K Number
K151604
Date Cleared
2015-09-11

(88 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Adjustable Chandelier Illuminator family of ophthalmic illuminators is for wide angle illumination during ophthalmic surgery.

Device Description

The 23ga and 25ga Adjustable Chandelier Illuminators are made to work with the Bausch and Lomb surgical system and Bausch and Lomb Valved ESA (Entry Site Alignment) systems (K012435). These devices are to be manufactured with distally located "Infusion Cannulas" that adequately secure themselves within 23ga and 25ga Bausch and Lomb Valved ESA Systems. The 23ga and 25ga Adjustable Chandelier Illuminators consist of a PMMA (Polymethylm-ethacrylate with Flourinated Polymer Cladding) acrylic fiber (for light transmission), white acetal handle assembly with finger slide for adjustability, PTFE (Polytetrafluor-oethylene) jacket (for fiber) and Stellaris style light connector to fit into the Bausch and Lomb Stellaris PC Vision Enhancement System (K133486). The 23ga and 25ga Adjustable Chandelier Illuminators are to be provided sterile to the user, is a single-use device and is packaged in double Tyvek-to-poly pouches to facilitate introduction to the sterile environment. These provide illumination for eye surgery.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Peregrine 23ga and 25ga Adjustable Chandelier Illuminator. The purpose of this notification is to demonstrate that the device is substantially equivalent to legally marketed predicate devices.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document lists various non-clinical performance tests conducted to evaluate the device. The reported performance is that all results were "acceptable" and the device "satisfactorily met the requirements." Specific numerical acceptance criteria or performance metrics are not provided in this summary.

TestTest MethodReported Device Performance
SterilityISO 11135-1Acceptable
Shelf-lifeASTM F1980Acceptable
BiocompatibilityISO 10993-01Acceptable (fully biocompatible device)
Light OutputInternal Test Method(ITP04-20)Satisfactorily met requirements (specific value of 3.7 lumen is provided in Table 5.2 for both)
Light FieldInternal Test Method (ITP04-30)Satisfactorily met requirements
Thermal InspectionInternal Test Method (ITP04-40)Satisfactorily met requirements
Bond StrengthInternal Test Method (ITP03-75)Satisfactorily met requirements
Cannula & Fiber InspectionInternal Test Method (ITP52-00)Satisfactorily met requirements

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample size for individual tests. However, it indicates that "Biocompatibility and sterility testing has been conducted on the 25ga Adjustable Chandelier Illuminators as a 'worst case device' and a family representative." This suggests a limited sample size, focusing on one representative (25ga).

The data provenance is internal, originating from Peregrine Surgical Ltd. It is retrospective, as these are bench tests conducted to support a premarket submission. No information about the country of origin of the data is provided beyond the company's location (New Britain, PA, USA).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is not an AI/ML diagnostic or prognostic tool that requires expert-established ground truth on a test set. The tests performed are bench tests evaluating physical and functional characteristics of a medical illuminator. The "ground truth" for these tests would be the established standards and specifications outlined in the referenced ISO and ASTM methods, or the company's internal test protocol specifications.

4. Adjudication Method for the Test Set

Not applicable, as this is not a diagnostic assessment requiring adjudication among experts. The tests are objective measurements against defined standards.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, and no MRMC study was performed.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device. The "standalone" performance refers to the device's functional characteristics as an illuminator, which were assessed through the listed bench tests.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the non-clinical performance data is the defined acceptance criteria and specifications within the referenced ISO, ASTM, and Internal Test Methods. For example:

  • Sterility: The device must meet the sterility assurance level (SAL) specified in ISO 11135-1.
  • Biocompatibility: The device materials must meet the biological safety requirements outlined in ISO 10993-01.
  • Light Output: The device's light output must meet the specified lumen value (3.7 lumen, as indicated in Table 5.2).

8. The Sample Size for the Training Set

Not applicable. This is a conventional medical device, not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. No training set was used.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them that resemble fabric or water.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 11, 2015

Peregrine Surgical Ltd. c/o Mr. Ryan O'Leary Product Development 51 Britain Dr. New Britain, PA 18901

Re: K151604

Trade Name: Peregrine 23 ga and 25 ga Adjustable Chandelier Illuminator Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: MPA Dated: June 5, 2015 Received: June 15, 2015

Dear Mr. O'Leary:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801) medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Y. Alexander -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K151604

Device Name

23ga and 25ga Adjustable Chandelier Illuminator

Indications for Use (Describe)

Adjustable Chandelier Illuminator family of ophthalmic illuminators is for wide angle illumination during ophthalmic surgery.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 5. 510(k) SUMMARY

