K Number
K151503

Validate with FDA (Live)

Date Cleared
2016-04-05

(306 days)

Product Code
Regulation Number
876.5980
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube device is inserted into the stomach through the nasal cavity for the purpose of supplying a nutritional source for those patients who are unable to feed themselves.

Device Description

The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube is an enteral feeding tube designed for continuous or intermittent use. It comprises of radio-opaque polyurethane tube with guide wire set and ENFit connector.

The guide wire set is composed of a rigid guide wire and a male threaded connector, which is inserted into the polyurethane tube. They're depth marks printed on the tube in order to provide reference about the depth of insertion to the practitioner.

The tube is to be inserted through the nasal or oral area and feed down into the alimentary canal system (i.e. stomach area, duodenum or small intestine as intended). The guide wire with male connector is removed when the insertion is done leaving the tube and ENFit connector in situ. Only a female ENFit enteral syringe can be attached to the ENFit connector when in use.

AI/ML Overview

This document describes the regulatory submission for the NUTRICARE – ISOSAF™ Naso-Gastric Feeding Tube, seeking substantial equivalence to a predicate device, rather than an AI-powered diagnostic device. Therefore, the questions related to AI-specific performance criteria (accuracy metrics, ground truth establishment, training/test sets, human reader studies, etc.) are not applicable to this submission.

The document focuses on demonstrating substantial equivalence based on:

  • Intended Use: Both the new device and the predicate device are for supplying nutritional sources through a naso-gastric route.
  • Technological Characteristics: Comparison of materials, design, sterilization, dimensions, and other physical properties.
  • Non-Clinical Performance Data: Biocompatibility testing (in accordance with ISO 10993), mechanical testing (e.g., tensile strength, leakage, resistance to separation), and flow rate testing.

There is no mention of an AI algorithm, image analysis, or diagnostic performance evaluation with acceptance criteria related to sensitivity, specificity, or AUC.

Therefore, I cannot provide a table of acceptance criteria and reported device performance related to AI-powered diagnostics or answer the specific questions about sample sizes for test/training sets, expert consensus, adjudication methods, or MRMC studies, as these types of studies were not conducted or required for this device's 510(k) submission.

Summary of what can be extracted from the document regarding "acceptance criteria" (non-AI related):

The acceptance criteria for this device are demonstrated by its successful performance in a series of non-clinical tests, aligning with relevant industry standards and showing comparable performance to the predicate device. These tests are physical and material property tests, not performance metrics for an AI algorithm.

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria CategorySpecific TestAcceptance Criteria (Implied: Pass)Reported Device Performance
BiocompatibilityCytotoxicity (ISO 10993)PassPassed
Sensitization (ISO 10993)PassPassed
Intracutaneous Reactivity (ISO 10993)PassPassed
Mechanical PerformanceFalling drop positive pressure liquid leakage (per AAMI/CN3:2014(PS))PassPass
Stress CrackingPassPass
Resistance to separation from axial loadPassPass
Resistance to separation from unscrewingPassPass
Resistance to overridingPassPass
Disconnection from unscrewingPassPass
Tensile testing of the catheterPassPass
Tensile testing of the connectorsPassPass
Corrosion testing of guide wirePassPass
Liquid leakagePassPass
Flow Rate Testing (EN 1618:1997)As per standardPassed
Mechanical Testing of connector (AAMI/CN3:2014(PS))As per standardPassed
Dimensional analysis (AAMI/CN3:2014(PS))As per standardPassed
Risk ManagementRisk Analysis (EN ISO 14971:2012), including misconnection riskIdentified & MitigatedConducted and reviewed
General ComplianceCompliance with applicable voluntary standards for biocompatibility & sterilizationYesComplies / Passed all testing

2. Sample sized used for the test set and the data provenance: Not applicable. This is a medical tubing device, not an AI diagnostic with a "test set" of images/data for AI performance evaluation. The "tests" are laboratory-based physical and material tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI performance (e.g., confirmed disease status) is not relevant here. Decisions are based on objective physical/chemical test results against ISO/EN standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. See above.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical device (feeding tube), not a diagnostic AI system requiring human reader studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): The "ground truth" for this device's acceptance is based on compliance with established international and national standards (ISO, EN, AAMI) for material properties and mechanical performance. This is not a diagnostic device where "ground truth" relates to clinical outcomes or expert labels.

