K Number
K151390
Date Cleared
2016-05-02

(342 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

Test Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time (Minutes)
Cisplatin, 1.0mg/ml>240
Cyclophosphamide (Cytoxan), 20mg/ml>240
Carmustine (BCNU), 3.3mg/ml1.0
Dacarbazine (DTIC), 10.0mg/ml>240
Doxorubicin Hydrochloride, 2.0mg/ml>240
Etoposide (Toposar), 20.0mg/ml>240
Fluorouracil, 50.0mg/ml>240
Paclitaxel (Taxol), 6.0mg/ml>240
Thiotepa, 10.0mg/ml31.3
Methotrexate ,25mg/ml>240
Mitomycin C, 0.5mg/ml>240
Vincristine Sulfate, 1.0mg/ml>240
Ifosfamide, 50.0mg/ml>240

Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml: 1.0 minute Thiotepa, 10.0mg/ml: 31.3 minutes

Device Description

These patient examination gloves are formulated using nitrile, non-sterile, powder free, meet all the requirements of ASTM D6319 and tested for use with chemotherapy drugs.

AI/ML Overview

This document does not describe a device that uses Artificial Intelligence (AI) or Machine Learning (ML). It is an FDA 510(k) premarket notification for Powder Free Nitrile Examination Gloves Tested For Use with Chemotherapy Drugs.

Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance for an AI/ML device is not applicable to this document.

The document focuses on the physical and chemical properties of the gloves, primarily their resistance to permeation by chemotherapy drugs, and compares them to established ASTM standards and a predicate device.

However, to address the acceptance criteria and performance data as presented in the document for the physical device (gloves):

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the gloves are primarily based on ASTM standards. The relevant performance data is provided in "Table 1" on page 4 of the document, comparing the applicant's device to a predicate device and relevant standards.

CharacteristicsStandardsAcceptance Criteria (from Standards or Predicate)Reported Device Performance (Current Device)
DimensionsASTM D6319-10Min 240mm (Predicate)Min 270mm
Physical PropertiesASTM D6319-10MeetMeet
Thickness - FingerASTM D6319-100.07-0.10mm (Predicate)Min 0.10mm
Thickness - PalmASTM D6319-100.07-0.09mm (Predicate)Min 0.10mm
Freedom From HolesASTM D6319-10 and ASTM D5151PassPass
Powder Free ResidueASTM D6319-10 and ASTM D6124MeetMeet
Resistance to Permeation by Chemotherapy Drugs (ASTM D6978-05)Minimum Breakthrough Detection Time (min)Minimum Breakthrough Detection Time (min)
Cisplatin, 1.0mg/ml>240 minutes>240 minutes
Cyclophosphamide (Cytoxan), 20mg/ml>240 minutes>240 minutes
Carmustine (BCNU), 3.3mg/ml15.0 minutes1.0 minute (with warning)
Dacarbazine (DTIC), 10.0mg/ml>240 minutes>240 minutes
Etoposide (Toposar), 20.0mg/ml>240 minutes>240 minutes
Doxorubicin Hydrochloride, 2.0mg/ml>240 minutes>240 minutes
Fluorouracil, 50.0mg/ml>240 minutes>240 minutes
Paclitaxel (Taxol), 6.0mg/ml>240 minutes>240 minutes
Thiotepa, 10.0mg/ml2.0 minutes31.3 minutes (with warning)
Methotrexate, 25mg/ml- (Not explicitly listed for predicate)>240 minutes
Mitomycin C, 0.5mg/ml- (Not explicitly listed for predicate)>240 minutes
Vincristine Sulfate, 1.0mg/ml>240 minutes>240 minutes
Ifosfamide, 50.0mg/ml>240 minutes>240 minutes
Biocompatibility - Primary Skin IrritationISO 10993PassPass
Biocompatibility - Dermal SensitizationISO 10993PassPass
MaterialASTM D6319-10NitrileNitrile
Texture-Finger TexturedFinger Textured

Note on Chemotherapy Drug Permeation: The "acceptance criteria" here are comparison points from the predicate device and the standard. The device performance for Carmustine and Thiotepa for the current device are lower than >240 minutes, but still considered acceptable with explicit warnings regarding their low permeation times, similar to how the predicate device also had drugs with low permeation times. For Thiotepa, the current device performance (31.3 minutes) is significantly better than the predicate (2.0 minutes). For Carmustine, the current device performance (1.0 minute) is lower than the predicate (15.0 minutes). These differences are noted and addressed by a warning statement.