510(k) SUMMARY for 23ga and 25ga Adjustable Chandelier Illuminator

    1. Submitter Information: Peregrine Surgical, Ltd. 51 Britain Drive New Britain, PA 18901
Contact Person:Ryan O'Leary
Telephone Number:(215) 348-0456
Fax Number:(215) 348-5526
  • Date Prepared: 05 June 2015
    1. Device Name:
Proprietary Name:23ga and 25ga Adjustable Chandelier Illuminator
Classification Name:Endoilluminator
CFR Number:876.1500
Device Class:II
Product Code:MPA
    1. Predicate Device:

Table 5.1 Predicate Devices

Predicate DeviceName510(k)Product CodeCompany Name
Peregrine TaperedDiffusion ProbeK980025MPAPeregrine Surgical,Ltd.
23 Gauge (ga)Curved IlluminatingLaser ProbeK122997HQB,HQF, MPAPeregrine Surgical,Ltd.
    1. Description of Device:
      The 23ga and 25ga Adjustable Chandelier Illuminators are made to work with the Bausch and Lomb surgical system and Bausch and Lomb Valved ESA (Entry Site Alignment) systems (K012435). These devices are to be manufactured with distally located "Infusion Cannulas" that adequately secure themselves within 23ga and 25ga Bausch and Lomb Valved ESA Systems. The 23ga and 25ga Adjustable Chandelier Illuminators consist of a PMMA (Polymethylm-ethacrylate with Flourinated Polymer Cladding) acrylic fiber (for light transmission), white acetal handle assembly with finger slide for adjustability, PTFE (Polytetrafluor-oethylene) jacket (for fiber) and Stellaris style light connector to fit into the Bausch and Lomb Stellaris PC Vision Enhancement System (K133486).

{4}------------------------------------------------

The 23ga and 25ga Adjustable Chandelier Illuminators are to be provided sterile to the user, is a single-use device and is packaged in double Tyvek-to-poly pouches to facilitate introduction to the sterile environment. These provide illumination for eye surgery.

  • ર. Indications for Use: Adjustable Chandelier Illuminator family of ophthalmic illuminators is for wide angle illumination during ophthalmic surgery.
    1. Substantial Equivalence: Technological Characteristics -
FeatureProposed DeviceAdjustable ChandelierIlluminatorPredicate DevicePeregrineTaperedDiffusion ProbePredicate Device23 Ga Curved Illuminating Laser Probe
23 Ga25 Ga
Image: Extended 25ga and Retracted 23gaImage: Peregrine Tapered Diffusion ProbeImage: 23 Ga Curved Illuminated Laser Probe
Note: The above photo reflects the chandlier inboth the extended and retracted state. Thisportion of both gauges is identical except forthe stainless infusion cannula which is slightlysmaller on the 25ga version.
510(k)To be assignedK980025K122997
Product CodeMPAMPAHGB, MPA, HQF
Indicationsfor UseAdjustable Chandelier Illuminator family ofophthalmic illuminators is for wide angleillumination during ophthalmic surgery.For wide angleillumination of theposterior segment duringophthalmic surgery.For photocoagulation andillumination duringophthalmic surgery. Thisdevice delivers illuminationas well as laser energy totarget tissue, causingcoagulation. Spot size canbe varied by altering thedistance between the tissueand probe tip.
LaserConnectionNo laser connecterNo laser connecterNi/Cu Stainless AlloyConnector
HandpieceConstructionAcetal handpieceAcetal handpieceAcetal handpiece

Table 5.2 Technological Characteristics

{5}------------------------------------------------

Optical fiberNo laser fiberNo Laser FiberGlass-Silica core - .006"(150 microns)
JacketMaterialPTFE(Polytetrafluoroethylene)PTFE(Polytetrafluoroethylene)PolyethyleneAcrylated Olefin
Jacket ColorBlackGreenBlackBlueBlackOrange
InnerDiameter(ID)Black PTFE Jacket 0.022"Green PTFE Jacket 0.052"Black PTFE Jacket 0.022"Blue PTFE Jacket 0.052"0.046"0.040"
OuterDiameter(OD)Black PTFE Jacket 0.046"Green PTFE Jacket 0.084"Black PTFE Jacket 0.046"Blue PTFE Jacket 0.084"0.062"0.070"
Length91"91"82"101"
Weight0.4198 oz.0.4198 oz.0.3704 oz0.9524 oz.
Gauge size23 ga25 ga20 ga23 ga
IlluminationFiber - OD0.015"0.0150"0.030"0.015"
IlluminationFiberMaterialPMMA(Polymethylmethylacrylatewith Fluorinated PolymerCladding)PMMA(Polymethylmethylacrylatewith Fluorinated PolymerCladding)PMMA(Polymethylmethylacrylatewith Fluorinated PolymerCladding)PMMA(Polymethylmethylacrylatewith Fluorinated PolymerCladding)
StretchCapacity-OD*Not stretchedNot stretchedNot stretched0.0095"
IlluminationConnectorAcetal with StainlessTubeAcetal with StainlessTubeAluminumAcetal with Stainless Tube
Light Output3.7 lumen3.7 lumen7 lumen3.4 lumen
TipAngulationTaperedTaperedTaperedStraight
NeedleMaterial304 stainless (infusioncannula) - OD .025"304 Stainless (infusioncannula) - OD .021"304 stainless304 stainless