8. The sample size for the training set: Not applicable. There is no AI training set.

9. How the ground truth for the training set was established: Not applicable. There is no AI training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing movement or progress.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 5, 2016

Gbuk Enteral Limited % Leeanne Baker Regulatory Consultant Advena Medical Pure Offices, Plato Close Warwick, CV34 6WE United Kingdom

Re: K151503

Trade/Device Name: NUTRICARE - ISOSAF™ Naso-Gastric Feeding Tube Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube and accessories Regulatory Class: Class II Product Code: PIF Dated: February 2, 2016 Received: February 12, 2016

Dear Leeanne Baker,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Herbert P. Lerner -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151503

Device Name

Nutricare ISO-SAF Naso-Gastric Feeding Tube

Indications for Use (Describe)

The NUTRICARE ISO-SAF Naso-gastric Feeding Tube device is inserted into the stomach through the nasal cavity for the purpose of supplying a nutritional source for those patients who are unable to feed themselves.

Type of Use (Select one or both, as applicable) ☑ Research Use (Per 21 CFR 201.66; Subject to Review) ☐ Other Commercial Use (21 CFR 201.66) ☑ Research Use (Per 21 CFR 201.66; Subject to Review) ☐ Other Commercial Use (21 CFR 201.66)
☑ Research Use (Per 21 CFR 201.66; Subject to Review) ☐ Other Commercial Use (21 CFR 201.66)

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

for

NUTRICARE – ISOSAF™ NASO-GASTRIC FEEDING TUBE

1. Submission Sponsor

  • GBUK Enteral Limited Blackwood Hall Business Park North Duffield Selby North Yorkshire, YO8 5DD United Kingdom Phone: +44 (0)1757 282945 Contact: Mark Thompson, Managing Director

2. Submission Correspondent

Advena Medical Pure Offices Plato Close Warwick CV34 6WE United Kingdom Cell Phone: +44 (0)7876 688344 Office Phone +44 (0)1926 800153 Contact: Leeanne Baker, Regulatory Consultant Email: Leeanne.baker@advenamedical.com

3. Date Prepared

20th May 2015

4. Device Identification

Trade/Proprietary Name:NUTRICARE – ISOSAFTM NASO-GASTRIC FEEDING TUBI
Common/Usual Name:Tube, feeding
Classification Name:Gastrointestinal tube and accessories
Classification Regulation:21 CFR 876.5980
Product Code:PIF
Device Class:Class II
Classification Panel:Gastroenterology/ Urology

5. Legally Marketed Predicate Device(s)

K083210 – CORFLO Anti-IV Enteral Feeding Tube Model Number 20-XXX

{4}------------------------------------------------

6. Device Description

The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube is an enteral feeding tube designed for continuous or intermittent use. It comprises of radio-opaque polyurethane tube with guide wire set and ENFit connector.

The guide wire set is composed of a rigid guide wire and a male threaded connector, which is inserted into the polyurethane tube. They're depth marks printed on the tube in order to provide reference about the depth of insertion to the practitioner.

The tube is to be inserted through the nasal or oral area and feed down into the alimentary canal system (i.e. stomach area, duodenum or small intestine as intended). The guide wire with male connector is removed when the insertion is done leaving the tube and ENFit connector in situ. Only a female ENFit enteral syringe can be attached to the ENFit connector when in use.

7. Indication for Use Statement

The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube device is inserted into the stomach through the nasal cavity for the purpose of supplying a nutritional source for those patients who are unable to feed themselves.

8. Substantial Equivalence Discussion

The following table compares the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.