This summary is based on the information provided, which details the device's compliance with established industry standards for patient examination gloves and specific testing for chemotherapy drug permeation.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with a flowing design beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 2, 2016

Maxter Glove Manufacturing SDN. BHD. Yap Peak Geeh QA & Regulatory Affairs Manager Lot 6070. Jalan Haji Abdul Manan, 6th Miles Off Jalan Meru. Klang, Selangor 41050 MALAYSIA

Re: K151390

Trade/Device Name: Powder Free Nitrile Examination Gloves Tested For Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZC. LZA Dated: March 25, 2016 Received: March 30, 2016

Dear Yap Geeh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K151390

Device Name

Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs

Indications for Use (Describe)

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

Test Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time (Minutes)
Cisplatin, 1.0mg/ml>240
Cyclophosphamide (Cytoxan), 20mg/ml>240
Carmustine (BCNU), 3.3mg/ml1.0
Dacarbazine (DTIC), 10.0mg/ml>240
Doxorubicin Hydrochloride, 2.0mg/ml>240
Etoposide (Toposar), 20.0mg/ml>240
Fluorouracil, 50.0mg/ml>240
Paclitaxel (Taxol), 6.0mg/ml>240
Thiotepa, 10.0mg/ml31.3
Methotrexate ,25mg/ml>240
Mitomycin C, 0.5mg/ml>240
Vincristine Sulfate, 1.0mg/ml>240
Ifosfamide, 50.0mg/ml>240

Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml: 1.0 minute Thiotepa, 10.0mg/ml: 31.3 minutes

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for MAXTER Glove Manufacturing SON BHO. The logo features a black panther in front of a gray circle on the left side. To the right of the panther is the word "MAXTER" in large, bold letters, with the words "GLOVE .MANUFACTURING SON BHO" below it. The text "(22°802-H)" is located below the company name, and the text "since 1989" is located in the bottom left corner.

Lot 6070, Jalan HaJiAbo

1.0 Submitter:

NameYap Peak Geeh
AddressLot 6070 Jalan Haji Abdul Manan,6th Miles off Jalan Meru,41050 Klang, Selangor, Malaysia.
Phone No.: 6-03-33929888
Fax No.: 60-03-33923328
Date of Summary Prepared:8th March 2016

2.0 Name of the device:

Powder Free Nitrile Examination Gloves Tested for use with Chemotherapy Drugs Common Name : Exam Gloves Classification Name: Patient Examination Gloves Speciality ( 21 CFR 880.6250) Product Code LZC, LZA)

  • 3.0 Identification Of The Legally Marketed Devices that equivalency is claimed:
Predicate
ManufacturerWRP Asia Pacific Sdn Bhd
Device NameDermagrip Powder Free Blue Nitrile PatientExamination Gloves Tested for Use withChemotherapy Drugs
510(k) NumberK141982
Regulation Number21 CFR 880.6250
Regulatory NamePatient Examination Glove
Regulatory ClassI
  • 4.0 Description of the Device:
    These patient examination gloves are formulated using nitrile, non-sterile, powder free, meet all the requirements of ASTM D6319 and tested for use with chemotherapy drugs.

  • 5.0 Intended Use of the Device:
    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands and finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MAXTER GLOVE MANUFACTURING SDN BHD. The logo features a black panther in front of a moon-like circle on the left side. To the right of the panther is the company name in bold, black letters, and below that is the company's full name and registration number.

6.0 Summary of the Technological Characteristics of the Device:

Below is the summary of the technological characteristics of the Powder Free Nitrile Examination Gloves compare to ASTM D6319 or equivalent standards.

Chemotherapy claim is similar to Predicate, which has a glove thickness below 0.10mm and is shorter than 270mm but complaint with the ASTM standards.

