K Number
K151406
Manufacturer
Date Cleared
2016-02-25

(275 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cogent Lateral Interbody System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to L5. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended for intervertebral body fusion, and are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.

Device Description

The Cogent Lateral Interbody System is an intervertebral body fusion device intended to stabilize the spinal segment to promote fusion. The implants are available in a variety of lengths and heights to accommodate varying patient anatomy. The Cogent Lateral Interbody System implants and instruments are supplied non-sterile. The implants in the The Cogent Lateral Interbody System are manufactured from PEEK, Ti-6Al-4V, and tantalum.

The Cogent Lateral Interbody consists of two PEEK spacers with axial voids to contain bone graft material, a titanium linkage that connects the PEEK spacers, angular anti-migration teeth, and tantalum x-ray markers.

The Cogent Lateral Interbody is available in various sizes to accommodate varying patient anatomy. The Cogent Lateral Interbody System is 18 or 22mm wide and includes five available lengths, 40, 45, 50, 55, and 60mm. The Cogent Lateral Interbody System is also available in three lordotic options, 0°, 6°, and 12°. The Cogent Lateral Interbody System is available in heights of 8mm to 16mm.

The Cogent Lateral Interbody System implants may be inserted via an open or minimally invasive approach and may be placed in a lateral or anterolateral orientation.

The Cogent Lateral Interbody System implants are non-sterile and are to be sterilized by the end user.

AI/ML Overview

This document is a 510(k) summary for the Cogent Lateral Interbody System, an intervertebral body fusion device. It primarily discusses the device's substantial equivalence to existing predicate devices. Consequently, it does not contain details about acceptance criteria, the study proving the device meets those criteria, or clinical performance metrics in the way typically required for AI/ML device evaluations.

Instead, the document focuses on:

  • Substantial Equivalence: The primary assertion is that the Cogent Lateral Interbody System is substantially equivalent to predicate devices (Cogent Med-LIF System and Cogent Med-LIF XL System) based on indications for use, technological characteristics, and performance testing.
  • Device Description: Materials (PEEK, Ti-6Al-4V, Tantalum), dimensions, and general design.
  • Non-Clinical Testing: A "Cadaveric Usability Study" is mentioned, but no details of acceptance criteria or performance results are provided.
  • Clinical Testing: Explicitly states, "No clinical studies were performed."

Therefore, based on the provided text, I cannot answer the specific questions regarding acceptance criteria, study methodologies, and performance metrics as they would apply to an AI/ML device or a device requiring new clinical efficacy data. The information requested (Table of acceptance criteria, sample sizes, ground truth establishment, MRMC studies, standalone performance, etc.) is not present in this 510(k) summary because the approval is based on substantial equivalence to existing devices that have already demonstrated safety and effectiveness, rather than novel clinical performance data for this specific device.

The only "study" mentioned is a "Cadaveric Usability Study" under non-clinical tests, and no details about its methodology, acceptance criteria, or results are provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 25, 2016

Cogent Spine LLC % Mr. Jude Paganelli Cor Medical Ventures LLC 101 North Acacia Avenue, Suite 106 Solana Beach, California 92075

Re: K151406

Trade/Device Name: Cogent Lateral Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: December 31, 2015 Received: January 4, 2016

Dear Mr. Paganelli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151406

Device Name Cogent Lateral Interbody System

Indications for Use (Describe)

The Cogent Lateral Interbody System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to L5. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended for intervertebral body fusion, and are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

SUBMITTER:

Submitted By:

Company Name:Cogent Spine LLC
Address:101 N. Acacia Ave., Suite 106Solana Beach, CA 92075
Telephone:(858) 774-7891

CONTACT PERSON: Jude Paqanelli

DATE PREPARED: February 16, 2016

TRADE NAME: Cogent Lateral Interbody System

COMMON NAME: Intervertebral Body Fusion Device

DEVICE CLASSIFICATION: Class II

CLASSIFICATION NAME: Intervertebral Body Fusion Device (21 CFR 888.3080)

PRODUCT CODE: MAX

SUBSTANTIALLY EQUIVALENT TO:

The Cogent Lateral Interbody System is substantially equivalent to the predicate, Cogent Med-LIF System (K132738) in all facets including: indications, technology, method of use and performance.

Primary Predicate Device: Cogent Med-LIF System (K132738)

Additional Predicate Device: Cogent Med-LIF XL System (K142193)

No reference devices were used in this submission.

The Cogent Lateral Interbody System is similar to the predicate devices in function, technical features, materials, and instrumentation. It differs from the predicate devices in sizing (primarily length and width) as well as intended surgical approach to the spine.

{4}------------------------------------------------

DESCRIPTION of the DEVICE:

The Cogent Lateral Interbody System is an intervertebral body fusion device intended to stabilize the spinal segment to promote fusion. The implants are available in a variety of lengths and heights to accommodate varying patient anatomy. The Cogent Lateral Interbody System implants and instruments are supplied non-sterile. The implants in the The Cogent Lateral Interbody System are manufactured from PEEK, Ti-6Al-4V, and tantalum.

The Cogent Lateral Interbody consists of two PEEK spacers with axial voids to contain bone graft material, a titanium linkage that connects the PEEK spacers, angular anti-migration teeth, and tantalum x-ray markers.

The Cogent Lateral Interbody is available in various sizes to accommodate varying patient anatomy. The Cogent Lateral Interbody System is 18 or 22mm wide and includes five available lengths, 40, 45, 50, 55, and 60mm. The Cogent Lateral Interbody System is also available in three lordotic options, 0°, 6°, and 12°. The Cogent Lateral Interbody System is available in heights of 8mm to 16mm.

The Cogent Lateral Interbody System implants may be inserted via an open or minimally invasive approach and may be placed in a lateral or anterolateral orientation.

The Cogent Lateral Interbody System implants are non-sterile and are to be sterilized by the end user.

MATERIALS:

The Cogent Lateral Interbody is manufactured from polyetheretherketone (PEEK) as per ASTM F2026 and contains titanium alloy (Ti-6Al-4V) per ASTM F1472 and tantalum per ASTM F560.

INDICATIONS FOR USE:

The Cogent Lateral Interbody System is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to L5. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended for intervertebral body fusion, and are intended to be

{5}------------------------------------------------

used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.

NON-CLINICAL TEST SUMMARY:

Cadaveric Usability Study

CLINICAL TEST SUMMARY:

No clinical studies were performed.

CONCLUSIONS:

The Cogent Lateral Interbody System device has shown to be substantially equivalent to a legally marketed predicate based on indications for use, technological characteristics, performance testing and comparison to predicate devices.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.