Table 5A - Comparison of Characteristics
ManufacturerGBUK ENTERAL LTDCORPAC INC.
Trade NameNUTRICARE ISO-SAF™ Naso-gastric Feeding TubeCorflo Anti IV Enteral FeedingTube
510(k) Number-K083210
Product CodePIFFPD
RegulationNumber21 CFR 876.598021 CFR 876.5980
Regulation NameGastrointestinal tubes andaccessoriesGastrointestinal tubes andaccessories
Indications forUseThe NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube device isinserted into the stomachthrough the nasal cavity for thepurpose of supplying anutritional source for thosepatients who are unable to feedthemselves.The CORFLO® ANTI IV EnteralFeeding tube is intended for usein those patients who requireintermittent or continuous tubefeedings via the nasogastric ornasoenteric route
MaterialPolyurethanePolyurethane

Table 54 - Comnarison of Characteristics

{5}------------------------------------------------

ManufacturerGBUK ENTERAL LTDCORPAC INC.
Trade NameNUTRICARE ISO-SAFT™ Naso-gastric Feeding TubeCorflo Anti IV Enteral Feeding Tube
SterileYes - Ethylene OxideRest unknownYes - Ethylene Oxide
Overall Design
Proximal ConnectorSingle connector ENFit/ISO 80369-1 compliantLuer slip two (2) ports
Distal feeding portDual openings on opposite sides of tube at offset levelsSingle opening
Distal TipContinuous with diameter of rest of tube. Extra radiopaque tipContoured larger than diameter of rest of tube
Fully Radiopaque throughout lengthYesYes
Cm MarkingsNumbered markings throughout the length of the tubeMarkings through the length of the tube
Overall length50cm - 120cm55cm - 140cm
GuidewireTwisted metal wire with sealed tip ends prior to the feeding port to prevent accidental emergenceBraided metal wireTip ends prior to the feeding port to prevent accidental emergence
Diameter6Fr to 12Fr5Fr to 12Fr
Duration of UseShort term (up to 30 days)Short term
Single-UseYesYes
Shelf Life3 years5 years
Latex FreeYesYes
Complies with ISO 10993-1CytotoxicitySensitizationIntracutaneous ReactivityUnknown but assumed to comply as has 510k
ISO 80369-1 ENFIT compliantYesCurrent product has luer slip connector

9. Non-Clinical Performance Data

The biocompatibility of the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube has been demonstrated through appropriate in vitro tests on the device. The device has been assessed in accordance with ISO 10993 and does not introduce any additional safety risk over the predicate device.

The following testing has been performed (per AAMI/CN3:2014(PS) where applicable) to support substantial equivalence:

  • Falling drop positive pressure liquid leakage pass ●
  • . Stress Cracking - pass
  • . Resistance to separation from axial load – pass
  • Resistance to separation from unscrewing – pass
  • . Resistance to overriding – pass
  • Disconnection from unscrewing pass .

{6}------------------------------------------------

  • . Tensile testing of the catheter – pass
  • . Tensile testing of the connectors – pass
  • . Corrosion testing of guide wire – pass
  • Liquid leakage - pass

A comparative study performed demonstrates that the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube performs as intended and is safe for its intended use when compared with the predicate device.

A Risk Analysis including analysis of the risk of misconnection has been conducted in accordance with EN ISO 14971:2012

As part of demonstrating safety and effectiveness of NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube and in showing substantial equivalence to the predicate device that are subject to this 510(k) submission, GBUK Enteral Limited completed a number of test as detailed below:

  • . Flow Rate Testing in accordance with EN 1618:1997
  • Mechanical Testing of the connector in accordance with AAMI/CN3:2014(PS) and AAMI/CN3:2014(PS), respectively
  • Dimensional analysis in accordance with AAMI/CN3:2014(PS)

The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube passed all testing stated above as shown by the acceptable results obtained.

The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube complies with the applicable voluntary standards for biocompatibility and sterilization. The device passed all the testing in accordance with national and international standards.

10. Clinical Performance Data

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.

11. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.

It has been shown in this 510(k) submission that the difference between the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube and the predicate devices do not raise any questions regarding its safety and effectiveness. The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube, as designed and manufactured, is determined to be substantially equivalent to the

{7}------------------------------------------------

referenced predicate devices.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.