Table 1

CharacteristicsStandardsDevice Performance
PredicateCurrent
ManufacturerWRP Asia PacificSdn BhdMaxter GlovesManufacturing Sdn Bhd
510 (K) NumberK141982K151390
DimensionsASTM D6319-10Min240mmMin270mm
Physical PropertiesASTM D6319-10MeetMeet
Thickness- Finger-palmASTM D6319-100.07-0.10mm0.07-0.09mmMin0.10mmMinO.10mm
Freedom FromHolesASTM D6319-10and ASTM D5151PassPass
Powder FreeResidueASTM D6319-10and ASTM D6124MeetMeet
Resistance toPermeation byChemotherapyDrugs:ASTM D6978-05
Test ChemotherapyDrugConcentrationMinimum Breakthrough Detection Time (min)
Cisplatin1.0mg/ml>240 minutes>240 minutes
Cyclophosphamide(Cytoxan)20mg/ml>240 minutes>240 minutes
Carmustine (BCNU)3.3mg/ml15.01.0 minute
Dacarbazine {DTIC)10.0mg/ml>240 minutes>240 minutes
I toposide (Toposar)20.0mg/ml>240 minutes>240 minutes
I oxorubicinHydrochloride2.0mg/ml>240 minutes>240 minutes
Fluorouracil50.0mg/ml>240 minutes>240 minutes
Paclitaxel (Taxol)6.0mg/ml>240 minutes>240 minutes
Thiotepa10.0mg/ml2.0 minutes31.3 minutes
Methotrexate25mg/ml->240 minutes
Mitomycin C0.5mg/ml->240 minutes
Vincristine Sulfate1.0mg/ml>240 minutes>240 minutes
Ifosfamide50.0 mg/ml>240 minutes>240 minutes
Warning StatementWarning: Pleasenote that thefollowing drugs haveextremely lowpermeation times:Carmustine(BCNU):15 Minutes andThiotepa: 2 minutesWarning: Please note thatthe following drugs havelow permeation time ofless than 240minutes:Carmustine (BCNU): 1.0Minute andThiotepa : 31.3 Minutes
BiocompatibilityPrimary SkinIrritation-ISO10993PassPass
DermalSensitization-ISO10993PassPass
Intended Use-A patientexamination glove isa disposable deviceintended for medicalpurposes that is wornon the examiner'shand or finger topreventcontaminationbetween patient andexaminer.A powder free patientexamination glove is adisposable device intendedfor medical purposes thatis worn on the examiner'shands and finger toprevent contaminationbetween patient andexaminer.In addition,these gloves were testedfor use with chemotherapydrugs in accordance withASTMD6978-05.
MaterialASTM D6319-10NitrileNitrile
Texture-Finger TexturedFinger Textured

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a black panther in front of a moon. The panther is looking to the right, and the moon is in the background. The text "SINCE 1965" is at the bottom left of the image. The image is a black and white drawing.

MAXTER

GIOVE MANUFACTURING SON BHD

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MAXTER GIOVE MANUFACTURING SON BHO (229862-H). The logo features a black panther in front of a circle on the left side. The text "MAXTER" is in large, bold letters on the right side of the logo.

lot 6070, Jaloo Hnji Abdul Manon
6tn Miles Off JalonIJ.eru
4H150 Klang, Sclongo<, Moloysio
Tet 603-33929868 {8 ines) Fax: 603-33923328
E-MAIL: m:utcrt'lrn.nct.my
'WYNt.maxter.com.my

SizeMedical GloveGuidance Manual-LabelingExtra SmallExtra Small
SmallSmall
MediumMedium
LargeLarge
Extra Lar_geExtra Large
Single UseMedical GloveGuidance Manual-LabelingSingle useSingle use

7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data

The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above (ASTM Requirements).

The device and the predicate share the same intended use, same material, same specifications for thickness and length, similar permeation rates for chemotherapy

Drugs, similar labeling according to the glove guidance, and same compliance with standards for physical properties, powder free, biocompatibility and watertightness.

8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data

Not applicable- Clinical data is not needed for gloves or for most devices cleared by the 510(K) process.

9.0 Conclusion

Based on intended uses, technological characteristics and non-clinical performance data, the Powder Free Nitrile Examination Gloves Tested for Use With Chemotherapy Drugs is substantially equivalent to the predicate device K141982.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.