The 23ga and 25ga Adjustable Chandelier Illuminators only transmit light energy, they do not control the intensity of the light output – this is controlled by the operating system to which it is attached. This is also true for both predicate devices.

The light connector for the 23ga and 25ga Adjustable Chandelier Illuminators is the same light connector utilized in the 23ga Curved Illumination Laser Probe (K122997) and is designed to fit into the light sources of the Bausch and Lomb Stellaris PC Vision Enhancement System (K133486).

The materials used to fabricate the acrylic fiber, handpiece and illumination connector used in the 23ga and 25ga Adjustable Chandelier Illuminators are identical to that of the 23ga Curved Illuminating Laser Probe as it was approved in K122997 on June 26, 2013 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).

{6}------------------------------------------------

The fiber used in the proposed device and both predicates is the exact same formulation (Polymethylmethacrylate (PMMA) with Flourinated Polymer Cladding) and is produced by the same supplier. The diameters of the fibers used in the 23ga and 25ga Adjustable Chandelier Illuminators and the 23ga Curved Illuminating Laser Probe are the same while the fiber used in the Tapered Diffusion is slightly bigger but still provides the same function.

The tapered fiber tip of the 23ga and 25ga Adjustable Chandelier Illuminators and the Tapered Diffusion Probe are manufactured in the same way and perform the same function.

4. Non-Clinical Performance Data

To evaluate the performance of the Adjustable Chandelier Illuminator, the following tests were conducted:

TestTest Method
SterilityISO 11135-1
Shelf-lifeASTM F1980
BiocompatibilityISO 10993-01
Light OutputInternal Test Method(ITP04-20)
Light FieldInternal Test Method (ITP04-30)
Thermal InspectionInternal Test Method (ITP04-40)
Bond StrengthInternal Test Method (ITP03-75)
Cannula & Fiber InspectionInternal Test Method (ITP52-00)

Biocompatibility and sterility testing has been conducted on the 25ga Adjustable Chandelier Illuminators as a "worst case device" and a family representative. All results were acceptable resulting in fully biocompatible device. The 23ga and 25ga Adjustable Chandelier Illuminators are made of the same materials as the predicate devices, Tapered Diffusion Probe (K980025) and 23ga Curved Illumination Laser Probe (K122997). The performance of the 23ga and 25ga Adjustable Chandelier Illuminators satisfactorily met the requirements of the non-clinical bench testing conducted to support substantial equivalence.

    1. Clinical Performance Data
      No data from human clinical studies has been included to support the substantial equivalence of the 23ga and 25ga Adjustable Chandelier Illuminator.
    1. Conclusion Regarding Substantial Equivalence
      The 23ga and 25ga Adjustable Chandelier Illuminators are intended to provide illumination for eye surgery. The 23ga and 25g Adjustable Chandelier Illuminators have the same intended use, incorporate the same fundamental technology, and have similar indications for use as the predicate devices. Peregrine Tapered Diffusion Probe cleared under premarket notification K980025 and the 23ga Curved Illumination Laser Probe

{7}------------------------------------------------

cleared under K122997. Physically, the 23ga and 25ga Adjustable Chandelier Illuminators share characteristics with both predicate devices. The Tapered Diffusion Probe (K980025) was chosen as the primary predicate because the fiber is tapered in the exact same manner as the Chandelier and produces a very similar light output pattern. The 23ga Curved Illuminating Laser Probe (K122997) was chosen as a secondary predicate because it uses the same connects to the same light source as the Chandelier. Test data to verify the performance of the 23ga and 25ga Adjustable Chandelier Illuminators has been provided including sterility, shelf-life, biocompatibility and functionality. The results of this testing, combined with the design and intended use comparison with the predicate devices, support substantial equivalence.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.