabilities have been expanded with the addition of a Soft Coag mode, a mode that . has been available in other electrosurgical generators such as the ERBE VIO® 300 D.
- . The LigaSure algorithm has been refined to improve clip detection and to respond more quickly to changes in tissue impedance.
- . The VLFT10 includes RFID technology to identify LigaSure instruments and to write usage information to the tags.
Performance Characteristics
Extensive verification and validation activities were successfully completed that establish that the VLFT10 performs as intended and is substantially equivalent to its predicate, ForceTriad. Testing included the following:
- Compliance with Electrical Safety and EMC standards: .
| IEC 60601-1:2005/A1:2012 Medical electrical equipment – Part 1: General requirements forbasic safety and essential performance – Edition 3.1 | |
|---|---|
| IEC 60601-1-2:2014 | Medical electrical equipment, Part 1-2: General requirementsfor basic safety and essential performance – Collateralstandard: Electromagnetic compatibility – Requirements andtests |
{5}------------------------------------------------
- IEC 60601-2-2:2009 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories -Edition 5.0
- . Ex vivo and in vivo testing using porcine tissue and a porcine model showed comparable performance with regard to thermal effects, and vessel sealing capabilities. Some incremental improvements in procedural speed were seen with the VLFT10.
- . System verification (Non-IEC electrical, instrument compatibility, basic functionality) showed that the VLFT10 has all required functionality and that it meets system specifications.
- . Software verification and validation testing was conducted and documentation provided in accordance with FDA's, Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices
- Usability testing in accordance with IEC 62366: 2007 Medical devices Application of . usability engineering to medical devices showed that representative users understood the user interface and could use the VLFT10 to complete typical surgical tasks.
The minor differences that were seen during testing are consistent with design changes (such as the improved tissue sensing capabilities) and do not raise safety or effectiveness concerns.
Clinical Studies
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
Conclusion
Verification and validation testing demonstrates the proposed Valleylab™ FT10 Energy Platform is substantially equivalent to the predicate, the ForceTriad™ Energy Platform. The two electrosurgical generators have the same intended use and fundamental technology. The new generator has similar performance with minor incremental improvements. The differences do not raise any new questions of safety and efficacy when compared with the predicate.
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.